K221873 · Energist Limited · GEI · Jan 12, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221873
Device Name
NeoGen PSR System
Applicant
Energist Limited
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 12, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions : - · Treatment of wrinkles and rhytids - · Superficial skin lesions - · Actinic Keratosis - Viral Papillomata - · Seborrheic Keratosis - · Acne Scars When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.
Device Story
NeoGen PSR System delivers controlled thermal energy to skin via nitrogen gas plasma; UHF energy (2450-2480 MHz) ionizes pressurized nitrogen gas within handpiece nozzle. System comprises generator, footswitch, handpiece, disposable nozzles (5mm or 25mm standoff), and nitrogen gas supply. Used in hospitals/clinics by qualified practitioners. Generator touchscreen allows setting energy (0.5-4.0J) and pulse duration (5-15ms). Plasma pulses (manual or automatic repeat) cause controlled epidermal non-viability and thermal modification of dermis; natural healing replaces damaged skin with new growth. System automatically detects nozzle type, adjusts energy display, and monitors nozzle life via internal counters. Benefits include improved appearance of dermatological conditions and controlled tissue coagulation.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI ES60601-1 (basic safety), IEC 60601-2-2 (HF surgical equipment), and IEC 60601-1-2 (EMC) demonstrated. System validation and V&V reports confirm functional performance requirements.
Technological Characteristics
Class II electrosurgical device. UHF energy source (2450-2480 MHz). Nitrogen gas plasma. Ultem patient-contacting material. 5mm and 25mm standoff nozzles. Type BF applied parts. Class 1 generator. Touchscreen GUI. Electronic shot counting/nozzle identification. Standalone system with nitrogen gas supply.
Indications for Use
Indicated for patients requiring treatment of wrinkles, rhytids, superficial skin lesions, actinic keratosis, viral papillomata, seborrheic keratosis, and acne scars (5mm nozzle), or tissue coagulation (25mm nozzle).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
January 12, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two main elements: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Energist Limited Graham Booth Chief Technical Officer 2 Park Pavilions, Clos Llyn Cwm Enterprise Park Swansea, SA6 8QY United Kingdom
Re: K221873
Trade/Device Name: NeoGen PSR System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: December 08, 2022 Received: December 13, 2022
Dear Graham Booth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
{1}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Colin K. Chen -S
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221873
#### Device Name
NeoGen PSR Plasma Skin Resurfacing Device
#### Indications for Use (Describe)
When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions :
- · Treatment of wrinkles and rhytids
- · Superficial skin lesions
- · Actinic Keratosis
- Viral Papillomata
- · Seborrheic Keratosis
- · Acne Scars
When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.
**Type of Use (Select one or both, as applicable)**
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221873: Energist Limited NeoGen PSR 510(k) Application - 510(k) Summary
| Revision: | B | Date: | 12 January 2023 |
|--------------------------------|-----------------------------------------------------------------------------------------------|-------|-----------------|
| Document Title: | 510(k) Summary (Per 21 CFR 807.92) | | |
| Subject: | NeoGen PSR System 510(k) Application | | |
| 510(k) Number | K221873 | | |
| Submitter Name: | Energist Limited | | |
| Establishment Registration No: | 3007353760 | | |
| Submitter Address: | 2 Park Pavilions, Clos Llyn Cwm, Swansea<br>Enterprise Park, Swansea, SA6 8QY, United Kingdom | | |
| Contact Person: | Graham Booth – Chief Technical Officer<br>Email: gbooth@energist.com | | |
| Telephone Number: | +44 1792 798768 | | |
| Revision Date: | 11 January 2023 | | |
| Device Trade Name: | NeoGen PSR System | | |
| Common Name: | Electrosurgical Device | | |
| Device Classification: | Electrosurgical cutting and coagulation device and<br>accessories Class II | | |
| Product Code: | GEI (21 CFR 878.4400) | | |
# DEVICE DESCRIPTION
### System Elements
The system comprises the following major elements:
- . The Energist NeoGen PSR generator. This incorporates a source of Ultra High Frequency (UHF) energy and the gas control system required to support creation of nitrogen plasma;
- . A Footswitch – single pedal type for activation of the plasma output.
- . A handpiece that connects to the generator via an umbilical, which incorporates electrical and gas connections. The handpiece incorporates a finger-operated button switch that can be used to activate the plasma as an alternative to the footswitch.
- . A Procedure Pack, comprising one or more limited life disposable tips (nozzles) through which the plasma is delivered. These are a snap fit to the distal end of the handpiece;
- A gas supply system - principally a nitrogen gas cylinder and a regulator assembly;
- . A bespoke trolley which provides a secure, mobile platform for the generator, handpiece, gas cylinder (plus a spare cylinder) and gas regulator assembly.
{4}------------------------------------------------
# Principle of Operation
The NeoGen PSR delivers controlled thermal energy to the outer layers of the skin so that part or all of the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis. The body's natural healing process causes the damaged skin to be replaced with new, fresh, growth – thus resulting in improvement of the appearance and effects of the indicated conditions.
The thermal energy emitted by the system is in the form of a nitrogen gas plasma. This is created by subjecting pressurised nitrogen gas to a high level of Ultra High Frequency (UHF) energy. This process takes place in the handpiece of the device, more specifically in a resonator tube, or "nozzle", the distal end of which is directed at the skin surface. The handpiece is connected to the main generator unit via an "umbilical", which incorporates a gas supply tube and a cable capable of conducting the UHF energy. It also incorporates a cable for control signals between the handpiece and the generator unit to support the action of a control button and gas solenoid valve, plus secure identification and "counter" elements in the nozzle and the handpiece itself. The plasma is created in short pulses, approximately 5 to 15 milli-seconds in duration. Automatic adjustment of the UHF power level and pulse duration to the handpiece enables control of tissue effects by altering the amount of energy delivered by the plasma to the skin surface for each pulse. In practice, the energy per pulse can be varied from 1.0 to 4.0 Joules when the system is used with the 5mm nozzle, and from 0.5 to 0.8 Joules when used with the 25mm nozzle. The system can be configured to automatically generate a series of pulses at set intervals to allow the operator to move the handpiece over the surface of the skin at a steady speed to achieve a uniform effect. Alternatively, the pulses can be manually triggered - either by pressing the control button on the handpiece, or by depressing a footswitch, according to operator preference.
The nozzles are limited-life components which are a snap-fit onto the distal end of the handpiece and are easily replaced. A stand-off incorporated in the nozzle allows the distance from the skin surface to be accurately determined. The separation distance influences the amount of energy impacting the skin surface. NeoGen PSR systems are capable of supporting the use of nozzles with 5mm standoff, as well as 25mm. This is as per the Rhytec Portrait PSR3, which was the legally marketed predicate device cited when the NeoGen PSR was originally cleared under K132754.
Handpiece usage is also monitored and the system limits use to guard against unexpected failure due to degradation with use over time. The handpiece is user replaceable.
The generator incorporates a large colour touchscreen display supporting a Graphical User Interface that clearly conveys information to the operator and allows settings to be made readliy and conveniently.
{5}------------------------------------------------
### Intended Use/ Indications For Use
When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions:
- . Treatment of wrinkles and rhytids
- Superficial skin lesions
- Actinic Keratosis .
- Viral Papillomata
- Seborrheic Keratosis .
- · Acne Scars
When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.
### Intended Use Environment
The system is intended for professional use by trained and duly qualified medical practitioners in hospitals and clinics. It must not be used in domestic, industrial or outdoor locations, or in critical care environments.
{6}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS
The technological characteristics of the current Energist NeoGen PSR system with 25mm nozzle option that is the subject of this K221873 submission and the predicate PSR system originally cleared under K132754 are summarised and compared in the table below:
| Parameter | K132754 Predicate NeoGen PSR system | K222873 Subject NeoGen PSR system with 25mm option |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Energy setting Displayed Values Precision | 0.1 Joule | 0.1 Joule |
| Max Output Energy | 4 Joules | 4 Joules |
| Treatment Pulse Repetition Rate | Repeat Pulse mode: 1.0 to 2.5 Hz<br>Single Pulse mode: User determined. | Repeat Pulse mode: 1.0 to 2.5 Hz<br>Single Pulse mode: User determined. |
| Output Pulse Width | 5.2 to 15.4 ms | 5.2 to 15.4 ms |
| Safety Classification<br>(per EN/ISO 60601-1) | Class 1 (the Generator requires connection to a Protective Earth). | Class 1 (the Generator requires connection to a Protective Earth). |
| FDA Regulatory Classification | Class II | Class II |
| FDA Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 |
| FDA Regulation Name | Electrosurgical cutting and coagulation device and accessories. | Electrosurgical cutting and coagulation device and accessories. |
| FDA Product Code | GEI | GEI |
| EU MDR Classification | Risk Class Ilb | Risk Class Ilb |
| Applied Parts | Type BF(not defibrillator proof) | Type BF(not defibrillator proof) |
| Duty Cycle | 20 seconds on, 10 seconds off, for a maximum of 50 plasma pulses at the 2.5Hz maximum rate | 20 seconds on, 10 seconds off, for a maximum of 50 plasma pulses at the 2.5Hz maximum rate |
| Operational Ambient Temperature Range | 10 to 30 °C. | 10 to 30 °C. |
| Operational Humidity | 30 to 75% RH | 30 to 75 RH |
| Atmospheric Pressure | 700 to 1060 hPa. | 700 to 1060 hPa. |
| Power<br>Requirements | Voltage: 100-230 Vrms, 50-60Hz, 1 ph.<br>Current: 6.5A max.<br>Power Consumption: ≤650 VA | Voltage: 100-230 Vrms, 50-60Hz, 1 ph.<br>Current: 6.5A max.<br>Power Consumption: ≤650 VA |
| Inlet Safety<br>Fuses | Quantity, Size: 2x, 20mm x 5mm<br>Type: Time Lag (T) type<br>Rating: 6.3A, 250 VAC<br>Approvals: UL Listed<br>Breaking Capacity: 1500A @250VAC | Quantity, Size: 2x, 20mm x 5mm<br>Type: Time Lag (T) type<br>Rating: 6.3A, 250 VAC<br>Approvals: UL Listed<br>Breaking Capacity: 1500A @250VAC |
| Parameter | K132754 Predicate NeoGen PSR system | K222873 Subject NeoGen PSR system with 25mm option |
| Mains Connection<br>Inlet type:<br>Power Cord:<br>Mains Switch: | IEC 60320 type C14<br>IEC 60320 type C13 (local plug a/r)<br>Double Pole, integral with inlet | IEC 60320 type C14<br>IEC 60320 type C13 (local plug a/r)<br>Double Pole, integral with inlet |
| RF Output Frequency | 2450 to 2480 MHz (2470 MHz typ).<br>No sub-harmonics | 2450 to 2480 MHz (2470 MHz typ).<br>No sub-harmonics |
| Modulation | Pulsed CW (Carrier Wave) | Pulsed CW (Carrier Wave) |
| Gas System | Commercially available medical<br>grade nitrogen is supplied via a<br>Regulator assembly included with<br>the system to provide a feed to the<br>Generator at 1.7Bar (nom.) | Commercially available medical<br>grade nitrogen is supplied via a<br>Regulator assembly included with<br>the system to provide a feed to the<br>Generator at 1.7Bar (nom.) |
| Footswitch | Medical type, rated 60V, 5A DC | Medical type, rated 60V, 5A DC |
| Trolley | Custom made 5-wheel trolley with<br>locking wheels accommodates the<br>Generator and 2 gas cylinders<br>(including a spare). Meets stability<br>requirements of IEC 60601-1. | Custom made 5-wheel trolley with<br>locking wheels accommodates the<br>Generator and 2 gas cylinders<br>(including a spare). Meets stability<br>requirements of IEC 60601-1. |
| Handpiece | Connecting to the Generator via a<br>flexible umbilical, the Handpiece<br>accommodates the nozzle, which is<br>where the plasma is generated.<br>Total number of shots is limited<br>(electronically) to 250,000 after<br>which the handpiece must be<br>replaced. | Connecting to the Generator via a<br>flexible umbilical, the Handpiece<br>accommodates the nozzle, which is<br>where the plasma is generated.<br>Total number of shots is limited<br>(electronically) to 250,000 after<br>which the handpiece must be<br>replaced. |
| Nozzles | 5mm standoff version allows for<br>delivery of plasma energy in the<br>range 1.0 to 4.0 Joules. | 5mm standoff version allows for<br>delivery of plasma energy in the<br>range 1.0 to 4.0 Joules.<br>25mm standoff option allows for<br>delivery of plasma energy in the<br>range 0.5 to 0.8 Joules. |
{7}------------------------------------------------
With the exception of the 25mm nozzle option (highlighted in red) – clearance of which is the subject of this 510(k) submission – the technological characteristics of the predicate and subject devices are identical.
{8}------------------------------------------------
# TESTS PERFORMED
A full description of tests carried out to confirm conformity to applicable standards is given in Energist document no. DD0254 NeoGen PSR Compliance Summary Report [USA], included with this submission.
Note that the 25mm nozzle has been included in the scope of the assessments conducted (it has been made available in other markets since 2014). However, the choice of 5mm or 25mm nozzle does not materially impact safety or EMC.
A summary is given below.
# Basic Safety
Representative samples of the NeoGen PSR system have been tested by an accredited thirdparty test organisation to demonstrate compliance with the following medical device safety standards: These tests have included assessment
# 60601-1 Basic Safety
| Standard | ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 &<br>A2:2010/(R)2012 – Medical electrical equipment Part 1: General<br>requirements for basic safety and essential performance. |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus<br>Standard Recognition<br>Number | 19-4 |
| Report Ref. | 071-75945748-000TRF_CE_E (inc. attachment 1) |
| Issued By | TÜV SÜD, United Kingdom |
### 60601-2-2 HF Surgical Equipment
| Standard | IEC 60601-2-2:2017 - Medical electrical equipment – Part 2-2:<br>Particular requirements for the basic safety and essential performance<br>of high frequency surgical equipment and high frequency surgical<br>accessories |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus<br>Standard Recognition<br>Number | 6-389 |
| Report Ref. | 071-75945748-000TRF_CE_E (attachment 2) |
| Issued By | TÜV SÜD, United Kingdom |
{9}------------------------------------------------
### Electro-Magnetic Compatibility
Representative samples of the NeoGen PSR system have been tested by an accredited thirdparty test organisation to demonstrate compliance with the following medical device EMC standards:
| Standard | IEC 60601-1-2 - Medical electrical equipment -- Part 1-2: General<br>requirements for basic safety and essential performance -- Collateral<br>Standard: Electromagnetic disturbances -- Requirements and tests |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus<br>Standard Recognition<br>Number | 19-8 |
| Report Ref. | R18-4514A Is. 2 |
| Issued By | Kiwa, United Kingdom |
# Performance Testing
Representative samples of the NeoGen PSR system have been tested by Energist Limited UK to demonstrate compliance with the user and functional performance requirements. Numerous documents are available, the "headline" ones being:
DD0133R NeoGen System Validation Report ;Overall System Requirements SD0618R NeoGen System V&V Report :Software determined functionality
In respect of the 25mm nozzle specifically, please refer to the following document:
DD0246 NeoGen 25mm Nozzle - Safety & Efficacy Review.
{10}------------------------------------------------
# TECHNOLOGY COMPARISON
This is a review of the differences between the 25mm standoff arrangements as cleared in K072394 for use with the Rhytec PSR3. (the original predicate as cited in the K132754 submission for the NeoGen PSR), and those of the current 25mm nozzle that is the subject of this K221873 submission.
# Rhytec 25mm Standoff
This took the form of an adapter that could be fitted over the standard nozzle. The below excerpt from Rhytec's instruction sheet for the standoff shows the fitting instructions, and also gives an idea of the physical appearance and characteristics.
Image /page/10/Picture/5 description: In the image, a person wearing purple gloves is holding a white nozzle. The top surface of the nozzle is visible. Above the nozzle, there is a handle with a purple design on it. The text in the image labels the top surface of the handle and nozzle.
Top surface of Stand-off
Orientate Stand-off to the Nozzle noting alignment of top surfaces of each part
Image /page/10/Picture/8 description: In this image, a hand wearing a purple glove is holding a white and purple device with two prongs extending from the bottom. The device appears to be some sort of medical or scientific instrument. The prongs are translucent and have a yellowish tint. The hand is gripping the device firmly, and the background is a plain, light-colored surface.
Engage the Stand-off onto the Nozzle
Image /page/10/Picture/10 description: In the image, a gloved hand is holding a white plastic device with yellow tubes extending from the bottom. The device appears to be some sort of medical or laboratory instrument. The hand is holding the device near the top, and the tubes are pointing downwards. A black arrow is pointing to the top of the white plastic device.
Stand-off correctly fitted onto the Nozzle
Align protrusion on the Stand-off in relation to the joint between the Handpiece and Nozzle
Essential characteristics of the Rhytec 25mm standoff are as follows:
| Standoff Distance: | 25mm |
|------------------------------|------------------------------------------|
| Service Life: | To be replaced after use with 10 nozzles |
| Energy Reduction Factor: | 0.65 |
| Patient Contacting Material: | Ultem |
| Spot Separation: | 12mm |
| Energy Range: | 0.65 to 1.0 Joule* |
* The lower limit is determined by the minimum output level of the Portrait PSR3 (1 Joule) and the Energy Reduction Factor (0.65). The maximum is that recommended in the IFU to prevent overheating of the tips, which could result in patient discomfort and damage to the standoff.
The Rhytec Portrait PSR was a relatively basic device and was not able to detect the presence or absence of the adapter. It was therefore necessary for the operator to perform their own calculation of the effective energy reaching the skin surface, based on the reading on the numeric display.
{11}------------------------------------------------
The range of energy outputs available were:
| PSR3 Energy Setting (J): | 1.0 | 1.1 | 1.2 | 1.3 | 1.4 | 1.5 |
|-----------------------------|------|-----|-----|------|-----|-----|
| Energy at Skin Surface (J): | 0.65 | 0.7 | 0.8 | 0.85 | 0.9 | 1 |
#### NeoGen 25mm Nozzle
On the NeoGen device, the standoff is an integral part of the nozzle. It attaches to the distal end of the handpiece in the same manner as the 5mm nozzle. A picture of the device is shown below.
Image /page/11/Picture/5 description: The image shows a black electrical plug with three prongs. The plug has a clear glass tube extending from one side, with a visible filament inside. The prongs are a light tan color and appear to be made of a flexible material. The plug is resting on a light-colored surface.
Essential characteristics of the NeoGen 25mm nozzle are as follows:
| Standoff Distance: | 25mm |
|------------------------------|----------------------------------|
| Service Life: | 9000 shots |
| Energy Reduction Factor: | N/A |
| Patient Contacting Material: | Ultem |
| Spot Separation: | 12mm |
| Energy Range: | 0.5 to 0.8 Joule (in 0.1J steps) |
The main practical differences between the NeoGen 25mm nozzle and the Rhytec solution are:
- . The NeoGen PSR automatically detects the presence of the 25mm nozzle and adjusts the on-screen energy settings to reflect the actual energy delivered to the skin surface. This reduces the burden on the operator, and the possibility of errors;
- It is not possible to exceed the maximum ratings of the nozzle in respect of enerqy level or ● rated lifetime;
- The possibility of the standoff being incorrectly fitted, or becoming detached, is reduced;
- Available energy levels differ. At such low levels the clinical differences are minimal however.
The main similarities are:
- Same spot size treatment techniques therefore remain the same; ●
- Same material used for standoff biocompatibility and thermal stability considerations ● therefore remain the same.
{12}------------------------------------------------
# CONCLUSION
The comparison of Technological Characteristics and the results of the performance and compliance testing conducted demonstrate that the subject device is Substantially Equivalent to the predicate device as cleared in K132754, and that the 25mm nozzle is Substantially Equivalent to that of the Rhytec nozzle/25mm standoff combination as cleared under K072394.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.