CompuFlo Epidural Computer Controlled Anesthesia System
K221702 · Milestone Scientific, Inc. · FMF · Feb 25, 2023 · General Hospital
Device Facts
Record ID
K221702
Device Name
CompuFlo Epidural Computer Controlled Anesthesia System
Applicant
Milestone Scientific, Inc.
Product Code
FMF · General Hospital
Decision Date
Feb 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The CompuFlo® Epidural Computer Control Anesthesia System is intended for use with an epidural needle for the realtime verification of needle placement in the lumbar or thoracic epidural space, inclusive of cervicothoracic junction (CT). It is intended for patients over age of 18 who are required to have epidural needle or catheter placement as part of a medically necessary, in-patient or out-patient procedure, as established by their Health Care Provider (HCP). The CathCheck is an additional feature for clinicians to use to assess proper catheter placement inside the epidural space by evaluating a pulsatile waveform. The CathCheck function utilizes the same technology as the CompuFlo® Epidural Computer Control Anesthesia System and should only be used in conjunction with standard clinical assessment practice. Once the Health Care Provider verifies the epidural needle or catheter placement in the epidural space, the HCP continues with the medical procedure.
Device Story
Device provides real-time verification of epidural needle placement; utilizes pressure-sensing technology to monitor resistance at needle tip. System includes primary unit, foot pedal, and single-use disposable kit (pressure sensor, syringe, tubing). Clinician operates device to navigate tissue; system provides visual/audible feedback based on pressure differentials. CathCheck feature evaluates pulsatile waveforms to confirm catheter placement within epidural space. Used in inpatient/outpatient settings by healthcare providers. Output assists clinicians in confirming needle/catheter location, reducing risk of accidental dural puncture, and facilitating successful epidural anesthesia administration.
Clinical Evidence
Prospective, controlled, mono-center, parallel-group randomized trial (n=133). Evaluated thoracic epidural placement in patients 18-75 years (BMI 18-50 kg/m²). Primary endpoint: successful thoracic epidural placement. CompuFlo success rate 96.7% vs. 91.2% for traditional LOR technique (no statistically significant difference). Safety profile comparable; two accidental dural punctures in each group. Hypotension reported but deemed unrelated to device.
Technological Characteristics
System includes primary unit, pneumatic foot pedal, and disposable kit (pressure sensor, 20mL syringe, tubing). Operates via AC mains or internal battery. Sensing principle: real-time pressure monitoring. Software-controlled. Sterilization: validated to SAL 1x10^-6. Packaging compliant with ISO 11607-1:2019 and ISO 11607-2:2019.
Indications for Use
Indicated for patients >18 years requiring epidural needle or catheter placement in lumbar or thoracic epidural space (including CTJ) for medically necessary procedures. Contraindications not specified.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
CompuFlo® Epidural Computer Controlled Anesthesia System (K161883)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2023
Milestone Scientific, Inc. % Kevin Go Consultant Ram+ 2790 Mosside Blvd. Monroeville, Pennsylvania 15146
Re: K221702
Trade/Device Name: CompuFlo Epidural Computer Controlled Anesthesia System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: January 25, 2023 Received: January 26, 2023
Dear Kevin Go:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James J. Lee -S
James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing. Respiratory and Anesthsia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221702
#### Device Name
CompuFlo® Epidural Computer Controlled Anesthesia System
#### Indications for Use (Describe)
The CompuFlo® Epidural Computer Control Anesthesia System is intended for use with an epidural needle for the realtime verification of needle placement in the lumbar or thoracic epidural space, inclusive of cervicothoracic junction (CT). It is intended for patients over age of 18 who are required to have epidural needle or catheter placement as part of a medically necessary, in-patient or out-patient procedure, as established by their Health Care Provider (HCP).
The CathCheck is an additional feature for clinicians to use to assess proper catheter placement inside the epidural space by evaluating a pulsatile waveform. The CathCheck function utilizes the same technology as the CompuFlo® Epidural Computer Control Anesthesia System and should only be used in conjunction with standard clinical assessment practice.
Once the Health Care Provider verifies the epidural needle or catheter placement in the epidural space, the HCP continues with the medical procedure.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
DATE PREPARED
February 19, 2023
## MANUFACTURER AND 510(k) OWNER
Milestone Scientific, Inc. 425 Eagle Rock Ave. Suite 403 Roseland, NJ 07068 Telephone: 973-535-1117 Official Contact: Arjan Haverhals, CEO
# REPRESENTATIVE/CONSULTANT
Kevin Go, MBA, RAC, CQA ROM+ 2790 Mosside Blvd Suite 800, Monroeville, PA 15146 Email: KGo@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: CompuFlo® Epidural Computer Controlled Anesthesia System Regulation Number: 21 CFR 880.5860 Class: Class II Product Code: FMF Review Panel: General Hospital
## PREDICATE DEVICE IDENTIFICATION
The CompuFlo® Epidural Computer Controlled Anesthesia System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Predicate/Reference |
|---------------|-------------------------------------------------------------|---------------------|
| K161883 | CompuFlo® Epidural Computer Controlled<br>Anesthesia System | Predicate |
| K201356 | Plastic LOR Syringe | Reference |
The predicate devices have not been subject to a design related recall.
# DEVICE DESCRIPTION
The device consists of: (1) the primary unit, (2) a pneumatic foot pedal control, (3) an AC power cord and (4) single-use disposable kit, which can include an external in-line fluid pressure sensor, a 20 mL plastic syringe, plastic tubing and an ID Adapter. Operation of the device is allowed when powered by the AC mains or by the internal battery source. The epidural needle and solution is not supplied in the disposable Kit
{4}------------------------------------------------
# INDICATIONS FOR USE
The CompuFlo® Epidural Computer Control Anesthesia System is intended for use with an epidural needle for the real-time verification of needle placement in the lumbar or thoracic epidural space, inclusive of cervicothoracic junction (CTJ). It is intended for patients over age of 18 who are required to have epidural needle or catheter placement as part of a medically necessary, inpatient or out-patient procedure, as established by their Health Care Provider (HCP).
The CathCheck is an additional feature for clinicians to use to assess proper catheter placement inside the epidural space by evaluating a pulsatile waveform. The CathCheck function utilizes the same technology as the CompuFlo® Epidural Computer Control Anesthesia System and should only be used in conjunction with standard clinical assessment practice.
Once the Health Care Provider verifies the epidural needle or catheter placement in the epidural space, the HCP continues with the medical procedure.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
This submission introduces several changes to the previously cleared submission including expanding the indications for use to include use in the thoracic space, compatibility with a catheter component (CathCheck program), modification to the packaging configuration for the disposable kits, and minor changes to the software. Milestone Scientific. Inc. believes that the CompuFlo® Epidural Computer Control Anesthesia System is substantially equivalent to the predicate devices based on the information summarized here:
The proposed CompuFlo® Epidural Computer Controlled Anesthesia System has the same intended use, environment of use, materials, operating principle, and fundamental technology as the predicate device. A clinical study has been conducted to demonstrate the safety and effectiveness of the device in the newly proposed thoracic indication. Software verification and validation have been completed to ensure the functionality of the new software features and address any residual risks. Finally, sterilization and packaging validation has been conducted to establish the acceptability of the new packaging configuration for the disposable kits. Therefore, the information provided in this submission supports safety and effectiveness of the proposed device for its intended use and demonstrates substantial equivalence to the predicate device.
# SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
| Performance Category | Testing |
|--------------------------|----------------------------------------------------------------------------------------------|
| Sterilization Validation | Sterilization validation has been completed on the worst-case construct to an SAL of 1x10-6. |
{5}------------------------------------------------
| Packaging Validation | Packaging performance validation, in accordance with ISO<br>11607-1:2019 and ISO 11607-2:2019, was performed on the<br>final packaging and sterilized device. |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software Verification<br>and Validation | Software verification and validation in accordance with the<br>appropriate Level of Concern as described in FDA Guidance:<br>Guidance for the Content of Premarket Submissions for<br>Software Contained in Medical Devices, issued May 11, 2005 |
The results of these tests indicate that the CompuFlo® Epidural Computer Controlled Anesthesia System is substantially equivalent to the predicate devices.
# SUMMARY OF CLINICAL TESTING
The purpose of this investigation study is to demonstrate that the safety and efficacy profile of the CompuFlo Epidural System for thoracic epidural placement is substantially equivalent to the traditional LOR technique. This study was a prospective, controlled, mono-center, parallel-group randomized trial in the US. A total of 133 subjects were enrolled at a single site. The study evaluated patients aged 18 - 75 years old with BMIs ranging from 18 -50 kg/m².
The primary endpoint of the study was successful performance of thoracic epidural placement. The CompuFlo group has a 96.7% success rate compared to a 91.2% success rate of the traditional LOR technique. The results of the primary efficacy endpoint demonstrated a success rate that was not statistically superior to the control. No significant difference on primary success rate was detected between the two study groups.
Regarding the safety endpoint, two (2) accidental dural punctures were observed in both groups. No other adverse events except for hypotension were reported during the study in the CompuFlo group. The overall observed hypotension was mild and patients were treated per routine protocol. Furthermore, the principal investigator judged the causal relationship as "not related" to the CT treatment since this adverse event was observed in both treatment groups.
# CONCLUSION
Based on the evaluations performed, including sterilization, packaging validation, software verification and validation and Milestone's clinical study, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate or reference devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed CompuFlo® Epidural Computer Controlled Anesthesia System are assessed to be substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.