SIRIUS Endoscope System

K221642 · Precision Robotics (Hong Kong) Limited · HET · Dec 12, 2022 · Obstetrics/Gynecology

Device Facts

Record IDK221642
Device NameSIRIUS Endoscope System
ApplicantPrecision Robotics (Hong Kong) Limited
Product CodeHET · Obstetrics/Gynecology
Decision DateDec 12, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1720
Device ClassClass 2

Indications for Use

It is intended for use for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. Not suitable for use for observation and treatment of the heart and must not contact the heart or its vicinity.

Device Story

SIRIUS Endoscope System is a robotic, fully integrated 3D laparoscopic camera system for minimally invasive surgery. Input: optical images captured by a stereo camera at the distal tip. Transformation: electromechanical control of the articulated tip (3 degrees of freedom, C/S-shaped bending) via motors in the handle, driven by a joystick. Output: 1080p 3D video displayed to the surgeon. Used in hospital operating theaters by trained surgeons and clinical staff. The system provides high-definition visualization to assist in surgical navigation and procedure performance. Benefits include enhanced maneuverability via robotic articulation and high-resolution 3D imaging.

Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogen), sterility (EO, SAL 10^-6), electrical safety/EMC (IEC 60601-1, 60601-2-18, 60601-1-2), light source safety (IEC 62471), and performance testing per ISO 8600 standards (field of view, direction of view, insertion width, water resistance). Usability evaluation conducted per IEC 62366-1.

Technological Characteristics

Robotic 3D endoscope; 10mm diameter, 340mm length. Materials: TPU, stainless steel, FLUORZ-UPF75-2. Articulation: electromechanical, 3 degrees of freedom. Optics: 1080p stereo camera, 120-degree FOV, 300-lumen light. Sterilization: Ethylene Oxide (ISO 11135). Compliance: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, IEC 62471, ISO 8600 series.

Indications for Use

Indicated for endoscopy and endoscopic surgery within thoracic and abdominal cavities, including female reproductive organs. Contraindicated for observation or treatment of the heart or contact with the heart/vicinity.

Regulatory Classification

Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 12, 2022 Precision Robotics (Hong Kong) Limited % Sharon Hsu Consultant Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong 0000 China Re: K221642 Trade/Device Name: SIRIUS Endoscope System Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope And Accessories Regulatory Class: Class II Product Code: HET. GCJ. FGB Dated: November 9, 2022 Received: November 9, 2022 Dear Sharon Hsu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Carr -S for Long Chen, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K221642 Device Name Sirius Endoscope System #### Indications for Use (Describe) It is intended for use for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. Not suitable for use for observation and treatment of the heart and must not contact the heart or its vicinity. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Precision Robotics. The logo consists of a stylized blue atom-like symbol on the left, followed by the words "PRECISION" and "ROBOTICS" stacked on top of each other in a sans-serif font, also in blue. The overall design is clean and modern. # 510(k) Summary # 1. SUBMITTER | Name : | Precision Robotics (Hong Kong) Limited | |----------------|-----------------------------------------------------------------------------------------------| | Address: | Suite 611-612 Lakeside 2 (10W), Hong Kong Science Park, Shatin, New<br>Territories, Hong Kong | | Contact Name: | Sai Yan Ng | | Telephone No.: | +852 3974 6970 | | Fax: | +852 3974 6971 | | Email Address: | s.ng@precisionrobotics.com | | Date Prepared: | 2022-12-03 | # 2. DEVICE | Device Name/ Trade<br>Name: | SIRIUS Endoscope System | |-----------------------------|-------------------------------------------------------------------------------------------| | Common Name | Endoscope system | | Classification Name: | 876.1500 Endoscope and Accessories<br>884.1720 Laparoscope, Gynecologic (And Accessories) | | Classification Panel: | General & Plastic Surgery<br>Obstetrics/Gynecology | | Product Code: | HET, GCJ, FGB | | Device Class: | II | ### 3. PREDICATE DEVICE Olympus LTF-190-10-3D, Endoeve Flex 3D Deflectable Videoscope, K123365: MAJ-Y1054 3D Processor, Olympus CV-190 EVIS Extra III Video system center ## 4. DEVICE DESCRIPTION The SIRIUS Endoscope System as a robotic endoscope consists of Video Processor, Laparoscope Handle, Laparoscope Head and Joystick. The SIRIUS Endoscope System is developed for minimal invasive surgery application. {4}------------------------------------------------ The SIRIUS Endoscope System is a fully integrated compact 3D laparoscopic camera system with a flexible tip that can change its viewing direction. The Laparoscope Head is made of biocompatible materials. The articulated tip has three degrees of freedom enabling C and S shaped bending. The insertion is 10 mm diameter and 340 mm working length. Stereo camera with 1080 high-definition resolution. It has 120 degrees field of view, 10-100mm depth of view and bright light with 300 lumen. The SIRIUS Endoscope System is to be used only under the supervision of a trained surgeons and professional clinical staff with trained use of the device. The device is intended for use in Hospital operating theatres only. ### 5. INDICATION FOR USE lt is intended for use for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. Not suitable for use for observation and treatment of the heart and must not contact the heart or its vicinity. ### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The SIRIUS Endoscope System is substantially equivalent to the predicate device in terms of intended use and technological characteristics. The differences between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. The following are comparisons between subject device and the predicate device. | | Subject Device (K221642) | Predicate Device<br>(K123365) | Comments | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | SIRIUS Endoscope System | Olympus LTF-190-10-3D,<br>Endoeye Flex 3D<br>Deflectable Videoscope<br>MAJ-Y1054 3D Processor<br>Olympus CV-190 EVIS Extra<br>III Video system center | -- | | Manufacturer | Precision Robotics (Hong<br>Kong) Limited | Olympus Medical System<br>Corp. | -- | | Device Class | Class II | Class II | Same | | Product Code | HET, GCJ, FGB | HET, GCJ, NWB, FGB | Similar | | Regulation number | 876.1500<br>884.1720 | 876.1500<br>884.1720 | Same | | Regulation Name | Endoscope and Accessories | Endoscope and Accessories | Same | | Intended Use/<br>Indications for Use | It is intended for use for<br>endoscopy and endoscopic<br>surgery within the thoracic<br>and abdominal cavities | This instrument is<br>indicated for use within<br>the thoracic and<br>abdominal cavities | Same | | | | | | | | including female<br>reproductive organs. Not<br>suitable for use for<br>observation and treatment<br>of the heart and must not<br>contact the heart or its<br>vicinity. | including female<br>reproductive organs.<br>This instrument must not<br>be used for observation or<br>treatment of the heart and<br>must not contact the heart<br>or any area near the<br>heart. In addition, this<br>instrument must not come<br>into contact with any<br>device or therapeutic<br>accessory that contacts<br>the heart or any area near<br>the heart. | | | Operating Principles | Tendon driven articulating<br>3D video endoscope driven<br>electromechanically using<br>motors in the handle. | Cable steered articulating<br>3D video endoscope driven<br>manually using mechanical<br>levers in the scope handle. | Different<br>The laparoscope<br>head movement of<br>subject device is<br>controlled<br>electromechanically.<br>It shas validated and<br>compliance to<br>IEC60601-1, IEC<br>62304 and ISO8600. | | Anatomical Access | Thoracic and abdominal<br>cavities including female<br>reproductive organs. | Thoracic and abdominal<br>cavities including female<br>reproductive organs. | Same | | Direction of View | 0° camera angle | 0° camera angle | Same | | Shaft diameter (OD) | 10 mm | 10 mm | Same | | Tip Articulation | $\pm$ 90° for each up-down<br>joint allowing retroflexion<br>$\pm$ 45° for left-right joint | $\pm$ 100° | Similar<br>Angle of deflection<br>meet the<br>requirement of ISO<br>8600-1:2015. | | Working Length | 340 mm | 370 mm | Different<br>Physical difference. | | Shaft material | TPU, Stainless steel,<br>FLUORZ-UPF75-2 | Carbon Fiber, covered with<br>heat shrink sheathing | Different<br>Differences are<br>addressed through<br>biocompatibility<br>testing per<br>ISO10993. | | Optics Type | Color | Color | Same | | Resolution | 1920 x 1080 | 1080 x 601 | Subject has higher<br>resolution | | Single Use | Yes | No | Different<br>The sterilization<br>validation has been<br>performed by ISO | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Precision Robotics. The logo consists of a blue atom-like symbol on the left, followed by the words "PRECISION ROBOTICS" in a sans-serif font. The word "PRECISION" is stacked on top of the word "ROBOTICS", and both words are in blue. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for Precision Robotics. The logo consists of a blue atom-like symbol on the left, followed by the words "PRECISION" and "ROBOTICS" stacked on top of each other in a sans-serif font, also in blue. The logo is simple and modern, conveying a sense of technology and precision. | Sterilization | EO sterilization | Product is provided<br>non-sterile | Different<br>The sterilization<br>validation has been<br>performed by ISO<br>11135:2014. | |------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Biocompatibility | Patient contacting<br>components meet<br>ISO10993 standard | Patient contacting<br>components meet<br>ISO10993 standard | Same | | Electrical Safety and<br>EMC | complies with IEC 60601-1,<br>IEC60601-2-18 standards<br>for safety and IEC<br>60601-1-2 standard for<br>EMC | complies with IEC 60601-1,<br>IEC60601-2-18 standards<br>for safety and IEC<br>60601-1-2 standard for<br>EMC | Same | #### Discussion: The tip articulation of subject device is within the range of predicate device. The specification of tip articulation of subject device fulfils the requirement of ISO 8600-1:2015. The difference does not affect the effectiveness and safety of the device. The subject device has shorter working length than predicate device, this physical different does not impact the effectiveness and safety of the device. The shaft material of subject device and predicate device are made of different materials, but both are biocompatible materials. The biocompatibility tests conducted demonstrated the safety of subject device. The subject device is singe use device with EO sterilization validated by ISO 11135:2014. This difference does not impact the effectiveness and safety of the device. Therefore, the differences between the subject device and its predicate do not affect substantially equivalent on safety and effectiveness. ### 7. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. #### Biocompatibility testing The biocompatibility evaluation for the Laparoscope Head was conducted in accordance with International Standard ISO 10993-1:2005, ISO 10993-5:2009, ISO 10993-10:2010 and ISO 10993-11:2017 as recognized by FDA. The evaluation included the following tests: - Cytotoxicity - Sensitization {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Precision Robotics. The logo consists of a blue atom-like symbol on the left and the words "PRECISION ROBOTICS" in blue on the right. The words are stacked on top of each other. - Irritation - Acute systemic toxicity . - . Pyrogen ### Sterility The sterilization of the product is achieved using ethylene oxide sterilization. Sterilization condition is validated per ISO 11135: 2014 overkill half-cycle approach. The sterility assurance level (SAL) is 10-6. The amount of ethylene oxide and chlorohydrin residual levels are within the limit and in compliance with ISO 10993-7: 2008 requirement. ### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on SIRIUS endoscope system. The device complies with IEC 60601-1, IEC60601-2-18 standards for safety and the IEC 60601-1-2 standard for EMC. ### Light source safety Light source safety was conducted on SIRIUS endoscope system. The lamp is certified as Risk group 2 and complies with IEC 62471: 2006. ### Software Verification and Validation Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. ### Usability Evaluation The Usability evaluation for SIRIUS endoscope system is conducted in accordance with ANSI AAMI IEC 62366-1:2015+AMD1:2020, and, therefore, by extension IEC 60601-1-6 Edition 3.2. The usability verification and validation are well defined with criteria specified in the Usability Engineering Report. Usability testing results and improvement action demonstrated that residual risk regarding usability is minimized and acceptable. It is concluded that the usability evaluation is well performed and acceptable. ### Performance testing The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include: - ISO 8600-1:2015, Endoscopes Medical endoscopes and endotherapy devices -- Part 1: General requirements - · ISO 8600-3:2019, Endoscopes Medical endoscopes and endotherapy devices Part 3: {8}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the logo for Precision Robotics. The logo consists of a blue atom-like symbol on the left, followed by the words "PRECISION ROBOTICS" in a stylized, sans-serif font. The word "PRECISION" is stacked on top of the word "ROBOTICS". The color of the text is the same blue as the atom symbol. Determination of field of view and direction of view of endoscopes with optics. - ISO 8600-4:2014, Endoscopes Medical endoscopes and certain accessories Part 4: Determination of maximum width of insertion portion - ISO 8600-7:2012, Endoscopes Medical endoscopes and endotherapy devices -. Part 7: Basic requirements for medical endoscopes of water-resistant type # 8. CONCLUSIONS The SIRIUS Endoscope System is substantially equivalent to the predicate device and present no new questions of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...