The retrospective clinical dataset was used to validate the performance (sensitivity, specificity, and time-to-notification) of the BriefCase AI algorithm in a real-world clinical setting.
Retrospective, blinded, multicenter study of 921 cases from 5 US-based clinical sites.
>1 (senior board-certified radiologists)
Indications for Use
BriefCase is a radiological computer-aided triage and notification software indicated for use in the analysis of frontal chest X-ray (CXR) images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina. Findings are flagged when the ETT distal tip is assessed as being more than 5 cm above the carina, less than 2 cm above the carina, or when it is below the carina (i.e., in the right or left mainstem bronchus). The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device does not provide results when the carina is not well-visualized on the x-ray image. The device does not discriminate types of ETTs, as such a properly positioned double lumen ETT may trigger a false prioritization alert. BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on the user's professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Device Story
BriefCase is a radiological computer-aided triage and notification software; operates in parallel to standard-of-care (SOC) image interpretation. Input: DICOM frontal chest X-ray images. Processing: AI/deep-learning algorithm analyzes images to detect vertical position of ETT distal tip relative to carina; flags cases where tip is >5cm above, <2cm above, or below carina. Output: Notifications sent to Aidoc Desktop Application; includes compressed, unannotated, low-quality preview images for informational purposes only. Used in hospital networks by radiologists/clinicians. Workflow: Does not alter original images; does not remove/deprioritize cases from SOC queue. Clinicians use notifications to prioritize review of full images in PACS. Benefit: Reduces turnaround time for identifying malpositioned ETTs, facilitating faster clinical intervention.
Clinical Evidence
Retrospective, blinded, multicenter study (5 US sites, N=921). Performance compared to ground truth (majority vote of 3 radiologists). Sensitivity: 86.81% (95% CI: 81.02%-91.36%); Specificity: 92.02% (95% CI: 89.20%-93.87%). Secondary endpoint: Time-to-notification (4.93 min) vs. SOC time-to-exam-open (157.0 min), demonstrating significant triage efficiency.
Technological Characteristics
Software-based triage system. Components: Aidoc Hospital Server (AHS), Aidoc Cloud Server (ACS), Aidoc Desktop Application. Connectivity: DICOM 3.0 compliant. Operating System: Microsoft Windows Server 2012 64bit. Database: PostgreSQL. Algorithm: Deep-learning AI. Deployment: Standalone application integrated with PACS/radiology workstations.
Indications for Use
Indicated for analysis of frontal chest X-ray (CXR) images in adults and transitional adolescents (≥18 years) to assist hospital networks and medical specialists in triage by flagging suspected vertically malpositioned endotracheal tubes (ETT) relative to the carina. Contraindications: carina not well-visualized; cannot detect esophageal intubation; does not discriminate ETT types.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Aidoc Medical, Ltd. % John J. Smith Partner Hogan Lovells US LLP 555 Thirteenth Street, N.W. WASHINGTON DC 20004
# Re: K221330
November 18, 2022
Trade/Device Name: BriefCase Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: October 11, 2022 Received: October 11, 2022
Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K221330
Device Name
#### BriefCase
Indications for Use (Describe)
BriefCase is a radiological computer-aided triage and notification software indicated for use in the analysis of frontal chest X-ray (CXR) images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina.
Findings are flagged when the ETT distal tip is assessed as being more than 5 cm above the carina, less than 2 cm above the carina, or when it is below the carina (i.e., in the right or left mainstem bronchus).
The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device does not provide results when the carina is not wellvisualized on the x-ray image. The device does not discriminate types of ETTs, as such a properly positioned double lumen ETT may trigger a false prioritization alert.
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on the user's professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human
Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Aidoc. The logo is in blue and consists of the word "aidoc" in lowercase letters. The "i" in "aidoc" is dotted with an orange circle. The logo is simple and modern.
#### 510(k) Summary Aidoc Medical, Ltd.'s BriefCase K221330
| Submitter: | Aidoc Medical, Ltd.<br>3 Aminadav St.<br>Tel-Aviv, Israel<br>Phone: +972-73-7946870 |
|---------------------------|-------------------------------------------------------------------------------------|
| Contact Person: | Amalia Schreier, LLM |
| Date Prepared: | November 17, 2022 |
| Name of Device: | BriefCase |
| Classification Name: | Radiological computer-assisted triage and notification software<br>device |
| Regulatory Class: | Class II |
| Product Code: | QAS (21 C.F.R. 892.2080) |
| Primary Predicate Device: | BriefCase (for iPE triage, K203508) |
#### Device Description
BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition: (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.
DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the desktop application, thereby facilitating triage and prioritization by the user. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.g., a chest radiologist may choose to filter out alerts on VO cases, and a neuro-radiologist would opt to divert vertically malpositioned ETT in relation to the carina alerts. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out alerts by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc desktop application.
The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.
{4}------------------------------------------------
Presenting the users with worklist prioritization facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
#### Intended Use / Indications for Use
BriefCase is a radiological computer-aided triage and notification software indicated for use in the analysis of frontal chest X-ray (CXR) images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina.
Findings are flagged when the ETT distal tip is assessed as being more than 5 cm above the carina, less than 2 cm above the carina, or when it is below the carina (i.e in the right or left mainstem bronchus).
The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device does not provide results when the carina is not well-visualized on the x-ray image. The device does not discriminate types of ETTs, as such a properly positioned double lumen ETT may trigger a false prioritization alert.
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on the user's professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
# Comparison of Technological Characteristics
The subject BriefCase for vertically malpositioned ETT in relation to the carina triage and primary predicate BriefCase for iPE triage (K203508) are identical in all aspects and defer only with respect to the training of the algorithm on vertically malpositioned ETT in relation to the images.
Both devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporating software packages for use with DICOM 3.0 compliant x-ray scanners, PACS, and radiology workstations. The predicate BriefCase for iPE evaluates images from CT scanners and the proposed Briefcase for vertically malpositioned ETT in relation to the carina evaluates images from X-ray scanners.
The proposed device for vertically malpositioned ETT in relation to the carina triage has identical technology and design as the predicate device, and similar indications for use, i.e., both devices are intended to aid in prioritization and triage of time-critical radiological images. The subject and predicate BriefCase devices raise the same types of safety and effectiveness questions, namely,
{5}------------------------------------------------
accurate detection of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor deprioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases. A table comparing the key features of the subject and the predicate devices is provided below.
| | Predicate Device<br>Aidoc Briefcase for iPE Triage<br>(K203508) | Subject Device<br>Aidoc Briefcase for Vertically<br>Malpositioned ETT in Relation to the<br>Carina Triage (K221330) |
|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of contrast enhanced chest CTs (but not dedicated CTPA protocol). The device is intended to assist hospital networks and trained medical specialists in workflow triage by flagging and communication of suspect positive cases of incidental Pulmonary Embolism (iPE) pathologies.<br><br>BriefCase uses an artificial intelligence algorithm to analyze images and flag suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.<br><br>The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for | BriefCase is a radiological computer- aided triage and notification software indicated for use in the analysis of frontal chest X-ray (CXR) images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina.<br><br>Findings are flagged when the ETT distal tip is assessed as being more than 5 cm above the carina, less than 2 cm above the carina, or when it is below the carina (i.e in the right or left mainstem bronchus).<br><br>The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device does not provide results when the carina is not well-visualized on the x-ray image. The device does not discriminate types of ETTs, as such a properly positioned double lumen ETT may trigger a false prioritization alert.<br><br>BriefCase uses an artificial intelligence algorithm to analyze images and |
| | Predicate Device<br>Aidoc Briefcase for iPE Triage<br>(K203508) | Subject Device<br>Aidoc Briefcase for Vertically<br>Malpositioned ETT in Relation to the<br>Carina Triage (K221330) |
| | viewing full images per the standard of<br>care. | highlight cases with detected findings<br>on a standalone application in parallel<br>to the ongoing standard of care image<br>interpretation. The user is presented<br>with notifications for cases with<br>suspected findings. Notifications<br>include compressed preview images<br>that are meant for informational<br>purposes only and not intended for<br>diagnostic use beyond notification. The<br>device does not alter the original<br>medical image and is not intended to be<br>used as a diagnostic device.<br>The results of BriefCase are intended to<br>be used in conjunction with other<br>patient information and based on the<br>user's professional judgment, to assist<br>with triage/prioritization of medical<br>images. Notified clinicians are<br>responsible for viewing full images per<br>the standard of care. |
| User population | Hospital networks and appropriately<br>trained medical specialists | Hospital networks and appropriately<br>trained medical specialists |
| Anatomical<br>region<br>of<br>interest | Chest | Chest |
| Data<br>acquisition<br>protocol | Contrast-enhanced chest CTs (but not<br>dedicated CTPA protocol) | Frontal Chest X-ray (CXR) |
| Notification-<br>only<br>(/notification<br>alerts), parallel<br>workflow tool | Yes | Yes |
| Images<br>format | DICOM | DICOM |
| | Predicate Device<br>Aidoc Briefcase for iPE Triage<br>(K203508) | Subject Device<br>Aidoc Briefcase for Vertically<br>Malpositioned ETT in Relation to the<br>Carina Triage (K221330) |
| Interference<br>with standard<br>workflow | No. No cases are removed from<br>Worklist or deprioritized. | No. No cases are removed from<br>desktop app or deprioritized |
| Inclusion/<br>Exclusion<br>criteria | Inclusion criteria<br>• Contrast-enhanced chest CTs (but<br>not dedicated CTPA protocol.<br>• Single energy exams.<br>• Scans performed with a 64 slice or<br>greater number of detectors.<br>• Scans performed on adults/<br>transitional adults ≥ 18 years of age.<br>• Slice thickness: 0.5mm - 2.0mm<br>axial.<br>Exclusion Criteria<br>• All studies that are technically<br>inadequate, including studies with<br>motion artifacts, severe metal<br>artifacts, or inadequate field of view. | Inclusion criteria<br>• Frontal Chest X-ray protocol.<br>• Images performed on adults/<br>transitional adults ≥ 18 years of age.<br>Exclusion Criteria<br>• All x-ray images that have<br>inadequate field of view. The device<br>does not provide results when the<br>carina is not well-visualized on the<br>x-ray image. |
| Algorithm | Artificial intelligence algorithm with<br>database of images. | Artificial intelligence algorithm with<br>database of images. |
| Structure | • AHS module (image acquisition);<br>• ACS module (image processing);<br>• Aidoc Worklist application for<br>workflow integration (worklist and<br>non-diagnostic Image Viewer). | • AHS module (orchestrator, image<br>acquisition);<br>• ACS module (image processing);<br>• Aidoc Desktop application for<br>workflow integration (feed and non-<br>diagnostic Image Viewer). |
# Table 1. Key Feature Comparison
{6}------------------------------------------------
{7}------------------------------------------------
# Performance Data
# Pivotal Study Summary
Aidoc conducted a retrospective, blinded, multicenter, study with the BriefCase software to evaluate the software's performance in identifying chest X-ray (CXR) images of vertically malpositioned endotracheal tube (ETT) in relation to the carina in 921 cases from 5 US-based clinical sites. The software's performance was compared to the ground truth as determined by 2 out of 3 majority voting senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm.
Primary endpoints were sensitivity and specificity with an 80% performance goal.
{8}------------------------------------------------
Secondary endpoints were BriefCase time-to-notification compared to the predicate device, Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed.
#### Primary Endpoint
Sensitivity and specificity exceeded the 80% performance goal. Sensitivity was 86.81% (95% C1: 81.02%-91.36%) and Specificity was 92.02% (95% Cl: 89.20%-93.87%).
### Secondary Endpoint
In addition, the time-to-notification metric observed for the BriefCase software in the five medical centers was compared to the equivalent metric of the predicate devices.
- -The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.
The BriefCase time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device) and is given in Table 2 below.
The secondary measure of time-to-notification analysis demonstrated that standard of care time-toexam-open (157.0 minutes: 95% CI: 155.0-159.0; Median: 50.0 IQR: 156.0) was substantially longer than the parallel time-to-notification of the BriefCase for Malpositioned Endotracheal Tube (ETT) device (4.93 minutes, 95% Cl: 4.64-5.22; Median: 4.13 IQR: 0.76). The mean difference of 152.07 minutes between these two measures quantifies effective triage by the device.
| Time-to-notification | Mean<br>Estimate | N | 95%<br>Lower CL | 95%<br>Upper CL | Median | IQR |
|---------------------------------------|------------------|-----|-----------------|-----------------|--------|-------|
| Standard of Care<br>Time-To-Exam-Open | 157.0 | 197 | 155.0 | 159.0 | 50.0 | 156.0 |
| BriefCase Time-to-<br>notification | 4.93 | 182 | 4.64 | 5.22 | 4.13 | 0.76 |
Table 2. Time Saving Data (Sec)
NPV was 98.43% (95% Cl: 97.74% - 98.92%), 96.54% (95% Cl: 95.05% 97.59%) and 94.21% (95% CI: 91.81% 95.94%) at 10%, 20% and 30% positive prevalence.
PPV was 54.71% (95% Cl: 48.45% 60.83%), 73.11% (95% Cl: 67.89%- 77.75%) and 82.33% (95% CI: 78.38%- 85.70%) at 10%, 20% and 30% positive prevalence.
PLR was 10.874 (95% Cl: 8.458-13.979) and NLR was 0.143 (95% Cl: 0.099- 0.208).
Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase for vertically malpositioned ETT in relation to the carina triage the radiologists may have the same benefit in time saving as with the BriefCase for iPE triage.
{9}------------------------------------------------
As can be seen in Table 3 the mean age of patients whose were reviewed for vertically malpositioned ETT in relation to the carina was 62.5 years, with standard deviation of 17.0 years. Gender distribution was 58.52% male, and 41.37% female (Table 4). X-ray scanner distribution can also be found in Table 5 below.
| | Mean | Std | Min | Median | Max | N |
|----------------|------|------|------|--------|-----|-----|
| Age<br>(Years) | 62.5 | 17.0 | 18.0 | 64.0 | 90 | 921 |
#### Table 3. Descriptive Statistics for Age
| Ground<br>Truth<br>Results | Gender | | Male | | Female | | All | |
|----------------------------|----------|---|------|-------|--------|------|-----|-------|
| | | N | % | N | % | N | % | |
| | Positive | | 112 | 12.16 | 70 | 7.60 | 182 | 19.76 |
#### Table 4. Frequency Distribution of Gender *
* The gender of 1 case remained unknown due to incomplete DICOM tag information
311
381
33.77
41.37
738
920*
80.13
100.0
Negative
All
427
439
46.36
58.52
Table 5. Frequency Distribution of Manufacturer
| Manufacturer | N | % |
|--------------|-----|------|
| Carestream | 358 | 38.9 |
| Canon | 223 | 24.2 |
| Siemens | 186 | 20.2 |
| Philips | 154 | 16.7 |
| Total | 921 | 100% |
Tables 6-8 below detail the distribution of pathologies with and without ETT in negative cases, distribution of ETT in negative cases, distribution of other (no ETT) tubes, lines and life support devices in positive cases.
{10}------------------------------------------------
| Pathology | N<br>(w/o<br>ETT) | %<br>(w/o ETT) | N<br>(w/ ETT) | %<br>(w/ ETT) |
|-----------------------------------------------------------|-------------------|----------------|---------------|---------------|
| Fully Negative | 84 | 11.4% | 0 | 0.0% |
| Other (non ETT ) tubes,lines<br>and life support devices* | 220 | 29.8% | 164 | 22.2% |
| Inflammatory | 185 | 25.1% | 57 | 7.7% |
| Post op | 133 | 18.0% | 58 | 7.9% |
| Cardiovascular | 94 | 12.7% | 45 | 6.1% |
| Lung pathologies | 46 | 6.2% | 7 | 0.9% |
| Pneumothorax | 36 | 4.9% | 5 | 0.7% |
| Infectious | 33 | 4.5% | 6 | 0.8% |
| Trauma | 14 | 1.9% | 2 | 0.3% |
| Neoplastic | 13 | 1.8% | 3 | 0.4% |
| Chronic diseases | 2 | 0.3% | 0 | 0.0% |
| None of the above | 28 | 3.8% | 7 | 0.9% |
Table 6. Distribution of Pathologies with and without ETT in Negative Cases
{11}------------------------------------------------
| Pathology | N | % |
|----------------------------------|-----|-------|
| With correctly positioned<br>ETT | 199 | 26.9% |
| With no ETT | 540 | 73.1% |
### Table 7. Distribution of ETT in Negative Cases
# Table 8. Distribution of Other (no ETT) Tubes, Lines and Life Support Devices in Positive Cases
| Pathology | N | % |
|---------------------------------------------------|-----|-------|
| Other tubes, lines and life<br>support devices | 144 | 60.0% |
| No other tubes, lines and<br>life support devices | 96 | 40.0% |
Other (non-ETT) tubes, lines & life support devices' includes a variety of tubes, lines, catheters, monitoring devices and other life support devices. For example: enteric tube (nasogastric/orogastric), chest tube, mediastinal drains, internal jugular central venous catheter, PICC, ECMO cannula, EKG leads, Mediport, etc.
Clinical Subgroups And Confounders:
- -Pathologies present in negative cases: Fully Negative; Other (non ETT) tubes, lines and life support devices; Inflammatory; Post op; Cardiovascular; Lung pathologies; Pneumothorax; Infectious; Trauma; Neoplastic; Chronic diseases; None of the above.
- -Tip position sub-types: Too high; Too low (Trachea); Too low (Bronchus); Too low - right mainstem bronchus; Too low - left mainstem bronchus. An additional enriched dataset was collected for Tip position sub-types analysis on 327 positive cases.
# Conclusions
The subject BriefCase for vertically malpositioned ETT in relation to the carina triage and the predicate BriefCase for iPE triage are intended to aid in prioritization and triage of radiological images for the indications of malpositioned endotracheal tube and incidental pulmonary embolism, respectively. Both devices are software packages with the same technological characteristics and principles of operation,
{12}------------------------------------------------
incorporating deep learning Al algorithms that process images, and software to send notifications and display unannotated compressed low-quality preview images. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.
The subject BriefCase device for vertically malpositioned ETT in relation to the carina triage is thus substantially equivalent to the primary predicate BriefCase for iPE triage.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.