The Dental Air Polishing Handpiece, AP-H Plus is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances. The Dental Air Polishing Handpiece, AP-H Plus is intended for patients suffering from periodontal disease. The Dental Air Polishing Handpiece, AP-H Plus is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment.
Device Story
Air-driven dental handpiece; connects to standard dental unit air turbine; projects mixture of water, air, and prophylaxis powder (glycine, sodium bicarbonate, or erythritol) onto tooth surfaces. Used by trained dental professionals in clinical settings for supragingival and subgingival cleaning/polishing. Components include supragingival handpiece (AP-1 Plus), subgingival handpiece (AP-2 Plus), spray head, nozzle, powder storage chamber, and tail cord connector. Device removes soft deposits, discoloration, and plaque; prepares surfaces for restorative/orthodontic procedures. Benefits include improved hygiene and surface preparation for dental treatments. Reusable; requires autoclave sterilization between uses.
Clinical Evidence
No clinical data.
Technological Characteristics
Air-powered dental handpiece; continuous operation mode. Compatible with glycine, sodium bicarbonate, and erythritol powders. Materials comply with ISO 10993-1. Sterilization via moist heat (autoclave) per ANSI/AAMI ST79 and ISO 17665-1. Performance standards: ISO 20608 (powder jet handpieces), ISO 9168 (hose connectors), IEC 62366, ISO 15223-1, ISO 17664. Weight: ~0.125kg.
Indications for Use
Indicated for patients with periodontal disease requiring non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment; also for cleaning/polishing teeth, removing soft deposits/discoloration, preparing teeth for restorative procedures (fillings, inlays, veneers), cleaning implant abutments, and cleaning prior to shade matching, fluoridation, bleaching, or placement of crowns, bridges, and orthodontic appliances.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 28, 2022
Guilin Woodpecker Medical Instrument Co., Ltd. % Charles Mack Principal Engineer IRC 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K221548
Trade/Device Name: Dental Air Polishing Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: August 31, 2022 Received: August 31, 2022
Dear Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K221548
Device Name Dental Air Polishing Handpiece
#### Indications for Use (Describe)
The Dental Air Polishing Handpiece, AP-H Plus is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances.
The Dental Air Polishing Handpiece, AP-H Plus is intended for patients suffering from periodontal disease.
The Dental Air Polishing Handpiece, AP-H Plus is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for Woodpecker, a company that specializes in dental equipment. The logo consists of a stylized red woodpecker icon on the left, with the word "WOODPECKER" in red, sans-serif font to the right of the icon. The woodpecker icon is a simplified, abstract representation of the bird, with a long beak and a curved body.
| K221548 | 510(k) SUMMARY |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Preparation Date: | September 27, 2022 |
| Manufacturer's Name and Address: | Guilin Woodpecker Medical Instrument<br>Co., Ltd<br>Information Industrial Park, Guilin<br>National High-Tech Zone, Guilin City,<br>Guangxi Province, China 541004 |
| Corresponding Official: | Charles Mack |
| Telephone Number: | 931-625-4938 |
| Email Address: | charliemack@irc-us.com |
| Trade Name: | Dental Air Polishing Handpiece<br>AP-H Plus |
| Common Name(s): | handpiece, air-powered, dental |
| Requlation Name(s): | Dental handpiece and accessories. |
| Regulation Number(s): | 21CFR872.4200 |
| Product Code: | EFB |
| Device Class: | Class II |
| Predicate Device: | E.M.S. Electro Medical Systems S.A.<br>K151912 |
{4}------------------------------------------------
# Device Description:
The Dental Air Polishing Handpiece, AP-H Plus is an air driven dental handpiece for the use by a trained professional in the field of supragingival and the subgingival treatment. The devices are air-powered handpieces that are reusable.
The proposed Dental Air Polishing Handpiece, AP-H Plus device is similar in design to the predicates AIR-FLOW handy 3.0 PLUS (K151912) . The proposed Dental Air Polishing Handpiece, AP-H Plus device and the predicates connect to a standard air turbine connection on a dental operative unit and deliver a mixture of water, air, and prophylaxis powder to a treatment site. The subject Dental Air Polishing Handpiece is comprised of a supragingival handpiece (model: AP - 1 Plus) a subgingival handpiece (model: AP - 2 Plus), spray head, nozzle, powder storage chamber, and tail cord connector.
# Indications for Use
The Dental Air Polishing Handpiece, AP-H Plus is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances.
The Dental Air Polishing Handpiece, AP-H Plus is intended for patients suffering from periodontal disease.
The Dental Air Polishing Handpiece, AP-H Plus is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment.
{5}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device
| Characteristics | Subject Device | Predicate Device | Discussion |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Dental Air Polishing Handpiece, AP-H Plus | AIR-FLOW handy 3.0 PLUS | - |
| 510K Applicant | Guilin Woodpecker Medical Instrument<br>Co., Ltd. | E.M.S. ELECTRO MEDICAL<br>SYSTEMS S.A. | - |
| 510(K) Number | Pending | K151912 | - |
| Regulation<br>Number | CFR872.4200 | CFR872.4200 | Identical |
| Product Code | EFB | EFB | Identical |
| Classification<br>Name | Air-powered dental handpiece | Air-powered dental handpiece | Identical |
| OTC or<br>Prescription | Prescription Use | Prescription Use | Identical |
| Medical Specialty | Dental | Dental | Identical |
| Characteristics | Subject Device | Predicate Device | Discussion |
| Indication for Use | The Dental Air Polishing Handpiece, AP-H Plus is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances. | The AIR-FLOW handy 3.0 PLUS is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and EMS prophylaxis powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances. | Identical |
| | The Dental Air Polishing Handpiece, AP-H Plus is intended for patients suffering from periodontal disease. | The AIR-FLOW handy 3.0 PLUS is intended for patients suffering from periodontal disease. | |
| | The Dental Air Polishing Handpiece, AP-H Plus is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment. | The AIR-FLOW handy 3.0 PLUS is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment. | |
| Characteristics | Subject Device | Predicate Device | Discussion |
| Treatment Site | Supragingival and Subgingival | Supragingival and Subgingival | Identical |
| Compatible<br>Prophylaxis<br>Powders | Glycine<br>Sodium bicarbonate<br>Erythritol | Glycine | The subject device including two<br>additional compatible prophylaxis<br>powders; They confirm to the<br>same performance requirement<br>as predicate device |
| Function | Air-polishing | Air-polishing | Identical |
| Mechanism of<br>Action | Projection of water/air/powder mixture | Projection of water/air/powder<br>mixture | Identical |
| Operation Mode | Continuous operation | Continuous operation | Identical |
| Service Pressure<br>to the Turbine<br>Connection | Water:<br>1 to 2.2 bar (1000-2200 hPa)<br>Air:<br>Static pressure 2.7 to 3.5 bar (2700-<br>3500 hPa) | Water:<br>1 to 2.2 bar (1000-2200 hPa)<br>Air:<br>Static pressure 2.7 to 3.5 bar<br>(2700-3500 hPa) | Identical |
| Operating<br>Conditions | +10°C to +40°C, 30% to 75% relative<br>humidity,<br>700 hPa to 1060 hPa air pressure, | +10°C to +40°C, 30% to 75%<br>relative humidity,<br>700 hPa to 1060 hPa air<br>pressure, | Identical |
| Usage | Re-useable | Re-useable | Identical |
| Weight | Approx. 0.125kg | Approx. 0.150kg | Similar |
| How Supplied | Non-sterile; Subject to be sterilized<br>before each use | Non-sterile; Subject to be<br>sterilized before each use | Identical |
| Applicable<br>Sterilization | Moist heat - Autoclave sterilization | Moist heat - Autoclave<br>sterilization | Identical |
| Performance | ANSI/AAMI ST79<br>ANSI/AAMI/ISO 17665-1<br>IEC 62366<br>ISO 15223-1<br>ISO 17664 | ANSI/AAMI ST79<br>ANSI/AAMI/ISO 17665-1<br>IEC 62366<br>ISO 15223-1<br>ISO 17664 | Identical |
| Biocompatibility | Complies with ISO 10993-1 | Complies with ISO 10993-1 | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Performance Testing
Performance testing was provided in support of the substantial equivalence determination and to validate and verify that the Dental Air Polishing Handpiece, AP-H Plus met all requirements of related international standards. The results of these tests demonstrate compliance with the requirements of the consensus standards noted below.
# Non-clinical Testing
#### Performance Testing
- ISO 20608 First edition 2018-04 Dentistry Powder jet handpieces and powders
- ISO 9168 Dentistry Hose connectors for air driven dental handpieces i
## Biocompatibility Information
The subject device is classified as surface device and contact mucosal membrane for limited contact (duration ≤ 24 h). The following tests were performed to ensure compliance with biocompatibility requirements.
- । ISO10993-1: BIOLOGICAL EVALUATION OF MEDICAL DEVICES
- ISO10993-5 Cytotoxicity Tests -
- ISO10993-10 Skin Irritation test -
- ISO10993-10 Oral Mucosa Irritation test -
- ISO10993-10 Sensitization Test -
# Sterility Information
The handpiece is supplied non-sterile and subject to be sterilized by autoclave prior to use. Fractional pre-vacuum cycle at 132°C for 4 min and a drying time of 20 min same as predicate device.
Recommended sterilization method which has been validated as below: Sterilization validation for the handpieces was performed in accordance with the standards noted below:
- ANSI AAMI ST79:2017, Comprehensive quide to steam sterilization and । sterility assurance in health care facilities
- । ISO 17665-1:2006, Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices (Sterility)
The cleaning method has been validated for the handpiece and the powder chamber per AAMI TIR30 and the FDA Guidance Reprocessing Medical Devices in Health
{9}------------------------------------------------
Care Settings: Validation Methods and Labeling.
The intermediate-level disinfection method has been validation for the powder chamber and cord per AAMI TIR12:
{10}------------------------------------------------
## Clinical Test:
No clinical study is included in this submission.
### Shelf Life:
The device is not delivered sterile and low likelihood of time-dependent product degradation, therefore, the shelf-life is not applicable.
#### Conclusions:
The differences between the predicate and the subject device do not raise any new or different questions of substantial equivalence. The Dental Air Polishing Handpiece, AP-H Plus is substantially equivalent to the E.M.S. Electro Medical Systems S.A., Inc. AIR-FLOW handy 3.0 PLUS cleared under K151912 with respect to the indications for use, target populations, treatment method, and technological characteristics.
-END-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.