← Product Code [NKB](/productcode/NKB) · K221745

# Sterile Posterior Spinal Fixation System (K221745)

_Shanghai Reach Medical Instrument Co, Ltd. · NKB · Aug 31, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221745

## Device Facts

- **Applicant:** Shanghai Reach Medical Instrument Co, Ltd.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Aug 31, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Sterile Posterior Spinal Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

## Device Story

Single-use posterior spinal fixation system; consists of pedicle screws, rods, and interconnecting devices; provides stabilization and fusion of two or more spinal segments; used in open (RS8 LEGEND) or minimally invasive (RS8 LONG) surgical procedures; implanted by surgeons; manufactured from titanium alloy; provided sterile via gamma sterilization; implants removed after stabilization function complete; subject device is a design change from predicate K201737, specifically transitioning from non-sterile to sterile-provided implants.

## Clinical Evidence

No clinical data provided. Bench testing only, including validation of sterility (ISO 11137-1:2006, AAMI TIR29:2012), biocompatibility, pyrogenicity, package integrity, and shelf-life.

## Technological Characteristics

Titanium alloy implants; pedicle screws, rods, and interconnecting devices; gamma sterilization (SAL 10^-6); validated per ISO 11137-1:2006 and AAMI TIR29:2012; single-use; non-active; mechanical fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Posterior Spinal Fixation System ([K201737](/device/K201737.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 31, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shanghai REACH Medical Instrument Co., Ltd. % Huifang Zhao Consultant Sinow Medical AS Hoyteknologisenteret, Thormohlens gate 55 Bergen, 5006 Norway

Re: K221745

Trade/Device Name: Sterile Posterior Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: August 19, 2022 Received: August 24, 2022

Dear Huifang Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K221745

#### Device Name

Sterile Posterior Spinal Fixation System

#### Indications for Use (Describe)

The Sterile Posterior Spinal Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| X   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K221745

#### 1. Contact Information

#### 1.1. Applicant

Date prepared: Aug.19,2022 Applicant: Shanghai REACH Medical Instrument Co., Ltd Address: Building 13, No.999 Jiangyue Road, Minhang District, 201114 Shanghai, PEOPLE'S REPUBLIC OF CHINA Contact Person: Mingsha Ye Title: Registration Manager Telephone: (86) 021-54840582 E-mail: yems(@reach-med.com 1.2. Consultant Company: Sinow Medical AS Address: Høyteknologisenteret, Thormøhlens gate 55, 5006 Bergen, Norway Contact Person: Huifang Zhao Telephone: +47 41412078 Email: huifang.zhao@sinowmedical.no

## 2. Device information

Trade Name: Sterile Posterior Spinal Fixation System Common Name: Thoracolumbosacral Pedicle Screw System Regulation Number: 21 CFR 888.3070 Classification Name: Thoracolumbosacral Pedicle Screw System Classification: II Product Code: NKB Review Panel: Orthopedic

## 3. Legally Marketed Primary Predicate Device

Posterior Spinal Fixation System (K201737)

#### 4. Indications for use

The Sterile Posterior Spinal Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and

{4}------------------------------------------------

radiographic studies); Spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

#### 5. Device Description

The Sterile Posterior Spinal Fixation System is a single-use system for surgical fixation of the spine. It is made up of pedicle screws, rods, and interconnecting devices. The implanted parts confer stabilization and fusion of two or more spinal segments and are to be removed once their stabilizing function is no longer required. The implantable devices are manufactured from titanium alloy and are provided sterile.

The Sterile Posterior Spinal Fixation System includes the following sets:

- RS8 LEGEND (designed for open surgery)
- RS8 LONG (designed for minimally invasive surgery)

Each set comprises of different rods, screws, set screws, and accessory parts, including various lengths and diameters of reduction screws. Both sets are designed for internal posterior thoracolumbar fixation of the spine. Patient diagnosis and individual conditions should be taken into consideration when selecting the surgical option.

Surgical instruments are provided with the subject device and predicate device.

The subject of this submission is a design change to Posterior Spinal Fixation System(K201737). The Posterior Spinal Fixation System(K201737) is provided non-sterile and the subject device is provided sterile.

#### 6. Substantially Equivalent (SE) Comparison

The subject device has the same indications for use/intended use, design, and materials as the primary predicate device. The only change proposed in this submission is the method used to sterilize the subject device.

## 7. Description of the change

The modification has been made only to sterilization method of the implants. The predicate device provides non-sterile implants and surgical instruments which must be cleaned and sterilized by the end user prior to use. For the subject device the implants are sterilized with gamma sterilization by manufacture and the sterilization process was validated according to ISO 11137-1:2006 and AAMI TIR29:2012. The implants will provide a sterility assurance level (SAL) of 10-6.

{5}------------------------------------------------

## 8. Performance data

The evaluation of biocompatibility, sterility, pyrogenicity, package integrity, and shelf-life were performed in support of the proposed change.

No additional pre-clinical or clinical data is required to support the substantial equivalence determination for this change.

## 9. Conclusion

The information presented in this 510(k) submission establishes that the subject device is substantially equivalent to the primary predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K221745](https://fda-staging.innolitics.com/device/K221745)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).