K221648 · Iconic Medicare Sdn Bhd · LYY · Aug 25, 2022 · General Hospital
Device Facts
Record ID
K221648
Device Name
Iconic Latex Glove
Applicant
Iconic Medicare Sdn Bhd
Product Code
LYY · General Hospital
Decision Date
Aug 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Iconic Latex Glove is a non-sterile, powder-free, natural rubber latex patient examination glove. Device is worn by clinicians or users to provide a barrier against contamination between patient and examiner. Available in sizes Small, Medium, Large, and X-Large. Device is manufactured to meet ASTM D3578-19 standards for physical dimensions, tensile strength, and elongation. It is a single-use, disposable device. No electronic, software, or algorithmic components are involved.
Clinical Evidence
No clinical data. Bench testing only. Device demonstrated compliance with ASTM D3578-19 (dimensions, physical properties), ASTM D5151-19 (freedom from holes), ASTM D6124-06 (residual powder), and ASTM D5712-15 (extractable protein). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), ISO 10993-11 (systemic toxicity), and ISO 10993-23 (irritation).
Technological Characteristics
Material: Natural rubber latex. Non-sterile, powder-free. Dimensions and physical properties meet ASTM D3578-19. Freedom from holes per ASTM D5151-19. Residual powder < 2.0 mg/glove per ASTM D6124-06. Extractable protein < 50 µg/dm² per ASTM D5712-15. Biocompatibility per ISO 10993-5, 10, 11, 23.
Indications for Use
Indicated for use as a disposable, non-powdered patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Latex Examination Gloves Powder Free Careglove Global SDN BHD (K161833)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2022
Iconic medicare sdn bhd % A.C. Thirumaran Official Correspondent Integrated Assessment Services Pvt Ltd 1495, Manasarovar, 16th Main road, Anna Nagar West Chennai, Tamil Nadu 600040 India
Re: K221648
Trade/Device Name: Iconic Latex Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LYY Dated: May 20, 2022 Received: June 7, 2022
Dear A.C. Thirumaran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bifeng Qian, M.D., Ph.D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221648
Device Name Iconic Latex Glove
Indications for Use (Describe)
This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1.0 Submitter:
Mr. Tan Cho Chia Managing Director Company Name: Iconic Medicare Sdn Bhd. Company Address: PMT 798, LINGKARAN CASSIA SELATAN. TAMAN PERINDUSTRIAN BATU KAWAN, 14110 BANDAR CASSIA, PULAU PINANG. MALAYSIA Email: cctan@iconic.com.my Telephone: 60 4 504159 Date of Summary Prepared: 15th March 2022
#### 2.0 Subject Device Identification:
Trade Name / Proprietary Name: Iconic Latex Glove Device Common Name: Non-powder Patient Examination gloves. Device Classification Name: Natural Rubber Latex Patient Examination gloves Device Classification: 1 Regulation Number: 21 CFR 880.6250 Product Code: LYY
#### 3.0 Official Correspondent
Mr. A.C. Thirumaran Integrated Assessment Services Private Limited No.1495, Manasarovar, 16th Main road, Anna Nagar west, Chennai- 600040, India. Telephone: 91-44-26162670 Email: iasfda16@gmail.com
### 4.0 Identification of the Legally Marketed Device:
Predicate Device: Careglove Global SDN BHD 510k Number: - K161833 Device Name: Latex Examination Gloves Powder Free Classification Name: Natural Rubber Latex Patient Examination Gloves Device Classification: 1 Regulation Number: 21 CFR 880.6250
#### 5.0 Device Description
The subject device in this 510(k) Notification is Iconic Latex Glove - Powder Free Latex Examination Glove. The subject device is a patient examination glove made from Latex compound, Natural White color, powder free and non-sterile (Per 21 CFR 880.6250 - class I). The device meets the specifications in ASTM D3578-19 Standard specification for Rubber Examination Gloves. The available sizes of the subject devices are Small, Medium, Large, X-Larqe.
### 6.0 Indications for Use
{4}------------------------------------------------
lconic Latex Glove. This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
### 7.0 Technological characteristics Comparison for the proposed and predicate devices
The Iconic Latex Glove is a Powder Free Latex Examination Glove, Non-sterile, with the following summarized technological characteristics, in comparison to the Predicate device and specifications in ASTM D3578-19.
| Characteristics | Acceptance Criteria | Subject device:<br>Iconic Latex Glove<br>(Small, Medium, Large,<br>X-Large) | Predicate Device Latex<br>Examination Gloves<br>Powder Free Careglove<br>Global SDN BHD | Comparison<br>Analysis |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| 510(k) Number | - | K221648 | K161833 | - |
| Product Code | LYY | LYY | LYY | same |
| Intended use | A powder free<br>patient examination<br>glove is a<br>disposable device<br>intended for<br>medical<br>purposes that is<br>worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. The<br>device is for over-<br>the-counter use. | This powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand to<br>prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the- counter use. | A patient examination<br>glove is a disposable<br>device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | same |
| Material used | Latex - Natural<br>Rubber | Latex - Natural Rubber | Latex - Natural Rubber | same |
| Color | N/A | Natural White | Natural White | same |
| Sterility | Sterile/Non-sterile | Non sterile | Non sterile | same |
| Single use | Single use | Single use | Single use | same |
| | Overall Length<br>(mm) Min 230mm | Meets ASTM D3578-19 | Meets ASTM D3578-19 | |
| Dimensions | Width (±10mm)<br>Small - 80<br>Medium- 95<br>Large-110<br>X-large-120 | Meets ASTM D3578-19 | Meets ASTM D3578-19 | same |
| Characteristics | Acceptance Criteria | Subject device:<br>Iconic Latex Glove<br>(Small, Medium, Large,<br>X-Large) | Predicate Device Latex<br>Examination Gloves<br>Powder Free Careglove<br>Global SDN BHD | Comparison<br>Analysis |
| Physical<br>Properties | Thickness at Palm<br>(mm) Min; 0.05 mm | Meets ASTM D3578- 19 | 0.06-0.09mm | same |
| | Thickness at Finger<br>Tip (mm) Min 0.05<br>mm | Meets ASTM D3578- 19 | 0.07-0.10mm | |
| Physical<br>Properties | Before Aging<br>Tensile Strength =<br>18 MPa, min.<br>Ultimate<br>Elongation = 650<br>% min Stress at<br>500 % Elongation<br>(MPa)<br>After Accelerated<br>Aging<br>Tensile Strength<br>=14 MPa, min.<br>Ultimate<br>Elongation = 500<br>% min | Meets ASTM D3578- 19 | Meets ASTM D3578- 19 | same |
| Freedom from<br>pinholes | AQL 2.5 Inspection<br>Level G-1 | Meets ASTM D5151- 19 | Meets ASTM D5151- 19 | same |
| Residual<br>Powder | < 2.0 mg/pc | Meets ASTM D6124- 06 | Meets ASTM D6124- 06 | same |
| Bio-<br>Compatibility | ISO 10993-23:2010<br>Biological<br>evaluation of<br>medical devices:<br>Tests for irritation | Under the conditions of<br>this study, the test article<br>was a non- irritant. | N/A | same |
| | ISO 10993-10:2021<br>Biological<br>evaluation of<br>medical devices -<br>Part<br>10: Tests For skin<br>sensitization | Under the conditions of<br>this study, the test article<br>was a non- sensitizer. | N/A | same |
| | ISO 10993-5:2009<br>Biological<br>evaluation of<br>medical devices -<br>Part<br>5: Tests for in vitro<br>cytotoxicity | Complies with the<br>requirement of this<br>standard | N/A | same |
| Characteristics | Acceptance Criteria | Subject device:<br>Iconic Latex Glove<br>(Small, Medium, Large,<br>X-Large) | Predicate Device Latex<br>Examination Gloves<br>Powder Free Careglove<br>Global SDN BHD | Comparison<br>Analysis |
| | ISO 10993-11:2017<br>Biological<br>evaluation of<br>medical devices —<br>Part<br>11: Tests for<br>systemic toxicity | Not induce systemic<br>toxicity | N/A | Same |
| Extractable<br>Protein | Water Extractable<br>Protein, Maximum<br>50 µg/dm²<br>ASTM D 5712-<br>15 | Small = 10.3 µg/dm²<br>Medium = 8.7 µg/dm²<br>Large = 9.9 µg/dm²<br>X Large = 8.8 µg/dm² | < 50 µg/dm² | Same |
{5}------------------------------------------------
# 510(k) SUMMARY 510(k) No: K221648
{6}------------------------------------------------
# 510(k) SUMMARY 510(k) No: K221648
## 8.0 Summary of non-clinical testing results
lconic Nitrile Glove was tested and found in conformance with the following standards:
| ASTM D3578-19 | Standard Specification for Rubber Examination Gloves for Medical Application |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D5151-19 | Standard Test Method for detection of Holes in Medical Gloves |
| ASTM D6124-06 | Standard Test Method for Residual Powder on Medical Gloves |
| ASTM D5712 -15 | Standard Test Method for Analysis of Aqueous Extractable Protein in Latex,<br>Natural Rubber, and Elastomeric Products Using the Modified Lowry Method |
| ISO 10993-5:2009 | Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity |
| ISO 10993-10:2021 | Biological evaluation on medical device Part 10: Test for Skin Sensitization |
| ISO 10993-11:2017 | Biological evaluation of medical devices Part 11: Tests for systemic toxicity |
| ISO 10993-23:2021 | Biological evaluation of medical devices Part 23: Tests for irritation |
| Test<br>Methodology | Purpose | Acceptance<br>Criteria | Average Results | | | | Final<br>Results |
|---------------------|--------------------------------------------|---------------------------------------------------------------------------|-----------------|-------------|-------------|-------------|------------------|
| | | | SMALL | Medium | Large | X Large | |
| | Sterility | - | Non sterile | Non sterile | Non sterile | Non sterile | - |
| ASTM D3578-<br>19 | Freedom from<br>hole - ASTM<br>D5151-19 | AQL 2.5 | Pass | Pass | Pass | Pass | Pass |
| | | Overall Length<br>(mm) Min<br>230mm. | 230.8 | 240.5 | 237.5 | 236.8 | Pass |
| | Dimension - width,<br>Length,<br>Thickness | Width (±10mm)<br>Small - 80<br>Medium - 95<br>Large -111<br>X-large - 120 | 83 | 96 | 108.4 | 118.7 | Pass |
| | | Thickness at Palm & fingertip Min: 0.08 mm | | | | | |
{7}------------------------------------------------
## 510(k) SUMMARY 510(k) No: K221648
| Test<br>Methodology | Purpose | Acceptance<br>Criteria | Average Results | | | | Final |
|---------------------|--------------------------------------------------------------------|--------------------------------------|-----------------|--------|--------|---------|---------|
| | | | SMALL | Medium | Large | X Large | Results |
| | | Palm | 0.14 | 0.13 | 0.12 | 0.14 | Pass |
| | | Fingertip | 0.10 | 0.10 | 0.09 | 0.14 | Pass |
| | | a. Before Aging | | | | | |
| | | Tensile<br>Strength=18<br>MPa, min. | 20.8 | 23.4 | 21.4 | 20.7 | Pass |
| | | Ultimate<br>Elongation=<br>650% min | 1018.7 | 878.9 | 1127.9 | 1097.6 | Pass |
| | Physical<br>properties before<br>aging, after<br>accelerated aging | Stress at 500%<br>Elongation | 3.3 | 3.9 | 3.7 | 3.3 | Pass |
| | | b. After Accelerated Aging | | | | | |
| | | Tensile<br>Strength=14<br>MPa, min. | 15.3 | 17.3 | 16.9 | 17.4 | Pass |
| | | Ultimate<br>Elongation= 500<br>% min | 856.4 | 890.8 | 1077.3 | 956.8 | Pass |
| | Powder-free<br>Residue exceeds<br>maximum limit -<br>ASTM D6124-06 | < 2.0 mg per<br>glove | 0.12 | 0.26 | 0.38 | 0.18 | Pass |
| | Extractable<br>Protein content | $50 µg/dm2$ | 10.3 | 8.7 | 9.6 | 8.8 | Pass |
| ISO 10993-5 | Test for Invitro<br>cytotoxicity | Cytotoxic<br>Characteristics | | | | | Pass |
| ISO 10993-10 | Test for irritation<br>and Skin<br>Sensitization | Non - Skin<br>Sensitized | | | | | Pass |
| ISO 10993-11 | Tests for systemic<br>toxicity | Not induce<br>systemic toxicity | | | | | Pass |
| ISO 10993-23 | Tests for irritation | Non-Irritant | | | | | Pass |
### 9.0 Summary of clinical Performance data
Not applicable - Clinical data was not used to assess performance of the subject device.
### 10.0Conclusion
The Conclusion drawn from the Non-Clinical test demonstrates that the subject device- Iconic Latex Glove is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K161833.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.