K221750 · ABBOTT MEDICAL · DQY · Jun 30, 2022 · Cardiovascular
Device Facts
Record ID
K221750
Device Name
CPS AIM Universal II Slittable Inner Catheter
Applicant
ABBOTT MEDICAL
Product Code
DQY · Cardiovascular
Decision Date
Jun 30, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system.
Device Story
The CPS Aim™ Universal II is a slittable inner catheter used as an introducer for left heart lead delivery. It functions as a conduit for contrast medium and medical devices within the coronary venous system. The device features a braid-reinforced PEBAX shaft, PTFE-lined inner diameter, and a hydrophilic-coated exterior for lubricity. It includes a proximal hub with a hemostasis valve, sideport, and 3-way stopcock. Used in clinical settings by physicians, it is guided fluoroscopically via an embedded marker band on the distal tip. The device is a line extension of the predicate, featuring a modified hydrophilic coating and a paired-wire braid design for improved kink resistance. It is supplied sterile for single-procedure use.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included biocompatibility (ISO 10993-1), design verification (physical/dimensional characteristics, functional testing, particulate testing, coating lubricity/durability), and 24-month accelerated aging shelf-life testing.
Technological Characteristics
Braid-reinforced PEBAX shaft; PTFE-lined inner diameter; hydrophilic-coated exterior; molded proximal hub with hemostasis valve, sideport, and 3-way stopcock; embedded radiopaque marker band. Sterilized via ethylene oxide. Dimensions: 59 cm and 65 cm lengths. Paired-wire braid design for kink resistance.
Indications for Use
Indicated for intracardiac access of the coronary sinus and subselection of the venous system of the heart to facilitate delivery of contrast medium and implantable devices (e.g., guidewires, left heart leads). Not studied in pediatric patients.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, while the text on the right is written in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
June 30, 2022
Abbott Medical Indira Verdekel Regulatory Affairs Project Manager 15900 Valley View Court Sylmar, California 91342
Re: K221750
Trade/Device Name: CPS AIM Universal II Slittable Inner Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: June 15, 2022 Received: June 16, 2022
Dear Ms. Verdekel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K221750
Device Name CPS AIM™ Universal II Slittable Inner Catheter
Indications for Use (Describe)
The CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
### Date Prepared
June 15, 2022
### Submitter
Abbott Medical (formerly St. Jude Medical) 15900 Valley View Court Sylmar, CA 91342 USA Phone: (818) 362-6822
Establishment Registration: 2017865
#### Contact Person
Indira Verdekel Regulatory Affairs Project Manager Phone: (818) 493-3360 Indira.Valdivia@abbott.com
Natasha Loghmanpour Manager Regulatory Affairs Phone: (818) 493-3015 Natasha.Loghmanpour(@abbott.com
#### Device
| Name of Device: | CPS Aim™ Universal II Slittable Inner Catheter |
|-----------------------|------------------------------------------------------------------------------------------------|
| Model Numbers: | DS2N100-59, DS2N100-65, DS2N101-59, DS2N101-65, DS2N102-59, DS2N102-65, DS2N103-65, DS2N104-65 |
| Common or Usual Name: | Percutaneous catheter |
| Classification Name: | Catheter Introducer (21 CFR 870.1250) |
| Regulatory Class: | Class II |
| Code: | DQY |
#### Predicate Device
The line extension. CPS Aim™ Universal II Slittable Inner Catheter, is substantially equivalent in intended use and method of operation to the following predicate manufactured by Abbott Medical.
- CPS Aim™ Universal Slittable Inner Catheter (K130252, cleared on March 20, 2013) -
{4}------------------------------------------------
### Device Description
The CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) is used to facilitate left heart lead delivery procedures. It is an introducer that is used to cannulate the coronary venous system and act as a conduit for contrast medium, implantable coronary leads, or other devices.
The CPS Aim™ Universal II Slittable Inner Catheter design is based on the currently marketed CPS Aim™ Universal Slittable Inner Catheter design and components. Three modifications were made:
- 1. Modified hydrophilic coating: the subject device uses a different hydrophilic coating material as compared to the currently marketed CPS Aim Universal catheters.
- 2. Use of a modified braid design pattern to provide improved kink resistance: the predicate device braid is a single braided wire, and the subject braid contains paired braided wires.
- 3. Reduction in shelf-life from three (3) years to two (2) years.
The CPS Aim™ Universal II Slittable Inner Catheter will be added as a line extension to the currently marketed family of CPS catheters. New. unique model numbers will distinguish the CPS Aim™ Universal II inner catheter models from existing CPS catheter models.
The CPS Aim™ Universal II Slittable Inner Catheters will be available in the same working lengths, 59 cm and 65 cm, as the predicate device, and will be available with the same shape of curves as the predicate device. The key design features of the CPS Aim™ Universal II Slittable Inner Catheters have only undergone a minimal change to support the coating and braid change. There is no change to the accessories, including the slitter and valve bypass tool (VBT).
The key design features CPS Aim Universal and CPS Aim Universal II Slittable Inner Catheters and their accessory, the CPS Valve Bypass Tools include:
- Braid reinforced, varying durometer PEBAX catheter shaft with molded proximal hub.
- . Inner diameter of the catheter is PTFE lined.
- . Atraumatic distal soft tip.
- Embedded marker band on soft tip for fluoroscopic visibility. ●
- . Outside surface of the catheter shaft is coated with a hydrophilic coating to provide lubricity during use.
- . Hub of the catheter is fitted with a retention cap and an integrated valve that provides hemostasis and facilitates contrast injection. The hub also includes a sideport assembly with 3-way stopcock.
- The catheters are available in cannulator and sub-selector models. .
- Accessories such as valve bypass tool (VBT) used to assist the insertion of Abbott Medical . devices (leads, guidewires, inner catheters, etc.).
{5}------------------------------------------------
Similar to the existing CPS catheters, the CPS Aim™ Universal II Slittable Inner Catheters are supplied sterile (via ethylene oxide) in one package and are intended for single procedure.
There is no change in accessory product or process for the different models as a result of this line extension. The proposed indications for use is consistent with the cleared indications.
### Pediatric Use
The CPS Aim™ Universal II Slittable Inner Catheter products have not been studied in pediatric patients.
### Indications for Use
The are no changes to the Indications for Use as a result of this submission. The Indications for Use for the CPS Aim™ Universal II Slittable Inner Catheter is as follows:
The CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) can work with outer guide catheters as a system.
# Comparison of Technological Characteristics with Predicate Device
The CPS Aim™ Universal II Slittable Inner Catheter has the same technological characteristics as the currently marketed CPS Aim™ Universal Slittable Inner Catheter with minor changes to the hydrophilic coating and braid. Where differences exist between the subject device and the predicate device, performance testing demonstrated that the CPS Aim™ Universal II Slittable Inner Catheter performs in a substantially equivalent manner to the currently marketed predicate device.
# Performance Data
The CPS Aim™ Universal II Slittable Inner Catheter is substantially equivalent to the predicate devices based on comparisons of the intended use, device functionality, and technological characteristics. The following performance testing was conducted to demonstrate that the device meets its design specifications and is as safe and effective as the predicate devices:
{6}------------------------------------------------
### Biocompatibility Testing
The biocompatibility evaluation for the CPS Aim™ Universal II Slittable Inner Catheter was conducted in accordance with ISO 10993-1 for biological evaluation and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued September 4, 2020. For biological testing in animal models, testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulations 21 CFR 58. Based on the ISO and FDA guidelines, the nature of patient contact for the CPS Aim™ Universal II Inner Catheter are externally communicating device with limited (224 hours) circulating blood contact. The Valve Bypass Tools are externally communicating, devices with limited (≤24 hour) indirect blood contact.
The testing included the following:
- Cytotoxicity ●
- Sensitization, Irritation ●
- Materials-Mediated Pyrogenicity
- Acute Systemic Toxicity
- Hemocompatibility (Direct Contact and Extract)
- Complement Activation Assay - SC5b-9
- Partial Thromboplastin Time
- . Platelet & Leukocyte Counts
- Particulate Matter per USP <788> ●
- Surface Characterization ●
#### Bench Testing
- Design Verification ●
- Physical and Dimensional Characteristics o
- Functional Characteristics o
- Particulate Testing o
- Hydrophilic Coating Lubricity and Durability O
- Shelf-Life Testing ●
- o Accelerated Aging 24-month (device)
#### Conclusions
The resulting evidence obtained from the design verification testing and the risk analysis demonstrated that the subject device is as safe and effective as the predicate device, the CPS Aim™ Universal Slittable Inner Catheter (K130252).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.