← Product Code [OBR](/productcode/OBR) · K221369

# FBCC Night Guard (K221369)

_Fbcc, Inc. · OBR · Nov 18, 2022 · DE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221369

## Device Facts

- **Applicant:** Fbcc, Inc.
- **Product Code:** [OBR](/productcode/OBR.md)
- **Decision Date:** Nov 18, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE
- **Attributes:** Therapeutic

## Indications for Use

The FBCC Night Guard is indicated for protection of teeth and restorations against grinding and clenching.

## Device Story

FBCC Night Guard is a patient-removable mouth guard designed to prevent tooth damage caused by bruxism (grinding and clenching). The device acts as a physical barrier between upper and lower tooth surfaces, providing disocclusion. It is intended for over-the-counter use. The device is manufactured via thermoforming using thermoplastic polyurethane or ethylene-vinylacetate. It is worn by the patient during sleep to protect teeth and restorations from wear and tear.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by indications for use, technological characteristics, and non-clinical performance testing, including durability testing and manufacturing validation of dimensional accuracy.

## Technological Characteristics

Materials: Thermoplastic polyurethane or ethylene-vinylacetate. Manufacturing: Thermoforming. Form factor: Patient-removable mouth guard for maxillary or mandibular teeth. Mechanism: Disocclusion. Sterility: Non-sterile.

## Regulatory Identification

To protect the teeth and reduce damage caused by bruxing or nighttime grinding and prevention of the noise associated with bruxing and grinding.

## Predicate Devices

- Bright Guard ([K181099](/device/K181099.md))

## Reference Devices

- JS Dental Lab Mouth Guard ([K210011](/device/K210011.md))
- Thermoformed Mouthguards/Nightguards ([K121365](/device/K121365.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

November 18, 2022

FBCC Inc % Elisabeth Miller Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043

Re: K221369

Trade/Device Name: FBCC Night Guard Regulatory Class: Unclassified Product Code: OBR Dated: September 19, 2022 Received: September 19, 2022

Dear Elisabeth Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bobak Shirmohammadi -S

For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K221369

Device Name FBCC Night Guard

Indications for Use (Describe)

The FBCC Night Guard is indicated for protection of teeth and restorations against grinding and clenching.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### Section 5. 510(k) Summary K221369

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| Submitter:                | FBCC Inc                                                 |
|---------------------------|----------------------------------------------------------|
| Company Contact Person:   | Pablo Osorio                                             |
| Address:                  | 6711 S Sepulveda Blvd, Unit 602<br>Los Angeles, CA 90045 |
| Phone:                    | (323) 596-6459                                           |
| Email:                    | pablosorio6@gmail.com                                    |
| Submission Correspondent: | Elisabeth Miller, Regulatory Affairs Consultant          |
| Address:                  | 1321 Upland Dr. Suite 6792<br>Houston, TX 77043          |
| Phone:                    | (586) 242-7718                                           |
| Email:                    | emiller@primepathmedtech.com                             |
| Date Prepared:            | November 2022                                            |
| Proprietary Name:         | FBCC Night Guard                                         |
| Common Name:              | Mouth Guard                                              |
| Product Code:             | OBR                                                      |
| Device Classification:    | Unclassified                                             |
| Predicate Device:         | Bright Guard                                             |
| Reference Device:         | JS Dental Lab Mouth Guard                                |

{4}------------------------------------------------

#### Device Description:

The FBCC Night Guard is a mouth guard used as a barrier between teeth for nighttime teeth grinding by creating physical separation between upper and lower tooth surfaces preventing tooth damage caused by bruxism (e.g., grinding and clenching).

#### Indications for Use:

The FBCC Night Guard is indicated for protections against grinding and clenching and clenching.

{5}------------------------------------------------

## Comparison to Predicate Devices:

| Specification              | Subject Device:<br>FBCC Night<br>Guard                                                                                                                                      | Predicate<br>Device:<br>Bright Guard<br>(K181099)                                                                          | Reference Device:<br>JS Dental Lab Mouth<br>Guard<br>(K210011)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison<br>Result |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Product Code               | OBR                                                                                                                                                                         | OBR                                                                                                                        | MQC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same as<br>predicate |
| Classification<br>Panel    | Dental                                                                                                                                                                      | Dental                                                                                                                     | Dental                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                 |
| Device Class               | Unclassified                                                                                                                                                                | Unclassified                                                                                                               | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                 |
| OTC or Rx                  | ОТС                                                                                                                                                                         | OTC                                                                                                                        | RX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same as<br>predicate |
| Material                   | Thermoplastic<br>Polyurethane or<br>Ethylene-<br>Vinylacetate                                                                                                               | Thermoplastic<br>resin Propylene<br>Elastomer:<br>ethylene-<br>vinylacetate<br>[Elvax] No Flavor;<br>No Color<br>Additives | Thermoplastic<br>Polyurethane, Ethyl Vinyl<br>acetate (Erkodur or Splint<br>Biocryl, Erkoloc-Pro,<br>Erkoflex-95, EVA Based<br>Clear Mouthguard<br>Material and BIOPLAST)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar              |
| Method of<br>Manufacturing | Thermoforming                                                                                                                                                               | Injection<br>Molding                                                                                                       | Thermoforming                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as<br>reference |
| Anatomical Sites           | Worn on<br>maxillary or<br>mandibular teeth                                                                                                                                 | Worn on<br>maxillary teeth                                                                                                 | Worn on maxillary or<br>mandibular teeth                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same as<br>reference |
| Sterile                    | Non-Sterile                                                                                                                                                                 | Non-Sterile                                                                                                                | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                 |
| Device<br>Description      | The FBCC Night<br>Guard is a mouth<br>guard used as a<br>barrier between<br>teeth for<br>nighttime teeth<br>grinding by<br>creating physical<br>separation<br>between upper | Flexible,<br>moldable mouth<br>guard used as a<br>barrier between<br>teeth for<br>nighttime teeth<br>grinding.             | The JS Dental Lab Mouth<br>Guard (herein referred to<br>as Subject Device) is a<br>patient contact protective<br>custom-fit mouth guard<br>that covers the upper<br>teeth and lower teeth to<br>prevent premature tooth<br>wear and noise caused<br>due to bruxism and teeth                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar              |
|                            |                                                                                                                                                                             |                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                      |
|                            | and lower tooth<br>surfaces<br>preventing tooth<br>damage caused<br>by bruxism (e.g.,<br>grinding and<br>clenching).                                                        |                                                                                                                            | grinding. It fits over upper<br>teeth or lower teeth<br>during sleep. The Subject<br>Device 510(k) Summary<br>K210011Ampower Dental<br>Laboratories LLC<br>Confidential Page 3 of 6 Al<br>response for K210011 can<br>offset the effects of<br>bruxing or teeth grinding<br>while protecting teeth<br>from daily wear and tear.<br>The Subject Device is<br>created based on the<br>user's teeth impression<br>and manufactured using<br>the biocompatible<br>material, equivalent to the<br>Thermoformed<br>Mouthguards/Nightguards<br>(herein referred to as<br>Predicate Device,<br>K121365). The Subject<br>Device contains<br>biocompatible materials,<br>namely, Erkodur or Splint<br>Biocryl, Erkoloc-Pro,<br>Erkoflex-95, EVA Based<br>Clear Mouthguard<br>Material and BIOPLAST<br>and is available in four<br>different variations. All<br>the biocompatible<br>materials used are ISO-<br>certified, BPA-free, and<br>are cleared for dental use<br>in humans. |                      |
| Patient<br>Removable?      | Yes                                                                                                                                                                         | Yes                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                 |
| Indication for<br>Use      | The FBCC Night<br>Guard is<br>indicated for<br>protection of<br>teeth and<br>restorations<br>against                                                                        | Bright Guard is a<br>mouth guard<br>intended to<br>protect against<br>grinding and<br>clenching.                           | The Subject Device (The JS<br>Dental Mouth Guard) is<br>intended for protection<br>against bruxism and teeth<br>grinding. They create a<br>barrier between the upper                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar              |
|                            | grinding and<br>clenching.                                                                                                                                                  |                                                                                                                            | protect the patient's<br>overall occlusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                      |
| Mechanism of<br>Action     | Disocclusion                                                                                                                                                                | Disocclusion                                                                                                               | Disocclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                 |

Table 1. Predicate and Reference Device Comparison

{6}------------------------------------------------

{7}------------------------------------------------

## Comparison of Indications for Use to Predicate Device:

The indications for use are similar for the subject and reference devices. The subject and predicate devices are indicated for the protection of teeth against grinding and clenching.

## Comparison of Technological Features to Predicate Devices:

The FBCC Night Guard uses the same mechanism of action as the predicate product. It is intended to be worn on the teeth to protect teeth against grinding and clenching.

As summarized above, the main differences between the subject device (FBCC Night Guard) and predicate device (Bright Guard):

- Material
- Method of Manufacturing ●

Overall, the differences are minor and do not impact risk to the patient or user of the product.

#### Material

The subject device is manufactured using thermoplastic polyurethane, whereas the predicate device is manufactured using a thermoplastic resin. While these are not identical polymers, each material is a known dental thermoplastic polymer with similar characteristics. The use of thermoplastic polyurethane does not introduce or reduce any risks when compared to the other.

## Method of Manufacturing

The subject device is thermoformed while the predicate device is injection molded. While these are not identical processes, both are commonly used to create mouth guards and other dental devices. The subject and reference device (JS Dental Lab Mouth Guard) both use thermoforming as the method of manufacturing. The use of thermoforming does not increase the risk associated with manufacturing.

{8}------------------------------------------------

#### Non-Clinical Performance Testing

The use of thermoplastic materials Night Guards indicated for protection of teeth against grinding and bruxing has been well documented in scientific literature. However, durability testing was completed on these night guards. Real world use was simulated to ensure that the Night Guard material and manufacturing process produced Night Guards that were suitable for the intended period of use.

An internal manufacturing validation was performed to test the dimensional accuracy of the manufacturing process for the FBCC Night Guards. The robustness of the process was demonstrated from 3D printing through thermoforming.

## Clinical Performance Testing

The technological characteristics, indications for use, material, and manufacturing and processes are the same or similar to the predicate device and therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.

## Conclusion

Based on the similarities in indications for use, technological characteristics, and nonclinical performance testing the FBCC Night Guards are substantially equivalent to the predicate device Bright Guard (K181099).

---

**Source:** [https://fda-staging.innolitics.com/device/K221369](https://fda-staging.innolitics.com/device/K221369)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com