← Product Code [FAJ](/productcode/FAJ) · K221690

# OES CYSTONEPHRO FIBERSCOPE (OLYMPUS CYF-5), OES CYSTONEPHRO FIBERSCOPE (OLYMPUS CYF-5R), LIGHT GUIDE ADAPTER (MAJ-1413) (K221690)

_Olympus Medical Systems Corp. · FAJ · Dec 29, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221690

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Dec 29, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

OES CYSTONEPHROFIBERSCOPEs OLYMPUS CYF-5 and CYF-5R have been designed to be used with an Olympus light source, documentation equipment, display monitor. EndoTherapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra and kidney.

## Device Story

The OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 and CYF-5R are flexible fiberscopes used for endoscopic diagnosis and treatment within the bladder, urethra, and kidney. The device consists of a control section, insertion section, and eyepiece section. It functions by transmitting visual information via an optical fiber system to an external display monitor when connected to an Olympus light source and documentation equipment. The device is operated by physicians in a healthcare facility or hospital setting. It allows for the passage of EndoTherapy accessories through an instrument channel for diagnostic or therapeutic procedures. The device is reusable and requires reprocessing (cleaning, disinfection, and sterilization) between uses. It provides direct visualization to assist clinicians in identifying pathologies and performing interventions, potentially improving diagnostic accuracy and treatment outcomes for urological conditions.

## Clinical Evidence

No clinical study was performed to demonstrate substantial equivalence. Evidence is based on non-clinical bench testing, including reprocessing validation, biocompatibility testing (ISO 10993-1), electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-2-18), and performance testing (thermal safety, durability, color performance, photobiological safety, image uniformity, resolution, and field/direction of view).

## Technological Characteristics

Reusable flexible fiberscope. Materials: Fluorine resin (insertion tube), Fluoro Rubber (bending section), Polyphenylsulfone (distal end), Stainless Steel (instrument channel components), PTFE (instrument channel), Silicone (seals). Optical sensing via fiber optics. Dimensions: 5.5mm insertion tube diameter, 4.6mm distal end diameter, 380mm working length. 120-degree field of view, 3-50mm depth of field. Reprocessing: AER, manual, ETO gas, or STERRAD NX. Connectivity: Compatible with Olympus light sources/monitors.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE VA2, CYF TYPE V2R ([K133538](/device/K133538.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 29, 2022

Olympus Medical Systems Corp. % Teffany Hutto Regulatory Affairs Consultant Olympus Corporation of the Americas 3500 Corporate Parkway, PO Box 610 Center Valley, PA 18034-0610

Re: K221690

> Trade/Device Name: OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5R Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: November 29, 2022 Received: November 29, 2022

Dear Teffany Hutto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K221690

### Device Name

OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5R

### Indications for Use (Describe)

OES CYSTONEPHROFIBERSCOPEs OLYMPUS CYF-5 and CYF-5R have been designed to be used with an Olympus light source, documentation equipment, display monitor. EndoTherapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra and kidney.

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| ×   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. A small registered trademark symbol is located to the upper right of the letter "S".

## 510(k) Summary

### 1. GENERAL INFORMATION

| ■ 510(k) Submitter: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan |
|---------------------|-----------------------------------------------------------------------------------------|
|---------------------|-----------------------------------------------------------------------------------------|

- Contact Person: Teffany Hutto Consultant for Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 512-508-6550 Fax: 484-896-7128 Email: teffany.hutto(@olympus.com
- . Manufacturing site: Shirakawa Olympus Co., Ltd. 3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawagun, Fukushima 961-8061, Japan

Aizu Olympus Co., Ltd., 3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

### 2. DEVICE IDENTIFICATION

- Device Name: OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5R
- Model Name: CYF-5, CYF-5R
- Common Name: CYSTO-NEPHRO FIBERSCOPE
- Regulation Number: 876.1500
- Regulation Name: Endoscope and accessories
- Regulatory Class: Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. A registered trademark symbol is located to the right of the last letter, "S". The background is white.

- FAJ (Cystoscope And Accessories, Flexible/Rigid) ■ Product Code:
- Classification Panel: Gastroenterology/Urology

### 3. PREDICATE DEVICE

### ■ Predicate device

| Device name                                                                                                        | 510(k) Submitter                 | 510(k) No. |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------|------------|
| VISERA CYSTO-NEPHRO VIDEOSCOPE<br>OLYMPUS CYF TYPE V2, CYF TYPE<br>VA2, CYF TYPE V2R<br>(EVIS EXERA II 180 SYSTEM) | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K133538    |

### 4. DEVICE DESCRIPTION

### ■ General Description of the subject device

The subject device has been designed to be used with an Olympus light source, documentation equipment, display monitor, EndoTherapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra and kidney.

### ■ Principle of Operation

The subject device consists of three parts: the control section, the insertion section, the eyepiece section. The basic principle including user interface and operation for the procedure of the subject device are identical to that of the predicate device.

{5}------------------------------------------------

#### ■ List of device components

The OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 and CYF-5R will be packed and offered together as Table below. They can be used with commercially available Olympus devices as described within the Instruction Manual.

| Model No. | Device Name                           |
|-----------|---------------------------------------|
| BW-411B   | Single Use Combination Cleaning Brush |
| MAJ-1413  | Light guide adapter                   |
| MB-156    | ETO cap                               |

### 5. INDICATIONS FOR USE

OES CYSTONEPHROFIBERSCOPEs OLYMPUS CYF-5 and CYF-5R have been has been designed to be used with an Olympus light source, documentation equipment, display monitor, EndoTherapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra and kidney.

### 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

The subject device has the same technological characteristics and design as the predicate device except for the following features:

- Optical mechanism
- Optical performance

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

A side-by-side comparison of the subject device and the predicate device is provided below.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "S".

|                                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                    | OLYMPUS CYF-5, CYF-5R                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for<br>Use                  | OES<br>CYSTONEPHROFIBERSCOPEs<br>OLYMPUS CYF-5 and CYF-5R<br>have been been designed to be used<br>with an Olympus light source,<br>documentation equipment, display<br>monitor, EndoTherapy accessories<br>and other ancillary equipment for<br>endoscopic diagnosis and treatment<br>within the bladder, urethra and<br>kidney.                                                                            | This instrument has been designed<br>to be used with an Olympus video<br>system<br>center, light source, documentation<br>equipment, monitor, EndoTherapy<br>accessories, and other ancillary<br>equipment for endoscopic<br>diagnosis and treatment within the<br>bladder, urethra, ureter, and<br>kidney.                                                                                                             |
| Regulation<br>Number                    | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                               | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulation Name                         | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                    | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulatory Class                        | Class II                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code                            | FAJ                                                                                                                                                                                                                                                                                                                                                                                                          | FAJ                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Classification<br>Panel                 | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                     | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                                |
| Use Environment                         | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                 | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                            |
| Reprocessing                            | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                  | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                             |
| Single-Use/Reuse                        | Reusable                                                                                                                                                                                                                                                                                                                                                                                                     | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                |
| Duration and<br>type of contact         | Surface-contacting device in contact<br>with mucosal membranes. The<br>contact duration is limited exposure<br>(<24 hours).                                                                                                                                                                                                                                                                                  | Surface-contacting device in<br>contact with mucosal membranes.<br>The contact duration is limited<br>exposure (<24 hours).                                                                                                                                                                                                                                                                                             |
| Depth of Field                          | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                       | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Direction of View                       | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                          | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                                     |
| Field of View                           | 120°                                                                                                                                                                                                                                                                                                                                                                                                         | 120°                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Outer Diameter of<br>Distal End         | ⌀ 4.6mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                       | ⌀ 4.8mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                                  |
| Diameter of<br>Insertion Tube           | ⌀ 5.5mm                                                                                                                                                                                                                                                                                                                                                                                                      | ⌀ 5.4mm                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                        |
| Item                                    | OLYMPUS CYF-5, CYF-5R                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                           |
| Bending Section<br>Angulation           | Up 210° / Down 120°                                                                                                                                                                                                                                                                                                                                                                                          | Up 210° / Down 120°                                                                                                                                                                                                                                                                                                                                                                                                     |
| Working Length                          | 380mm                                                                                                                                                                                                                                                                                                                                                                                                        | 380mm                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Instrument<br>Channel inner<br>diameter | ⌀ 2.4 mm                                                                                                                                                                                                                                                                                                                                                                                                     | ⌀ 2.2 mm                                                                                                                                                                                                                                                                                                                                                                                                                |
| Observation mode                        | WLI                                                                                                                                                                                                                                                                                                                                                                                                          | WLI, NBI                                                                                                                                                                                                                                                                                                                                                                                                                |
| Patient contacting<br>material          | Insertion tube: Forced dry-hard type<br>of fluorine resin coat<br>Bending section: Fluoro Rubber<br>Glue: Epoxy glue<br>Distal End: Polyphenylsulfone<br>Objective lens/ Light guide lens:<br>Glass<br>Objective lens frame/ Instrument<br>channel pipe/ Instrument channel<br>joint/ Instrument channel port:<br>Stainless Steel<br>Instrument Channel:<br>Polytetrafluoroethylene<br>Rubber seal: Silicone | Insertion tube: Forced dry-hard<br>type of fluorine resin coat<br>Bending section: Fluoro Rubber<br>Glue: Epoxy glue<br>Distal End: Polyphenylsulfone<br>Objective lens/ Light guide lens:<br>Glass<br>Solder: AuSn Solder<br>Instrument channel pipe/<br>Instrument channel joint/ Junction/<br>Instrument channel port: Stainless<br>Steel<br>Instrument Channel:<br>Polytetrafluoroethylene<br>Rubber seal: Silicone |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is slightly angled upwards from left to right. There is a registered trademark symbol to the right of the letter "S".

### 7 PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

### 1) Reprocessing validation testing

Reprocessing instruction and reprocessing method validation testing were conducted and documentations were provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".

### 2) Biocompatibility testing

Biocompatibility testing were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

{8}------------------------------------------------

- Cytotoxicity Study Using the ISO Elution Method
- ISO Intracutaneous Study in Rabbits
- ISO Guinea Pig Maximization Sensitization Test
- ISO Intracutaneous Study in Rabbits
- USP Rabbit Pyrogen Study
- ISO Acute Systemic Toxicity Study in Mice

### 3) Electrical safety and electromagnetic compatibility (EMC)

Electrical safety testing was conducted in accordance with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety.

### 4) Performance testing - Bench

Bench testing as listed below were conducted to ensure that the subject device performs as intended and meet design specifications.

- Thermal Safety
- Composite Durability
- Color performance
- Photobiological Safety
- Image Intensity Uniformity
- Resolution
- Field of View / Direction of View

### 5) Performance testing - Animal

No animal study was performed to demonstrate substantial equivalence.

### 6) Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

### 7) Risk management

Risk management was performed in accordance with ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk management.

### 8. CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 and CYF-5R raise no new issue of

safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness and performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K221690](https://fda-staging.innolitics.com/device/K221690)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
