K221653 · International Biomedical · IMD · Oct 6, 2022 · Physical Medicine
Device Facts
Record ID
K221653
Device Name
InfaTherm Disposable Infant Warming Mattress
Applicant
International Biomedical
Product Code
IMD · Physical Medicine
Decision Date
Oct 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5710
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
The InfaTherm™ Disposable Infant Warming Mattress is an instant warm pack intended to provide warmth during transport of an infant within the hospital or between hospitals. It is recommended for full-term infants. It is a single use, nontoxic, non-sterile, disposable device.
Device Story
InfaTherm™ is a single-use, disposable, non-sterile infant warming mattress. Device consists of a rectangular pouch containing a supersaturated solution of food-grade sodium acetate, water, and a metal catalyst disk. Operation: user flexes the internal catalyst disk to initiate an exothermic crystallization reaction, generating heat. The mattress reaches 100-102°F when activated at 75°F, providing warmth to the infant during transport. Used in hospital settings or during patient transfer; operated by clinical staff. Output is passive thermal energy. Benefits include maintenance of infant body temperature during transport, reducing risk of hypothermia.
Clinical Evidence
Bench testing only. Testing included temperature profile analysis (activation, surface, duration, and skin contact temperatures), shelf-life stability (20 months), seal strength (80 psi cross seal/50 psi inline seal), and burst testing (40 lbs for 15 minutes). Biocompatibility evaluation leveraged from K936084 per ISO-10993 standards.
Technological Characteristics
Materials: Polyethylene LLDPE pouch with polyester (PET) bottom. Principle: Exothermic chemical crystallization of supersaturated sodium acetate solution triggered by a metal catalyst disk. Dimensions: ~10" x 16". Single-use, non-sterile. No electronics, software, or external energy source.
Indications for Use
Indicated for full-term infants requiring warmth during hospital or inter-hospital transport.
Regulatory Classification
Identification
A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.
Predicate Devices
Rapid Aid Corp. – Infant Transport Mattress Warmer with Disc (K163295)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.
October 6, 2022
International Biomedical Amy Pieper Director of Regulatory Affairs 8206 Cross Park Drive Austin, Texas 78754
Re: K221653
Trade/Device Name: InfaTherm™ Disposable Infant Warming Mattress Regulation Number: 21 CFR 890.5710 Regulation Name: Hot Or Cold Disposable Pack Regulatory Class: Class I, reserved Product Code: IMD Dated: June 7, 2022 Received: June 7, 2022
Dear Amy Pieper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K221653
Device Name
InfaTherm™ Disposable Infant Warming Mattress
Indications for Use (Describe)
The InfaTherm™ Disposable Infant Warming Mattress is an instant warm pack intended to provide warmth during transport of an infant within the hospitals. It is recommended for full-term infants. It is a single use, nontoxic, non-sterile, disposable device.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### Submitter Information:
### Regulatory Affairs Contact:
International Biomedical 8206 Cross Park Drive Austin, TX 78754 U.S.A.
Amy Pieper Director of Regulatory Affairs (512) 873-0033 - phone (512) 873-9090 - fax
Date Summary Prepared: October 6, 2022
### Device Identification:
Trade Names: InfaTherm™ Disposable Infant Warming Mattress Common Name: Infant Warming Mattress Regulatory Class: I Regulatory Name: Hot or cold disposable pack Regulatory Number: (21 CFR 890.5710) Product Code: IMD 510(k) number: K221653
## Predicate Device:
Rapid Aid Corp. – Infant Transport Mattress Warmer with Disc – K163295
## Reference Device:
Omni Therm Inc.- Omni Warm Gel Packs - K936084
## Device Description:
The InfaTherm™ Disposable Infant Warming Mattress is a single use, disposable pouch that is filled with a super saturated solution of food grade sodium acetate and water and a catalyst disk. When the catalyst disk inside the packet is flexed by the user, the catalyst disk reacts with the nontoxic solution and initiates a chemical exothermic crystallization of the sodium acetate, generating heat. When activated at 75°F, the Infant warming mattress will reach between 100-102°F and then gradually decrease in temperature over time.
The InfaTherm™ Disposable Infant Warming Mattress is rectangular in shape. Immediately following activation of the mattress, the user places the infant on the mattress.
{4}------------------------------------------------
# Indications for Use:
The InfaTherm™ Disposable Infant Warming Mattress is an instant warm pack intended to provide warmth during transport of an infant within the hospital or between hospitals. It is recommended for full-term infants. It is a single use, nontoxic, non-sterile, disposable device.
## Substantial Equivalence:
The substantial equivalence of the International Biomedical InfaTherm™ Disposable Infant Warming Mattress to the predicate is shown by similarity in intended use, indications for use, materials and performance. The table below provides a comparison of the technological characteristics of the subject device to the predicate.
| System | International Biomedical | Rapid Aid Corp | Comparison |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Specification | InfaTherm™ Disposable Infant | Infant Transport Mattress | |
| | Warming Mattress | Warmer with Disc k163295 | |
| | Subject Device | Predicate Device | |
| Indications for<br>Use | The InfaTherm™ Disposable<br>Infant Warming Mattress is an<br>instant warm pack intended to<br>provide warmth during<br>transport of an infant within the<br>hospital or between hospitals. It<br>is recommended for full-term<br>infants. It is a single use,<br>nontoxic, non-sterile, disposable<br>device. | The Rapid Aid Infant<br>Transport Mattress Warmer<br>with Disc provides warmth<br>during transport of an infant<br>within the hospital or<br>between hospitals. It is<br>recommended for full-term<br>infants. | Same |
| Environment of<br>Use | Hospitals or between hospitals | Hospitals or between<br>hospitals | Same |
| Prescriptive | Yes | Yes | Same |
| Pouch Material | Polyethylene LLDPE with white<br>Polyester extrusion (PET)<br>material on bottom of bag | Polyethylene/60g polyamide<br>(PE/PA) with woven cloth<br>material on bottom outside<br>of bag | Similar – both devices utilize<br>polyethylene pouches with a cloth<br>material on the bottom. The<br>predicate device does not provide<br>enough detail on the cloth<br>material to claim a complete<br>match. Comparison between two<br>devices does not indicate a<br>material difference that would<br>introduce any new issues of safety<br>and effectiveness. |
| Solution Material | Sodium acetate (food grade),<br>water, activation disc;<br>supersaturated solution | Sodium acetate and water,<br>thickener, activation disc;<br>supersaturated solution | Similar – the subject device<br>omitted the thickening agent that<br>is used in the predicate. The<br>thickening agent only serves to |
| System<br>Specification | International Biomedical<br>InfaTherm™ Disposable Infant<br>Warming Mattress<br>Subject Device | Rapid Aid Corp<br>Infant Transport Mattress<br>Warmer with Disc k163295<br>Predicate Device | Comparison |
| | | | make the gel more viscous when activated. Upon activation, the liquid becomes viscous by nature of the chemical reaction, a thickening agent is redundant. Omitting the thickening agent doesn't introduce any new issues of safety and effectiveness. |
| Activation<br>Method | Activating Disc triggers the exothermic reaction | Activating Disc triggers the exothermic reaction | Same |
| Average<br>Device<br>Maximum<br>Surface<br>Temperature | 100-102°F | 101-104°F | Similar - Subject Device average temperature has a slightly lower peak temperature. Lower peak temperature doesn't introduce any new issues of safety and effectiveness. |
| Average Skin<br>Surface<br>Temperature | 101°F | Not Publicly Available | Similar – The predicate device does not publish the skin surface temperature, but the subject device temperature is in line with or slightly lower than the predicate for device surface temperature and there is a strong correlation in this type of product between the device surface temperature and the skin surface temperature. Therefore the stated skin surface temperature of the subject device doesn't introduce any new issues of safety and effectiveness. |
| Size of Pouch | Approximately 10" x 16" (160 sq in) | 15.7" x 9.6" (151 sq in) | Similar - subject device is <0.5" larger. Slight change in size does not introduce any new issues of safety and effectiveness. |
| Sterility, Number<br>of Uses | Non-sterile, Single Use, Disposable | Non-sterile, Single Use, Disposable | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for International Biomedical. The logo consists of a stylized symbol resembling a plant or abstract figure on the left, followed by the words "international" in a smaller font size above the word "BIOMEDICAL" in a larger font size. A registered trademark symbol is placed to the right of the word "BIOMEDICAL".
#### Bench Testing:
The following tests/evaluations were performed
- Temperature Profile Testing
{6}------------------------------------------------
- Temperature profile testing, including skin contact temperature, was performed O on the International Biomedical InfaTherm™ Disposable Infant Warming Mattress and compared to the date from the predicate device. Testing included activation temperature testing, surface temperature, temperature duration testing and skin temperature. All tests were a pass.
- Stability Testing ●
- Stability Testing was performed on the International Biomedical InfaTherm™ o Disposable Infant Warming Mattress in conjunction with the temperature profile testing to confirm its shelf life of 20 months.
- Seal Strength Testing ●
- The Subject device has been evaluated for seal strength testing of 80 psi pressure O (pull test) for a cross seal and 50 Psi Pressure for an inline seal by the pouch manufacturer. The filled and sealed pouch is then subjected to a 40 pound for 15 minute burst test. We consider the burst and seal integrity testing to be safe and effective.
- Biocompatibility Testing ●
- The biocompatibility evaluation for the International Biomedical InfaTherm™ O Disposable Infant Warming Mattress was leveraged from k936084. Therefore, the evaluation is considered to be in accordance with the FDA guidance on Biocompatibility on the International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing".
# Conclusion:
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate, the International Biomedical InfaTherm™ Disposable Infant Warming Mattress has been shown to be substantially equivalent to the predicate device identified, and does not present any new issues of safety or effectiveness
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.