Nightwear Aligners are indicated for the alignment of permanent teeth during orthodontic treatment of malocclusion.
Device Story
Nightwear Aligners are thermoformed, patient-specific, transparent polyurethane plastic trays used for orthodontic tooth movement. Input consists of digital dental scans provided by a dental professional. A treatment plan is generated and approved by the clinician, resulting in a sequence of 3D-printed models used to thermoform the aligners. The device is worn by the patient for 9-12 hours daily, removed for eating and cleaning. Aligners apply gentle, continuous force to teeth to achieve incremental movement toward a final position. Clinicians monitor treatment progress from the first to the final aligner. The device benefits patients by providing a removable, aesthetic orthodontic treatment option for malocclusion.
Clinical Evidence
Clinical study conducted on 20 patients under IDE G200298. Results demonstrated that tooth alignment occurs with 9-12 hours of daily wear for one week per aligner stage in patients with mild malocclusions as determined by a treating dentist.
Technological Characteristics
Thermoformed transparent polyurethane plastic aligners. Patient-specific dimensions. Biocompatibility per ISO 10993-1, -5, -10, -11. Material testing per ASTM D638 (tensile) and ASTM D790 (flexural). Manufacturing involves 3D printing of models followed by thermoforming.
Indications for Use
Indicated for alignment of permanent teeth during orthodontic treatment of malocclusion in patients requiring mild malocclusion correction.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2023
Smylio Inc. % Jennifer Day Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043
Re: K221537
Trade/Device Name: Nightwear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 26, 2022 Received: May 27, 2022
Dear Jennifer Day:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221537
Device Name Nightwear Aligners
Indications for Use (Describe)
Nightwear Aligners are indicated for the alignment of permanent teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Attachment A: 510(k) Summary
### 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information for this submission in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Smylio Inc.<br>48890 Milmont Dr<br>Suite 101D<br>Fremont CA, 94538 |
|---------------------------|----------------------------------------------------------------------------------------|
| Company Contact Person: | Ren Menon, CEO |
| Phone: | 510-868-2184 |
| Email: | rmenon@orthofx.com |
| Submission Correspondent: | Jennifer Day, Regulatory Affairs Consultant<br>Breanne Butler, PhD |
| Company: | Prime Path Medtech Inc. |
| Address: | 1321 Upland Dr, Ste 6792 Houston, TX 77043 |
| Phone: | 860-810-5594 |
| Email: | jday@primepathmedtech.com |
| Date Prepared: | May 26, 2022 |
| Proprietary Name: | Smylio Nightwear Aligners |
| Common Name: | Orthodontic plastic bracket. |
| Product Code: | NXC – Orthodontic plastic bracket. |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Primary Predicate Device: | iSMILE Aligners (K200908) |
| Reference Predicates: | Smylio Invisible Clear Aligners (K212660)<br>Smylio Invisible Clear Aligners (K172784) |
{4}------------------------------------------------
#### Device Description:
The Smylio Nightwear Aligners are thermoformed plastic aligners designed to be worn in sequence to facilitate the movement to the final desired position. The sequential aligners introduce incremental movements that move teeth by way of gentle continuous force. The aligners are to be worn at least 9 to 12 hours a day and are to be removed for eating and for cleaning.
Smylio Nightwear Aligners are designed from digital scans of a patient's dentition submitted by a dental health professional (e.g., dentist or orthodontist). Using the scan, sequential dental models are designed and approved by the treating physician prior to manufacturing.
Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The aligners are shipped to the treating clinician's office where they are then distributed to the patient in sequential stages. The patients' dental health professional then monitors their treatment of the first aligner to the delivery of the final aligner.
#### Indications for Use:
Nightwear Aligners are indicated for the alignment of permanent teeth during orthodontic treatment of malocclusion.
#### Comparison to Predicate Devices:
Smylio Nightwear Aligners are functionally equivalent to the following predicate device: iSMILE Aligners (K200908). The following table demonstrates the functional specifications of Smylio Nightwear Aligners are substantially equivalent to the predicate devices.
| Specification | Subject Device: Smylio<br>Nightwear Aligners | Predicate Device: iSMILE<br>Aligners (K200908) | Comparison<br>Result |
|------------------------|----------------------------------------------|------------------------------------------------|----------------------|
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | Same |
| Classification<br>Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket | Same |
| Product Code | NXC | NXC | Same |
| Classification | Class II | Class II | Same |
Table 1. Predicate Device Comparison Table
{5}------------------------------------------------
| Specification | Subject Device: Smylio<br>Nightwear Aligners | Predicate Device: iSMILE<br>Aligners (K200908) | Comparison<br>Result |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| OTC or Rx | Rx | Rx | Same |
| Material | Thermoplastic Polymer | Thermoplastic Polymer | Same |
| Material<br>Properties | Demonstrates sufficient<br>tensile strength, elasticity,<br>ductility, chemical resistance,<br>and clarity for use as a clear<br>tray aligner. | Demonstrates sufficient<br>tensile strength, elasticity,<br>ductility, chemical resistance,<br>and clarity for use as a clear<br>tray aligner. | Same |
| Biocompatible | Yes | Yes | Same |
| Specification | Subject Device: Smylio<br>Nightwear Aligners | Predicate Device: iSMILE<br>Aligners (K200908) | Comparison<br>Result |
| Device<br>Description | Nightwear Aligners are a<br>sequence of transparent<br>aligners created from a<br>digital orthodontic treatment<br>plan. The Nightwear Aligners<br>are fabricated of a<br>transparent thermoformed<br>polyurethane plastic. Each<br>aligner delivers a unique<br>combination of minor forces<br>to create the planned tooth<br>movement. The digital<br>orthodontic treatment plan<br>is created by a dental health<br>professional.<br>The Nightwear Aligners<br>incorporate the use of<br>attachments to create spaces<br>and force points in order to<br>cause minor tooth<br>movement. These force<br>points are located in specific<br>areas and positioned in such<br>a way that they provide a<br>continuous force which<br>slowly dissipates over time<br>on the tooth to be moved for<br>as long as the aligner is worn. | The iSMILE is a sequence of<br>transparent aligners created<br>from a digital orthodontic<br>treatment plan. The iSMILE<br>aligner is fabricated of a<br>transparent thermoformed<br>polyurethane plastic. Each<br>aligner delivers a unique<br>combination of minor forces<br>to create the planned tooth<br>movement. The digital<br>orthodontic treatment plan<br>is created by a dental health<br>professional.<br>The iSMILE Aligner<br>incorporates the use of<br>attachments to create spaces<br>and force points in order to<br>cause minor tooth<br>movement. These force<br>points are located in specific<br>areas and positioned in such<br>a way that they provide a<br>continuous force which<br>slowly dissipates over time<br>on the tooth to be moved for<br>as long as the aligner is worn. | Same |
| Anatomy<br>Location | Mouth; mucosal | Mouth; mucosal | Same |
| Size | Patient Specific | Patient Specific | Same |
| Specification | Subject Device: Smylio<br>Nightwear Aligners | Predicate Device: iSMILE<br>Aligners (K200908) | Comparison<br>Result |
| Manufacturing<br>Method | Thermoforming | Thermoforming | Same |
| Patient<br>Removable? | Yes | Yes | Same |
| Indication for<br>Use | Nightwear Aligners are<br>indicated for the alignment<br>of permanent teeth during<br>orthodontic treatment of<br>malocclusion. | iSMILE is indicated for the<br>alignment of teeth during<br>orthodontic treatment of<br>malocclusion. | Same |
| Mode of<br>Action | Orthodontic tooth<br>movement occurs through<br>forces applied by the<br>appliance to the dentition as<br>each tooth follows the<br>programmed displacement<br>based on a doctor's<br>prescription | Orthodontic tooth<br>movement occurs through<br>forces applied by the<br>appliance to the dentition as<br>each tooth follows the<br>programmed displacement<br>based on a doctor's<br>prescription | Same |
{6}------------------------------------------------
{7}------------------------------------------------
#### Comparison of Indications for Use to Predicate Devices:
The indications for use of the Smylio Nightwear Aligners are the iSMILE Aligners (K200908) as they are both indicated for use in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces. Thus, the Smylio Nightwear Aligners can be considered substantially equivalent to its predicate device.
#### Comparison of Technological Characteristics to Predicate Devices:
Based on the above comparison, the design, construction, and performance characteristics of the Smylio Nightwear Aligners is similar to that of the iSMILE Aligners (K200908). Thus, the Smylio Nightwear Aligners can be considered substantially equivalent to its predicate device.
#### Non-clinical performance testing:
The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces.
{8}------------------------------------------------
However, durability testing was conducted on the Smylio Nightwear Aligners Real world use was simulated to ensure that the aligner material and manufacturing process produced aligners that were suitable for their prescribed period of use.
Material property testing was performed to investigate elongation, tensile and flexural properties of the material in accordance with ASTM D638 and ASTM D790.
An internal manufacturing validation was performed to test the manufacturing process for Smylio Nightwear Aligners. The robustness of the process was demonstrated from 3D printing through thermoforming.
The thermoplastic material used for Smylio Nightwear Aligners is broadly used in many aligners. Biocompatibility testing for the aligner material, the only patient contacting material, was in accordance with the International Standard "Biological evaluation of medical devices" ISO 10993-1, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-10:2010, and ISO 10993-11:2017.
#### Clinical performance testing:
Clinical performance testing was conducted on 20 patients for the Smylio Nightwear Aligners under an FDA cleared IDE application (G200298). The study results demonstrated that tooth alignment may occur for patients with at least 9 to 12 hours of wear time per day for one week prior to moving to the next aligner for patients with mild malocclusions as determined by the treating dentist.
#### Conclusion:
Based on similarities in indications for use, technological characteristics, non-clinical performance testing, and clinical performance testing, Smylio Nightwear Aligners are substantially equivalent to the iSMILE Aligners (K200908).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.