IPL Hair Remover, Model: G993, G996, G998 and G885
K221679 · Shenzhen Mywin Technology Co., Ltd. · OHT · Jul 7, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221679
Device Name
IPL Hair Remover, Model: G993, G996, G998 and G885
Applicant
Shenzhen Mywin Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Jul 7, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.
Device Story
Handheld IPL hair removal device (models G993, G996, G998, G885) for over-the-counter use. Device uses a Xenon lamp to emit intense pulsed light (510-1100 nm) to heat hair roots. User operates device by applying it to skin; integrated skin proximity sensor prevents pulse emission unless full skin contact is detected. Features 5 intensity levels (1.8-4.5 J/cm²) controlled via finger switch. Powered by AC adapter. Intended for large areas including legs, arms, back, and abdomen. Benefits include home-based removal of unwanted hair.
Clinical Evidence
No clinical data provided. Safety and efficacy established via bench testing, electrical safety (IEC 60601-1 series), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10) leveraged from the predicate device. Usability study performed to support OTC use.
Technological Characteristics
Intense Pulsed Light (IPL) technology using a Xenon arc flashlamp. Wavelength: 510-1100 nm. Max fluence: 4.5 J/cm². Spot size: 4.3 cm². Pulse duration: 3 ms. Powered by external AC adapter (100-240V). Dimensions: 188 x 76 x 49 mm. Weight: 215.1g. Includes skin proximity sensor for safety. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, and IEC 60601-1-2.
Indications for Use
Indicated for the removal of unwanted body and/or facial hair in adults.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 7, 2022
Shenzhen Mywin Technology Co., Ltd. % Jet Li Regulation manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K221679
Trade/Device Name: IPL Hair Remover, Model: G993, G996, G998 and G885 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: June 8, 2022 Received: June 9, 2022
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221679
Device Name
IPL Hair Remover, model: G993, G996, G998 and G885
Indications for Use (Describe)
IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K221679
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
## 1 Submitter Information
Company Name: Shenzhen Mywin Technology Co., Ltd.
Address: 2nd Floor, Building B, New Age gongrong Industrial Zone, No. 2, Shihuan Road,
Shilong community, Shiyan Street, Baoan District, Shenzhen 518103, China
Phone: +86-18902842323
Contact Person (including title): Meiyun Liu (General Manager)
E-mail: Szmywin@163.com
Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd E-mail: med-jl@foxmail.com Phone: 86-18588874857 Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China
# 2 Subject Device Information
Type of 510(k) submission: Special 510(k) Device Modification Common Name: IPL Hair Remover Model: G993, G996, G998 and G885 Common Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Requlation Number: 878.4810
{4}------------------------------------------------
Requlation Class: 2
### 3 Predicate device Information
### Legally existing device K211368
Company Name:
Common Name: IPL Hair Remover
Model: G993, G996, G998 and G885
Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect
Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal
Effect
Review Panel: General & Plastic Surgery
Product Code: ONF
Regulation Number: 878.4810
Regulation Class: 2
### Reference device K211185
Common Name: IPL Home Use Hair Removal Device Model: D-1150, D-1171, D-1153, D-1155, D-1156, D-1126, D-1178, D-1187 Common Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 878.4810 Regulation Class: 2
### 4 Device Description
IPL Hair Remover Device (Model: G993, G996, G998 and G885), is an over-the-counter use device for removal of unwanted body and/or facial hair based on Intense Pulsed Light
{5}------------------------------------------------
(IPL) treatment. The device package includes IPL DEVICE, Power supply and User manual. It is supplied by AC power supply (Input 100-240 V AC). The weight of the device is 215.1g, and the size is 188 x 76 x 49mm. The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.
The device is equipped with a Xenon Lamp and a skin proximity sensor to detect appropriate skin contact. If the IPL Hair Remover Device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered for pulse emitting. The device can be used for large areas on legs, arms, back and abdomen. The product can flash for 100,000 times. .
#### Intended Use ട്
IPL Hair Remover is an over-the-counter device intended for removal of unwanted body and/or facial hair.
#### 6 Complied Standards
IPL Hair Remover G993, G996, G998 and G885 complies with the following FDA recognized consensus standards:
- 网 Electrical safety test according to IEC 60601-1, IEC60601-1-11 and IEC 60601-2-57 standards
- 风 Electromagnetic compatibility test according to IEC 60601-1-2 standard
- 风 ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
- 区 ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
# 7 Device modification description
The only modification of subject device is Intended Use. The device has changed to overthe-counter use. Therefore, the User Manual and Box labelling has been adding
{6}------------------------------------------------
corresponding use instructions for lay person user.
#### 8 Performance Testing
As the modification of subject device as above, results in no technological characteristics changes, the tests and data utilized to demonstrate safety and efficacy of the predicate device (legally existing device) are suitable for use in the assessment of the subject devices.
As there have been no changes to the performance of the subject device from the predicate device, this submission leverages performance and electrical testing provided in previous submission.
#### Biocompatibility 9
All the modified device materials which come in direct contact with the patient skin are biocompatible and identical to the materials used in the predicate device manufacturing. No biocompatibility test report is provided in this submission.
# 10 Clinical performance
Clinical performance is not deemed necessary.
# 11 Comparison with predicate device
Compare with predicate device (IPL Hair Remover K211368), the subject device is same in design principle, functions, material and the applicable standards. And the intended use of subject device is same to Reference device K211185. The differences between subject device and predicate devices do not raise any new questions of safety or effectiveness.
| Item | Subject Device | Predicate Device | Reference Device |
|--------------|-----------------------------------------|-----------------------------------------|------------------------------------------------------------------|
| Manufacturer | Shenzhen Myw in<br>Technology Co., Ltd. | Shenzhen Myw in<br>Technology Co., Ltd. | Shenzhen Beauty Every<br>Moment intelligent electric<br>Co.,Ltd. |
| K number | TBD | K211368 | K211185 |
{7}------------------------------------------------
| | IPL Hair Remover | IPL Hair Remover | IPL Home Use Hair Removal<br>Device |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | G993, G996, G998 and<br>G885 | G993, G996, G998 and<br>G885 | D-1150, D-1171, D-1153, D-<br>1155, D-1156, D-1126, D-<br>1178, D-1187 |
| Regulation &<br>Classification | Classification Name: Laser<br>surgical instrument for use in<br>general and plastic surgery<br>and in dermatology<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: OHT<br>Regulation Number:<br>878.4810<br>Regulation Class: 2 | Classification Name:<br>Powered Light Based Non-<br>Laser Surgical Instrument<br>With Thermal Effect<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: ONF<br>Regulation Number:<br>878.4810<br>Regulation Class: 2 | Classification Name: Laser<br>surgical instrument for use in<br>general and plastic surgery<br>and in dermatology<br>Review Panel: General &<br>Plastic Surgery<br>Product Code: OHT<br>Regulation Number:<br>878.4810<br>Regulation Class: 2 |
| Indications for<br>Use | IPL Hair Remover is an over-<br>the-counter device intended<br>for removal of unwanted<br>body and/or facial hair. | The IPL Hair Remover<br>Device, Model: G993, G996,<br>G998 and G885 is indicated<br>for the removal of unwanted<br>hair under the direction of a<br>physician, after training by a<br>healthcare professional. The<br>device is also indicated for<br>the permanent reduction in<br>hair regrowth, defined as the<br>long-term, stable reduction in<br>the number of hairs re-<br>growing when measured at<br>6, 9 and 12 months after the | IPL Home Use Hair Removal<br>Device is an over-the-<br>counter device intended for<br>removal of unwanted body<br>and/or facial hair. |
{8}------------------------------------------------
| | | completion of a treatment<br>regime. The device is used<br>for adults. | |
|---------------------------------|---------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------|
| | OTC Use | Prescription Use | OTC Use |
| Source Energy | Supplied by external adapter | Supplied by external adapter | Supplied by external adapter |
| Device Type | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light |
| | | | Regular window : 530- |
| Wavelength<br>(nm) | 510nm ~ 1100nm | 510nm ~ 1100nm | 1100nm, 590-1100nm |
| | | | Filter window: 600-1100nm |
| Max. Fluence | Max 4.5 J/cm² | Max 4.5 J/cm² | Max 4.5 J/cm² |
| Spot Size | 4.3 cm² | 4.3 cm² | 4.3 cm² |
| | Level 1: 1.8 J/cm² | Level 1: 1.8 J/cm² | |
| | Level 2: 2.3 J/cm² | Level 2: 2.3 J/cm² | |
| Light Intensity | Level 3: 3.2 J/cm² | Level 3: 3.2 J/cm² | 2.0 to 4.5 J/cm² |
| | Level 4: 4.4 J/cm² | Level 4: 4.4 J/cm² | |
| | Level 5: 4.5 J/cm² | Level 5: 4.5 J/cm² | |
| Pulse duration | 3 ms | 3 ms | 7.5~12 ms |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Pulsing Control | Finger switch | Finger switch | Finger switch |
| Number of<br>Output<br>Channels | One channel | One channel | One channel |
| Output Intensity<br>Level | 5 levels | 5 levels | / |
| Software<br>Control? | Yes | Yes | / |
| Weight | 215.1g | 215.1g | / |
| Dimensions | 188*76*49 mm (H*W*D) | 188*76*49 mm (H*W*D) | / |
| Electrical | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| safety, EMC | IEC 60601-1-11<br>IEC 60601-1-2<br>IEC 60601-2-57 | IEC 60601-1-11<br>IEC 60601-1-2<br>IEC 60601-2-57 | IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 |
| Biocompatibility | ISO10993-5<br>ISO10993-10 | ISO10993-5<br>ISO10993-10 | ISO10993-5<br>ISO10993-10 |
{9}------------------------------------------------
Sponsor: Shenzhen Mywin Technology Co., Ltd. Subject Device: IPL Hair Remover, Model: G993, G996, G998 and G885
### Note
Although the indication for use of subject device is different to predicate device K211368, but it is only to simplify the statement of the legally existing predicate device and remove the prescription use statement, which is not involved with technical specifications. The revised indication for use of subject device in this submission also had been covered by the indication for use of the legally existing predicate device; which is same to Reference device K211185. And the subject device had been verify by usability study for its OTC use. So the difference does not affect the safety and effectiveness.
### 12 Summary Prepared Date
28 Jun. 2022
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.