Spine & Trauma Navigation, Spine Navigation, Navigation Software Spine & Trauma, Alignment System Spine, Alignment Software Spine, Cirq Arm System
K221618 · Brainlab AG · OLO · Aug 30, 2022 · Neurology
Device Facts
Record ID
K221618
Device Name
Spine & Trauma Navigation, Spine Navigation, Navigation Software Spine & Trauma, Alignment System Spine, Alignment Software Spine, Cirq Arm System
Applicant
Brainlab AG
Product Code
OLO · Neurology
Decision Date
Aug 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data. Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy. As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to support the surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.
Device Story
Spine & Trauma Navigation is an intraoperative image-guided surgery system. It inputs preoperative or intraoperative 2D/3D image data (CT, MR, fluoroscopy) and tracks surgical instruments via an optical infrared camera and passive markers. The system registers patient anatomy to image space, allowing surgeons to plan and navigate screw and interbody device placement. The Alignment System Spine accessory, featuring the Cirq Arm and Robotic Motor Unit, provides automated fine alignment for spinal screw placement; the surgeon manually pre-aligns the arm, then the software controls the motor for final positioning. Used in ORs by neuro/ortho/spine surgeons. Output is displayed on a mobile monitor cart or ceiling-mounted display, showing real-time instrument position relative to patient anatomy. This assists in precise surgical execution, potentially improving accuracy in spinal procedures.
Clinical Evidence
No clinical testing was needed. Evidence consists of bench testing, including software verification and validation, hardware endurance and safety testing, and system accuracy testing (2D/3D positional and angular navigation accuracy). Electrical safety and EMC testing were performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Optical tracking system using infrared cameras and passive markers. Software-based navigation on mobile/fixed IGS platforms. Components include surgical instruments, Cirq Arm System, and Robotic Motor Unit. Connectivity via IGS platform. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Software is major level of concern.
Indications for Use
Indicated for patients requiring stereotactic surgery where rigid anatomical structures (skull, pelvis, long bone, vertebra) can be referenced to preoperative or intraoperative 2D/3D images (CT, MR, fluoroscopy) or anatomical models. Used for open or minimally invasive spinal screw placement.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
August 30, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Brainlab AG Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str.9 Munich, 81829 Germany
Re: K221618
Trade/Device Name: Spine & Trauma Navigation, Spine Navigation, Navigation Software Spine & Trauma, Alignment System Spine, Alignment Software Spine, Cirq Arm System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: May 31, 2022 Received: June 3, 2022
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's
Page 2
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K221618
#### Device Name
Spine & Trauma Navigation, Navigation Software Spine & Trauma, Spine Navigation Alignment System Spine, Alignment Software Spine, Cirq Arm System
#### Indications for Use (Describe)
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to support the surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is an abstract design featuring a pattern of triangles in varying shades of blue and white. The triangles are arranged in a way that creates a sense of depth and movement. The darker blue triangles are concentrated on the left side of the image, gradually transitioning to lighter shades of blue and white on the right side. The overall effect is modern and visually appealing.
Image /page/3/Picture/2 description: The image shows the logo for Brainlab. The logo consists of a stylized image of three interconnected shapes resembling a brain, followed by the word "BRAINLAB" in bold, sans-serif font. The logo is simple and modern, with a focus on the company's name and its connection to the brain.
# 510(k) Summary
#### August 30, 2022
| General Information | |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Name | Spine & Trauma Navigation<br>Navigation Software Spine & Trauma<br>Spine Navigation<br>Alignment System Spine<br>Alignment Software Spine<br>Cirq Arm System |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Product Code | OLO |
| Regulation Number | 882.4560 |
| Regulatory Class | Class II |
| Panel | Orthopedic |
| Predicate Device(s) | Primary Predicate: K212245 Spine & Trauma Navigation System<br>Secondary Predicate: K110204 Brainlab Trauma<br>Third Predicate: K202320 Cirq Robotic Alignment Module |
| Contact Information | |
| Primary Contact | Alternate Contact |
| Chiara Cunico<br>Manager Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: regulatory.affairs@brainlab.com | Regulatory Affairs Brainlab<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 5033<br>Email: regulatory.affairs@brainlab.com |
## 1. Indication for Use
#### Spine & Trauma Navigation
Spine & Trauma Navigation is intended as an intraoperative image-quided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
{4}------------------------------------------------
K221618
Image /page/4/Picture/1 description: The image is an abstract design featuring a pattern of interconnected triangles. The triangles vary in size and shade, transitioning from a darker blue on the left to a lighter blue on the right, eventually fading into white. The white lines form the edges of the triangles, creating a geometric and fragmented appearance. The overall effect is modern and visually dynamic.
Image /page/4/Picture/2 description: The image shows the word "BRAINLAB" in bold, black letters. To the left of the word is a stylized logo that appears to be three curved lines stacked on top of each other. The logo is also in black and is smaller than the word "BRAINLAB".
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
### Alignment System Spine
As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to support the surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.
### 2. Device Description
The Spine & Trauma Navigation is an image guided surgery system for navigated treatments in the fields of spine and trauma surgery, whereas the user may use image data based on CT, MR, 3D fluoroscopic image reconstruction (cone beam CT) or 2D fluoroscopic images coming from the compatible imaging device LoopX. It offers different patient image registration methods and instrument selection and calibration to allow surgical navigation by using optical tracking technology.
The software is installed on a mobile or fixed Image Guided Surgery (IGS) platform to support the surgeon in clinical procedures by displaying tracked instruments in patient's image data. The IGS platforms comprise of a mobile Monitor Cart or a fixed ceiling mounted display and an infrared camera for image guided surgery purposes.
The Spine &Trauma Navigation consists of the following components:
- . Software enabling instrument selection, different registration methods (e.g. surface matching) as well as navigation in different types of images.
- . IGS platforms
- . Surgical instruments for navigation, patient referencing and registration
The Alignment System Spine is an accessory to the Spine & Trauma Navigation. It serves as a holding and positioning system to support the surgeon in reaching a pre-defined screw with surgical instruments. The device needs to be first manually pre-aligned to the region of interest by opening the brakes of the Cirq Arm System. Then tracking information provided by the optical camera is used by the Alignment Software Spine 2.0 to control the movement of the Cirq Robotic Motor Unit to perform the final automatic fine alignment. Once the alignment to the planned screw is done, the Alignment System Spine maintains its position during the rest of the procedure and Spine & Trauma 3D Navigation takes over the navigation of the instruments. The Alignment System Spine consists of the following components:
- Alignment Software Spine 2.0 ●
- Cirg Arm System with multiple structural hardware components ●
- Cirq Robotic Motor Unit
- Cirq robotic instruments
{5}------------------------------------------------
K221618
Image /page/5/Picture/1 description: The image shows a blue geometric design on the left side and the word "BRAINLAB" in black on the right side. The geometric design is made up of triangles of different shades of blue. Above the word "BRAINLAB" is the alphanumeric string "K221618". The background of the image is white.
The operator's profile for the Subject Device are Neuro / Ortho / Spine / Trauma surgeons or their assistants having a 3D image acquisition system (such as a CT or a 3D C-arm),or utilizing preoperatively acquired CT / CT-like (and potentially fused MR) imaging data in combination with a Brainlab navigation.
# 3. Substantial Equivalence
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Item | Predicate device | Product Code | New/modified<br>device |
|----------------|--------------------------------------------------|--------------|-----------------------------------------------------------------------------|
| 1st Predicate: | Spine & Trauma<br>Navigation System<br>(K212245) | OLO | Spine & Trauma<br>Navigation 2.0 |
| 2nd Predicate | Brainlab Trauma<br>(K110204) | OLO | Spine & Trauma<br>Navigation<br>(specifically for Fluoro<br>Navigation 1.0) |
| 3rd Predicate | Cirq Robotic<br>Alignment Module<br>(K202320) | OLO | Alignment<br>System<br>Spine |
At a high level, the main similarities and differences between the subject and predicate devices are:
For Spine &Trauma Navigation:
| Same or similar aspects | Differences |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use Localization technology: Optical System accuracy in 2D and 3D images Registration and calibration of instruments Supported image modalities Supported IGS platforms Surgical instruments Programming language | GUI technology Main functionality is the same, but tools and view layouts are organized slightly differently Compared to K212245: Integrated interface to the Alignment System Spine Compared to K110204: Fluoro Navigation supports only 2D images from the compatible imaging device LoopX and different registration method |
For Alignment System Spine:
| Same or similar aspects | Differences |
|-------------------------|------------------|
| • Indications for use | • GUI technology |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Brainlab logo and some text. The text on the left describes the guided positioning, robotic movement, same positioning arm and hardware components, same instruments, localization technology, and supported IGS platforms. The text on the right describes the functionality to calibrate and navigate instruments that has been removed and is now performed with Spine & Trauma Navigation, and the integrated interface to the Spine & Trauma Navigation.
#### Performance Data 4.
The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate devices:
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
#### Hardware verification
Hardware verification was carried out to ensure all requirements for the platforms are met. It was also ensured that the platforms are able to work as intended during their defined lifetime by performing endurance testing, the safety of the cleaning and disinfection process and the biological safety of the materials and surfaces which may become in contact with the user.
#### System accuracy testing
The 2D and 3D positional and angular navigation accuracy of the Spine &Trauma Navigation including the software, the platforms and the instruments was evaluated considering a realistic clinical setup and representative worst-case scenarios. The results show the following acceptance criteria are fulfilled:
- Mean Positional Error of the placed instrument's tip ≤ 2 mm
- . Mean Angular Error of the placed instrument's axis ≤ 2°
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety, RFID and EMC testing was conducted on the Subject device according to the standards:
IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance. Including US national deviations.
{7}------------------------------------------------
K221618
Image /page/7/Picture/1 description: The image shows a blue geometric design on the left side and the word "BRAINLAB" on the right side. The geometric design is made up of triangles of different shades of blue. The word "BRAINLAB" is in black and is written in a sans-serif font. The image also contains the text "K221618" in the upper right corner.
AIM 7351731 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
No clinical testing was needed for the Subject Device.
### 5. Conclusion
The comparison of the Subject Device with the predicate devices shows that the Spine & Trauma Navigation and its accessory the Alignment System Spine have similar functionality, intended use and technological characteristics as the predicate device(s). Based on the comparison to the predicates and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.