K221655 · Vizaramed, Inc. · DYB · Nov 14, 2022 · Cardiovascular
Device Facts
Record ID
K221655
Device Name
Multiflex Steerable Sheath
Applicant
Vizaramed, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Nov 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Device Story
Multiflex Steerable Sheath is a sterile, single-use, percutaneous introducer; facilitates placement of cardiovascular catheters into the heart. Device features steerable shaft capable of 3D shapes; handle with lockable steering controls; Tuohy-Borst adapter with hemostasis valve; side port; atraumatic soft distal tip. Shaft is radiopaque for fluoroscopic visualization. Used in clinical settings by physicians; provides access path for catheters. Benefits include controlled navigation to target cardiac sites. Device is mechanical; no software or algorithms involved.
Clinical Evidence
No clinical data. Evidence consists of bench testing including dimensional verification, tensile testing, leak testing, and biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility). Usability assessment conducted via physician questionnaire in a simulated heart model.
Technological Characteristics
Deflectable, percutaneous steerable sheath with dilator. Available in 12.5F and 15.5F sizes. Features lockable steering handle, Tuohy-Borst adapter, hemostasis valve, and radiopaque shaft. Sterilization via ethylene oxide (EO) per ISO 11135. Limited contact (≤24 hrs) externally communicating device.
Indications for Use
Indicated for patients requiring cardiovascular catheter introduction into the heart, including access to the left side of the heart via the interatrial septum.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
St. Jude Medical Agilis Steerable Introducer (K081645)
Reference Devices
Cook, Inc. Extra Large Check-Flow Introducer (K203670)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2022
VizaraMed, Inc. Jack P Douglas, PhD Vice President, Regulatory Affairs 1914 O'Toole Way San Jose, California 95131
Re: K221655
Trade/Device Name: Multiflex Steerable Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: June 3, 2022 Received: June 7, 2022
Dear Dr. Jack Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
S
Ankurita Datta - Digitally signed by Ankurita Datta -S Date: 2022.11.14 14:35:55 -05'00'
for Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221655
Device Name Multiflex Steerable Sheath
Indications for Use (Describe)
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 1. Date prepared | June 1, 2022 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Manufacturer | VizaraMed, Inc. |
| 3. Contact | Jack P Douglas, PhD<br>Vice President, Regulatory Affairs<br>VizaraMed, Inc.<br>1914 O'Toole Way<br>San Jose, CA 95131<br>(510) 792-7477 |
| 4. Contract Manufacturer Name | Novel Cath<br>(A Cirtec Company)<br>90 Great Oaks Blvd.<br>San Jose, CA 95119 |
| 5. Device Identification | Trade Name: Multiflex Steerable Sheath<br>Common Name: Introducer, Catheter<br>Classification Name: Catheter Introducer<br>Class: Class II, 21 CFR 870.1340<br>Product Code: DYB<br>Classification Panel: Cardiovascular |
# 510(k) SUMMARY
#### 6. Device Description
The VizaraMed Multiflex Steerable Sheath is a deflectable, sterile, single-use, percutaneous steerable sheath with dilator used to facilitate placement of various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
The sheath is available in two sizes and consists of a steerable shaft capable of threedimensional shapes, a handle with lockable steering controls, a Tuohy-Borst adapter with a hemostasis valve, a side port, and atraumatic soft distal tip. The shaft is radiopaque for visualization under fluoroscopy. The accompanying dilator is packaged with the sheath in a pouch for removal.
The 12.5F sheath can accommodate devices of 3F(0.039") - 12.5F (0.162"). The 15.5F sheath can accommodate devices sized 3F (0.039") - 15.5F (0.203").
{4}------------------------------------------------
- 7. Intended Use
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
- 8. Predicate Device
- St. Jude Medical Agilis Steerable Introducer [K081645]
- 9. Reference Device
- Cook, Inc. Extra Large Check-Flow Introducer [K203670] ●
- 10. Characteristics of Substantial Equivalence
The VizaraMed Multiflex Steerable Sheath is substantially equivalent to the predicate device based on the following:
- Indications for Use
- Specifications ●
- Product features ●
- Technological characteristics ●
- Product design
- Materials
- Sterilization method
- Mechanical testing
- Biocompatibility testing ●
The subject device was subjected to applicable testing to ensure reliable design and performance under testing parameters. The testing performed is described below:
{5}------------------------------------------------
## 11. Mechanical Properties
Mechanical testing, including dimensional verification, tensile testing, and leak testing, were conducted to ensure the Multiflex Steerable Sheath meets design requirements to support substantial equivalence to the predicate.
## 12. Sterilization and Packaging
Sterilization validation via ethylene oxide (EO) per ISO 11135 was conducted to ensure design requirements and to support substantial equivalence to the predicate. Sterilization validation used fractional and full cycles, and appropriate sample size of final preconditioned, packaged product that incorporated the use of biological indicators (BI) for the inoculated product (IP).
### 13. Usability Assessment
Subjective and objective evidence was obtained via questionnaire from physicians to assess critical tasks in a simulated heart model. Training was provided followed by user evaluation while manipulating the device.
### 14. Biocompatibility
The Vizaramed Sheath is considered to be a limited contact (≤24 hrs.), externally communicating device used in circulating blood. Testing was conducted for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity and hemocompatibility. All tests were found to meet acceptance criteria, supporting substantial equivalence to the predicate.
#### 15. Conclusions
The results of the tests support the conclusion that the VizaraMed Steerable Sheath meets design input requirements based on the intended use and that it is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.