Aviator® The Aviator® Anterior Cervical Plating System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: • Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) • Trauma (including fractures) • Tumors • Deformities or curvatures (including kyphosis, lordosis, or scoliosis) • Pseudarthrosis • Failed previous fusion • Decompression of the spinal cord following total or partial cervical vertebrectomy • Spondylolisthesis • Spinal stenosis LITe® Plate System The LITe® Plate System Universal, Sacral, 2 Screw and 4 Screw Plates are indicated for use via a lateral or anterolateral surgical approach above the bifurcation of the great vessels in the treatment of the thoracic and thoracolumbar (T1-L5) spine or via an anterior approach below the bifurcation of the great vessels in the treatment of lumbar and lumbosacral (L1-S1) spine. The system is intended to provide additional support during fusion in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: • Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); • Pseudoarthrosis; • Spondylolysis; • Spondylolisthesis; • Spinal stenosis; • Tumors; • Trauma (i.e. Fractures or Dislocation); • Deformities (i.e. Scoliosis, Kyphosis or Lordosis); • Failed Previous Fusion The LITe® Plate System Buttress Plate is intended to stabilize the allograft or autograft at one level (T1-S1) as an aid to spinal fusion and to provide temporary stabilization and augment development of a solid spinal fusion. It may be used alone or with other anterior, anterolateral, or posterior spinal systems made of compatible materials. This device is not intended for load bearing applications. DynaTran™ The DynaTran™ Anterior Cervical Plating System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: • Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) • Trauma (including fractures) • Tumors • Deformities or curvatures (including kyphosis, lordosis, or scoliosis) • Pseudoarthrosis • Failed previous fusion • Decompression of the spinal cord following total or partial cervical vertebrectomy • Spondylolisthesis • Spinal stenosis Reflex™ Hybrid The Reflex™ Hybrid ACP System is intended for anterior intervertebral screw fixation of the cervical spine from C2 – T1. These systems are indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: • Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) • Decompression of the spinal cord following total or partial cervical vertebrectomv • Trauma (including fractures) • Tumors • Deformities or curvatures (including kyphosis, lordosis or scoliosis) • Pseudoarthrosis • Failed previous fusions • Spondylolisthesis • Spinal stenosis UniVise™ Spinous Process Fixation Plate The UniVise™ Spinous Process Fixation Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The UniVise™ Spinous Process Fixation Plate is intended for use with bone graft and not intended for stand-alone use.
Device Story
Spinal plate systems (Aviator, LITe, DynaTran, Reflex, UniVise) provide temporary stabilization/supplemental fixation during spinal fusion; used in cervical, thoracic, thoracolumbar, and lumbosacral regions. Implants consist of plates and screws; surgically implanted by orthopedic/neurosurgeons. Submission establishes MR Conditional labeling; no changes to device design or materials. Benefits include spinal stability during bone healing.
Clinical Evidence
Bench testing only. MR Compatibility testing performed per ASTM F2503. No clinical data presented.
Technological Characteristics
Spinal intervertebral body/interlaminal fixation orthosis. Metallic plate and screw systems. MR Conditional per ASTM F2503. No changes to materials or design from predicate devices.
Indications for Use
Indicated for skeletally mature patients requiring spinal stabilization/fusion due to degenerative disc disease, trauma (fractures/dislocations), tumors, spinal deformities (scoliosis, kyphosis, lordosis), pseudarthrosis, failed previous fusion, spondylolisthesis, spondylolysis, or spinal stenosis. Specific levels: C2-T1 (cervical systems), T1-L5/L1-S1 (thoracic/lumbar systems). Contraindications: Not for load-bearing (Buttress plate) or stand-alone use (UniVise).
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
CAYMAN Thoracolumbar and Buttress Plate Systems (K182473)
Submission Summary (Full Text)
{0}------------------------------------------------
October 19, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Stryker Corporation Katie McNeil Senior Regulatory Affairs Specialist 2 Pearl Court Allendale. New Jersev 07401
# Re: K221728
Trade/Device Name: Aviator® Anterior Cervical Plating (ACP) System, LITe® Plate System, DynaTran™ Anterior Cervical Plating (ACP) System, Reflex™ Hybrid ACP System, UniVise™ Spinous Process Fixation Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: September 22, 2022 Received: September 22, 2022
Dear Katie McNeil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
## 510(k) Number (if known) K221728
#### Device Name
Aviator® Anterior Cervical Plating (ACP) System
Indications for Use (Describe)
The Aviator® Anterior Cervical Plating System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
· Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Trauma (including fractures)
- Tumors
- · Deformities or curvatures (including kyphosis, lordosis, or scoliosis)
- · Pseudarthrosis
- Failed previous fusion
- · Decompression of the spinal cord following total or partial cervical vertebrectomy
- · Spondylolisthesis
- · Spinal stenosis
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K221728
Device Name LITe® Plate System Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
## Indications for Use (Describe)
The LITe® Plate System Universal, Sacral, 2 Screw and 4 Screw Plates for use via a lateral or anterolateral surgical approach above the bifurcation of the great vessels in the thoracic and thoracolumbar (T1-L5) spine or via an anterior approach below the bifurcation of the great vessels in the treatment of lumbosacral (L1-S1) spine. The system is intended to provide additional support during fusion in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:
· Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies);
- · Pseudoarthrosis;
- · Spondylolysis:
- · Spondylolisthesis;
- · Spinal stenosis;
- Tumors:
- · Trauma (i.e. Fractures or Dislocation)
- · Deformities (i.e. Scoliosis, Kyphosis or Lordosis)
- · Failed Previous Fusion
The LITe® Plate System Buttress Plate is intended to stabilize the allograft at one level (T1-S1) as an aid to spinal fusion and to provide temporary stabilization and augment of a solid spinal fusion. It may be used alone or with other anterior, anterolateral, or posterior spinal systems made of compatible materials. This device is not intended for load bearing applications.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
### 510(k) Number (if known) K221728
#### Device Name
DynaTran™ Anterior Cervical Plating (ACP) System
Indications for Use (Describe)
The DynaTran™ Anterior Cervical Plating System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
· Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Trauma (including fractures)
- Tumors
- · Deformities or curvatures (including kyphosis, lordosis, or scoliosis)
- · Pseudarthrosis
- Failed previous fusion
- · Decompression of the spinal cord following total or partial cervical vertebrectomy
- · Spondylolisthesis
- · Spinal stenosis
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K221728
Device Name Reflex™ Hybrid ACP System
## Indications for Use (Describe)
The Reflex™ Hybrid ACP System is intended for anterior intervertebral screw fixation of the cervical spine from C2 -T1. These systems are indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
· Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Decompression of the spinal cord following total or partial cervical vertebrectomv
- · Trauma (including fractures)
- Tumors
- · Deformities or curvatures (including kyphosis, lordosis or scoliosis)
- · Pseudoarthrosis
- · Failed previous fusions
- · Spondylolisthesis
- · Spinal stenosis
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K221728
Device Name UniVise™ Spinous Process Fixation Plate
### Indications for Use (Describe)
The UniVise™ Spinous Process Fixation Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (TI-S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); trauma (i.e., fracture or dislocation); and/or tumor. The UniVise™ Spinous Process Fixation Plate is intended for use with bone graft and not intended for stand-alone use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 100%;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 100%;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
| 510(k) Summary: Stryker Spine Plate Systems | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 |
| Contact Person : | Name: Katie McNeil<br>Phone: 201-749-8518<br>Email: katie.mcneil@stryker.com |
| Date Prepared: | October 19, 2022 |
| Trade Names: | 1. Aviator® Anterior Cervical Plating (ACP) System<br>2. LITe® Plate System<br>3. DynaTran™ Anterior Cervical Plating (ACP) System<br>4. Reflex™ Hybrid ACP System<br>5. UniVise™ Spinous Process Fixation Plate |
| Common Name: | 1. Aviator®<br>Anterior Cervical Plate System<br>2. LITe® Plate System<br>Appliance, fixation, spinal intervertebral body<br>3. DynaTran™<br>Anterior Cervical Plate System<br>4. Reflex™ Hybrid<br>Anterior Cervical Plate System<br>5. UniVise™ Spinous Process Fixation Plate<br>Spinous Process Fixation Plate |
| Proposed Class: | Class II |
| Classification Name: | 1. Aviator®<br>Spinal intervertebral body fixation orthosis; 21 CFR §888.3060<br>2. LITe® Plate System<br>Spinal intervertebral body fixation orthosis; 21 CFR §888.3060<br>3. DynaTran™<br>Spinal intervertebral body fixation orthosis; 21 CFR §888.3060<br>4. Reflex™ Hybrid<br>Spinal intervertebral body fixation orthosis; 21 CFR §888.3060<br>5. UniVise™ Spinous Process Fixation Plate<br>Spinal interlaminal fixation orthosis; 21 CFR §888.3050 |
| Product Code: | 1. Aviator®<br>KWQ<br>2. LITe® Plate System<br>KWQ<br>3. DynaTran™<br>KWQ |
| 510(k) Summary: Stryker Spine Plate Systems | |
| | 4. Reflex™ Hybrid<br>KWQ |
| | 5. UniVise™ Spinous Process Fixation Plate<br>PEK |
| Predicate Devices: | Primary Predicate: Aviator® (K142237) |
| | Additional Predicates: |
| | LITe® Plate System (K150449) |
| | DynaTran™ (K083020) |
| | Reflex™ Hybrid (K062310, K063430) |
| | UniVise™ Spinous Process Fixation Plate (K132968) |
| | PYRENEES, BLUE RIDGE and OZARK Cervical Plate Systems and the CAYMAN |
| | Thoracolumbar and Buttress Plate Systems (K182473) |
| Device Description: | The previously cleared devices consist of a variety of plate systems designed |
| | to provide support across implanted levels in the cervical, thoracolumbar, |
| | and lumbosacral spine until fusion is achieved.<br>The primary purpose of this submission is to establish an MR Conditional |
| | labeling claim for these implants. |
| Intended Use: | Aviator®<br>The Aviator® Anterior Cervical Plating System is intended for anterior<br>intervertebral screw fixation of the cervical spine at levels C2-T1. The system<br>is indicated for temporary stabilization of the anterior spine during the<br>development of cervical spine fusions in patients with the following<br>indications:<br>• Degenerative Disc Disease (as defined by neck pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic<br>studies)<br>• Trauma (including fractures)<br>• Tumors<br>• Deformities or curvatures (including kyphosis, lordosis, or scoliosis)<br>• Pseudarthrosis<br>• Failed previous fusion<br>• Decompression of the spinal cord following total or partial cervical<br>vertebrectomy<br>• Spondylolisthesis<br>• Spinal stenosis |
| | LITe® Plate System<br>The LITe® Plate System Universal, Sacral, 2 Screw and 4 Screw Plates are<br>indicated for use via a lateral or anterolateral surgical approach above the<br>bifurcation of the great vessels in the treatment of the thoracic and<br>thoracolumbar (T1-L5) spine or via an anterior approach below the<br>bifurcation of the great vessels in the treatment of lumbar and lumbosacral<br>(L1-S1) spine. The system is intended to provide additional support during<br>fusion in skeletally mature patients in the treatment of the following acute<br>and chronic instabilities or deformities: |
| 510(k) Summary: Stryker Spine Plate Systems | |
| • Degenerative Disc Disease (defined as back pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic<br>studies);<br>• Pseudoarthrosis;<br>• Spondylolysis;<br>• Spondylolisthesis;<br>• Spinal stenosis;<br>• Tumors;<br>• Trauma (i.e. Fractures or Dislocation)<br>• Deformities (i.e. Scoliosis, Kyphosis or Lordosis)<br>• Failed Previous Fusion | |
| The LITe® Plate System Buttress Plate is intended to stabilize the allograft or<br>autograft at one level (T1-S1) as an aid to spinal fusion and to provide<br>temporary stabilization and augment development of a solid spinal fusion. It<br>may be used alone or with other anterior, anterolateral, or posterior spinal<br>systems made of compatible materials. This device is not intended for load<br>bearing applications. | |
| DynaTran™<br>The DynaTran™ Anterior Cervical Plating System is intended for anterior<br>intervertebral screw fixation of the cervical spine at levels C2-T1. The system<br>is indicated for temporary stabilization of the anterior spine during the<br>development of cervical spine fusions in patients with the following<br>indications: | |
| • Degenerative Disc Disease (as defined by neck pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic<br>studies)<br>• Trauma (including fractures)<br>• Tumors<br>• Deformities or curvatures (including kyphosis, lordosis, or scoliosis)<br>• Pseudoarthrosis<br>• Failed previous fusion<br>• Decompression of the spinal cord following total or partial cervical<br>vertebrectomy<br>• Spondylolisthesis<br>• Spinal stenosis | |
| Reflex™ Hybrid<br>The Reflex™ Hybrid ACP System is intended for anterior intervertebral screw<br>fixation of the cervical spine from C2 – T1. These systems are indicated for<br>temporary stabilization of the anterior spine during the development of<br>cervical spine fusions in patients with the following indications: | |
| • Degenerative disc disease (as defined by neck pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic<br>studies)<br>• Decompression of the spinal cord following total or partial cervical vertebrectomv | |
| 510(k) Summary: Stryker Spine Plate Systems | |
| | • Trauma (including fractures)<br>• Tumors<br>• Deformities or curvatures (including kyphosis, lordosis or scoliosis)<br>• Pseudoarthrosis<br>• Failed previous fusions<br>• Spondylolisthesis<br>• Spinal stenosis |
| | UniVise™ Spinous Process Fixation Plate<br>The UniVise™ Spinous Process Fixation Plate is a posterior, non-pedicle<br>supplemental fixation device, intended for use at a single level in the non-<br>cervical spine (T1-S1). It is intended for plate fixation/attachment to spinous<br>processes for the purpose of achieving supplemental fusion in the following<br>conditions: degenerative disc disease (defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by history and radiographic<br>studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or<br>tumor. The UniVise™ Spinous Process Fixation Plate is intended for use with<br>bone graft and not intended for stand-alone use. |
| Summary of the<br>Technological<br>Characteristics | The devices in this submission possess the same technological<br>characteristics as their predicate devices; no changes have been made to any<br>of the devices. Therefore, the fundamental scientific technology of the<br>subject devices is the same as previously cleared devices. |
| Summary of the<br>Performance Data | MR Compatibility testing per ASTM F2503 was performed. The test results<br>demonstrate that the subject devices performance met the prescribed<br>acceptance criteria and are substantially equivalent to the predicate devices. |
| Conclusion | The subject devices possess the same intended use and technological<br>characteristics as the predicate devices. Therefore, the subject devices are<br>substantially equivalent. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.