K221725 · Medacta International S.A. · LLZ · Jan 20, 2023 · Radiology
Device Facts
Record ID
K221725
Device Name
Advanced MyHip Planner
Applicant
Medacta International S.A.
Product Code
LLZ · Radiology
Decision Date
Jan 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The MyHip Planner software is intended for image processing and preoperative planning of acetabular cup and femoral stem positioning for Total Hip Arthroplasty (THA). The user is assisted in producing a preoperative plan, making decisions based on the leg offsets, the patient's potential impingement and, optionally, the spino-pelvic interaction. Through the software, the user can request 3D Printed Patient Specific Bone Models not intended for diagnostic use, but only intended for a physical representation of the 3D anatomical models visualized in the software. The 3D Printed Patient Specific Bone Models are provided non-sterile.
Device Story
Advanced MyHip Planner is a software-based preoperative planning tool for Total Hip Arthroplasty (THA). It processes patient CT scans and X-ray images to generate 3D anatomical models. Surgeons use the software to plan acetabular cup and femoral stem positioning, evaluate leg length/offset, assess potential impingement, and analyze spino-pelvic mobility/deformities. The software provides an optional service to produce non-sterile 3D printed bone models as visual references. Operated by surgeons in a clinical setting, the software assists in optimizing implant orientation to improve THA stability. It runs on standard Windows or Mac OS PCs. The output informs surgical decision-making by providing a patient-specific plan and visual anatomical aids, potentially improving surgical outcomes by addressing patient-specific biomechanics.
Clinical Evidence
No clinical studies were conducted. Performance was established through non-clinical software verification and validation, including segmentation validation. Analysis of automatic segmentation and landmark picking performance showed no statistical divergence from manual segmentation using a two-sided student's t-distribution.
Technological Characteristics
Software-based medical image management and processing system. Compatible with Windows and Mac OS. Inputs: CT scans and X-rays. Features: Automatic segmentation (user-verified), 3D anatomical modeling, spino-pelvic evaluation, and optional 3D printing of bone models. Connectivity: Data transfer via hard disk (CD/USB) or wireless through MySolution website.
Indications for Use
Indicated for patients undergoing Total Hip Arthroplasty (THA) requiring preoperative planning of acetabular cup and femoral stem positioning, including assessment of leg offsets, potential impingement, and spino-pelvic interaction.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 3973 Delp Street MEMPHIS, TENNESSEE 38118
Re: K221725
Trade/Device Name: Advanced MyHip Planner Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: December 22, 2022 Received: December 23, 2022
January 20, 2023
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K221725
Device Name Advanced MyHip Planner
### Indications for Use (Describe)
The MyHip Planner software is intended for image processing and preoperative planning of acetabular cup and femoral stem positioning for Total Hip Arthroplasty (THA). The user is assisted in producing a preoperative plan, making decisions based on the leg offsets, the patient's potential impingement and, optionally, the spino-pelvic interaction. Through the software, the user can request 3D Printed Patient Specific Bone Models not intended for dragnostic use, but only intended for a physical representation of the 3D anatomical models visualized in the software. The 3D Printed Patient Specific Bone Models are provided non-sterile.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221725 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Sr. Director of Quality, Regulatory, and Clinical Research, Medacta USA Date Prepared: June 13, 2022 Date Revised: January 19, 2023
#### II. Device
| Device Proprietary Name: | Advanced MyHip Planner |
|--------------------------|------------------------------------------------|
| Common or Usual Name: | Advanced MyHip Planner |
| Classification Name: | Medical image management and processing system |
| Primary Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following primary predicate device:
- A MyHip Planner, K200350, Medacta Interational SA
In addition, the following reference device is cited within the submission:
- A My Knee Cutting Blocks, K093806, Medacta International SA
#### IV. Device Description
The MyHip Planner is a software whose output is a patient-specific preoperative plan based on CT scans and aimed at evaluating the effects of different device choices and positioning options on the patient's hip joint biomechanics in terms of leg length and offset.
It is intended to be used in Primary Hip Arthroplasty and it is compatible with Windows and Mac OS operating system.
The subject Advanced version of the MyHip Planner additionally allows to evaluate patient's spinopelvic deformities and pelvic tilt starting from preoperative X-ray images, to help the surgeon to
{4}------------------------------------------------
understand the implications of spinopelvic mobility on THA stability and optimize implant components orientation.
Through the software, the user can request non-sterile 3D Printed Patient Specific Bone Models intended to be used as an additional visual reference of the patients' specific anatomy.
#### V. Indications for Use
The MyHip Planner software is intended for image processing and preoperative planning of acetabular cup and femoral stem positioning for Total Hip Arthroplasty (THA). The user is assisted in producing a preoperative plan, making decisions based on the leg offsets, the patient's potential impingement and, optionally, the spino-pelvic interaction.
Through the software, the user can request 3D Printed Patient Specific Bone Models not intended for diagnostic use, but only intended for a physical representation of the 3D anatomical models visualized in the software. The 3D Printed Patient Specific Bone Models are provided non-sterile.
| Parameters | Advanced MyHip Planner<br>(Subject device) | MyHip Planner<br>(Predicate device K200350) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>use | The MyHip Planner software is<br>intended for image processing and<br>preoperative planning of acetabular<br>cup and femoral stem positioning<br>for Total Hip Arthroplasty (THA).<br>The user is assisted in producing a<br>preoperative plan, making decisions<br>based on the leg offsets, the<br>patient's potential impingement<br>and, optionally, the spino-pelvic<br>interaction.<br><br>Through the software, the user can<br>request 3D Printed Patient Specific<br>Bone Models not intended for<br>diagnostic use, but only intended for<br>a physical representation of the 3D<br>anatomical models visualized in the<br>software. The 3D Printed Patient<br>Specific Bone Models are provided<br>non-sterile. | The MyHip Planner software is<br>intended for image processing and pre-<br>operative planning of acetabular cup<br>and femoral stem positioning for Total<br>Hip Arthroplasty (THA). The device<br>assists the user in assessing potential<br>leg length and offset differences as<br>well as potential range of motion<br>impingement. |
| Operative<br>system | Windows or Mac OS | Same |
| Hardware | Laptop and/or PC | Same |
| Input images | <br>• CT scan images | CT scan images |
#### VI. Comparison of Technological Characteristics
{5}------------------------------------------------
| Parameters | Advanced MyHip Planner<br>(Subject device) | MyHip Planner<br>(Predicate device K200350) |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Input images<br>acquisition<br>method | X-rays Through hard disk (CD/USB) or wireless through MySolution website | Same |
| User interface | GUI | Same |
| Workflow | CT images upload Segmentation Planning including spino-pelvic evaluation Bone models production and shipping (optional) | CT images upload Segmentation Planning |
| Segmentation | Automatic but the user has to check, eventually modify, and confirm it | Same |
| Available<br>implants | Medacta hip implants cleared to FDA by Medacta | Same |
| Measurements<br>output | Cup positioning Stem positioning ROM evaluation Spino-pelvic evaluation | Cup positioning Stem positioning ROM evaluation |
The subject Advanced MyHip Planner and the predicate device are substantially equivalent with regards to the following characteristics:
- operative system compatibility;
- design concept/principle of operation;
- . user interface;
- images importation method;
- segmentation and landmark acquisition method; and
- . available implants.
The subject Advanced MyHip Planner differs from the predicate device with respect to:
- input images;
- workflow; and
- measurements output.
# Discussion
The slightly different input images and workflow of the subject device with respect to the predicate does not affect the safety and effectiveness of the Advanced MyHip Planner since it is only related to the addition of new input images (x-rays) and workflow steps strictly related to the newly developed feature allowing spino-pelvic evaluation and the manufacturing of bone models, both optional steps. Identically to the bone models included in the reference device clearance (K093806), the subject bone models are manufactured through the software starting from the patient's CT as an additional visual reference of the patients' specific anatomy, thus no new issue of safety and effectiveness arise.
{6}------------------------------------------------
Also measurements output differences are strictly related to the new spino-pelvic evaluation feature and does not raise any new issue of safety and effectiveness since the other output measurements are identical to the ones of the predicate device (K200350).
The comparison of technological characteristics and validation provided within this submission, supports the substantial equivalence of the subject devices respect to the predicate devices.
#### VII. Performance Data
Based on the risk analysis, verification activities were conducted to written protocols. The following software verification and validation are provided in support of the substantial equivalence determination:
# Non-Clinical Studies
- . Software verification and validation including segmentation validation. No statistical divergence between the algorithm and the manual segmentation has been revealed by an analysis of the automatic segmentation and landmark picking performance on a two-sided students t-distribution.
# Clinical Studies:
- No clinical studies were conducted. ●
#### Conclusion VIII.
The information provided above supports that the Advanced MyHip Planner is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.