Essenz HLM is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
Device Story
Modular heart-lung machine; provides extracorporeal circulation support during surgery. Inputs: sensor data (flow, pressure, temperature, bubble detection, blood level). Operation: central console base with power supply, control units, and user interface (Cockpit); manages roller/centrifugal pumps and venous occluders. Output: real-time monitoring, automated pump control, and alarm generation. Used in clinical settings (e.g., OR) by perfusionists/physicians. System components include pumps, sensors, clamps, and data management cabinet. Benefits: maintains patient circulation and oxygenation during bypass; enhances safety via automated monitoring and emergency clamping (e.g., air bubble detection, arterial line clamping).
Clinical Evidence
No clinical or animal testing was required. Substantial equivalence is supported by non-clinical verification and validation, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability engineering (IEC 62366-1), alarm system testing (IEC 60601-1-8), software lifecycle processes (IEC 62304), and risk management (ISO 14971).
Technological Characteristics
Modular heart-lung machine console. Components: roller/centrifugal pumps, sensors (bubble, level, temperature, pressure, flow), venous occluders, and control units. Connectivity: local CAN bus, data interface to external management systems with encryption. Power: electrical. Software: embedded control and monitoring. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62366-1, IEC 60601-1-8, IEC 62304, ISO 14971.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures lasting six hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
March 9, 2023
LivaNova Deutschland GmbH Julia Leslie Senior Director, Regulatory Affairs Lindberghstr. 25 Munich, Bavaria 80939 Germany
Re: K221373
Trade/Device Name: Essenz HLM Regulation Number: 21 CFR 870.4220 Regulation Name: Cardiopulmonary Bypass Heart-Lung Machine Console Regulatory Class: Class II Product Code: DTQ, DWA, DWF Dated: January 18, 2023 Received: January 27, 2023
Dear Julia Leslie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Rachel E. Neubrander -S
for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221373
Device Name Essenz HLM
Indications for Use (Describe)
Essenz HLM is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-size: 10pt;"> <b> <span style="font-family: Symbol;">☑</span> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="font-size: 10pt;"> <b> <span style="font-family: Symbol;">☐</span> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221373
# 510(k) SUMMARY
# I. SUBMITTER
| Name: | LivaNova Deutschland GmbH |
|--------------------------------------|-------------------------------------------------|
| Address: | Lindberghstrasse 25<br>D-80939 München, Germany |
| Establishment<br>Registration Number | 9611109 (Manufacturer: LivaNova Deutschland) |
| Contact Person: | Julia E. Leslie, PhD |
| Phone: | +49 89 323 01 141 |
| Email: | julie.leslie@livanova.com |
| Secondary Contact: | Florian Goetz |
| Phone: | +49 89 323 01 236 |
| Email: | florian.goetz@livanova.com |
| Date Prepared: | March 3, 2023 |
# II. DEVICE
| Proprietary Name: | Essenz HLM |
|-----------------------------|-----------------------------------------------------|
| Common Name: | Heart-Lung Machine |
| Classification Name: | Console, Heart-Lung Machine, Cardiopulmonary Bypass |
| Classification Panel: | 74 Cardiovascular |
| Regulation Number: | 21CFR870.4220 |
| Product Code: | DTQ |
| Secondary Product<br>Codes: | DWA, DWF |
| Device Class: | Class II |
{4}------------------------------------------------
## III. PREDICATE DEVICE INFORMATION
Primary predicate:
- Stöckert S5 System, Heart-Lung Machine (K210130) ●
Secondary predicates:
- Sorin Centrifuqal Pump 5 (CP5) (K112225) ●
- . Electrical Venous Occluder (EVO) (K082344)
# IV. INDICATIONS FOR USE
Essenz HLM is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
# V. DEVICE DESCRIPTION
Essenz HLM is a modular heart-lung machine like its primary predicate Stöckert S5 System. The device consists of a central console base for support, positioning mobility and power supply, roller and centrifuqal pumps, user interface displays, controls, clamps and sensors for the monitoring of extracorporeal perfusion.
The Essenz HLM is configurable to user needs with different system components. The main configurable and optional system components consist of:
## Console
The Console provides a mobile chassis containing the power supply and several control units.
## Cockpit
The Essenz HLM Cockpit contains the display and control modules for monitoring, control and measuring devices and is the central user interface between the operator and the Essenz HLM.
## Control Units/ Console Control Units
Control and independent supervision of connected pump drive or EVO via local CAN, incl. dual (alternative) actuator management for leftmost CCU.
#### Pumps
Pumps may be roller or centrifugal pump type. Pumps provide speedcontrolled pumping of flow in the ECC (Extra Corporeal Circulation) using a peristaltic (positive displacement) pump or using a roto-dynamic pump (nonocclusive), automatic clamping and pump control according to measured flow rate and direction
## Bubble sensor
Monitoring device that detects air bubbles and microbubbles in the ECC - if detected, a visual and acoustic alarm is triggered, and the pump stops.
{5}------------------------------------------------
#### Level sensor
The level monitor controls the blood level in the oxygenator/reservoir. Display, alarm generation and pump speed requlation based on detection of blood level in a venous reservoir within the ECC.
#### Temperature sensor
The temperature monitor allows the simultaneous measurement and display of up to four temperatures, as measured by connected temperature probes. Display and alarm generation based on measurement of temperature of flow within the ECC.
#### Pressure sensor
The pressure sensor module is used to measure and display the pressure in the extracorporeal circuit.
Display, alarm generation and pump speed regulation based on measurement of pressure in the ECC.
#### Flow Sensor
The flow sensor monitors flow in tubing. Display, alarm generation and pump regulation based on measurement of flow in the ECC.
#### Manual Venous Occluder
Provides a separate control unit and line clamp.
#### Electrical venous occluder (EVO)
The clamp closes automatically when the stop link function to the arterial pump is activated.
#### Arterial Clamp/ Electric Remote Control
Clamp arterial line upon centrifugal pump stop / min rpm to fully stop the flow, prevent gravitational backflow in tubing and prevent air delivery to the patient
#### EP-Pack/ Power Pack
Provide power to system components
#### Cabinet (Enclosure)
Data interface between external devices and Data Management System, incl. data encryption
#### Mast
Provides structural stability and mounting points for system components and disposables
{6}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
LivaNova's Essenz HLM and its predicates have the same indications for use, intended use, use environment, intended user, target patient population, principles of operation and technological characteristics. The minor technological differences, including minor updates to pumps, and new software and graphical user interface, between the cleared and modified devices do not raise different questions of safety or effectiveness.
# VII. PERFORMANCE DATA
# NON-CLINICAL TESTING
In accordance with 21 CFR 820.30, LivaNova Deutschland GmbH. has conducted the following verification and validation testing of the Essenz HLM to ensure that it can provide all the capabilities necessary to operate safely and effectively:
- Electrical safety .
- Electromagnetic compatibility (EMC)
- 트 Performance testing
- " Software verification and validation
- Human Factors testing .
- . Mechanical testing
- Performance testing of shipping containers I
In support of the determination of substantial equivalence of the Essenz HLM to its predicate devices, the following recognized Standards have been used:
| Standard | Title | FDA<br>recognition<br>number |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| IEC 60601-1 2005<br>A1:2012 | Medical electrical equipment - Part 1:<br>General requirements for basic safety and<br>essential performance | 19-4 |
| IEC 60601-1-2 Edition 4.0<br>2014 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Electromagnetic disturbances<br>Requirements and tests | 19-8 |
| IEC 62366-1 Edition 1.0<br>2015 | Medical devices — Part 1: Application of<br>usability engineering to medical devices | 5-114 |
| 60601-1-8:2006 and<br>A1:2012 | Medical electrical equipment - Part 1-8:<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>General requirements, tests and guidance | 5-76 |
{7}------------------------------------------------
| | for alarm systems in medical electrical<br>equipment and medical electrical systems | |
|---------------------------------|-------------------------------------------------------------------------------------|-------|
| IEC 62304 Edition 1.1<br>2015 | Medical device software - Software life<br>cycle processes | 13-79 |
| ISO 14971 Third Edition<br>2019 | Application of risk management to medical<br>devices | 5-125 |
# CLINICAL TESTING
None required.
# ANIMAL TESTING
None required.
# VIII. SUBSTANTIAL EQUIVALENCE
The Essenz HLM is as safe and effective as the predicate S5 System. The Essenz HLM has the same intended use and indications, similar technological characteristics, and the same principles of operation as its predicate devices. The minor technological differences between the cleared and modified devices do not raise different questions of safety or effectiveness. Performance and validation data demonstrate that the subject Essenz HLM is as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.