K221818 · Morpheus AG · HAE · Oct 7, 2022 · Neurology
Device Facts
Record ID
K221818
Device Name
Rebellion; Phantom Multi-Bite Kerrison Rongeur
Applicant
Morpheus AG
Product Code
HAE · Neurology
Decision Date
Oct 7, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4840
Device Class
Class 2
Indications for Use
The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.
Device Story
Manual bone punch (Kerrison rongeur) used in orthopedic, neuro, and spine surgery; operated by surgeons. Device features integrated Luer-Lock connection for NaCl 0.9% rinsing solution to prevent tube clogging; posterior tube connects to bone/tissue trap or standard suction system. Removes bone/tissue via manual biting action; suction facilitates debris removal. Single-use, sterile device; 160 mm length; 2-4 mm bite sizes; 90° up-biting angulation. Benefits include efficient bone removal and reduced clogging via integrated irrigation/suction.
Clinical Evidence
No clinical performance testing submitted. Substantial equivalence is based on non-clinical performance data and comparison to the predicate device.
Technological Characteristics
Manual rongeur; stainless steel (420, 304, 302) and polymers (PVC, ABS, PA6 GF60, TPE, fluoropolymer). Features integrated Luer-Lock for irrigation and suction port for bone/tissue trap. 160 mm shaft length; 14 mm jaw opening; 2-4 mm bite size; 90° up-biting. Sterile, single-use.
Indications for Use
Indicated for cutting and removing bone, vertebral bodies, and tissue in orthopedic, neuro, and spine surgery involving the skull or spinal column.
Regulatory Classification
Identification
A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2022
Morpheus AG Timo Rack, CEO Bahnhofstraße 18 Spaichingen, Baden Würrtemberg 78549 Germany
Re: K221818
Trade/Device Name: Rebellion; Phantom Multi-Bite Kerrison Rongeur Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: August 30, 2022 Received: September 8, 2022
Dear Timo Rack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221818
Device Name Rebellion; Phantom Multi-Bite Kerrison Rongeur
Indications for Use (Describe)
The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Morpheus AG Rebellion 510(k) Premarket Notification
Image /page/3/Picture/2 description: The image is a logo for Morpheus AG, a company that specializes in inspiring medical technology. The logo features a stylized head with a pixelated effect on the right side, followed by the company name "MORPHEUS AG" in bold, sans-serif font. Below the company name, the tagline "INSPIRING MEDICAL TECHNOLOGY" is written in a smaller, sans-serif font.
#### DATE OF APPLICATION: 10/07/2022
APPLICANT:
Morpheus AG Bahnhofstrasse 18 78549 Spaichingen Germany Tel: + 49 (0) 7461 70382-0 E-Mail: info@morpheus.com.ag
CONTACT PERSON:
Timo Rack CEO Tel.: + 49 (0) 7461 70382-11 E-Mail: t.rack@morpheus.com.ag
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Morpheus AG. The logo consists of a stylized head in profile on the left, with a series of squares emanating from the back of the head. To the right of the head is the text "MORPHEUS AG" in bold, followed by the text "INSPIRING MEDICAL TECHNOLOGY" in a smaller font.
#### 1 Device Name
| Trade Name: | Rebellion;<br>Phantom Multi-Bite Kerrison Rongeur |
|-----------------------------|---------------------------------------------------|
| Common Name: | Kerrison Rongeur |
| Device Classification Name: | Manual rongeur |
#### Classification / Product Code 2
Rebellion can be classified according to following device name and product code:
| Device | Regulation<br>Description | Regulation<br>Medical Specialty | Product Code | Regulation<br>Number | Device<br>Classification |
|-----------|---------------------------|---------------------------------|--------------|----------------------|--------------------------|
| Rebellion | Manual rongeur | Neurology | HAE | 21 CFR<br>882.4840 | II |
#### Predicate Device / Reference Device 3
| Subject Device | Predicate Device | Reference Device | 510(k) number | 510(k) Holder |
|----------------|------------------------------------------------------|------------------|---------------|--------------------------------------|
| Rebellion | Rebellion;<br>Phantom Multi-Bite<br>Kerrison Rongeur | -- | K200768 | Morpheus AG |
| | -- | Fehling-punches | K153243 | Fehling Instruments<br>GmbH & Co. KG |
#### 4 Device Description
The Rebellion is a Bone Punch and it is available in the length 160 mm with tree different tip sizes (2, 3 and 3 mm). The Rebellion can be connected to a rinsing solution (NaCl 0,9%) via integrated Luer-Lock connection. The tubes are flushed with the rinsing solution, so a clogging of the tube is prevented. At posterior tube the Rebellion can be connected to the Morpheus Bone and Tissue Trap or to a standard suction system via sterile tubing.
Image /page/4/Picture/12 description: The image shows a tool that is labeled as "Figure 1: Rebellion". The tool has a long, white barrel with a black handle and trigger. There is also a gray cable coming out of the handle.
Figure 1: Rebellion
{5}------------------------------------------------
Morpheus AG Rebellion 510(k) Premarket Notification
Image /page/5/Picture/2 description: The image shows the logo for Morpheus AG, a company focused on inspiring medical technology. The logo features a stylized silhouette of a human head in profile, with a pixelated or fragmented design emanating from the back of the head. To the right of the head is the company name, "MORPHEUS AG," in bold, sans-serif font, with the tagline "INSPIRING MEDICAL TECHNOLOGY" underneath in a smaller font size.
#### Intended Use 5
The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.
#### Technological Characteristics e
Morpheus Rebellion possesses similar technological characteristics as compared to the predicate device K200768 and reference device K153243. Different characteristics do not raise different questions of safety and effectiveness, and scientific methods were applied to evaluate different characteristics' effects on safety and effectiveness.
| Description | Subject Device | Predicate Device<br>Morpheus AG<br>(Rebellion) | Reference Device<br>Fehling Instruments GmbH<br>& Co. KG<br>(Fehling-punches) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | -- | K200768 | K153243 |
| Product Code | HAE | HAE | HAE |
| Class | II | II | II |
| Regulation # | 882.4840 | 882.4840 | 882.4840 |
| Classification<br>Name | Manual, Rongeur | Manual, Rongeur | Manual, Rongeur |
| Indication for Use | The Rebellion is indicated to cut<br>and remove bone, vertebral<br>body and tissue in the<br>orthopedic, neuro and spine<br>surgery involving the skull or<br>spinal column. | The Rebellion is indicated to<br>cut and remove bone,<br>vertebral body and tissue in<br>the orthopedic, neuro and<br>spine surgery involving the<br>skull or spinal column. | Fehling rongeurs (bone<br>punches) are manually<br>operated instruments<br>indicated for cutting or<br>biting bone during surgery<br>involving the skull or<br>spinal column. |
| Anatomical<br>location | Skull and spine | Skull and spine | Skull and spine |
| Sterility | Sterile | Sterile | Non-Sterile |
| Re-Use | No | No | Yes |
| Material | Stainless Steel: 420, 304 and<br>302<br>Polymer: PVC, ABS, PA6 GF60<br>(Grivory GV-6H); TPE<br>(TM5MED) and Fluoropolymer<br>(Altera MT1000A) | Stainless Steel: 420, 304 and<br>302<br>Polymer: PVC, ABS, PA6 GF60<br>(Grivory GV-6H); TPE<br>(TM5MED) and Fluoropolymer<br>(Altera MT1000A) | 420 and 304 Stainless<br>Steels;<br>Coating: CERAMO®<br>(TiAIN)<br>Polymer: Silicone |
| Patient<br>Contacting | Stainless Steel: 420, 304<br>Polymer: Fluoropolymer (Altera<br>MT1000A) | Stainless Steel: 420, 304<br>Polymer: Fluoropolymer<br>(Altera MT1000A) | 420 and 304 Stainless<br>Steels; CERAMO® (TiAIN) |
| Design features | Manual Rongeur | Manual Rongeur | Manual Rongeur |
| Bone Removal | Suctional function | Suctional function | Manual |
| Shaft-length | 160 mm | 200 - 250 mm | 110 - 400 mm |
| Jaw opening | 14 mm | 14 mm | 9 - 19 mm |
| Bite size | 2 - 4 mm | 2 - 4 mm | 0.8 - 8 mm |
| Description | Subject Device | Predicate Device<br>Morpheus AG<br>(Rebellion) | Reference Device<br>Fehling Instruments GmbH<br>& Co. KG<br>(Fehling-punches) |
| Cutting<br>angulation | 90° up-biting | 90° up-biting | 40° and 90° up/down bi-<br>ting |
## 6.1 Device Characteristics Table
{6}------------------------------------------------
## Morpheus AG Rebellion 510(k) Premarket Notification
Image /page/6/Picture/2 description: The image shows the logo for Morpheus AG, a company focused on inspiring medical technology. The logo features a stylized silhouette of a human head with a pixelated effect emanating from the back of the head. To the right of the head is the company name "MORPHEUS AG" in bold, uppercase letters, with the tagline "INSPIRING MEDICAL TECHNOLOGY" underneath in a smaller font.
#### Performance Data 7
# 7.1 Clinical performance testing
Clinical performance testing was not submitted in this 510(k).
# 7.2 Non-Clinical performance testing
The subject device does not represent a new worst-case when compared to the previously cleared Morpheus Rebellion Rongeur (K200768).
#### Conclusion 8
In accordance with 21 CFR Part 807 and based on the information provided in this premarket notification, Morpheus AG concludes that the Rebellion Rongeur is as safe and as effective for their intended use as the predicate device, the Rebellion Rongeur (K200768).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.