K221660 · Olympus Medical Systems Corp. · EOB · Dec 28, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K221660
Device Name
RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE-XP
Applicant
Olympus Medical Systems Corp.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Dec 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is intended to be used with an Olympus video system center, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis. RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is indicated for use within the nasal and nasopharyngeal lumen.
Device Story
The Rhino-Laryngofiberscope Olympus ENF Type XP is a flexible fiberscope used for endoscopic diagnosis of the nasal and nasopharyngeal lumen. It functions by transmitting light from an external source through optical fiber bundles to the distal end to illuminate the target area. The reflected image is captured by an objective lens and transmitted via image fiber bundles to an eyepiece. The observer can view the image directly through the eyepiece or via a monitor when connected to a video processor and camera head. The device is operated by clinicians in a clinical setting. It provides visual access to the nasal and nasopharyngeal anatomy, aiding in diagnosis. The device is a passive optical instrument; it does not contain software or automated analysis algorithms.
Clinical Evidence
No clinical data was provided. Substantial equivalence was supported by non-clinical bench testing, including thermal safety, durability, color performance, image intensity, resolution, photobiological safety, and verification of optical specifications (FOV, direction of view). Biocompatibility was evaluated per ISO 10993-1.
Technological Characteristics
Flexible fiberscope; optical fiber bundles for illumination and image transmission; 0° forward viewing; 75° field of view; 2.5-50 mm depth of field; 1.8 mm distal end diameter; 2.2 mm insertion tube diameter; 300 mm working length; 130° up/down angulation. Sterilization via EO gas. No software or electronic components integral to the scope itself.
Indications for Use
Indicated for use within the nasal and nasopharyngeal lumen for endoscopic diagnosis in patients requiring such procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 28, 2022
Olympus Medical Systems Corp. % Steven Keenan Program Manager, Regulatory Affairs Olympus Corporation of the Americas 800 West Park Drive Westborough, Massachusetts 01581
Re: K221660
Trade/Device Name: Rhino-Laryngofiberscope Olympus ENF Type XP Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: November 21, 2022 Received: November 21, 2022
Dear Steven Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221660
Device Name
RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP
Indications for Use (Describe)
RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is intended to be used with an Olympus video system center, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis. RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is indicated for use within the nasal and nasopharyngeal lumen.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the word "OLYMPUS" in a bold, sans-serif font. The text is a dark blue color, which contrasts with the white background. A registered trademark symbol is located to the right of the word.
### K221660 - 510(k) Summary
For
## RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP
| General Information | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan<br>192-8507<br>Phone: (+81) 42-642-2694<br>Fax: (+81) 42-642-2307<br>Establishment Registration Number: 8010047 |
| Manufacturer: | Aizu Olympus Co., Ltd.,<br>500 Muranishi, Niidera, Monden-machi,<br>Aizuwakamatsu-shi, Fukushima 965-8520,<br>Japan<br>Phone: (+81) 242-28-2111<br>Fax: (+81) 242-26-4234<br>Establishment Registration Number: 9610595 |
| 510(k) Submitter: | Olympus Corporation of the Americas<br>800 West Park Drive<br>Westborough, MA 01518 USA |
| Establishment Registration Number: | 2429304 |
| Contact Person: | Steven Keenan<br>Program Manager<br>Mobile: (508)-523-5004<br>Email: Steven.Keenan@olymus.com |
| Date Prepared: | June 3, 2022 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. A registered trademark symbol is located to the right of the letter "S".
### Device Description
| Model No. | Device/Trade Name | Product Classification |
|-----------|------------------------------------------------|------------------------|
| ENF-XP | RHINO-LARYNGOFIBERSCOPE<br>OLYMPUS ENF TYPE XP | EOB (874.4760) |
Classification Name: Generic/Common Name: Regulation Number: Regulatory Class: Product Codes: Review Panel:
Nasopharyngoscope (flexible or rigid) and accessories Rhino-Laryngo Fiberscope 874.4760 Class II EOB Ear Nose & Throat
### Predicate Device
| Predicate Device | 510(k) No. |
|------------------------------------------|------------|
| RHINO-LARYNGO FIBERSCOPE OLYMPUS ENF-GP2 | K181240 |
### Product Description
RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is intended to be used with an Olympus video system center, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis. RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP is indicated for use within the nasal and nasopharyngeal lumen.
### Comparison of Technological Characteristics
Table 5-1 compares RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP to the predicate device with respect to intended use, technological characteristics, and principle of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
### Table 5-1: Comparison of the technological characteristics of ENF-XP to predicate device
| Feature/<br>Technological<br>Characteristics | Subject Device<br>ENF-XP | Predicate Device<br>ENF-GP2 | Comparison |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory | | | |
| Device Name | RHINO-<br>LARYNGOFIBE<br>RSCOPE<br>OLYMPUS ENF<br>TYPE XP | Rhino-Laryngo<br>Fiberscope ENF-<br>GP2 | Different model names. |
| Regulatory Decision | This submission | K181240 | N/A |
| Feature/<br>Technological<br>Characteristics | Subject Device<br>ENF-XP | Predicate Device<br>ENF-GP2 | Comparison |
| Product Code | EOB | EOB | Same as predicate. |
| Regulation Number | 874.4760 | 874.4760 | Same as predicate. |
| Regulation Name | Nasopharyngosco<br>pe (flexible or<br>rigid) and<br>accessories | Nasopharyngosco<br>pe (flexible or<br>rigid) and<br>accessories | Same as predicate. |
| Intended Use | RHINO-<br>LARYNGOFIBE<br>RSCOPE<br>OLYMPUS ENF<br>TYPE XP is<br>intended to be<br>used with an<br>Olympus video<br>system center,<br>documentation<br>equipment,<br>display monitor,<br>and other<br>ancillary<br>equipment for<br>endoscopic<br>diagnosis.<br>RHINO-<br>LARYNGOFIBE<br>RSCOPE<br>OLYMPUS ENF<br>TYPE XP is<br>indicated for use<br>within the nasal<br>and<br>nasopharyngeal<br>lumen. | This instrument is<br>intended to be<br>used with an<br>Olympus light<br>source,<br>documentation<br>equipment,<br>display monitor<br>and other<br>ancillary<br>equipment for<br>endoscopic<br>diagnosis. This<br>instrument is<br>indicated for use<br>within the nasal<br>and<br>nasopharyngeal<br>lumen. | Similar to predicate. Device name<br>and model number are now<br>included in the Indications for Use<br>statement. The intended use is<br>identical to predicate. |
| Feature/<br>Technological<br>Characteristics | Subject Device<br>ENF-XP | Predicate Device<br>ENF-GP2 | Comparison |
| Mode of Action | When observing<br>nasal or<br>nasopharyngeal<br>lumen, the light<br>emitted from light<br>source enters into<br>the fiberscope via<br>apparatus on<br>connector section<br>and is eventually<br>transmitted and<br>released from<br>examination light<br>outlet in the distal<br>end to supply<br>light by optical<br>fiber bundles.<br>Subsequently the<br>illuminated image<br>reflected into<br>objective lens is<br>sent to the<br>eyepiece section<br>by image fiber<br>bundles. The<br>observer can<br>observe the image<br>directly in the<br>eyepiece section<br>or alternatively,<br>the image can be<br>observed on a<br>monitor generated<br>by a video<br>processor and a<br>camera head<br>attached on the<br>eyepiece. | When observing<br>nasal or<br>nasopharyngeal<br>lumen, the light<br>emitted from light<br>source enters into<br>the fiberscope via<br>apparatus on<br>connector section<br>and is eventually<br>transmitted and<br>released from<br>examination light<br>outlet in the distal<br>end to supply light<br>by optical fiber<br>bundles.<br>Subsequently the<br>illuminated image<br>reflected into<br>objective lens is<br>sent to the<br>eyepiece section<br>by image fiber<br>bundles. The<br>observer can<br>observe the image<br>directly in the<br>eyepiece section<br>or alternatively,<br>the image can be<br>observed on a<br>monitor generated<br>by a video<br>processor and a<br>camera head<br>attached on the<br>eyepiece. | Same as predicate. |
| Optical System Parameters | | | |
| Field of View | 75° | 85° | Although the field of view is<br>different from PD, the area within<br>the view can be adjusted by the<br>advance or retreat of the endoscope.<br>Therefore, the difference in FOV<br>does not affect the safety and<br>effectiveness of the subject device. |
| Direction of View | 0° (Forward<br>viewing) | 0° (Forward<br>viewing) | Same as predicate. |
| Feature/<br>Technological<br>Characteristics | Subject Device<br>ENF-XP | Predicate Device<br>ENF-GP2 | Comparison |
| Depth of Field | 2.5-50 mm | 5-50 mm | The increased range of depth of<br>field covers the PD's depth of field.<br>In addition, although the subject<br>device is compatible with various<br>processors, light sources, and<br>camera heads, this specification is<br>not affected by the variation of the<br>systems. |
| | Bending Section | | |
| Angulation range | Up 130° / Down<br>130° | Up 130° / Down<br>130° | Same as predicate. |
| | Insertion section | | |
| Insertion Tube<br>Diameter – Distal<br>End | 1.8 mm | 3.1 mm | The difference in tube diameter of<br>the distal end for the subject<br>device does not alter or change the<br>indications for use or result in a<br>new intended use. |
| Insertion Tube<br>Diameter –<br>Flexible Outer<br>Diameter | 2.2mm | 3.5mm | The difference in flexible outer<br>diameter of the insertion tube for<br>the subject device does not alter or<br>change the indications for use or<br>result in a new intended use. |
| Insertion Section<br>Working Length | 300 mm | 300 mm | Same as predicate. |
| | Connection to Light Source | | |
| Configuration | Light Guide (LG)<br>cable is not<br>detachable | LG cable is<br>detachable | Cable configuration does not<br>directly affect the safety and<br>effectiveness of the subject device. |
| | Venting Connector | | |
| Venting<br>Connector<br>Position | On LG connector | On control body | Venting connector position does<br>not directly affect the safety and<br>effectiveness of the subject device. |
| | Sterilization | | |
| EO | Available | Available | Same as predicate. |
| | Other | | |
| Suction Function | Not provided | Not provided | Same as predicate. |
| Total Length | 530 mm | 550 mm | Due to the difference in the shape<br>of the control section, the subject<br>device has a 20mm shorter total<br>length compared with the predicate<br>device. This difference does not<br>alter or change the indications for<br>use or result in a new intended use. |
| Feature/<br>Technological<br>Characteristics | Subject Device<br>ENF-XP | Predicate Device<br>ENF-GP2 | Comparison |
| Compatible<br>Processor | OTV-S200/S300<br>(with light source<br>function) | OTV-SC/SC2<br>OTV-S7V<br>CV-170<br>OTV-S7Pro<br>OTV-S190 | Compatibility of ENF- XP with OTV-<br>S200/S300 was demonstrated with<br>bench performance testing in Section<br>18 confirm that these additional<br>processors do not raise any new<br>questions of safety or effectiveness. |
| Compatible Light<br>Source | OTV-S200/S300<br>(with light source<br>function) | CLV-S190<br>CLV-S40Pro<br>CLH-SC<br>CLL-V1<br>CLL-S1<br>CV-170 | Compatibility of ENF- XP with OTV-<br>S200/S300 (with light source<br>function) was demonstrated with<br>bench performance testing in Section<br>18 confirm that these additional light<br>sources do not raise any new<br>questions of safety or effectiveness. |
| Compatible<br>Monitor | OEV-262H<br>LMD-X310ST*<br>*This can only be<br>combined with<br>OTV-S300. | OEV-141<br>OEV-142<br>OEV-143<br>OEV-201<br>OEV-202<br>OEV-203<br>OEV-181H<br>OEV-191<br>OEV-191H<br>OEV-261H | Compatibility of ENF- XP with OEV-<br>262H and LMD-X310ST was<br>demonstrated with bench performance<br>testing in Section 18 confirm that<br>these additional monitors do not raise<br>any new questions of safety or<br>effectiveness. |
| Compatible<br>Camera Head | CH-S200-XZ-EA<br>CH-S200-XZ-EB | OTV-S7H-1NA-<br>10E/12E<br>OTV-S7H-<br>1N/1NA<br>OTV-S7ProH-<br>HD-10E/12E | Compatibility of ENF- XP with CH-<br>S200-XZ-EA and CH-S200-XZ-EB<br>camera heads was demonstrated with<br>bench performance testing in Section<br>18 confirm that these additional<br>camera heads do not raise any new<br>questions of safety or effectiveness. |
{5}------------------------------------------------
# OLYMPUS®
#### Traditional 510(k) Notification RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP
{6}------------------------------------------------
# OLYMPUS®
#### Traditional 510(k) Notification RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A registered trademark symbol is located to the right of the word.
{8}------------------------------------------------
#### Traditional 510(k) Notification RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP
### Indications for Use
RHINO-LARYNGOFIBROSCOPE OLYMPUS ENF TYPE XP is intended to be used with an Olympus video system center, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis. RHINO-LARYNGOFIBROSCOPE OLYMPUS ENF TYPE XP is indicated for use within the nasal and nasopharyngeal lumen.
### Compliance to Voluntary Standards
The following voluntary standards have been applied to the subject devices respectively:
| Standard | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI ES 60601-<br>1:2005+A1:2012 | Medical electrical equipment – Part 1: General<br>requirements for basic safety and essential performance |
| IEC 60601-2-18:2009 | Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment |
| IEC 62471:2006 | Photobiological safety of lamps and lamp systems |
| ISO 10993-1:2009 | Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
| ISO 10993-5:2009 | Biological evaluation of medical devices – Part 5: Tests for in vitro cytotox…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.