← Product Code [MQV](/productcode/MQV) · K221726

# Uni-FuZe-C Bone Strip (K221726)

_Zavation Medical Products, LLC · MQV · Jan 18, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221726

## Device Facts

- **Applicant:** Zavation Medical Products, LLC
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Jan 18, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Uni-FuZe-C Bone Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Uni-FuZe-C Bone Strip resorbs and is replaced with bone during process.

## Device Story

Bioactive, osteoconductive, resorbable bone graft substitute in strip form; composed of purified bovine collagen, beta-tricalcium phosphate (Beta-TCP), and Bioglass 45S5. Applied directly to bone defect sites by surgeons; hydrated with patient's whole blood prior to implantation. Functions as a scaffold to fill bony voids/gaps; resorbs over time and is replaced by natural bone during the healing process. Used in clinical settings for orthopedic procedures involving extremities and pelvis. Benefits patient by facilitating bone regeneration in non-load-bearing defects.

## Clinical Evidence

No human clinical data. Bench testing included biocompatibility (ISO 10993-1, ISO 10993-6), bioactivity (ISO 23317), and packaging/sterilization validation. In vivo performance evaluated in a rabbit femoral critical cancellous bone defect model (n=not specified) comparing subject device to MasterGraft Putty. Endpoints included radiographic, micro-CT, and histological analysis at 1 day, 6 weeks, and 13 weeks. Results showed normal osteoconductive healing, absence of adverse inflammatory reactions, and performance substantially equivalent to the predicate.

## Technological Characteristics

Matrix of Type I bovine collagen (ASTM F2212), Beta-TCP (ASTM F1088), and Bioglass 45S5 (ASTM F1538). Form factor: strip. Sterile, single-use. Sterilization: VDmax method per ANSI/AAMI/ISO 11137-2 (25 kGy). Bioactive per ISO 23317. Packaging validated per ASTM F88/F88M and ASTM F2096.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- MasterGraft Putty ([K051386](/device/K051386.md))
- MasterGraft Strip ([K082166](/device/K082166.md))
- Uni-FuZe-P Bone Putty ([K201781](/device/K201781.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 18, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zavation Medical Products, LLC % Robert A. Poggie, Ph.D. President BioVera, Inc. 65 Promenade Saint Louis Notre-Dame-de-L'lle-Perrot, QC J7W3J6 Canada

Re: K221726

Trade/Device Name: Uni-FuZe-C Bone Strip Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: June 10, 2022 Received: June 14, 2022

Dear Dr. Poggie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Sara S. Thompson -S

For

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221726

Device Name Uni-FuZe-C Bone Strip

#### Indications for Use (Describe)

Uni-FuZe-C Bone Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Uni-FuZe-C Bone Strip resorbs and is replaced with bone during process.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <line stroke="black" stroke-width="2" x1="0" x2="16" y1="0" y2="16"></line> <line stroke="black" stroke-width="2" x1="16" x2="0" y1="0" y2="16"></line> </svg> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div>                                                                                                                                                          |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY for Zavation Uni-FuZe-C Bone Strip – K221726

In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of Zavation Uni-FuZe-C Bone Strip.

## A. SUBMITTERS INFORMATION

| Submitter Name:     | BioVera, Inc.                                                                   |
|---------------------|---------------------------------------------------------------------------------|
| Submitter Address:  | 65 Promenade Saint-Louis, Notre-Dame-de-L'Ile-Perrot<br>Québec, J7W 3J6, CANADA |
| Contact Person:     | Robert A. Poggie, PhD                                                           |
| Phone & Fax Number: | 514-901-0796                                                                    |
| Date of Submission: | January 5, 2023                                                                 |

## B. DEVICE IDENTIFICATION & MANUFACTURER

| Manufacturer Name:           | Zavation Medical Products, LLC                  |
|------------------------------|-------------------------------------------------|
| Manufacturer Address:        | 3670 Flowood Drive,<br>Flowood, MS 39232 USA    |
| Registration Number:         | 3008583793                                      |
| Contact Name:                | Elke Carter                                     |
| Title:                       | Regulatory Manager                              |
| Device Trade Name:           | Uni-FuZe-C Bone Strip                           |
| Device Common Name:          | Bone Void Filler                                |
| Classification Name:         | Filler, bone void, calcium compound             |
| Classification Code:         | MQV                                             |
| Classification Panel:        | Orthopedic                                      |
| Regulation Number:           | 21 CFR sections 888.3045                        |
| C1. PRIMARY PREDICATE DEVICE |                                                 |
| K051386                      | Medtronic Sofamor Danek USA, MasterGraft Putty  |
| C2. PREDICATE DEVICES        |                                                 |
| K082166                      | Medtronic Sofamor Danek USA, MasterGraft Strip  |
| K201781                      | Zavation Medical Products Uni-FuZe-P Bone Putty |

{4}------------------------------------------------

# D. DEVICE DESCRIPTION

Zavation Medical's Uni-FuZe-C Bone Strip is a bioactive osteoconductive, resorbable, biocompatible bone graft substitute in strip form. The product is composed of a matrix of purified bovine collagen per ASTM F2212, beta tricalcium phosphate (Beta-TCP per ASTM F1088), and Bioglass 45S5 per ASTM F1538.

# E. INDICATIONS FOR USE

Uni-FuZe-C Bone Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created of the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Uni-FuZe-C Bone Strip resorbs and is replaced with bone during the healing process.

# F. TECHNOLOGICAL CHARACTERISTICS

Zavation Medical's Uni-FuZe-C Bone Strip is a bioactive, osteoconductive, resorbable, biocompatible bone graft substitute in strip form. The product is composed of a matrix of purified bovine collagen per ASTM F2212, beta tricalcium phosphate (Beta-TCP per ASTM F1088), and Bioglass 45S5 per ASTM F1538. The Uni-FuZe-C Bone Strip can be applied directly to the defect site with whole blood hydration. The table below compares characteristics of the subject and predicate devices.

|                           | Subject Device<br>Zavation Uni-FuZe-C<br>Bone Strip                                           | Primary Predicate<br>K051386, Medtronic<br>MasterGraft Putty               | Predicate Device<br>K082166, Medtronic<br>MasterGraft Strip            | Predicate Device<br>K201781, Zavation<br>Uni-FuZe-P Bone Putty                                                             |
|---------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Regulatory<br>Class, Code | 888.3045, MQV                                                                                 | 888.3045, MQV                                                              | 888.3045, MQV                                                          | 888.3045, MQV                                                                                                              |
| Materials                 | Beta-TCP, ASTM F1088<br>Type I bovine collagen,<br>ASTM F2212<br>Bioglass 45S5, ASTM<br>F1538 | Biphasic ceramic that<br>is15% HA & 85% beta TCP<br>Type I bovine collagen | Biphasic ceramic of 15%<br>HA & 85% beta TCP<br>Type I bovine collagen | Beta-TCP, ASTM F1088<br>Type I bovine collagen,<br>ASTM F2212<br>Bioglass 45S5, ASTM<br>F1538<br>Polyethylene glycol (PEG) |
| Physical form             | Strip                                                                                         | Putty                                                                      | Strip                                                                  | Putty                                                                                                                      |
| Dosage                    | 3, 5, 6, 10, 13, 20 cc                                                                        | NA                                                                         | 5, 10, 20 cc                                                           | 2.5, 5.0, 10.0 cc                                                                                                          |
| Resorbable                | Yes                                                                                           | Yes                                                                        | Yes                                                                    | Yes                                                                                                                        |
| Porosity                  | Highly porous                                                                                 | Highly porous                                                              | Highly porous                                                          | Highly porous                                                                                                              |
| Sterile, single<br>use    | Yes                                                                                           | Yes                                                                        | Yes                                                                    | Yes                                                                                                                        |

The subject Uni-FuZe-C Bone Strip and predicate devices are comprised of similar materials (collagen, HA/TCP, bioglass), resorb quickly, similar in size, and can be gently manipulated into bone-voids. Mastergraft putty, the primary predicate device, was used as the control product in the functional animal model for the subject device; the study results support substantial equivalence. The Uni-FuZe-C Bone Strip has similar clinical indications for use as the primary predicate MasterGraft Putty and the same indications as the predicate Uni-FuZe-P Bone Putty.

{5}------------------------------------------------

# G. PERFORMANCE DATA

Assessment of biocompatibility of the Uni-FuZe-C Bone Strip was performed according to ISO 10993-1:2018 and biological effects were considered per the FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Material-Mediated Pyrogenicity, Cytotoxicity, Irritation, Systemic reaction, and Implantation tests demonstrated acceptable biological safety profiles.

Assessment of biocompatibility through implantation was performed per ISO 10993-6 at time points of 4 and 13 weeks. Bone healing and biological response were evaluated using an established rabbit functional femoral critical cancellous bone defect model at time points of 1 day and 6 and 12 weeks following implantation for the subject device Uni-FuZe-C Bone Strip and the primary predicate device MasterGraft Putty. Biological performance was measured using radiographic images, micro-CT, and histological analyses. The macro observations of the implant sites demonstrated healthy tissue absent of adverse inflammatory reactions for Uni-FuZe-C Bone Strip. Radiographic and microCT analyses indicated no adverse reactions and a normal progression in healing over time. Histopathology assessment showed normal osteoconductive healing response. The study confirmed the biocompatibility and normal osteoconductive healing response associated with the Uni-FuZe-C and demonstrated substantially equivalent in vivo performance to the primary predicate device across all endpoints.

The 45S5 bioglass, collagen, and ß-TCP constituents of the subject device were tested and / or certified as meeting the requirements of ASTM F2212 for Type I bovine collagen, ASTM F1538 for Bioglass 45S5, and ASTM F1088 for the β-TCP.

The subject device was determined to be bioactive per the requirements of ISO 23317. The bioactive glass granules in the Uni-FuZe-C strip precipitated apatite during the in-vitro assessment, while immersed in simulated body fluid per the definition in ISO 23317.

Packaging seal strength and integrity was validated via peel strength (per ASTM F88/F88M) and bubble emission (per ASTM F2096); acceptance criteria were met. Shipping and handling validations was performed per ASTM D7386; acceptance criteria were met. Shelf life was validated for one-year.

Sterilization was validated to SAL of 10° using method VDmax per ANSI/AAMI/ISO 11137-2: 2013/(R)2019 (VDmax25) with 25 kGy minimum dose. Limulus Amebocyte Lysate (LAL) testing demonstrated the maximum dose for one surgery has less than 20 endotoxin units (< 20 EU).

# H. CONCLUSIONS

Uni-FuZe-C Bone Strip has similar indications for use, technological characteristics, and principles of operation as the predicate devices. The minor technological differences between Uni-FuZe-C Bone Strip and the predicate devices do not raise new issues of safety or effectiveness. The performance data demonstrate that Uni-FuZe-C Bone Strip is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K221726](https://fda-staging.innolitics.com/device/K221726)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).