AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K221564 · Feb 23, 2023
Brainomix 360 e-ASPECTS
Brainomix Limited
Retrospective NCCT scans from 8 clinical institutions; Patient medical records/clinical data
Retrospective clinical data was used to conduct stand-alone performance testing and an MRMC reader study to validate the device's ability to assist clinicians in ASPECTS scoring compared to unassisted reading.
Brainomix 360 e-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data. The software automatically registers images and uses an Atlas to segment and analyze ASPECTS regions. Brainomix 360 e-ASPECTS extracts image data from individual voxels in the image to provide analysis and computer and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) Score. Brainomix 360 e-ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. Brainomix 360 e-ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 6 hours from time last known well). Brainomix 360 e-ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. Brainomix 360 e-ASPECTS additionally provides a visualization of the voxels contributing to the automated ASPECTS score and the voxels excluded from the automated ASPECTS score.
Device Story
Brainomix 360 e-ASPECTS is a standalone software device for clinical/hospital use by neuroradiologists/clinicians. It processes non-contrast CT (NCCT) brain images to assist in stroke assessment. The device registers images to an atlas, segments ASPECTS regions, and uses a random forest machine learning algorithm to analyze voxel-wise data for ischemic changes. It outputs an automated ASPECTS score, highlights affected regions, and provides heatmaps of contributing/excluded voxels. Clinicians review, verify, and may manually edit the automated score. The device integrates with PACS via DICOM for image transfer and reporting. By providing standardized, automated analysis, it aims to improve diagnostic accuracy and consistency, particularly for less experienced clinicians, facilitating timely patient management decisions for neurothrombectomy or other interventions.
Clinical Evidence
Clinical evidence includes a multiple reader multiple case (MRMC) cross-over study (n=54 retrospective NCCT scans) and stand-alone performance testing (n=256 scans). MRMC study showed a statistically significant improvement in reader AUC (0.02 increase, p=0.028), with sensitivity increasing from 66% to 70% and specificity at 96%. Stand-alone testing demonstrated an AUC of 83% (95% CI: 81-85), sensitivity of 68%, and specificity of 97%. Digital phantom testing confirmed heatmap correlation (r>=0.95) with synthetic hypodensities.
Technological Characteristics
Software-only device; no physical materials or energy delivery. Uses random forest machine learning algorithm for voxel-wise classification of NCCT brain images. Integrates with DICOM-compliant PACS/servers. Validated for use on Siemens Somatom Definition scanners. Operates via automated registration to an anatomical atlas. No electrical, thermal, or radiation safety requirements due to software-only nature.
Indications for Use
Indicated for evaluation of patients aged >21 presenting for diagnostic imaging workup with known MCA or ICA occlusion to assess extent of ischemic brain tissue injury within 6 hours of time last known well. Contraindicated in patients with excessive motion artifacts, hemorrhagic transformation, or hematoma. Not for primary interpretation.
Regulatory Classification
Identification
A radiological computer-assisted diagnostic software for lesions suspicious of cancer is an image processing prescription device intended to aid in the characterization of lesions as suspicious for cancer identified on acquired medical images such as magnetic resonance, mammography, radiography, or computed tomography. The device characterizes lesions based on features or information extracted from the images and provides information about the lesion(s) to the user. Diagnostic and patient management decisions are made by the clinical user.
Special Controls
A radiological computer-assisted diagnostic (CADx) software for lesions suspicious for cancer must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, and algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will improve reader performance as intended. iii. Results from performance testing protocols that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, results, and cybersecurity). 2. Labeling must include: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended reading protocol. iii. A detailed description of the intended user and recommended user training. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations, including situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. Detailed instructions for use. viii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders (e.g., lesion and organ characteristics, disease stages, and imaging equipment).
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will improve reader performance as intended.
(iii) Results from performance testing protocols that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain sufficient numbers of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Standalone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; and description of verification and validation activities including system level test protocol, pass/fail criteria, results, and cybersecurity).(2) Labeling must include:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and recommended user training.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations, including situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Detailed instructions for use.
(viii) A detailed summary of the performance testing, including: Test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders (
*e.g.,* lesion and organ characteristics, disease stages, and imaging equipment).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
Brainomix Limited % Gwilym Owen Head of Quality and Regulatory Suite 11-14, Suffolk House, 263 Banbury Road Oxford. Oxfordshire OX2 7HN UNITED KINGDOM
February 23, 2023
### Re: K221564
Trade/Device Name: Brainomix 360 e-ASPECTS Regulation Number: 21 CFR 892.2060 Regulation Name: Radiological computer-assisted diagnostic software for lesions suspicious of cancer Regulatory Class: Class II Product Code: POK Dated: January 20, 2023 Received: January 23, 2023
Dear Gwilym Owen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K221564
Device Name Brainomix 360 e-ASPECTS
#### Indications for Use (Describe)
Brainomix 360 e-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.
The software automatically registers images and uses an Atlas to segment and analyze ASPECTS regions. Brainomix 360 e-ASPECTS extracts image data from individual voxels in the image to provide analysis and computer and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) Score.
Brainomix 360 e-ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. Brainomix 360 e-ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 6 hours from time last known well).
Brainomix 360 e-ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. Brainomix 360 e-ASPECTS additionally provides a visualization of the voxels contributing to the automated ASPECTS score and the voxels excluded from the automated ASPECTS score.
Limitations:
1. Brainomix 360 e-ASPECTS is not intended for primary interpretation of CT images. It is used to assist physician evaluation.
2. Brainomix 360 e-ASPECTS has been validated in patients with known MCA or ICA occlusion prior to ASPECTS scoring.
3. Brainomix 360 e-ASPECTS is not suitable for use on brain scans displaying neurological pathologies other than acute stroke, such as tumours or abscesses, haemorrhagic transformation and hematoma.
4. Use of Brainomix 360 e-ASPECTS Module in clinical settings other than brain ischemia within 6 hours from time last known well, caused by known ICA or MCA occlusions has not been tested.
5. Brainomix 360 e-ASPECTS has only been validated and is intended to be used in patient populations aged over 21.
6. Brainomix 360 e-ASPECTS has been validated and is intended to be used on Siemens Somatom Definition scanners.
7. Brainomix 360 e-ASPECTS is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
Contraindications/Exclusions/Cautions:
· Patient motion: Excessive patient motion leading to artifacts that make the scan technically inadequate.
- · Haemorrhagic Transformation, Hematoma.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in bold, sans-serif font. The brain icon is a gradient of teal and white, while the letters of the word "BRAINOMIX" are dark gray, except for the "AI" which is teal.
# 510(K) Summary of Safety and Effectiveness
| Date of summary: | 23rd February 2023 |
|--------------------------|-----------------------------------------------------------------------------------------------------|
| Submitter's name: | Brainomix Limited |
| Submitter's address: | Brainomix Limited, Suites 11-14 Suffolk House, 263 Banbury Road,<br>Oxford. OX2 7HN. United Kingdom |
| Submitter's contact: | Gwilym Owen |
| Telephone number: | +44 (0)1865 582730 |
| Device Proprietary Name: | Brainomix 360 e-ASPECTS |
| Device Common Name(s): | CADx |
| Classification Name: | Class II |
| Product Code: | POK |
| Regulation No: | 21 C.F.R. §892.2060 |
Brainomix 360 e-ASPECTS is Substantially Equivalent to the following Legally Marketed device:
#### Predicate Devices
| 510(k) Number | Trade Name | Manufacturer |
|---------------|---------------|-----------------|
| K200760 | Rapid ASPECTS | iSchemaView Inc |
# Indications for Use
Brainomix 360 e-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.
The software automatically registers images and uses an Atlas to segment and analyze ASPECTS regions. Brainomix 360 e-ASPECTS extracts image data from individual voxels in the image to provide analysis and computer analytics and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) Score.
Brainomix 360 e-ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. Brainomix 360 e-ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 6 hours from time last known well).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is teal, and the "A" in "BRAINOMIX" is also teal, while the rest of the letters are dark gray. The overall design is modern and clean.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
Brainomix 360 e-ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. Brainomix 360 e-ASPECTS additionally provides a visualization of the voxels contributing to the automated ASPECTS score and the voxels excluded from the automated ASPECTS score.
Limitations:
- Brainomix 360 e-ASPECTS is not intended for primary interpretation of CT images. It is used 1. to assist physician evaluation.
- Brainomix 360 e-ASPECTS has been validated in patients with known MCA or ICA occlusion 2. prior to ASPECTS scoring.
- 3. Brainomix 360 e-ASPECTS is not suitable for use on brain scans displaying neurological pathologies other than acute stroke, such as tumours or abscesses, haemorrhagic transformation and hematoma.
- 4. Use of Brainomix 360 e-ASPECTS Module in clinical settings other than brain ischemia within 6 hours from time last known well, caused by known ICA or MCA occlusions has not been tested.
- 5. Brainomix 360 e-ASPECTS has only been validated and is intended to be used in patient populations aged over 21.
- 6. Brainomix 360 e-ASPECTS has been validated and is intended to be used on Siemens Somatom Definition scanners.
- 7. Brainomix 360 e-ASPECTS is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
Contraindications/Exclusions/Cautions:
- Patient motion: Excessive patient motion leading to artifacts that make the scan technically inadequate.
- . Haemorrhagic Transformation, Hematoma.
# Device Description
Brainomix 360 e-ASPECTS is a stand-alone software device which uses machine learning algorithms to automatically process NCCT (Non-contrast CT scans) brain image data to provide an output ASPECTS score based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines.
The post-processing image results and ASPECTS score are identified based on regional imaging features and overlayed onto brain scan images. e-ASPECTS provides an automatic ASPECTS score based on the input CT data for the physician. The score includes which ASPECTS regions are identified based on regional imaging features derived from non-contrast computed tomography (NCCT) brain image data. The results are generated based on the Alberta Stroke Program Early CT
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in bold, sans-serif font. The brain icon is teal, and the "AI" in "BRAINOMIX" is also teal, creating a visual connection between the brain and the concept of artificial intelligence.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
Score (ASPECTS) guidelines and provided to the clinician for review and verification. At the discretion of the clinician, the scores may be adjusted based on the clinician's judgment.
Brainomix 360 e-ASPECTS can connect with other DICOM-compliant devices, for example to transfer NCCT scans from a Picture Archiving and Communication System (PACS) to Brainomix 360 e-ASPECTS software for processing.
Results and images can be sent to a PACS via DICOM transfer and can be viewed on a PACS workstation or via a web user interface on any machine contained and accessed within a hospital network and firewall and with a connection to the Brainomix 360° e-ASPECTS software (e.g. a LAN connection)
Brainomix 360 e-ASPECTS notification capabilities enable clinicians to preview images through via email notification with result image attachments.
lmages that are previewed via e-mail are compressed, are for preview purposes only, and not intended for diagnostic use beyond notification.
Brainomix 360 e-ASPECTS is not intended for mobile diagnostic use. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.
Brainomix 360 e-ASPECTS provides an automated workflow which will automatically process image data received by the system in accordance with pre-configured user DICOM routing preferences.
Once received, image processing is automatically applied. Once any image processing has completed, notifications are sent to pre-configured users to inform that the image processing results are ready. User can then access and review the results and images via the Web User Interface case viewer or PACS viewer.
Brainomix 360 e-ASPECTS principal workflow for NCCT includes the following key steps:
- NCCT Image Loading.
- . Automated image analysis and processing to identify and visualize the voxels which have been included in the ASPECTS score and the voxels which have been excluded from the ASPECTS score (Also referred to as a 'heat map').
- Automated image analysis and processing to register the subject image to an atlas to segment and highlight ASPECTS regions and to display whether or not each region is qualified as contributing to the ASPECTS score.
- . Notifications and alerts to users.
- Generation of a summary results report.
- Presentation of results for review and analysis by users.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain image on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "A" in "BRAINOMIX" is colored in a gradient from light blue to dark blue, while the rest of the letters are dark gray.
Suites 11-14 Suffolk House 263 Banburv Road Oxford OX2 7HN
Once the physician has been notified of availability of the ASPECTS score, the system requires that the physician confirms that the case in question is for an ICA occlusion. The ASPECTS results, including the ASPECTS score, indication of affected side, affected ASPECTS regions and voxel-wise analysis (shown as a heatmap of voxels 'contributing to e-ASPECTS score' and a heat map of voxels 'excluded from e-ASPECTS score') can be exported as a report and/or sent to the Picture Archiving and Communications System (PACS).
# Technological Characteristics
Brainomix 360 e-ASPECTS is a machine learning implementation using the processing pipeline below. Brainomix 360 e-ASPECTS provides an automatic ASPECT score based on the case input file for the physician. The score includes which ASPECT regions are identified based on a voxel-wise of non-contrast computed tomography (NCCT) brain image data based on the random forest machine learning technique. The results are generated based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines and provided to the clinician for review and verification.
ASPECTS regions are generated as an overlay by registering the subject NCCT scan to an atlas. The Brainomix 360 e-ASPECTS algorithm analyses the voxels in the subject NCCT scan to identify voxels contributing to, or excluded from, the e-ASPECTS score. The regions then contribute towards the determination of an ASPECTS score which must be reviewed by the physician.
ASPECTS regions identified as contributing to the ASPECTS score are highlighted and labelled for the physician. The voxel-wise analysis is visualized for the physician as shaded heat maps showing voxels contributing to the e-ASPECTS score and voxels excluded from the e-ASPECTS score.
Physicians can, at their discretion and depending on their level of agreement with the processing results, manually edit the ASPECTS score by manually selecting ASPECTS regions to be included or excluded from the calculation of the ASPECTS score.
The following figure provides the general layout of the ASPECTS display as presented by Brainomix 360 e-ASPECTS
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of light blue to dark blue, and the word "BRAINOMIX" is in a dark gray color, with the "AI" in a gradient of light blue to dark blue.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
Image /page/8/Picture/3 description: The image shows a Brainomix 360 scan with a patient ID of EX127 and a case ID of 7. The scan date is 2018-12-27 09:01:27. The ASPECTS score is 7, with the right side showing M1, Lentiform, and Insula. The image also shows an unannotated scan and results with highlighted regions.
# Substantial Equivalence Discussion
Both proposed and predicate devices have the same intended use. Both devices are computer-aided diagnosis (CADx) software devices used to assist clinicians in the assessment and characterization of brain tissue abnormalities using CT image data.
Both devices use an atlas to segment and identify ASPECTS regions and use artificial intelligence algorithms to analyse image data to facilitate the evaluation of extent of disease and to identify and count the affected ASPECTS regions to generate an ASPECTS score. Both devices display and highlight the affected ASPECTS regions along with the automatically generated ASPECTS score
Both devices are not intended for primary interpretation of CT images and both require that cases processed meet pre-requisite criteria such that a primary diagnosis of ICA or MCA occlusion has already been made by clinicians prior to accessing the processing results and the resulting ASPECTS score.
Both devices are indicated for cases within 6 hours from time last known well.
Both devices include a gating condition requiring users to confirm that the case in question is for a confirmed ICA or MCA occlusion before processing results are made available.
Both devices include pre-processing steps to straighten NCCT scans and to register the NCCT scans to an atlas of ASPECTS regions in order to identify and segment ASPECTS regions on the subject NCCT scan.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in bold, sans-serif font. The brain icon is teal, and the "AI" in "BRAINOMIX" is also teal, creating a visual connection between the brain and the concept of artificial intelligence.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
In both cases, the algorithms identify signs indicative of ischemic damage and relate these to the segmented ASPECTS regions in order to determine whether or not individual ASPECTS regions are included as contributing to a reduced ASPECTS score.
In both cases, the segmented ASPECTS regions are labelled and presented to users in the software User Interface with the regions highlighted depending on whether they have been determined to affect the e-ASPECTS score or not. In both cases, users can manually edit and override the automated result for one or more ASPECTS regions by selecting the regions.
Both predicate and proposed devices use trained machine learning AI algorithms to analyse NCCT scans to identify signs indicative of ischemic damage. Both devices use a random forest machine learning technique to classify tissue in brains scans to identify both early/acute ischemic changes and old/non-acute ischemic changes to differentiate old or prior signs of stroke from the acute stroke for which the ASPECTS methodology is intended. In both devices, signs of non-acute ischemic change are masked to ensure that these do not contribute towards the determination of the output ASPECTS score.
The following principal difference exists between Brainomix 360 e-ASPECTS and the predicate device:
In addition to highlighting affected ASPECTS regions which is a common technical characteristic of both subject and predicate devices, the Brainomix 360 e-ASPECTS device includes the visualization of the analysis of individual voxels underlying the scoring of ASPECTS regions and the overall ASPECTS score as a heat map.
| Characteristic/Parameter | Brainomix 360 e-ASPECTS -<br>Proposed Device | iSchemaView Rapid ASPECTS -<br>Predicate Device (K200760) |
|--------------------------|----------------------------------------------|-----------------------------------------------------------|
| Product Code | POK | POK |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "A" in "BRAINOMIX" is a gradient of blue and green, while the rest of the letters are gray.
Suites 11-14 Suffolk House 263 Banburv Road Oxford OX2 7HN
#### Indications for Use
Brainomix 360 e-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.
The software automatically registers images and uses an Atlas to segment and analyze ASPECTS regions. Brainomix 360 e-ASPECTS extracts image data from individual voxels in the image to provide analysis and computer analytics and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) Score.
Brainomix 360 e-ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score.
Brainomix 360 e-ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 6 hours from time last known well).
Rapid ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data. The Software automatically registers images and segments and analyzes ASPECTS Regions of Interest (ROIs). Rapid ASPECTS extracts image data for the ROI(s) to provide analysis and computer analytics based on morphological characteristics. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) Score. Rapid ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. This device provides information that may be useful in the characterization of early ischemic brain tissue injury during image interpretation (within 6 hours). Rapid ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment using the ASPECTS atlas definitions and atlas display including highlighted ROIs and numerical scoring.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "BRAINOMIX" with a logo to the left. The logo is a teal-colored graphic that resembles a brain. The word "BRAINOMIX" is written in a sans-serif font, with the "A" in "BRAINOMIX" being teal and the rest of the letters being gray.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
Brainomix 360 e-ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. Brainomix 360 e-ASPECTS additionally provides a visualization of the voxels contributing to the automated ASPECTS score and the voxels excluded from the automated ASPECTS score.
Limitations:
- Brainomix 360 e-ASPECTS 1. is not intended for primary interpretation of CT images. It is used to assist physician evaluation.
- 2. Brainomix 360 e-ASPECTS has been validated in patients with known MCA or ICA occlusion prior to ASPECTS scoring.
- 3. Brainomix 360 e-ASPECTS is not suitable for use on brain scans displaying neurological pathologies other than acute stroke, such as tumors or abscesses, hemorrhagic transformation and hematoma.
- 4. Use of Brainomix 360 e-ASPECTS Module in clinical settings other than brain ischemia within 6 hours from time last known well, caused by known ICA or
Limitations:
1. Rapid ASPECTS is not intended for primary interpretation of CT images, it is used to assist physician evaluation.
2. Rapid ASPECTS has been validated in patients with known MCA or ICA Occlusion prior to ASPECTS scoring.
3. Use of the Rapid ASPECTS Module in clinical settings other than early brain ischemia (within 6 hours) caused by known ICA or MCA occlusions has not been tested.
4. Rapid ASPECTS has been validated and is intended to be used on GE Lightspeed VCT Scanners.
Contraindications/Exclusions/Cauti ons:
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate.
- Hemorrhagic Transformation, Hematoma
- Very thin or no Ventricles
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is teal, and the letters "A" and "I" in "BRAINOMIX" are also teal, while the rest of the letters are gray.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
| Characteristic/Parameter | Brainomix 360 e-ASPECTS –<br>Proposed Device | iSchemaView Rapid ASPECTS -<br>Predicate Device (K200760) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| | MCA occlusions has not<br>been tested.<br>5. Brainomix 360 e-ASPECTS<br>has been validated and is<br>intended to be used on<br>Siemens Somatom<br>Definition scanners.<br>6. Brainomix 360 e-ASPECTS<br>has only been validated<br>and is intended to be used<br>in patient populations<br>aged over 21 years.<br>7. Brainomix 360 e-ASPECTS<br>is not intended for mobile<br>diagnostic use. Images<br>viewed on a mobile<br>platform are compressed<br>preview images and not<br>for diagnostic<br>interpretation.<br><br>Contraindications/Exclusions/Cautions: | |
| | • Patient motion: Excessive<br>patient motion leading to<br>artifacts that make the<br>scan technically<br>inadequate<br>• Hemorrhagic<br>Transformation,<br>Hematoma | |
| Environment of use | Clinical/Hospital environment | Clinical/Hospital environment |
| Environment of use | Clinical/Hospital environment | Clinical/Hospital environment |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Energy used and/or<br>delivered | None - software only application.<br>The software application does not<br>deliver or depend on energy<br>delivered to or from patients | None – software only application.<br>The software application does not<br>deliver or depend on energy<br>delivered to or from patients |
| Primary Users | Neuroradiologist/Clinician | Neuroradiologist/Clinician |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "A" in "BRAINOMIX" is colored in a gradient from blue to green, while the rest of the letters are in a dark gray color.
| Characteristic/Parameter | Brainomix 360 e-ASPECTS –<br>Proposed Device | iSchemaView Rapid ASPECTS -<br>Predicate Device (K200760) |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Clinical<br>Application/Anatomical Region | Stroke/Head | Stroke/Head |
| Standard of Care<br>Representation | ASPECTS Scoring | ASPECTS Scoring |
| Design: Modalities for<br>image processing and<br>visualization | CT | CT |
| Technical<br>Implementation | ML/AI/Random Forest | ML/AI/Random Forest |
| Image Overlay | ASPECTS regions, highlighted by<br>algorithms. Voxel-wise analysis<br>visualized as a heat map. | ASPECTS Atlas ROIs, highlighted by<br>algorithms. |
| Gating Conditions | Users must confirm ICA or MCA<br>occlusion prior to accessing<br>Brainomix 360 e-ASPECTS results | Users must confirm LVO prior to<br>accessing Rapid ASPECTS results |
| Design: PACS<br>functionality | View process and analyze medical<br>images. performs standard PACS<br>functions with respect to querying<br>and listing | same |
| Design: DICOM<br>compliance | Yes | Yes |
| Design: Computer<br>Platform | Standard off-the-shelf server or<br>virtual server | same |
| Design: Data<br>acquisition | Acquires medical image data from<br>DICOM compliant imaging devices<br>and modalities | same |
| Alters Standard of<br>Care Workflow | In parallel to | In parallel to |
| Materials | N/A – Software only device | same |
| Biocompatibility | N/A – Software only device | same |
| Sterility | N/A – Software only device | same |
| Electrical Safety | N/A – Software only device | same |
| Mechanical Safety | N/A – Software only device | same |
| Chemical Safety | N/A – Software only device | same |
| Thermal Safety | N/A – Software only device | same |
| Radiation Safety | N/A – Software only device | same |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of blue and green, while the word "BRAINOMIX" is in a dark gray color, with the "A" in "BRAINOMIX" having a blue gradient.
Brainomix Limited Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
# Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Software Verification and Validation Testing
Brainomix 360 e-ASPECTS complies with DICOM (Digital Imaging and Communications in Medicine) – Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20
Brainomix 360 e-ASPECTS is tested against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission. Algorithm validation testing has additionally been performed using phantom data with known properties.
The results of software verification and validation and algorithmic testing demonstrate that Brainomix 360 e-ASPECTS has met all design requirements and specifications associated with the intended use of the software.
#### Stand-alone Performance Testing
Stand-alone performance testing was conducted to comply with special controls for this device type. Stand-alone performance testing demonstrated that the proposed device provides accurate representation of key processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the stand-alone performance testing demonstrates the generalized performance for a range of typical patient demographics and a typical range of cases with confounding factors and including radiologically normal cases which is the same as for the predicate device. The standalone performance dataset was additionally enriched to include a greater number of cases within the low ASPECTS and late time-window sub-groups to demonstrate generalizability across these groups.
The stand-alone performance testing dataset comprised 256 non-contrast CT scans from patients with suspected and later confirmed acute ischemic stroke with available follow up imaging at 24 hours, from 8 different USA institutions_Patients were admitted between March 2014 and March 2020. The median age was 71 years, and 45% were female. Median NIHSS was 18 with an interquartile range of 12 to 22. 141 patients presented within 6 hours of time last seen well, 62 presented between 6 and 16 hours and 53 were unknown onset time. 180 patients had a proximal vessel occlusion (internal carotid artery or M1 segment or middle cerebral artery) and 76 were nonproximal. Ground truth ASPECTS score was <7 in 43 patients and >6 in 213 patients.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of light blue to white. The letters "BRAI" are dark gray, the letters "NO" are a gradient of light blue to dark gray, and the letters "MIX" are dark gray.
Suites 11-14 Suffolk House 263 Banburv Road Oxford OX2 7HN
Overall performance in 256 patients showed an area under the curve of 83% (81-85, 95% Cl), with a sensitivity of 68% (57-72) and a specificity of 97% (86-98).
This stand-alone performance testing demonstrates that e-ASPECTS performance generalises to a range of image acquisition and reconstruction parameters encountered in multiple clinical sites and for a range of scanner types and manufacturers.
Stand-alone performance results were generated to demonstrate generalizable performance for a range of typical patient demographics (including patient sex, age, site of occlusion, and ASPECTS score) and within individual ASPECTS regions. Consistent performance was observed in subgroups dichotomised by median age, and subgroups defined by sex. Performance was slightly lower in the non-proximal vessel occlusion subgroup (AUC, 78% versus 84%) although this was not statistically significant. Consistent performance was also observed between the grouped cortical ASPECTS regions compared to grouped basal ganglia ASPECTS regions, although at the level of the individual ASPECTS region performance did vary. Performance was lower in the M4, M6, and internal capsule regions, although this difference was not statistically significant.
#### Digital phantom testing
Specific validation has been carried out to test the heatmaps. This testing made use of synthetic digital phantom data created to provide a ground truth against which the performance of the heatmaps can be assessed and demonstrated. For clarity, we report that this testing showed a correlation between the volumes of the e-ASPECTS heatmaps and the volumes of synthetic hypodensities generated in the phantom dataset.
The heatmaps are generated and visualized to provide transparency of the basis of the ASPECTS score. The phantom testing results show a correlation (r>=0.95) between the volumes of e-ASPECTS heatmaps and the volume of synthetic hypodensities generated in the 200 digital phantom dataset.
#### Clinical Studies
A multiple reader multiple case (MRMC) cross-over study was conducted to assess performance of Brainomix 360 e-ASPECTS. The MRMC study evaluated the performance of 10 clinical readers (4 neurologists and 6 radiologists) when scoring the ASPECTS for 54 clinically representative NCCT retrospective scans with and without the aid of Brainomix 360 e-ASPECTS. Data truthing was performed by three expert neuroradiologists with access to clinical data and follow up imaging.
The MRMC study dataset comprised 54 non-contrast CT scans from patients with acute ischemic stroke affecting the middle cerebral artery territory and eligible for acute reperfusion therapy. Patients must have had no evidence of intracranial hemorrhage and imaging data had to be of adequate quality with associated demographic and follow up imaging available. Patients aged 21 years and younger were excluded. The median age was 72 years, and 48% were female. Median NIHSS was 17 with an interquartile range of 12 to 21. 36 patients presented within 6 hours of time
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of light blue to white, and the letters "AI" in the word "BRAINOMIX" are also in a light blue gradient, while the rest of the letters are in dark gray.
Suites 11-14 Suffolk House 263 Banburv Road Oxford OX2 7HN
last seen well, 10 presented between 6 and 16 hours and 8 were unknown onset time. 40 patients had a proximal vessel occlusion (internal carotid artery or M1 segment or middle cerebral artery) and 14 were non-proximal. Ground truth ASPECTS score was <7 in 12 patients and >6 in 42 patients.
Comparison of the area under the curve (AUC) for each reader with and without the support of the e-ASPECTS tool showed a statistically significant improvement of 0.02 from 0.81 to 0.83 (p=0.028) which was the primary endpoint for the study. When comparing reader performance with the ground truth, the improved AUC in the primary endpoint was driven by an increase in both the sensitivity (positive percentage agreement, from 66 to 70%) and a small improvement in specificity (negative percentage agreement, 96%) when assisted by e-ASPECTS compared to unassisted reading. Overall percentage agreement (accuracy) also improved in line with the AUC result from 93 to 94%. Subgroup analysis based on the clinical training of the reader (radiologist versus neurologist) demonstrates a consistent impact of Brainomix 360 e-ASPECTS across reader groups. Analysis of individual reader ROC curves showed greater magnitude of increases in AUC were observed in users with lower unassisted performances ("lower performers"), and smaller changes in readers with higher unassisted performance ("high performers"). The overall ASPECTS score was more consistent between readers with use of e-ASPECTS. The range in AUC between users was also narrower with Brainomix 360 e-ASPECTS than unassisted indicating a reduction in the variation of performance between different readers when Brainomix 360 e-ASPECTS outputs are available.
Subgroup analysis of the individual regions within the ASPECTS scoring system showed that the impact of e-ASPECTS is consistent in both deep and cortical ASPECTS regions.
As the study used retrospective data, there were no specific safety endpoints in the reader study.
This study demonstrates substantial equivalence of the Brainomix 360 e-ASPECTS software for improving reader accuracy, compared to the predicate device. The results showed statistically significant improvement in the agreement between the readers and a reference standard when using the Brainomix 360 e-ASPECTS software compared to routine clinical practice.
| Risk Benefit Summary | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Benefits | This device provides a systematic, automated<br>analysis of NCCT scans of the head to provide a<br>standardized, automated ASPECT score for<br>Stroke workup. |
| | The clinical reader study, which included 4<br>neurologists and 6 neuroradiologists, |
# Risk Benefit Analysis
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "AI" in "BRAINOMIX" is a gradient of blue, while the rest of the letters are gray.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
demonstrated a statistically significant improvement in the readers' AUC (0.02 increase, with 0.04 sensitivity improvement and 0.01 specificity improvement) when scoring was performed in conjunction with the e-ASPECTS output.
In a subgroup analysis, the benefit of the software was most substantial among the "lower performers" (clinical readers with a lower unaided performance compared to the ground truth) whereas "high performers" were minimally affected by the device output. Both cohorts included a mix of US board certified neurologist and neuroradiologist readers, indicating potential benefit for less experienced clinicians evaluating NCCT for ASPECTS.
It is important to note that ASPECTS evaluations can be challenging, particularly for less experienced clinicians, and studies have shown varying levels of interobserver agreement for ASPECTS evaluation. Overall, the system should provide more consistent and timely benefit of aided reads for lessexperienced or less confident users in the evaluation of overall ASPECTS.
Timely, consistent and accurate assessments of acute ischemic patients with known MCA and ICA occlusions can help ensure that physicians determine appropriate patient management at the right time. ASPECTS may be used in the assessment of patients for consistency with patient selection for neurothrombectomy and other acute stroke therapeutic interventions. Proper and timely patient selection increases the likelihood for successful therapeutic interventions proven to reduce long term disability. The duration of the effect is lifetime.
Based on the data provided it is not possible to accurately predict the exact likelihood of
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in bold, sans-serif font. The brain icon and the "AI" in BRAINOMIX are in a gradient of light blue to dark blue, while the rest of the letters are in dark gray.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
| | patient benefit. The study demonstrates<br>improvement in the assessment for readers<br>interpreting NCCT scans for overall ASPECTS<br>assessment.<br><br>This device additionally also provides<br>transparent Al processing enabling users tobeen determined and enables users to perform<br>a more technically informed review of the<br>automated output. This is a significant benefit<br>which may help mitigate risks related to false<br>negatives or false positives in the automated<br>output, particularly for more challenging cases. |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of the Risks | There are minimal potential risks associated<br>with the use of the device. |
| | Incorrect scoring which may result in false<br>positive results and result to incorrect patient<br>management with possible adverse effects such<br>as, unnecessary additional medical imaging<br>and/or unnecessary additional diagnostic<br>workup. |
| | Incorrect scoring which may result in false<br>negative results may lead to complications,<br>including incorrect diagnosis and delay in<br>disease management. |
| | The device could be misused to analyze images<br>from an unintended patient population or on<br>images acquired with incompatible imaging<br>hardware or incompatible image acquisition<br>parameters, leading to inappropriate diagnostic<br>information being displayed to the user. |
| | Device failure could lead to the absence of<br>results, delay of results or incorrect results,<br>which could likewise lead to inaccurate patient<br>assessment. |
| Summary of additional factors | However, based on the performance data and<br>the application of mitigating measures<br>(including but not limited to general controls<br>and special controls established for this device<br>type), use of the device is unlikely to decrease<br>diagnostic performance of the user and<br>possible misuse of the device does not present<br>additional risks compared with misuse of<br>other types of radiological image processing<br>devices. |
| | The device includes a gating condition in<br>software to help ensure that the software is not<br>used on stroke mimics or the incorrect patient<br>population. |
| | The physician may either confirm or overrule<br>the device |
| | Heatmap provides additional transparency on<br>regions evaluated by the ASPECTS, which may<br>be used by the clinician to qualify or correct<br>device outputs |
| | The study was enriched to cover the range of<br>ASPECT scores; and, the readers in practice may<br>not experience a significant improvement in<br>determining ASPECTS. |
| | This device may improve the diagnostic<br>accuracy of less experienced or less confident<br>physicians in rating NCCT for the ASPECT score<br>in the context of acute ischemic stroke caused<br>by ICA or MCA occlusion (confirmed using<br>vascular imaging such as CTA or MRA). |
| Conclusion | This device shows positive improvement in<br>diagnostic accuracy in comparison to the<br>reference standard. |
| | Although some performance data may suggest<br>that sensitivity for some ASPECTS regions may<br>be diminished or inconsistent in comparison to<br>other region however, sample size limitations |
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "A" in "BRAINOMIX" is a gradient of blue, while the rest of the letters are gray.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of blue and teal, while the letters "AI" in "BRAINOMIX" are also in a similar gradient, contrasting with the rest of the word which is in a dark gray color.
Suites 11-14 Suffolk House 263 Banburv Road Oxford OX2 7HN
limit conclusions in this regard. The clinical reader study evidence provided demonstrates improvement in the overall performance of aided readers compared to the unaided readers. The greatest benefit was observed for the "lower performers" – bringing them to the level of the "higher performers" while avoiding adversely impacting "high performers" accuracy. The software gating conditions and other warnings integrated in the software UI are expected to be helpful in ensuring appropriate use of the device and directing clinician attention to particular regions which may be more difficult to interpret, both for users and for the device.
# Prescriptive Statement
Caution: Federal law restricts this device to sale by of on the order of a physician.
# Conclusion
In conclusion, the predicate device has the same intended use and substantially similar indications for use and technological characteristics as Brainomix 360 e-ASPECTS. Performance data demonstrates that Brainomix 360 e-ASPECTS performs as intended within the same clinically relevant parameters for the intended use as the predicate device.
The predicate device was cleared based in part on the results of clinical studies, therefore a clinical study was required to support substantial equivalence. The clinical data demonstrates that Brainomix 360 e-ASPECTS shows a significant improvement in the agreement between the readers and a reference standard when using the Brainomix 360 e-ASPECTS software compared to routine clinical practice. Brainomix 360 e-ASPECTS is intended to improve reader performance in estimating an overall ASPECTS score but may be less sensitive for some ASPECTS regions such as the IC and M4-M6 regions based on the performance data disclosed. The risk/benefit profile of Brainomix 360 e-ASPECTS remains favorable based on the favorable impact on readers' overall aided performance in calculating an ASPECTS score (as measured in the reader study), and the inherent inter-reader variability in identifying individual affected ASPECTS regions.
Brainomix 360 e-ASPECTS has been designed, verified and validated in compliance with 21 CFR Part 820.30 requirements and has been designed to meet the requirements associated with ANSI AAMI
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The "A" in "BRAINOMIX" is stylized with a blue triangle in the middle. The brain icon is a gradient of blue and green.
Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN
ISO 14971:2007/(R)2016. Software verification and validation, performance testing and risk management demonstrates that Brainomix 360 e-ASPECTS is safe and effective for use as intended and described in its indications for use.
We propose that Brainomix 360 e-ASPECTS is substantially equivalent to the Rapid ASPECTS predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.