← Product Code [FXX](/productcode/FXX) · K221352

# Surgical Face Mask (K221352)

_Wuhan Huashida Protective Products Co., Ltd. · FXX · Mar 13, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221352

## Device Facts

- **Applicant:** Wuhan Huashida Protective Products Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Mar 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

## Device Story

Surgical face mask; 3-layer flat-pleated design; constructed of nonwoven polypropylene. Available in two models: HSD-TYP2R (ear loops) and HSD-TYP2RT (tie-on). Includes malleable aluminum nose clip for fit. Single-use, disposable, non-sterile. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against blood, body fluids, and particulates. Performance meets ASTM F2100-19 Level 3 standards. Provides protection by filtering microorganisms and resisting fluid penetration.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

3-layer nonwoven polypropylene construction; melt-blown polypropylene filter layer; malleable aluminum nose clip; spandex/polyester ear loops or spun-bond polypropylene tie tapes. Flat-pleated form factor. ASTM F2100-19 Level 3 compliant. Non-sterile, single-use. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K202903](/device/K202903.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

March 13, 2023

Wuhan Huashida Protective Products Co., Ltd. Mei Chen Manager No. 511 Weihu Rd, Shamao St, Hannan District Wuhan, Hubei 430090 China

Re: K221352

Trade/Device Name: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 27, 2023 Received: February 27, 2023

Dear Mei Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katherine D. Kavlock -S

for

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K221352

Device Name Surgical Face Mask

#### Indications for Use (Describe)

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

# 510(k) summary

# I Submitter

| Device submitter: | Wuhan Huashida Protective Products Co., Ltd.                     |
|-------------------|------------------------------------------------------------------|
|                   | No.511 Weihu Rd, Shamao St, Hannan district, Wuhan, Hubei, China |
| Contact person:   | Chen Mei                                                         |
|                   | Manager                                                          |
|                   | Phone: +86-18971272798                                           |
|                   | E-mail: Happy@huashida168.cn                                     |

#### II Proposed Device

| Trade/Device Name: | Surgical Face Mask |
|--------------------|--------------------|
| Regulation Number: | 21 CFR 878.4040    |
| Regulation Name:   | Surgical Apparel   |
| Regulatory Class:  | Class II           |
| Product code:      | FXX                |
| Review Panel:      | General Hospital   |

# III Predicate Devices

| 510(k) Number:     | K202903                                         |
|--------------------|-------------------------------------------------|
| Trade/Device Name: | Surgical Face Mask                              |
| Regulation Number: | 21 CFR 878.4040                                 |
| Regulation Name:   | Surgical Apparel                                |
| Classification:    | Class II                                        |
| Product Code:      | FXX                                             |
| Manufacturer       | Rizhao Sanqi Medical & Health Articles Co., Ltd |

#### IV Indication for use

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

#### V Device Description

The Surgical Face Mask is 3-layer, and it is two model is HSD-TYP2R (Ear loops) and HSD-TYP2RT (Tie-On). Flat-folded masks constructed of nonwoven polypropylene materials.

{4}------------------------------------------------

The mask is provided with ear loops (spandex and polyester) or tie tapes (spun-bond polypropylene). A malleable nose clip is placed in the layers of facemask for comfort and individualized fit. The surgical face mask color is blue. Product is Level 3 based on ASTM F2100-19. The surgical face mask is single-use, disposable devices, provided non-sterile.

| Item                    | Subject device                                                                                                                                                                                                                                                                                                                                                            | Predicate device (K202903)                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion   |                                         |           |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-----------------------------------------|-----------|
| Product name            | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                        | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                             | NA           |                                         |           |
| Product model           | HSD-TYP2R, HSD-TYP2RT                                                                                                                                                                                                                                                                                                                                                     | Level 3 Surgical Face Mask models: SQ-3001, SQ-3001H                                                                                                                                                                                                                                                                                                                                                                                                           | NA           |                                         |           |
| Classification          | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                     | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical    |                                         |           |
| Intended use            | The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile. | The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. When worn properly, these face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile. This is a single use, disposable device(s), provided non-sterile | Identical    |                                         |           |
| Component and Materials | Structure                                                                                                                                                                                                                                                                                                                                                                 | Materials                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Structure    | Materials                               | Different |
|                         | Outer layer                                                                                                                                                                                                                                                                                                                                                               | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                        | Outer layer  | Polypropylene                           |           |
|                         | Inner layer                                                                                                                                                                                                                                                                                                                                                               | polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Inner layer  | Polypropylene                           |           |
|                         | Filter layer                                                                                                                                                                                                                                                                                                                                                              | melt-blown polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                       | Filter layer | 3 Flat leated: Melt-blown Polypropylene |           |
|                         | Nose clip                                                                                                                                                                                                                                                                                                                                                                 | malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                                                                        | Nose clip    | Steel coated by polypropylene           |           |
|                         | Ear loops                                                                                                                                                                                                                                                                                                                                                                 | spandex and polyester                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ear loops    | Spandex                                 |           |
|                         | Tie Tapes                                                                                                                                                                                                                                                                                                                                                                 | spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                        | Tie Tapes    | Spunbond Polypropylene                  |           |

# VI Summary of Technological characteristics

{5}------------------------------------------------

| Item                                    | Subject device                                    | Predicate device<br>(K202903)                     | Discussion |
|-----------------------------------------|---------------------------------------------------|---------------------------------------------------|------------|
| Mask style                              | Flat-pleated                                      | Flat-pleated                                      | Identical  |
| Design<br>Features                      | Ear Loops, Tie-on                                 | Ear Loops, Tie-on                                 | Identical  |
| Mask style                              | 3 Flat Pleated                                    | 3 Flat Pleated<br>4 Flat Pleated                  | Similar    |
| Color                                   | Blue                                              | Blue, white, pink, green, yellow                  | Similar    |
| Dimension<br>(Length)                   | 175±10mm                                          | 175±5mm                                           | Similar    |
| Dimension<br>(Width)                    | 90±10mm                                           | 95±5mm                                            | Similar    |
| ASTM F2100<br>Level                     | Level 3                                           | Level 3                                           | Identical  |
| OTC use                                 | Yes                                               | Yes                                               | Identical  |
| Sterility                               | Non-Sterile                                       | Non-Sterile                                       | Identical  |
| Use                                     | Single Use, Disposable                            | Single Use, Disposable                            | Identical  |
| Particulate<br>filtration<br>efficiency | Meet ASTM F2299-17                                | Meet ASTM F2299-17                                | Identical  |
| Bacterial<br>filtration<br>efficiency   | Meet ASTM F2101-19                                | Meet ASTM F2101-19                                | Identical  |
| Differential<br>pressure                | Meet EN 14683: 2019, Annex C                      | Meet EN 14683: 2019, Annex C                      | Identical  |
| Flammability                            | Meet 16 CFR 1610                                  | Meet 16 CFR 1610                                  | Identical  |
| Fluid<br>resistance                     | Meet ASTM F1862-17                                | Meet ASTM F1862-17                                | Identical  |
| Biocompatibil<br>ity                    | Non-cytotoxic, Non-sensitizing,<br>non-irritating | Non-cytotoxic, Non-sensitizing,<br>non-irritating | Identical  |

#### Section 3-510(k) Summary

#### Different - Patient Contacting Material

The patient contacting material for the proposed device is different from the predicate devices. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the two predicate devices.

# VII Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the 

{6}------------------------------------------------

Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

- A 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;
- A ASTM F1862 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- > ASTM F2299 Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres
- A ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- > ASTM F2100 Standard Specification for Performance of Materials Used in Medical Face Masks
- > EN 14683 Medical face masks- Requirements and test methods
- A ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- A ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization

| Items                                | Purpose                                                                                                      | Acceptance Criteria              | Result |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------|--------|
| Fluid Resistance<br>Performance      | The purpose of the<br>performance testing<br>is to demonstrate<br>The functionality of<br>the subject device | 29 out of 32<br>pass at 160 mmHg | Pass   |
| Particulate Filtration<br>Efficiency |                                                                                                              | $\geq$ 98%                       | Pass   |
| Bacterial Filtration<br>Efficiency   | $\geq$ 98%                                                                                                   | Pass                             |        |
| Differential Pressure                |                                                                                                              | <6.0 mmH2O/cm²                   | Pass   |
| Flammability                         |                                                                                                              | Class 1                          | Pass   |

For Surgical Face Mask Performance

#### For Surgical Face Mask Biocompatibility

| Items         | Purpose                                                                               | Acceptance<br>Criteria | Result                                                               |
|---------------|---------------------------------------------------------------------------------------|------------------------|----------------------------------------------------------------------|
| Cytotoxicity  | The purpose of<br>the testing is to<br>demonstrate<br>safety of the<br>subject device | Non-Cytotoxic          | Under the conditions of the study, the<br>device is non-cytotoxic.   |
| Irritation    |                                                                                       | Non-Irritating         | Under the conditions of the study, the<br>device is non-irritating.  |
| Sensitization |                                                                                       | Non-Sensitizing        | Under the conditions of the study, the<br>device is non-sensitizing. |

{7}------------------------------------------------

## VIII Clinical Test Conclusion

No clinical study is included in this submission.

# IX Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device K202903.

---

**Source:** [https://fda-staging.innolitics.com/device/K221352](https://fda-staging.innolitics.com/device/K221352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).