K221319 · Care Fusion · PWH · Jul 21, 2023 · General Hospital
Device Facts
Record ID
K221319
Device Name
BD Alaris™ Pump Epidural Infusion Set
Applicant
Care Fusion
Product Code
PWH · General Hospital
Decision Date
Jul 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BD Alaris™ Pump Epidural Infusion Set is indicated for use by trained healthcare professionals within healthcare facilities through the epidural route for adults, pediatrics and neonates.
Device Story
Administration set for epidural medication delivery; interfaces with BD Alaris™ Pump Module. Components include bag spike, vent cap, drip chamber, tubing, safety clamp, roller clamp, and male NRFit™ connector. NRFit™ connector (ISO 80369-6) prevents misconnection with intravenous luer systems. Used in healthcare facilities by trained professionals. Tubing color-coded yellow for epidural identification. Single-use, sterile, non-pyrogenic. Eliminates risk of medication misconnection; ensures safe delivery of epidural anesthetics/narcotics.
Clinical Evidence
No clinical data.
Technological Characteristics
Administration set with bag spike, drip chamber, tubing, safety clamp, roller clamp, and male NRFit™ connector. Materials: ABS, PVC, silicone. Sterilization: Gamma irradiation (SAL 10^-6). Connectivity: Mechanical interface with BD Alaris™ Pump Module. Compliant with ISO 80369-6 (neuraxial connectors), ISO 8536-4, and ISO 8536-8.
Indications for Use
Indicated for use by trained healthcare professionals in healthcare facilities for epidural administration of medication in adults, pediatrics, and neonates.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
CADD® Yellow High Volume Administration Set with NRFit™ Connector (K172592)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
July 21, 2023
CareFusion Paulina Davis Staff Regulatory Affairs Specialist 10020 Pacific Mesa Blvd San Diego, California 92121
Re: K221319
Trade/Device Name: BD Alaris™ Pump Epidural Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: PWH Dated: June 21, 2023 Received: June 21, 2023
Dear Paulina Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
Page 2
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221319
Device Name BD Alaris™ Pump Epidural Infusion Set
Indications for Use (Describe)
The BD Alaris™ Pump Epidural Infusion Set is indicated for use by trained healthcare professionals within healthcare facilities through the epidural route for adults, pediatrics and neonates.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, it says "has joined BD" in a smaller font. The logo is simple and modern, and the colors are bright and eye-catching.
# K221319 - 510(K) SUMMARY
### Submitter Information
510(k) Submitter:
CareFusion 10020 Pacific Mesa Blvd San Diego, CA 92121, USA
Contact Person:
Paulina Davis Staff Regulatory Affairs Specialist Phone: 714-330-6037 Email Address: Paulina.Davis(@BD.com
Date of Summary Preparation:
July 21, 2023
# Subject Device Identification
Trade Name: Common Name:
Regulation Name: Classification Panel: Regulation Number: Regulatory Class: Product Code:
# Predicate Device Identification
Trade Name:
510(k) Number: 510(k) Clearance Date: Regulation Name: Classification Panel: Regulation Number: Regulatory Class: Product Code: Manufacturer:
BD Alaris™ Pump Epidural Infusion Set Intravascular Administration Set with Neuraxial Connector Intravascular Administration Set General Hospital 21 CFR 880.5440 Class II PWH
CADD® Yellow High Volume Administration Set with NRFit™ Connector K172592 May 17, 2018 Intravascular Administration Set General Hospital 21 CFR 880.5440 Class II PWH Smiths Medical ASD, Inc.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text to the right of the circle. Below the word "CareFusion" is the phrase "has joined BD" in a smaller font size.
# Reason for Submission
The purpose of this Traditional 510(k) is to obtain clearance to market the BD Alaris™ Pump Epidural Infusion Set which contains an NRFit™ connector. The NRFit™ connector conforms to ISO 80369-6, Smallbore connectors for liquids and gases in healthcare applications -- Part 6: Connectors for neuraxial applications. The NRFit™ connector is not compatible with the standard luer connector commonly used for intravenous infusion. The addition of the NRFit™ connector eliminates the risk of misconnection that may result in infusion of medications not intended for an epidural route of administration.
### Device Description
The BD Alaris™ Pump Epidural Infusion Set is an administration set consisting primarily of bag spike, vent cap, drip chamber, tubing (including pump segment), safety clamp, roller clamp, male NRFit™ connector, and male NRFit™ cap. The drip chamber cap located at the proximal end of the administration set, and the male NRFit™ cap, located at the distal end of the administration set, maintain sterility of the fluid path prior to usage. The spike located on the proximal end of the drip chamber is inserted into a prepared fluid container. The male NRFit™ connector enables connection of the administration set to a compatible female NRFit™ connector on the patient's epidural catheter. The BD Alaris™ Pump Epidural Infusion Set is intended to interface with the BD Alaris™ Pump Module. The device is supplied fluid-path sterile using gamma irradiation, in a perforated pouch, is non-pyrogenic, and is for single-use only. The device tubing is color-coded with a yellow stripe to signify administration of medication intended for an epidural route of administration. The BD Alaris™ Pump Epidural Infusion Set is not intended for gravity administration. The device is intended for prescription use only. DEHP or natural rubber latex are not part of the material formulation.
The BD Alaris™ Pump Epidural Infusion Set is designed to deliver medications intended for the epidural route of administration. The NRFit™ connector conforms to ISO 80369-6, Smallbore connectors for liquids and gases in healthcare applications -- Part 6: Connectors for neuraxial applications. The NRFit™ connector is not compatible with the standard luer connector commonly used for intravenous infusion. The addition of the NRFit™ connector eliminates the risk of misconnection that may result in infusion of medications not intended for an epidural route of administration.
### Indication for Use
The BD Alaris™ Pump Epidural Infusion Set is indicated for use by trained healthcare professionals within healthcare facilities through the epidural route for adults, pediatrics and neonates.
### Comparison of Technological Characteristics with Predicate Device
Table 5 - 1 compares the BD Alaris™ Pump Epidural Infusion Set to the predicate device, CADD® Yellow High Volume Administration Set with NRFit™ Connector, including the technological characteristics. The results of this comparison demonstrate that the design,
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The text "CareFusion" is written in gray next to the logo. Below the text, it says "has joined BD" in a smaller font.
technology, materials and composition of the BD Alaris™ Pump Epidural Infusion Set are substantially equivalent to the predicate CADD® Yellow High Volume Administration Set with NRFit™ Connector.
# Table 5 - 1. Comparison of BD Alaris™ Pump Epidural Infusion Set with Predicate CADD® Yellow High Volume Administration Set with NRFit™ Connector
| Characteristic | New Device | Predicate Device | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) | K221319 | K172592 | N/A |
| Applicant | CareFusion | Smiths Medical ASD, Inc. | N/A |
| Trade Name | BD Alaris™ Pump Epidural<br>Infusion Set | CADD® Yellow High<br>Volume<br>Administration Set<br>with NRFit™<br>Connector | N/A |
| Model Numbers | NRF2208-0007<br>NRF2206-0007 | 21-7684-24 | N/A |
| Classification<br>Regulation | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| Regulation Name | Intravascular Administration<br>Set | Intravascular Administration<br>Set | Same |
| Device Class | II | II | Same |
| Product Code | PWH | PWH | Same |
| Indications for Use | The BD Alaris™ Pump<br>Epidural Infusion Set is<br>indicated for use by trained<br>healthcare professionals<br>within healthcare facilities<br>through the epidural route<br>for adults, pediatrics and<br>neonates. | CADD® Yellow High<br>Volume<br>Administration Set<br>with NRFit™<br>connectors are<br>intended for use only<br>with NRFit™<br>components for<br>delivery of regional<br>anesthetics or<br>narcotics, excluding<br>subarachnoid / spinal block,<br>and are<br>designed for use with<br>CADD® pumps (see<br>CADD® pump | Epidural administration<br>(the administration route<br>for subject device) is a<br>type of regional<br>administration. The<br>subject device has the<br>same intended use as the<br>predicate device. Both<br>the subject device and<br>predicate device are<br>intended to be used with<br>infusion pumps and in<br>the same patient<br>populations (adults and<br>pediatrics (also includes<br>neonates). |
| Characteristic | New Device | Predicate Device | Comparison |
| | | Operator's Manual for<br>compatibility). | |
| | | The intended population is<br>pediatrics and adults. | |
| Rx or OTC? | Rx | Rx | Same |
| Utility | Disposable, for Single Use<br>Only | Disposable, Single Use Only | Same |
| Connection | ISO 80369-6 compliant male<br>connector | ISO 80369-6 compliant male<br>connector | Same |
| Bag Spike | Acrylonitrile<br>Butadiene Styrene (ABS) | Acrylonitrile<br>butadiene styrene<br>(ABS) - ABS - HI<br>121H-Natural<br>Granule 9700640000<br>Colorant: Yellow | Different; safety of<br>materials that comprise<br>the finished device was<br>demonstrated by the<br>passing results of ISO<br>10993 biocompatibility<br>testing. |
| Tubing | PVC, silicone | PVC | Different; safety of<br>materials that comprise<br>the finished device was<br>demonstrated by the<br>passing results of ISO<br>10993 biocompatibility<br>testing. |
| Male NRFit™<br>connector | ABS | Polycarbonate - Makrolon<br>RX-1805-<br>013771<br>Colorant: White | Different; safety of<br>materials that comprise<br>the finished device was<br>demonstrated by the<br>passing results of ISO<br>10993 biocompatibility<br>testing. |
| Check valve | No | Yes (part of male NRFit™<br>connector) | Different; subject device<br>does not have a check-<br>valve as part of the male<br>NRFit™ connector.<br>Check valve is not<br>required because no<br>additional infusion line |
| Characteristic | New Device | Predicate Device | Comparison |
| | | | will be connected for<br>epidural infusions. |
| Flowstop | Yes (safety clamp) | Yes | Different; Verification<br>testing was completed to<br>demonstrate that the<br>safety clamp performs as<br>intended |
| Set Length | NRF2208-0007: 117 inches<br>NRF2206-0007: 96 inches | 130 inches | Different; system-level<br>performance testing (flow<br>rate and occlusion time-<br>to-alarm) demonstrates<br>the subject device<br>performs as intended. |
| Priming Volume | NRF2208-0007: 14.4 mL<br>NRF2206-0007: 13 mL | 16-23 mL | Different; priming<br>volume of subject devices<br>is calculated based on<br>tubing length and subject<br>device components, and<br>verified through<br>performance testing. |
| Flow Type | Pump Operated | Pump Operated | Same |
| Packaging Type | Perforated Pouch | Tyvek Pouch | Different; microbial<br>barrier testing was<br>conducted as part of<br>shelf-life performance to<br>demonstrate the subject<br>device performs as<br>intended |
| Sterilization | Provided fluid path sterile by<br>gamma irradiation | Provided sterile by ethylene<br>oxide | Different; both Category<br>A sterilization methods.<br>CareFusion has validated<br>that the sterilization<br>method and parameters<br>are adequate to ensure a<br>Sterility Assurance Level<br>(SAL) of 10-6. |
| Shelf Life | 3 years | Unknown | Different; performance<br>testing on accelerated-<br>aged devices<br>demonstrates the subject<br>device performs as |
| Characteristic | New Device | Predicate Device | Comparison |
| | | | intended through its<br>labeled shelf life. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, it says "has joined BD" in a smaller font.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, it says "has joined BD" in a smaller font.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller font.
# Safety and Performance Testing
The BD Alaris™ Pump Epidural Infusion Set was evaluated via non-clinical safety and performance testing to demonstrate that the device met pre-established acceptance criteria and is substantially equivalent to the predicate device.
The safety and performance testing on the BD Alaris™ Pump Epidural Infusion Set was conducted according to the following FDA recognized and internally recognized consensus standards listed below:
Human Factors
- FDA Recognition Number 5-57, ANSI/AAMI HE75: 2009 (R2013), Human ● factors engineering- Design of medical devices
- FDA Recognition Number 5-114, IEC 62366-1: Edition 1.0 2015, Medical devices -Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
The following functional, bench and system testing have been performed:
- . FDA Recognition Number 6-318. ISO 8536-4: 2010/Amd1: 2013(E), Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed [Including: Amendment 1 (2013)]
- FDA Recognition Number 6-358, ISO 8536-8: 2015, Infusion Equipment for Medical Use - Part 8: Infusion Sets for Single Use with Pressure Infusion Apparatus
- FDA Recognition Number 5-121. ISO 80369-1: 2018. Small-bore connectors for liquids . and gases in healthcare applications - Part 1: General requirements
- FDA Recognition Number 5-108, ISO 80369-6: 2016, Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications
- FDA Recognition Number 5-97, ISO 80369-20: 2015, Small-bore connectors for liquids ● and gases in healthcare applications - Part 20: Common test methods
- Flow Rate Accuracy
- AAMI TIR101:2021 Fluid delivery performance testing for infusion pumps
- FDA guidance document, Infusion Pumps Total Product Life Cycle •
- Bolus Accuracy
- AAMI TIR101:2021 Fluid delivery performance testing for infusion pumps •
- FDA guidance document, Infusion Pumps Total Product Life Cycle
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for CareFusion. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is smaller text that says "has joined BD".
- Air-in-Line ●
- Upstream Occlusion Time-to-Alarm ●
- Post Occlusion Bolus & Downstream Occlusion Time to Alarm ●
- EN ISO 14971: 2012, Medical devices - Application of risk management to medical devices
# Biocompatibility
Biocompatibility tests for the subject device were performed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process as recognized by the FDA. Per ISO 10993-1: 2018 the device is categorized as:
- externally communicating, blood path indirect, and prolonged exposure (>24 hours . to 30 days).
- externally communicating, tissue contact (indirect), and prolonged exposure; and, .
- surface contacting, intact skin, and limited duration (≤ 24 hours). .
A biological and Chemical Risk Assessment for BD Alaris™ Pump Epidural Infusion Sets was conducted, and the following tests were addressed:
- Cytotoxicity: FDA Recognition Number 2-245, ISO 10994-5: 2009, Biological ● evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity
- Sensitization: FDA Recognition Number 2-174, ISO 10993-10: 2010, Biological ● evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- Irritation or Intracutaneous Activity: FDA Recognition Number 2-291, ISO 10993-23:2021, Biological evaluation of medical devices – Part 23: Tests for irritation
- Acute Systemic Toxicity: FDA Recognition Number 2-255, ISO 10993-11: 2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Hemocompatibility: FDA Recognition Number 2-250, ASTM F756-2017, Standard ● Practice for Assessment of Hemolytic Properties of Materials
- Hemolysis: FDA Recognition Number 2-248, ISO 10993-4: 2017, Biological ● evaluation of medical devices--Part 4: Selection of tests for interactions with blood
- USP 41, National Formulary 36 (USP), General Chapter <788> Particulate Matter In Injections
- Genotoxicity per ISO 10993-3:2014. Biological evaluation of medical devices -● Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- FDA Recognition Number 2-289, ISO 10993-12: 2021, Biological evaluation of ● medical devices - Part 12: Sample preparation and reference materials
- ISO 10993-17: 2002/(R)2008, Biological evaluation of medical devices -● Part 17: Establishment of allowable limits for leachable substances
- ISO 10993-18: 2020, Biological evaluation of medical devices Part 18: ●
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller, lighter gray font.
Chemical characterization of materials
- FDA Recognition Number 2-258, ISO 10993-1: 2018, Biological evaluation of ● medical devices - part 1: Evaluation and testing within a risk management process
- FDA Recognition Number 2-222. ISO 10993-2: 2006, Biological Evaluation of ● medical devices - Part 2: Animal welfare requirements
- USP 41, National Formulary 36 (USP), General Chapter <151>: 2018, Pyrogen Test (USP Rabbit Test)
Sterilization and Shelf Life/Packaging
- . FDA Recognition Number 14-528, ISO 11137-1: 2006, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
- FDA Recognition Number 14-409, ISO 11137-2: 2013, Sterilization of health care ● products - Radiation - Part 2: Establishing the sterilization dose
- FDA Recognition Number 14-514, ISO 11737-1: 2018, Sterilization of health care ● products - Microbiological methods - Part 1: Determination of a population of microorganisms on product
- FDA Recognition Number 14-541, ANSI/AAMI ST72: 2019, Bacterial endotoxins -● Test methods, routine monitoring, and alternatives to batch testing
- ASTM D4728-17, Random vibration testing of shipping containers ●
- FDA Recognition Number 14-499, ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM D5276-19, Standard Test Method for Drop Test of Loaded Containers by Free Fall
- FDA Recognition Number 14-497, ASTM F1980-16, Standard Guide for Accelerated ● Aging of Sterile Barrier Systems for Medical Devices
- USP 41, National Formulary 36 (USP), General Chapter <161>: 2018, Medical ● Devices- Bacterial Endotoxin and Pyrogen Tests
- USP 41, National Formulary 36 (USP), General Chapter <85>:2018, Bacterial ● Endotoxins Test
- . FDA Recognition Number 5-117, ISO 15223-1: 2016, Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements
All testing met pre-established acceptance criteria and successfully demonstrated that the device, the BD Alaris™ Pump Epidural Infusion Set, is safe for its intended use and performs as intended. Additionally, the BD Alaris™ Pump Epidural Infusion Set and the predicate CADD® Yellow High Volume Administration Set with NRFit™ Connector were tested to similar standards, including ISO 80369-6, and thus the subject device is considered substantially equivalent to the predicate device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller, lighter gray font.
# Clinical Data
There are no clinical data included in this submission.
# Conclusion
The information provided in this 510(k) submission, including the non-clinical safety and performance testing, is sufficient to demonstrate substantial equivalence of the BD Alaris™ Pump Epidural Infusion Set to the predicate device. The results of non-clinical safety and performance testing met the pre-established acceptance criteria. The BD Alaris™ Pump Epidural Infusion Set and its predicate have similar indications for use, the same intended use, and similar technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise new or different questions of safety and effectiveness. Additionally, the information provided demonstrates the subject device is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.