K221696 · Re Walk Robotics , Ltd. · PHL · Mar 2, 2023 · Physical Medicine
Device Facts
Record ID
K221696
Device Name
ReWalk P6.0
Applicant
Re Walk Robotics , Ltd.
Product Code
PHL · Physical Medicine
Decision Date
Mar 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3480
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K221696 · Mar 2, 2023
ReWalk P6.0
Re Walk Robotics , Ltd.
ReWalk Users Field Survey Study; Customer Relationship Management (CRM) database
The sponsor used retrospective survey data and CRM complaint/malfunction monitoring from European users to support the safety and effectiveness of the new 'stairs ascend and descend' functionality in real-world home and community settings.
Field Survey; Postmarket Surveillance; Real-world usage; CRM database
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
ReWalk P6.0 Active Users Field Survey Study- Stairs Usage; Retrospective field survey and CRM database analysis; Follow-up/Duration: Mean 26 months (range 1 month to 5 years); Study Period: 2015 to May 2022
47 users with spinal cord injury (T7-L5) using the device with stairs functionality; Sample Size: 47
Not applicable for this study
Safety (adverse events) and effectiveness (stairs usage)
Indications for Use
The ReWalk® P6.0 fits to the lower limbs and part of the upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory functions in home and community settings with supervision of a specially certified companion in accordance with the User assessment and training certification program. The device is also intended to enable individuals with spinal cord injury at levels T4 to T6 to perform ambulatory functions in rehabilitation institutions in accordance with the User assessment and training certification program. The ReWalk P6.0 is intended for indoor and outdoor use: including standing and walking on level surfaces and mild slopes, and ascending and descending stairs and curbs.
Device Story
Powered lower extremity exoskeleton; motorized frame fits lower limbs/upper body. Inputs: user-operated wireless remote controller (RC), tilt sensor, specific body movements. Operation: gears/motors at hip/knee joints execute gait. Output: ambulatory movement (standing, walking, stairs/curbs). Used in home/community (with certified companion) or rehabilitation (with therapist). Healthcare provider/companion monitors device status via RC/GUI. Enables mobility for SCI patients; improves independence. System includes exoskeleton, waistpack, straps, battery charger, laptop GUI, and crutches.
Clinical Evidence
Real-world evidence from 47 users (European market, 6-year monitoring) and human factors study (11 patients/11 companions). Field survey: 18,038 total stairs climbed; no device-related adverse events. Human factors study: 99% success rate in task completion; no adverse events reported. Bench testing confirmed mechanical integrity and walking speed.
Indicated for individuals aged 18+ with spinal cord injury (SCI) at levels T7-L5 for home/community use with certified companion, or T4-T6 for institutional rehabilitation use with certified therapist.
Regulatory Classification
Identification
A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
*i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
*e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
*e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
*e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 2, 2023
ReWalk Robotics Ltd. Miri Pariente VP QA, RA & Operations 3 Hatnufa St. POB Box-161 Yokneam, 2069203 Israel
Re: K221696
Trade/Device Name: ReWalk® P6.0 Regulation Number: 21 CFR 890.3480 Regulation Name: Powered lower extremity exoskeleton Regulatory Class: Class II Product Code: PHL Dated: June 10, 2022 Received: June 10, 2022
Dear Miri Pariente:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221696
Device Name ReWalk® P6.0
#### Indications for Use (Describe)
The ReWalk® P6.0 fits to the lower limbs and part ofthe upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory functions in home and community settings with supervision of a specially certified companion in accordance with the user assessment and training certification program. The device is also intended to enable individuals with spinal cord injury at levels T4 to T6 to perform ambulatory functions in rehabilitation institutions in accordance with the user assessment and training certification program. The ReWalk P6.0 is intended for indoor and outdoor use: including standing and walking on level surfaces and mild slopes and ascending and descending stairs and curbs.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the ReWalk Robotics logo. The word "ReWalk" is displayed in a combination of gray and yellow colors, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in gray, with a series of vertical lines to the left of the word. The logo is clean and modern, with a focus on the company's name and branding.
# ReWalk® P6.0 510(k) Summary K221696
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ReWalk Robotics. The word "ReWalk" is written in a combination of gray and yellow colors. Below the word "ReWalk" is the word "Robotics" written in gray. There are also several gray lines to the left of the word "Robotics".
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
#### 1. Submission Sponsor ReWalk Robotics Ltd. 3 Hatnufa St. Yokneam, Israel 2069203, POB: 161 Establishment Registration: 3007615665 Submission Correspondent 2. Miri Pariente, VP QA, RA & Operations Phone: (+972) 58-5605090 E-mail: miri.pariente@rewalk.com 3. Date Prepared June 09, 2022 4. Device Identification ReWalk® P6.0 Name of Device: Classification Name: Powered Exoskeleton 21 CFR §890.3480 Regulation: Regulatory Class: Class II Product Classification Code: PHIL Neurology Classification panel: 5. Legally Marketed Predicate Device
| Predicate Manufacturer: | ReWalk Robotics Ltd. |
|-------------------------|----------------------|
| Predicate Trade Name: | ReWalk P6.0 |
| Predicate 510(k): | K200032 |
#### 6. Legally Marketed Reference Device
| Reference Device Manufacturer: | ReWalk Robotics Ltd. |
|--------------------------------|----------------------|
| Reference Device Trade Name: | ReWalk |
| Reference Device 510(k): | K160987 |
#### 7. Device Description
The ReWalk® P6.0 Exoskeleton is a prescription device which enables individuals with spinal cord injuries to perform ambulatory functions and composed of an external, powered, motorized frame that fits to the lower limbs and part of the upper body.
The ReWalk is intended to enable certified users with spinal cord injuries at levels T7 to L5 to perform ambulatory functions in in home and community setting accompanied by a certified companion. Additionally, the ReWalk Personal Exoskeleton 6.0 is intended to enable certified users with spinal cord injuries at levels T4 to T6 to perform ambulatory functions in rehabilitation centers accompanied by a certified therapist.
Control of the device is achieved through a wrist-worn User-operated wireless remote controller (RC), tilt sensor and specific body movements. The gait movements are performed by a set of gears and motors at the knee and the hip joints. The ReWalk system includes Remote control (RC) Communicator, Exoskeleton (Inc. Rigid Frames, Waistpack and Straps), Battery charger and Laptop (GUI), and off the shelf crutches. All of the ReWalk components are mandatory, suitable for indoor and outdoor usage. The device is intended for indoor use: including standing and walking on level surfaces and mild slopes, and ascending and descending stairs and curbs for users who are at
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for ReWalk Robotics. The word "ReWalk" is written in a stylized font, with the "Re" in gray and the "Walk" in yellow. Below the word "ReWalk" is the word "Robotics" in gray. The logo is simple and modern, and it conveys the company's focus on robotics and walking.
least 18 years old.
#### 8. Intended Use / Indication for Use
The ReWalk® P6.0 fits to the lower limbs and part of the upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory functions in home and community settings with supervision of a specially certified companion in accordance with the User assessment and training certification program. The device is also intended to enable individuals with spinal cord injury at levels T4 to T6 to perform ambulatory functions in rehabilitation institutions in accordance with the User assessment and training certification program. The ReWalk P6.0 is intended for indoor and outdoor use: including standing and walking on level surfaces and mild slopes, and ascending and descending stairs and curbs.
#### 9. Technological Characteristics
The current submission introduces the stairs ascend and descend functionality to the ReWalk P6.0 device (already cleared for sit, stand, and walk under K160987 and K200032). The ReWalk P6.0 with stairs functionality is identical to the previously cleared version of the device ("Predicate Device") (K200032) in terms of design, material, physical & electrical components, energy source, and principles of operation.
Adding the Stairs ascend and descend functionality to the ReWalk P6.0 device does not raise different questions of safety and effectiveness, as confirmed by the company's clinical and performance testing.
# 10. Performance Data
## 10.1 Performance Testing - Bench
Both the Subject Device and Predicate Device are identical in physical and electrical components and software, with Stairs functionality disabled for the cleared device and enabled for the subject device. Therefore, testing submitted under (K160987 and K200032) remain applicable to the subject device: EMC (per IEC 60601-1-2), Electrical safety (per IEC 60601-1), Biocompatibility (per ISO 10993-1), Software validation & verification (per IEC 62304), Environmental testing (per IEC 60601-11).
The Subject Device has undergone additional bench and non-clinical testing to demonstrate that the ReWalk P6.0 device with stairs ascend and descend functionality meets all design requirements and is in compliance with all applicable standards and regulations including special controls in 21 CFR 890.3480. These tests include non-clinical performance testing, and user testing. The tests are summarized at a high level in the tables below.
| Test name | Test description | Result |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Stairs usage mechanical loading test | Device has sufficient mechanical integrity for<br>safe and effective use under worst-case<br>conditions | PASS |
| Walking Speed verification | Measure the maximal walking speed of ReWalk<br>P6.0 device with its maximal configuration (as<br>the maximal speed depends on device<br>configuration). | PASS |
### Table 1: Bench testing
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for ReWalk Robotics. The word "ReWalk" is written in a bold, sans-serif font, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in a smaller, gray font. There are also several gray lines below the word "Robotics".
| Test name | Test description | Result |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------|
| Stairs Validation | Device performs as intended when used by certified users for ascending and descending stairs and curbs in a variety of environments | PASS |
| Curbs Validation | Device performs as intended when used by certified users for ascending and descending stairs and curbs in a variety of environments | PASS |
| Outdoor Validation | Device performs as intended when used by certified users for ascending and descending stairs and curbs in a variety of environments | PASS |
# 10.2 Performance Testing - Clinical
## ReWalk P6.0 Active Users Field Survey Study- Stairs Usage
Use of the stairs functionality is supported by clinical data collected by the Company in the ReWalk Users Field Survey Study-Stairs Usage, which contains real world evidence collected through a survey of actual users in Europe (where ReWalk P6.0 device with stairs functionality has been commercially available on the European market since 2015). The safety and efficacy of ReWalk P6.0 device with stairs ascend and descend functionality, both indoor and outdoor use in home and community setting has been studied and monitored over a period of six (6) years. In addition, and unrelated to the survey, for the past seven (7) years, data regarding device malfunctions, injuries, and complaints had also been monitored using ReWalk's Customer Relation Management (CRM) database.
### Study subgroup analysis/ demographic characteristics
A total of 47 users have been evaluated. 87% of subjects were males and 13% females. All study subjects were at least 6 months after injury, who had a device with stairs functionality to ascend and descend stairs.
Note that information regarding users' race and ethnicity was not allowed to be collected due to local legislation laws in Europe.
### Safety and effectiveness
Between 2015 to May 2022, only three (3) Stairs-related adverse events (AE) were recorded in ReWalk's Customer Relationship Management system, covering all ReWalk users in Europe. It was concluded that the AEs were not device-related, but rather associated with erroneous mode selection , not-following the device Instructions for Use and short handle at the top step. No device deficiencies/malfunctions related to stairs usage were recorded.
### Study summary and conclusion
The study recruited 85 subjects, of which a total of 47 users used the device to climb stairs (ascending/ descending) indoor and outdoor use in home and community setting. Average age was 46 years; 87% of the stairs-mode users were male. On average the stairs-mode was used by the study participants for 26 months (ranging from 1 month to 5 years). Total number of stairs taken by the stairs-mode users in the study, ranged from 9 to 2,371 stairs per-user, with the mean value of 383 stairs (which is approximately equal to 22 floors, assuming 17 stairs per floor). This yields cumulatively 18,038 stairs in total for all study users over the entire duration of use. The study has successfully demonstrated that the ReWalk P6.0 device is safe and effective in enabling individuals with spinal cord injury to ascend stairs, both indoor and outdoor, in home and community settings. The use of the ReWalk P6.0 device is safe when it is
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for ReWalk Robotics. The word "ReWalk" is written in a combination of gray and yellow, with the "Re" in gray and the "Walk" in yellow. Below "ReWalk" is the word "Robotics" in gray. There are several gray lines to the left of the word "Robotics".
used according to its instructions for use under the supervision of a specially certified companion.
# 10.3 Performance Testing - Human Factor Study ReWalk P6.0 with Stairs Enabled Human Factors Engineering Study
ReWalk Users and companions human factors engineering study have been conducted under IRB Approved protocol, according to principles of Good Clinical Practice (GCP). The study objective was to evaluate patients' ability to safely and effectively interact with ReWalk P6.0 device with stairs feature user interface components, while assisted by their companion (in which advanced functions of ascending and descending stairs and curbs are introduced). The study was conducted in a representative real world use environment, included actual use test scenarios using the ReWalk P6.0 device with stairs enabled and comprehension questions about critical information regarding the use of the device focusing on the ascending/descending function.
The study consists of 11 ReWalk patients subjects and 11 companion's subjects who performed actual use scenarios using a working device with the Stairs feature enabled. The training during the usability testing was designed to provide realistic training environment that users would encounter in real world situations. The content and the method of delivery were identical to the training performed at the clinic.
## Study Summary and Conclusion
Despite the significant limitations and constraints imposed by the unique user characteristics, the study results demonstrate that after a short and partial training session, participants were largely able to learn and execute the sequence of steps required for stair use. In 99% of the times participants successfully completed the test tasks without any observed use difficulties, and the vast majority of participants were able to correctly answer the knowledge questions evaluated in scenarios. No adverse events were reported.
The study concluded that the ReWalk P6.0 stairs and curbs ascend and descend functions can be used safely by the intended user population in the intended use environments.
### 11. Conclusion
The Subject Device with stairs functionality and its Predicate Device have similar intended use, same indications, contraindications, principles of operation, and technological characteristics. The difference in the intended use does not present different questions of safety or effectiveness than the Predicate Device. Accepted scientific methods exist to evaluate the performance of the Subject Device compared to its Predicate Device. Clinical and performance data have demonstrated that the Subject Device is as safe and effective as its Predicate. Thus, the ReWalk P6.0 device with stairs functionality is substantially equivalent to the previously cleared ReWalk P6.0 device (K200032).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.