This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures. It is designed for use by a trained professional in the field of general dentistry.
Device Story
Air-powered dental handpiece; connects to dental unit via coupling; receives compressed air to drive high-speed turbine; receives cooling water and illumination light. Used by trained dental professionals in general dentistry clinics. Operator controls rotation speed via air flow adjustment. Device facilitates tooth preparation and surgical procedures; benefits patient through efficient removal of carious material and restoration of tooth surfaces. Reusable; sterilized via prevacuum steam autoclave.
Clinical Evidence
No clinical performance testing was performed. Substantial equivalence is supported by non-clinical bench testing verifying compliance with ISO 14457 (dentistry handpieces and motors) and ISO 9168 (hose connectors). Sterilization validated per ISO 17665-1; cleaning validation per AAMI TIR30/TIR12.
Technological Characteristics
Air-driven dental handpiece; materials include SUS303F stainless steel (hard chrome coated). Operates via compressed air (approx. 43.5 psi); high-speed turbine (360,000-440,000 rpm). Features water cooling port and illumination. Complies with ISO 14457 and ISO 9168. Sterilization via prevacuum steam autoclave.
Indications for Use
Indicated for removal of carious material, cavities, crown preparations, filings, tooth processing, surface restoration, and third molar removal in general dentistry patients.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
February 22, 2023
Saeshin Precision Co., Ltd. % Sanghwa Myung Regulatory Affair Consultant E&M D-1474. 230. Simin-daero, Dongan-gu Anyang, Gyeonggi-do 14067 South Korea
Re: K221741
Trade/Device Name: TRAUS Air Dental Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: November 18, 2022 Received: November 25, 2022
Dear Sanghwa Myung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221741
Device Name TRAUS Air Dental Handpiece
Indications for Use (Describe)
This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures. It is designed for use by a trained professional in the field of general dentistry.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">X</span> |
|----------------------------------------------|---------------------------------------------------|
| Over The Counter Use (21 CFR 801 Subpart C) | <span>☐</span> |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for SAESHIN. The logo consists of a blue geometric shape resembling a stylized cube or a series of interconnected squares. Below the shape, the word "SAESHIN" is written in a sans-serif font, also in blue. The overall design is clean and modern.
# 510(k) Summary
# K221741
Submitter: Saeshin Precision Co., Ltd. # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Republic of Korea Telephone: +82 53 587 2345 Fax: +82 53 580 0939 Email: ksqc@saeshin.com
Contact: E&M Requlatory Affair/Sang Hwa Myunq Telephone: +80-70-7807-0550 FAX: +82-31-388-9263 Cellphone: +82-10-4952-6638 E-mail: mshenmc@gmail.com Primary Contact: Sanghwa Myung
#### Date 510(k) summary prepared: February 21, 2023
Common Name: Handpiece, Air-Powered, Dental Trade Name: TRAUS Air Dental Handpiece Classification: I Product Code: EFB Classification Panel: Dental Requlation Numbers: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories
Description of Device:
The TRAUS Air Dental handpieces are air driven dental handpieces for the use by a trained dental professional. The devices are air-powered handpieces that are reusable and designed accordance with international standard. The devices can be sterilized by the prevacuum steam autoclave methods that have been validated. Through the coupling connected to a dental unit, the proposed dental handpiece receive air for functionality of the high-speed turbine. They also receive cooling water for cutting through one port and light for illumination through another port.
Using in combination with dental unit chair, it is possible to use by installing the instrument at air handpiece according to purpose. It is possible to change the rotation speed by adjusting the flowing air.
Indication for use:
This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures.
It is designed for use by a trained professional in the field of general dentistry.
{4}------------------------------------------------
#### Predicate Device:
Manufacturer: KaVo do Brasil Industria e Comercio Ltda
510(k) Number: K141576 Trade Name: Maxima PRO 45L Common Name: Handpiece, Air-Powered, Dental Regulation Name: Dental handpiece and accessories Regulation Numbers: 21 CFR 872.4200 Product Code: EFB Classification: Class I
#### Substantial Equivalence:
Comparison table is as follows.
#### Table 1: Substantial equivalence comparison 1) Predicate Device
| 1) Predicate Device<br>Contents | Subject Device | Predicate Device |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)Number | Not available | K141576 |
| Trade Name | TRAUS Air Dental Handpiece | Maxima PRO 45L |
| Indication for Use | This air-powered dental<br>handpiece is intended for<br>removal of carious material,<br>cavities and crown<br>preparations, removal of filings,<br>processing of tooth, restoration<br>of surfaces and as a surgical<br>tool for third molar removal<br>procedures.<br>It is designed for use by a<br>trained professional in the field<br>of general dentistry. | This air-powered dental<br>handpiece is intended for<br>removal of carious<br>material, cavities and crow<br>preparations, removal of<br>filings, processing of tooth,<br>restoration of surfaces and<br>as a surgical tool for third<br>molar removal procedures.<br>It is designed for use by a<br>trained professional in the<br>field of general dentistry. |
| Operation principal | Through the tube and the<br>coupling connected to a<br>dental unit, the airpowered<br>handpiece<br>receives the air for<br>operating the high speed<br>turbine, the cooling water<br>for cutting treatment<br>through one port and light<br>for illumination the<br>operation area. | Through the tube and the<br>coupling connected to a<br>dental unit, the airpowered<br>handpiece<br>receives the air for<br>operating the high speed<br>turbine, the cooling water<br>for cutting treatment<br>through one port and light<br>for illumination the<br>operation area. |
| Head Size | ATN400: Ø11.2 x13.55mm<br>CAB10LK: Ø11.2 x13.7m<br>CAB10LN: Ø11.2 x14.2m<br>CAB20ND M4: Ø11.2 x13.55m<br>CAB20ND B2: Ø11.2 x13.55m<br>ISO1797-2017<br>4.Classification | Head size-Height: 14,6<br>mm<br>Head size-Diameter: 12,5<br>mm |
| Type of chuck | -Type 3: FG(Friction Grip)<br>CAB10LN/CAB10LK/CAN10NN<br>/CAB20ND M4 : 4hole | Push Button |
| Water ports | CAB20ND B2: 1hole | 1 hole |
| Coupling Dimensions | Length with coupling: | Length with coupling: |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a blue cube-like shape on top of the word "SAESHIN" in blue block letters. The cube is made up of several smaller cubes, and the word "SAESHIN" is in a sans-serif font.
| | TRAUS CAN10NN + TRAUS<br>CP10NQ<br>Approx. 129.59 mm<br>TRAUS CAB10LN + TRAUS<br>CP10LN<br>Approx. 143.53 mm<br>TRAUS CAB10LK + TRAUS<br>CP10LK<br>Approx. 138.97 mm | Approx. 121 mm |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Chemical composition of<br>the patient- contacting<br>portions of the device | 303F | 303F |
| Light Intensity | Approximately 7,000 LUX | Approximately 25,000 LUX |
| Bur retention force | up to 24 Ncm | up to 24 Ncm |
| Operating Pressure | 43.5 psi (0.3Mpa)<br>recommended | 40 ± 1.45 psi<br>recommended |
| Idling Speed | 360,000 - 440,000 rpm | 380,000 - 420,000 rpm |
| Head angle | 90-degree | 45-degree |
| Compliance to Standards | Complies with ISO 14457<br>and ISO 9168 | Complies with ISO 14457<br>and ISO 9168 |
| Lubricant | Pana Plus Spray | KaVo Spray<br>Henry Schein Spray &<br>Clean |
Biocompatibility: Materials tested in accordance with 10993-1 are used and materials that have been confirmed to be biologically safe are used.
Patient contact parts of the TRAUS Air Dental Handpiece are as follows.
| Parts | Raw Material | Surface Treatment | Patient contact | Duration of contact by ISO 10993-1 |
|-------|--------------|--------------------|-----------------|------------------------------------|
| Head | SUS303F | Hard Chrome coated | Yes | Limited<br>(<24 hours) |
- 1) ISO 10993-1: 2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- 2) ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- 3) ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
- 4) ISO 10993-12:2012 Biological evaluation of medical devices Part 12: Sample preparation and reference materials
{6}------------------------------------------------
### Non-clinical Performance Data:
- . The bench tests were performed in order to ensure the performance of the TRAUS Dental Air Handpieces (ATN400,CAB10LK,CAB10LN,CAB20ND M4,CAB20ND B2) verify conformity to ISO 14457 and demonstrate substantial equivalence to the predicates. ATN400,CAB10LK,CAB10LN,CAB20ND M4,CAB20ND B2 samples were compliant with ISO 14457: 2017 Dentistry -Handpieces and Motors and demonstrated substantial equivalence to the predicates.
- . Sterilization has been validated in conformance to the FDA recognized consensus standard ISO 17665-1:2006 Sterilization of health care products moist heat - Part 1: requirements for the development, validation and routine control of a sterilization process for medical devices. Cleaning validation testing is performed in accordance with recommended evaluations as listed in AAMI TIR30, AAMI TIR12, and Guidance for Industry and FDA Staff -Processing/Reprocessing Medical Devices in Health Care Settings.
- . ISO 9168:2009 Dentistry— Hose connectors for air driven dental handpieces, device's hose connection has been verified by standard of ISO 9168.
Clinical Data: No clinical performance testing was performed.
## Conclusion
The performance of TRAUS Air Dental Handpiece meets the requirements of the non-clinical bench testing conducted to support substantial equivalence. Based on the available information, the subject device and the predicates are similar indication for use, operational principal, performance data.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that we conclude that substantially equivalent with predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.