The E3 and Profile system is an electrodiagnostic device used to generate photic signals and to measure and display the electrical signals generated by the retina and the visual nervous system. It displays digitized electroretinogram (ERG), visual evoked potential (VEP) and sensory electro-oculogram (EOG) signals, power spectra and topographic maps. These functions are controlled and interpreted by trained medical professionals, for use with patients who are experiencing visual system abnormalities.
Device Story
Electrodiagnostic system; generates photic stimuli via LEDs or monitor; records electrical signals (ERG, VEP, EOG) via external EEG electrodes; used in hospitals, clinics, and physician offices by ophthalmologists and trained medical professionals. Device transforms raw input signals into digitized waveforms, power spectra, and 3D topographical maps; software performs spatial filtering and artifact rejection. Real-time display provided during 1-20 minute acquisition. Output assists clinicians in diagnosing visual system abnormalities by quantifying retinal and visual nervous system function.
Clinical Evidence
No clinical data. Bench testing only. Includes technical hardware/firmware/software verification, functional testing per ISCEV recommendations, compatibility testing with electrodes, and safety/performance testing per IEC 60601-1, IEC 60601-1-2, IEC 62366-1, IEC 62304, and ISO 10993 series.
Indicated for patients with ophthalmic conditions experiencing visual system abnormalities; used for quantification of electrophysiological response of retina and visual cortex.
Regulatory Classification
Identification
An evoked response photic stimulator is a device used to generate and display a shifting pattern or to apply a brief light stimulus to a patient's eye for use in evoked response measurements or for electroencephalogram (EEG) activation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Diagnosys LLC Jeffrey Farmer CEO 55 Technology Drive, Suite 100 Lowell, Massachusetts 01851
Re: K221471
Trade/Device Name: E3 and Profile Regulation Number: 21 CFR 882.1890 Regulation Name: Evoked Response Photic Stimulator Regulatory Class: Class II Product Code: GWE, HLT Dated: October 27, 2022 Received: October 28, 2022
Dear Jeffrey Farmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elvin Y. I
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221471
Device Name E3 and Profile
# Indications for Use (Describe)
The E3 and Profile system is an electrodiagnostic device used to generate photic signals and to measure and display the electrical signals generated by the retina and the visual nervous system. It displays digitized electroretinogram (ERG), visual evoked potential (VEP) and sensory electro-oculogram (EOG) signals, power spectra and topographic maps. These functions are controlled and interpreted by trained medical professionals, for use with patients who are experiencing visual system abnormalities.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="white-space: nowrap;"> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Diagnosys. The logo features the word "diagnosys" in a lowercase, sans-serif font, with the tagline "leading the wave" underneath in a smaller font. Above the word "diagnosys" is a stylized blue wave graphic, which adds a visual element to the logo.
510(k) SUMMARY for the E3 and Profile PER 21 CFR 807.92
Date prepared: October 24, 2022
| Company name: | Diagnosys LLC |
|---------------|--------------------------------|
| | 55 Technology Drive, Suite 100 |
| | Lowell, MA 01851 |
| | USA |
| | Telephone: 1-978-458-1600 |
| | Fax: 1-978-458-1755 |
Contact Person: Jeffrey Farmer, CEO Email: jeff(@diagnosysllc.com
Name of the device: E3, Profile
Classification name: 21 CFR 882.1890, Evoked response photic stimulator
Regulatory class: Class II
Product Code: GWE Subsequent product code, based on other device functions in Class II: HLT
# A. Legally Marketed Predicate Device
The E3 and Profile (also known as the Espion Ophthalmic Electrophysiology System) is substantially equivalent to the Vision Monitor MonPackONE system manufactured by MetroVision 510(k) number K211643. They are both hardware and software products. The E3 and Profile is substantially equivalent to the predicate devices with regard to device features and specifications as well as intended use. All devices are visual evoked response test systems with similar operating requirements that are based on standard clinical procedures. Devices consist of hardware and software to provide photic stimulations and analysis of the evoked response data that is collected.
# B. Device Description
Photopic stimuli are presented to the patient using LEDs or a monitor, using various number of elements in separately stimulated fields. Various modes are available for preferential stimulation of different retinal mechanism and isolation of signal from different retinal lavers. Data is recorded by up to 5 recording channels using conventional EEG electrodes (not provided with the device).
During the period of time that the system is acquiring data (1-20 minutes), there is a real time display of the raw and processed data presented to the user. Once the resulting individual waveforms are acquired,
.diagnosysllc.com
US: Diagnosys U.C.: 55 Technology Drive Suite 100 Lowell MA 01851 978-458-1600 Sa
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Diagnosys. The logo features the word "diagnosys" in a lowercase, sans-serif font, with a blue wave-like line above it. Below the word "diagnosys" is the tagline "leading the wave" in a smaller, italicized font. The overall design is simple and modern, with a focus on the company name and its message.
the signals are analyzed by software using algorithms for spatial filtering and artifact rejection. Data may be presented in a number of forms, including waves recorded at each of the points tested, color plots, or 3D topographical representation.
# C. Intended Use/Indications for Use
The E3 and Profile is an electrodiagnostic device used to generate photic signals and to measure and display the electrical signals generated by the retina and the visual nervous system. It displays digitized electroretinogram (ERG), visual evoked potential (VEP) and sensory electro-oculogram (EOG) signals, power spectra and topographic maps. These functions are controlled and interpreted by trained medical professionals, for use with patients who are experiencing visual system abnormalities.
| Substantial equivalence<br>Attribute | Submission device E3, Profile<br>Diagnosys LLC | MonPackONE<br>Metrovision |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|---------------------------|
| Intended use:<br>Generate photic signals and<br>measure and display the<br>electrical response signals<br>generated by the retina and<br>the visual nervous system | YES | YES |
| Intended users:<br>Ophthalmologists and<br>trained medical technicians<br>and professionals | YES | YES |
| Indications for use:<br>Quantification of the<br>electrophysiological<br>response of the retina and<br>visual cortex | YES | YES |
| Intended population:<br>Patients with ophthalmic<br>conditions | YES | YES |
| Intended use environment:<br>Hospitals, clinics and<br>physician offices | YES | YES |
| Physiological data<br>collected:<br>ERG, VEP, EOG waveforms | YES | YES |
| Compliance with<br>recognized standards:<br>ISO/EN 60601-1-2 | YES | YES |
#### Substantial equivalence
US: Diagnosys II C: 55 Technology Drive Suite 100 Lowell MA 01851: 978-458-1600; sales@diagnosys.llc.com
EU: Diagnosys Vision Ltd; Office 117, DOC Buildin, K67 E5A0, Ireland; +44 (0) 1223 52069; mail@diagnosysuis
www.diagnosysllc.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Diagnosys. The logo features the word "diagnosys" in a lowercase, sans-serif font, with the tagline "leading the wave" underneath in a smaller, italicized font. Above the word "diagnosys" is a curved, blue line that resembles a wave, reinforcing the tagline.
| Materials | Aluminum, molded parts, machined parts, PCBAs, PC, LCD monitor, LED and LCD light sources | Aluminum, molded parts, machined parts, PCBAs, PC, LCD monitor, LED and LCD light sources |
|-----------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Design | LCD display, Ganzfeld stimulator, amplifier, built in camera, desktop and electric stand option, PC | LCD display, Ganzfeld stimulator, amplifier, built in camera, desktop and electric stand option, PC |
| Energy Source | LCD monitor with LED backlight; LEDs | LCD monitor with LED backlight; LEDs |
| Other design features | Protocols to run all ISCEV clinical protocols; movable head and chin rests | Protocols to run all ISCEV clinical protocols; movable head and chin rests |
# Safety
The stimulator and the amplifier comply with ISO/EN 60601-1-2
The patient eye is exposed to visual light and near infra-red light with levels of exposure that have been measured and represents no risk to the patient.
## Performance Data
The E3 and Profile was tested and/or assessed to the following standards:
1. IEC 60601-1:2012. Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
2. ISO 14971: 2012 Medical devices - Application of risk management to medical devices.
3. IEC 60601-1-2:2014. Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
4. IEC 62366-1: 2015. Medical Devices - Part 1: Application Of Usability Engineering To Medical Devices.
5. IEC 60601-1-6:2010. AMD1:2013. Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability.
6. IEC 60601-2-26:2012. Medical Electrical Equipment - Part 2-26: Particular Requirements For The Basic Safety And Essential Performance Of Electroencephalographs.
7. IEC 62304:2015. Medical Device Software - Software Life Cycle Processes.
8. ANSI Z80.36-2021 Light hazard protection for ophthalmic instruments.
9. ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
10. ISO 10993-5. Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. 11. ISO 10993-10. Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
US: Diagnosys U.C.: 55 Technology Drive Suite 100 Lowell MA 01851 978-458-1600: sal on Ltd; Office 117, DOC Building, Balheary Road, Swords, Dublin, K6 UK: Diagnosys UK Ltd: 5 Trust Court. Chivers Way. Vision Park Histon. Cambridge, CB24 9PW. UK: +44 (0) 1223 520699; majl@djagnosysuk.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Diagnosys. The logo features the word "diagnosys" in a bold, sans-serif font, with the tagline "leading the wave" underneath in a smaller font. Above the word "diagnosys" is a blue, curved line that resembles a wave. The color scheme is primarily blue and white.
12. ISO 10993-23. Biological evaluation of medical devices - Part 23: Tests for irritation.
The non-clinical testing completed for the E3 and Profile included these three major categories of testing:
#### Technical tests:
Internal hardware, firmware and software component tests, packaging and labeling tests. Software verification and validation have been tested against their specifications and according to IEC 62304: 2015.
## Functional tests:
The different functions needed to measure electroretinograms (ERG), visual evoked potentials (VEP) and sensory electro-oculograms (EOG) were tested according to the recommendations of the International Society of Clinical Electrophysiology of Vision (ISCEV).
#### Compatibility tests:
Compatibility of the equipment and software with the major types of electrodes was tested for the different exams performed by the equipment.
## Conclusion
The E3 and Profile and the Vision MonPackONE system have the same intended use and indications, technological characteristics and principles of operation. Any technological differences between the E3 and Profile and its predicate does not present any new issues of safety or effectiveness. Testing has demonstrated that the E3 and Profile is substantially equivalent to the predicate device.
.diagnosysllc.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.