Flowart

K221473 · Asset Medikal Tasarim As. · FPA · Sep 26, 2022 · General Hospital

Device Facts

Record IDK221473
Device NameFlowart
ApplicantAsset Medikal Tasarim As.
Product CodeFPA · General Hospital
Decision DateSep 26, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Extension sets are for administering fluids from container to a patient's vascular system through a vascular access device. Needleless valve allows the user to add medication into the without the use of a needle. Power Injection Rating and Flowrate (400 psi and 10 mL/sec).

Device Story

FlowArt is an intravascular extension set with an integrated needleless valve; facilitates fluid administration from containers to patient vascular systems; enables needle-free medication addition; supports power injection of contrast media at rates up to 10 mL/sec and pressures up to 400 psi; used in clinical settings by healthcare professionals; provides a closed system for fluid delivery; benefits patients by reducing needle-stick risks and enabling high-flow diagnostic imaging procedures.

Clinical Evidence

bench testing only

Technological Characteristics

Intravascular extension set with integrated needleless valve; supports power injection up to 400 psi and 10 mL/sec; Class II device (21 CFR 880.5440, Product Code FPA).

Indications for Use

Indicated for patients requiring fluid administration via vascular access device. Used for needleless medication delivery and power injection of contrast media up to 400 psi and 10 mL/sec.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Asset Medikal Tasarim AS. % Raymond Kelly Consultant Licensale Inc. 3422 Leonardo Lane New Smyrna Beach, Florida 32168 Re: K221473 Trade/Device Name: Flowart Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: August 17, 2022 Received: August 23, 2022 Dear Raymond Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K221473 Device Name FlowArt Indications for Use (Describe) Extension sets are for administering fluids from container to a patient's vascular system through a vascular access device. Needleless valve allows the user to add medication into the without the use of a needle. Power Injection Rating and Flowrate (400 psi and 10 mL/sec). | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------------------------| | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...