K221865 · BTL Industries, Inc. · ISA · Apr 14, 2023 · Physical Medicine
Device Facts
Record ID
K221865
Device Name
BTL-094
Applicant
BTL Industries, Inc.
Product Code
ISA · Physical Medicine
Decision Date
Apr 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5660
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
BTL-094 is indicated to be used for: - Relief of minor muscle aches and pains. - Temporary increase in local blood circulation. - Activation of connective tissue.
Device Story
BTL-094 is a non-invasive therapeutic device using acoustic waves to stimulate local biological responses in tissue. It generates extracorporeally induced massage pulses via a piezo transducer and coupling pad. The device features a main unit with a color touch screen for parameter adjustment (intensity, frequency, number of shocks) and a secondary screen on the applicator for status monitoring. It is operated by a clinician in a clinical setting. The device provides continuous vibration or single pulse modes. By delivering acoustic pressure waves, it aims to relieve muscle aches, increase circulation, and activate connective tissue, assisting in physical therapy and pain management.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and biocompatibility (ISO 10993-1, 5, 10).
Technological Characteristics
Non-invasive therapeutic massager; piezoelectric transducer with acoustic lens; coupling gel pads; 100-240V AC power; 1-25 Hz frequency; 35 mm max penetration depth; 0.01-0.11 mJ/mm2 energy flow density. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-36, ISO 14971, and ISO 10993 series. Software lifecycle per IEC 62304.
Indications for Use
Indicated for relief of minor muscle aches and pains, temporary increase in local blood circulation, and activation of connective tissue in patients requiring therapeutic massage.
Regulatory Classification
Identification
A therapeutic massager is an electrically powered device intended for medical purposes, such as to relieve minor muscle aches and pains.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 14, 2023
BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752
Re: K221865
Trade/Device Name: BTL-094 Regulation Number: 21 CFR 890.5660 Regulation Name: Therapeutic Massager Regulatory Class: Class I Product Code: ISA Dated: March 20, 2023 Received: March 20, 2023
Dear David Chmel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Amber T. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221865
Device Name BTL-094
Indications for Use (Describe) BTL-094 is indicated to be used for:
- Relief of minor muscle aches and pains.
- Temporary increase in local blood circulation.
- Activation of connective tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares, each nested inside the other.
## 510(k) Summary
## General Information
| Sponsor: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person: | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com |
| Summary Preparation<br>Date: | April 14, 2023 |
## Device Name
| Trade/Proprietary Name: | BTL-094 |
|------------------------------|--------------------------|
| Primary Classification Name: | Therapeutic Massager |
| Common/Usual Name: | Therapeutic Massager |
| Classification Regulation: | 21 CFR 890.5660, Class I |
| Classification Product Code: | ISA |
{4}------------------------------------------------
## Legally Marketed Predicate Device
The BTL-094 is a state-of-the-art device substantially equivalent to the following products, which are already cleared for distribution in the USA under the following 510(k) Premarket Notification number:
- D-Actor 200 (K173692) ●
- DolorClast® Radial (K220538) .
## Product Description
The BTL-094 is a non-invasive therapeutic device, which uses acoustic waves in order to stimulate a local biological response in the treated tissue.
The BTL-094 generates extracorporeally induced massage pulses to stimulate a local biological response in the treated tissue. The main source of acoustic waves is a piezo transducer inside the applicator. The energy transfer from the piezo transducer to the treated tissue is realized via coupling pad.
The main unit is equipped with a color touch screen, which considerably simplifies its operation. The onscreen information guides the user step through the entire therapy process. A secondary screen is placed on the applicator for the presentation of the actual state of main therapy parameters. The therapeutic parameters are adjustable via touch-screen, buttons and knob. The user is able to set following therapy parameters: intensity, frequency and number of shocks.
The mechanism of action and technological similarities and differences between the BTL-094 device and the predicate devices are described below in the comparison table. The differences do not raise any new types of safety or effectiveness questions.
## Indications For Use
BTL-094 is indicated to be used for:
- Relief of minor muscle aches and pains.
- Temporary increase in local blood circulation.
- Activation of connective tissue.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares, each containing the letters "B", "T", and "L" respectively.
## Non-clinical Testing
The BTL-094 device has been thoroughly evaluated for electrical safety. The device has been found to comply with the following applicable medical device safety standards:
| IEC 60601-1:2005/A1:2012<br>FR Recognition Number: 19-4 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2014<br>FR Recognition Number: 19-8 | Medical electrical equipment – Part 1-2: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-6:2010/A1:2013<br>FR Recognition Number: 5-89 | Medical electrical equipment - Part 1-6: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Usability |
| IEC 60601-2-36:2014<br>FR Recognition Number: 9-119 | Medical electrical equipment - Part 2-36: Particular requirements for<br>the basic safety and essential performance of equipment for<br>extracorporeally induced lithotripsy |
| IEC 62304:2006/A1:2015<br>FR Recognition Number: 13-79 | Medical device software – Software lifecycle processes |
| ISO 14971:2019<br>FR Recognition Number: 5-125 | Medical devices - Application of risk management to medical<br>devices |
| ISO 10993-1:2018<br>FR Recognition Number: 2-258 | Biological evaluation of medical devices – Part 1: Evaluation and<br>testing within a risk management process |
| ISO 10993-5:2009<br>FR Recognition Number: 2-245 | Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity |
| ISO 10993-10:2010<br>FR Recognition Number: 2-174 | Biological evaluation of medical devices – Part 10: Tests for<br>irritation and skin sensitization |
## Clinical Testing
Not aplicable
## Technological Characteristics
The BTL-094 is a non-invasive therapeutic device, which uses acoustic waves in order to stimulate a local biological response in the treated tissue.
The BTL-094 generates extracorporeally induced massage pulses to stimulate a local biological response in the treated tissue. The main source of acoustic waves is a piezo transducer inside the applicator. The energy transfer from the piezo transducer to the treated tissue is realized via coupling pad.
{6}------------------------------------------------
The mechanism of action and technological similarities and differences between the BTL-094 device and the predicate devices are described below in the comparison table. The differences do not raise any new types of safety or effectiveness questions.
| Comparison with the Predicate Device | | | |
|--------------------------------------|--|--|--|
|--------------------------------------|--|--|--|
| 510(k)<br>number | K221865 | K173692 | K220538 | |
|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | BTL-094 | D-Actor 200 | DolorClast®<br>Radial | |
| Company<br>name | BTL Industries Inc. | Storz Medical | EMS Electro<br>Medical SA | |
| Indications for<br>Use | Relief of minor<br>muscle aches and<br>pains<br>Temporary increase<br>in local blood<br>circulation<br>Activation of<br>connective tissue | *Relief of minor<br>muscle aches and<br>pains<br>*Temporary increase<br>in local blood<br>circulation<br>*Activation of<br>connective tissue | - Relief of minor<br>muscle aches and<br>pains.<br>- Temporary<br>increase in local<br>blood circulation.<br>- Activation of<br>connective tissue. | None |
| Product Code<br>and<br>Regulation | 21 CFR<br>890.5660<br>ISA | 21 CFR<br>890.5660<br>ISA | 21 CFR<br>890.5660<br>ISA | None |
| Modes of<br>Action | Extracorporeally<br>induced pressure<br>waves. | Radial pressure<br>waves, or<br>extracorporeal pulse<br>activation<br>respectively | Radial pressure<br>waves, or<br>extracorporeal<br>pulse activation<br>respectively | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Mechanisms of<br>Action | Vibrations generated<br>by electroacoustic<br>technology. | Pneumatically<br>generated vibrations | Pneumatically<br>generated<br>vibrations | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Maximum and<br>Minimum<br>intensity<br>settings | 5-30%<br>alternative output<br>power setting scale | 1-5bar | 1-4bar | Not Significantly<br>different<br>Differences are<br>caused by<br>alternative<br>technology of the<br>pressure wave<br>generation. |
| Number and<br>size of<br>treatment<br>applicator<br>heads | 2;<br>L Pad - 12.2 cm²<br>U Pad - 7.3 cm² | 4;<br>6mmOD, 15mmOD,<br>20mmOD, 35mmOD | 7;<br>5mm OD, 10 mm<br>OD, 15 mm OD, 25<br>mm OD, 40 mm OD | Not Significantly<br>different<br>The application<br>area of the pads in<br>contact with the<br>patient is within the<br>set range of the<br>predicate devices.<br>All materials in<br>contact with the<br>patient have been<br>assessed for<br>biocompatibility |
| Maximum and<br>minimum<br>displacements<br>of applicator<br>heads | N/A | 0.6 - 2.0mm | 0.18 - 0.28 mm | N/A<br>The BTL-094<br>applicator does not<br>have any moving<br>parts that come in<br>contact with the<br>patient. |
| Type of<br>application<br>(e.g.,<br>continuous<br>vibration at a<br>fixed<br>frequency); | Continuous vibration<br>at a fixed frequency<br>and single pulse<br>mode | Continuous vibration<br>at a fixed frequency | Continuous<br>vibration at a fixed<br>frequency | None |
| Maximum and<br>minimum<br>vibration<br>frequency | 1-25 Hz | 1-21Hz | 1-25 Hz | Not Significantly<br>different |
| Power Supply | 100 V to 240 V AC<br>50 - 60 Hz<br>150VA | 500VA | Mains power | None |
| Maximum<br>penetration<br>depth | 35 mm | 32.3mm | 40 mm | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Energy flow<br>density | In the range of<br>0.01 - 0.11 mJ/mm2 | Values of ultrasonic<br>pulse:<br>5bar/0.284mJ/mm2<br>3bar/0.176mJ/mm2 | 0.29 mJ/mm2 at 4<br>bar<br>0.14 mJ/mm2 at 2.4<br>bar | Not Significantly<br>different |
| Operating<br>mode | Continuous | Continuous | Continuous | None |
| Projectile<br>mass (g) | N/A | 3 | 3.2 | Not applicable<br>The BTL-094<br>applicator does not<br>contain a projectile. |
| Pulse repeat<br>rate (1/s) | 1-25 Hz | 1-21Hz | 1-25 Hz | Not Significantly<br>different |
| Number of<br>pulses (min<br>and max) | Variable in range<br>0 – 9999 | Variable | Variable,<br>Max<br>5000/treatment | Not Significantly<br>different |
| Maximum<br>operating<br>temperature | 10 - 35 °C | 10-40°C | 10 - 30 °C | Not Significantly<br>different |
| Type of<br>acoustic wave<br>generation | Electroacoustic<br>technology with<br>electroacoustic lens<br>system | Pneumatic/ballistic | Pneumatic/ballistic | Not Significantly<br>different<br>Differences are<br>caused by<br>alternative<br>technology of the<br>pressure wave<br>generation. |
| Positive peak<br>pressure<br>amplitude<br>(MPa) | In the range of<br>3 - 10 MPa | Values of ultrasonic<br>pulse: 5bar/18.5MPa<br>3bar/13.4MPa | 17 MPa at 4 bar<br>11.24 MPa at 2.4<br>bar | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Negative peak<br>pressure<br>amplitude<br>(MPa) | In the range of<br>3 - 10 MPa | Values of ultrasonic<br>pulse: 5bar/6.8MPa<br>3bar/5.0MPa | 10 MPa at 4 bar<br>7.2 MPa at 2.4 bar | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Derived focal<br>acoustic pulse<br>energy (mJ) | In the range of<br>0.1 – 1.4 mJ | Values of ultrasonic<br>pulse: 5bar/6.5mJ<br>3bar/2.4mJ | 5.9 mJ at 4 bar*<br>2.2 mJ at 2.4 bar*<br>*Averaged across<br>all applicator sizes. | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Derived<br>pulse-<br>intensity<br>integral,<br>integrated over<br>total temporal<br>integration<br>limits<br>(mJ/mm2) | In the range of<br>0.01 - 0.11 mJ/mm2 | Values of ultrasonic<br>pulse:<br>5bar/0.284mJ/mm2<br>3bar/0.176mJ/mm2 | 0.29 mJ/mm2 at 4<br>bar<br>0.14 mJ/mm2 at 2.4<br>bar | Not Significantly<br>different<br>Please see the<br>discussion and<br>conclusion below. |
| Rise time (ns) | 10 ns | Ultrasonic pulse:<br>2.5µs | 3.2 µs | Not Significantly<br>different |
| | | Sonic pulse: 25µs –<br>2.5ms | | Differences are<br>caused by<br>alternative<br>technology of the<br>energy generation. |
| Compressional<br>pulse duration<br>(μs) | 2µs | Ultrasonic pulse:<br>5.0µs Sonic pulse:<br>50µs – 5.0ms | 2.6 µs (1st peak)<br>62.7 µs (1st phase) | Not Significantly<br>different |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares with the letters "BTL" inside. To the right of the logo, the words "BTL Industries" are written in a modern, sans-serif font. The text is in a dark gray color.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested squares with the letters "BTL" inside the innermost square. To the right of the logo is the text "BTL Industries" in a sans-serif font.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the square logo is the text "BTL Industries" in a gray sans-serif font.
#### Modes of Action, Mechanisms of Action
The technology used in BTL-094 device differs from the predicate device. Nevertheless, the technology creates the same type of acoustic pressure delivered to the tissue. Similar devices that generate a pressure wave based on a different technological principle but with the same or very comparable mechanism of action have already been recognised as substantially equivalent in this device category.
#### Maximum and Minimum intensity settings
BTL-094 has an alternative type of technology from the predicate device and therefore the BTL-094 device has an alternative power setting scale - in percentage and not in bars as a comparative device has. In addition, for comparability with the predicate devices the BTL-094 device intended for the US market is limited to a maximum power level of 30% of maximum driving power.
#### Number and size of treatment applicator heads
The number of application pads are different from predicate device but the area of the subjected device pads that is in contact with the patient is comparable within a set range to the contact area of application tips of predicate devices. Additionally all materials in contact with patient have been assessed for biocompatibility and the maximal depth of pressure wave penetration is comparable to the predicate.
#### Maximum and minimum displacements of applicator heads
The predicate device is generating the pressure wave by the head displacement. Due to an alternative type of technology used for pressure wave generation the BTL-094 device does not have the ability of displacements of applicator heads/pads.
#### Maximum and minimum vibration frequency
BTL-094 device differs from one predicate device in the value of the maximum frequency, where the frequency value of the BTL-094 device can be up to 4Hz higher than the predicate device, but has available the same frequency range as its second predicate device.
#### Driving Power
BTL-094 device has an alternative type of technology from the predicate device and therefore the driving power parameter is not applicable. In addition, for comparability with the predicate devices the BTL-094 device intended for the US market is limited to a maximum power level of 30% of maximum driving power.
#### Energy flow density
BTL-094 device uses alternative technology and is able to better pulse targeting, a maximum driving power of the device has been limited at a maximum power level of 30%. The EFD of our device differs from the predicates but it is lower or within setting range of predicat devices.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three squares, each rotated 45 degrees and nested inside the other. The text "BTL Industries" is written in a sans-serif font.
#### Operating mode
The BTL-094 device has two modes of operation: single therapy mode or sequence mode. The two modes differ based on repetition of therapy sections and settings that can be done during the running therapy. Nevertheless, both the modes have a character of continual therapy.
#### Maximum penetration depth
The BTL-094 device is equipped with two different coupling pad sizes, which allow it to reach up to 35 mm depth. This penetration depth is comparable within setting range of predicate devices.
#### Pulse repeat rate (1/s)
BTL-094 device differs from one predicate device in the value of the maximum frequency, where the frequency value of the BTL-094 device can be up to 4Hz higher than the predicate device, but has available the same frequency range as its second predicate device.
#### Number of pulses (min and max)
BTL-094 device allows to set a variable number of pulses according to the user's needs. This parameter is similar as for the predicate device.
#### Type of acoustic wave generation
The subject device provides the therapy by the application of extracorporeally induced acoustic pressure waves that are generated by means of piezoelectric principle. The device is designed with planar piezoelement and acoustic lens quiding the waves. The generated wave is coupled to biological tissue via the coupling gel pads of the applicator.
This technology creates the same type of acoustic pressure wave with comparable parameters therefore it reach the same mechanism of action as the predicate devices.
#### Positive peak pressure amplitude (MPa)
A maximum driving power of the BTL-094 has been limited at a maximum power level of 30%. The device with this limitation in maximal output power fits with output parameters within or bellow setting range of predicat devices and is not significantly different from the predicate devices.
#### Negative peak pressure amplitude (MPa)
A maximum driving power of the BTL-094 has been limited at a maximum power level of 30%. The device with this limitation in maximal output power fits with output parameters within or bellow setting range of predicat devices and is not significantly different from the predicate devices.
#### Derived focal acoustic pulse energy (mJ)
A maximum driving power of the BTL-094 has been limited at a maximum power level of 30%. The device with this limitation in maximal output power fits with output parameters within or bellow setting range of predicat devices and is not significantly different from the predicate devices.
#### Derived pulse- intensity integrated over total temporal integration limits (m.J/mm2)
A maximum driving power of the BTL-094 has been limited at a maximum power level of 30%. The device with this limitation in maximal output power parameters within or bellow setting range of predicat devices and is not significantly different from the predicate devices.
#### Rise time (ns)
BTL-094 has an alternative type of technology from the predicate device and therefore the BTL-094 device is able to reach a faster rise time and shorter compressional pulse duration.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue interlocking squares on the left, with the letters "BTL" inside the squares. To the right of the squares is the text "BTL Industries" in a gray sans-serif font. The logo is simple and modern, and the use of blue and gray gives it a professional look.
#### Compressional pulse duration (µs)
BTL-094 has an alternative type of technology from the predicate device and therefore the BTL-094 device is able to reach a faster rise time and shorter compressional pulse duration.
## Substantial Equivalence
The BTL-094 is substantially equivalent to the predicate devices D-Actor (K173692) and DolorClast® Radial (K220538). The subject device is safe and effective for its intended use.
Any differences between the predicate devices and BTL-094 have no significant influence on the safety or effectiveness of the subject device. Therefore, the BTL-094 is substantially equivalent to the predicate devices.
## Conclusion
Based upon the intended use and known technical information provided in this pre-market notification, the BTL-094 device has been shown to be substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.