K221354 · Arthrex, Inc. · GAT · Sep 9, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221354
Device Name
Arthrex SutureTape
Applicant
Arthrex, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Sep 9, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex SutureTape is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft tissues are used for repair.
Device Story
Arthrex SutureTape is a non-absorbable surgical suture designed for soft tissue approximation and ligation. The device features a flat braided tape portion transitioning to a round-braided suture tail. It is used by surgeons in clinical settings to secure soft tissue or as a component in surgical constructs involving allograft or autograft tissues. The tape portion provides high tensile strength, while the tail facilitates handling. The device is supplied sterile in various configurations, including pre-cut lengths and swaged needles. By providing a secure, high-strength fixation method, the suture assists in tissue repair and healing. It is a line extension of existing SutureTape products, maintaining established surgical techniques and material compositions.
Clinical Evidence
Bench testing only. Performance data includes knot pull tensile strength testing for the tape portion (meeting USP size 2 standards) and the suture tail (meeting USP 2-0 standards). Needle pull tensile strength meets USP size 2-0 standards. Bacterial Endotoxins Test (BET) performed per ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, and EP 2.6.14 confirms compliance with pyrogen limit specifications.
Technological Characteristics
Non-absorbable surgical suture composed of UHMWPE, polyester, and nylon yarns. Includes cyanoacrylate and dyes (D&C Blue No. 6, Logwood Black). Features a flat braided tape section and round-braided tail. Supplied sterile. Testing conducted per USP standards for tensile strength and ANSI/AAMI ST72:2011/(R)2016 for endotoxins.
Indications for Use
Indicated for soft tissue approximation and ligation in surgical procedures, including those utilizing allograft or autograft tissues.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
September 9, 2022
Arthrex Inc. Lai Saeteurn Regulatory Affairs Specialist II 1370 Creekside Boulevard Naples, Florida 34108-1945
Re: K221354
Trade/Device Name: Arthrex SutureTape Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: June 17, 2022 Received: June 21, 2022
Dear Lai Saeteurn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Deborah Fellhauer Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221354
Device Name Arthrex SutureTape
Indications for Use (Describe)
The Arthrex SutureTape is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft tissues are used for repair.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="12" width="12"> <path d="M1 1 L11 1 L11 11 L1 11 Z" fill="none" stroke="black" stroke-width="1"></path> <path d="M2 2 L10 10 M2 10 L10 2" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;"> <svg height="12" width="12"> <path d="M1 1 L11 1 L11 11 L1 11 Z" fill="none" stroke="black" stroke-width="1"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| X | Prescription Use (Part 21 CFR 801 Subpart D)
_] Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Date Prepared | September 9, 2022 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 |
| Contact Person | Lai Saeteurn<br>Regulatory Affairs Specialist II<br>1-239-598-4302 Ext. 71764<br>Lai.Saeteurn@Arthrex.com |
| Name of Device | Arthrex SutureTape |
| Common Name | Suture |
| Product Code | GAT |
| Classification Name | 21 CFR 878.5000: Nonabsorbable poly(ethylene terephthalate) surgical suture |
| Regulatory Class | II |
| Predicate Device | K171296: Arthrex SutureTape |
| Reference Devices | K193575: Arthrex SutureTape<br>K122374: Arthrex Suture |
| Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain clearance for the<br>Arthrex 0.9 mm SutureTape sutures as a line extension to the Arthrex<br>SutureTape. |
| Device Description | The Arthrex SutureTape is comprised of non-absorbable sutures made of a<br>polyblend of Ultra High Molecular Weight Polyethylene (UHMWPE) and polyester<br>yarns but may also include nylon yarns.<br><br>The Arthrex 0.9 mm SutureTape is a 0.9 mm-wide tape suture with a round<br>braided suture tail that is smaller than the tape portion. The tape and tail<br>portion of the suture are implantable while the suture tail is not intended to be<br>knotted. The flat tape portion of the Arthrex 0.9 mm SutureTape suture meets<br>or exceeds USP performance standards for knot pull tensile strength for size 2<br>non-absorbable surgical sutures. The suture tail portions of the Arthrex 0.9 mm<br>SutureTape suture meet the performance standards for USP 2-0 non-absorbable<br>surgical suture, except for an oversize in diameter.<br><br>The Arthrex 0.9 mm SutureTape is supplied sterile, in pre-cut lengths, in various<br>loop configurations, and in some cases, with various swaged needles and with<br>stiffened ends. The Arthrex 0.9 mm SutureTape is available non-dyed, dyed, and<br>fully or partially striped. Dyes may include D&C Blue No. 6 and Logwood Black.<br>Suture strands that are dyed with Logwood black are made of nylon. Additional<br>material for the Arthrex 0.9 mm SutureTape includes cyanoacrylate. |
| Indications for Use | The Arthrex SutureTape is intended for use in soft tissue approximation and/or<br>ligation. The suture may be provided individually or be incorporated as a<br>component, into surgeries where constructs including those with allograft or<br>autograft tissues are used for repair. |
| Performance Data | The flat tape portion of the Arthrex 0.9 mm SutureTape suture meets or<br>exceeds USP performance standards for knot pull tensile strength for size 2 non-<br>absorbable surgical sutures. The suture tail portions of the Arthrex 0.9 mm<br>SutureTape suture meet the performance standards for USP 2-0 non-absorbable<br>surgical suture, except for an oversize in diameter. The attached needle at the<br>suture tail portion of the Arthrex 0.9 mm SutureTape suture meets or exceeds<br>USP performance standards for needle pull tensile strength for size 2-0 non-<br>absorbable surgical sutures. |
| | Bacterial Endotoxins Test (BET) was performed on the representative samples<br>utilizing the Kinetic Chromogenic Method in accordance with ANSI/AAMI<br>ST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14 to demonstrate that the<br>proposed device meets pyrogen limit specifications. |
| Technological<br>Comparison | The Arthrex 0.9 mm SutureTape suture is a line extension to the predicate device.<br>In comparison to the predicate device, the Arthrex 0.9 mm SutureTape suture<br>share the same basic design features (e.g., flat braided with round-braided suture<br>tails, non-USP size), fundamental scientific technology, intended use, materials<br>(e.g., yarns, dyes, additive), shelf life, surgical technique, manufacturing,<br>packaging, and sterilization processes as the predicate device. However, the<br>Arthrex 0.9 mm SutureTape suture has a tape width and suture tail diameter<br>range smaller than that of the predicate. |
| Conclusion | The Arthrex 0.9 mm SutureTape devices are substantially equivalent to the<br>predicate device in which the basic design features, materials, manufacturing,<br>and intended use are the same. Any differences between the proposed and<br>predicate devices are considered minor and do not raise questions concerning<br>safety or effectiveness.<br>Based on the indication or use, technological characteristics, and the tensile test<br>data submitted, Arthrex Inc. has determined that the proposed device is<br>substantially equivalent to the currently marketed predicate device. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.