K221312 · Guangzhou Chuang Zao Mei Technology Co., Ltd. · GEX · Jun 29, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221312
Device Name
Diode Laser Hair Removal Device
Applicant
Guangzhou Chuang Zao Mei Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Diode laser (808 nm) surgical device for dermatologic hair removal; utilizes laser handpiece and foot switch for activation. Energy density ranges: 1-77 J/cm² (EVOLUTION MEDICAL) or 1-70 J/cm² (M-I-X MEDICAL). Operated by clinicians in professional settings. Device delivers controlled laser pulses to skin to achieve permanent hair reduction. Safety and efficacy established via non-clinical bench testing and comparison to predicate device.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety, EMC, laser safety, and biocompatibility (cytotoxicity, sensitization, irritation).
Technological Characteristics
Diode laser (808 nm); Class IV laser. Handpiece tip materials: 6063 Aluminum and S1 Quartz. Fluence: 1-77 J/cm²; Pulse duration: 3-320 ms; Frequency: 1-10 Hz. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin, in patients requiring dermatological hair removal.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
June 29, 2022
Guangzhou CHUANG ZAO MEI Technology Co., Ltd % Asher No Last Name Provided Regulatory Affairs Manager AskWay Innovative Ltd. 4F, Yuehuayuan Building, 2008 Nanshan Avenue, Nanshan Street Nashan District Shenzhen, Guangdong 518000 China
Re: K221312
Trade/Device Name: Diode Laser Hair Removal Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 29, 2022 Received: May 5, 2022
Dear Asher No Last Name Provided:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221312
Device Name Diode Laser Hair Removal Device
Indications for Use (Describe)
The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|--|
| <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo for "PROF BEAUTY". To the left of the text is a circular graphic containing a line drawing of a woman's face. Below the text is the phrase "innovations creating the beauty".
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### l. SUBMITTER
Guangzhou CHUANG ZAO MEI Technology Co., Ltd. Dehe international A505, No.2, Qixinggang Industrial Road, JUNHE street, Baiyun District, Guangzhou Contact Person: KARPOV ALEKSANDR Position: General Manager Phone: 086-(0) 13147561925 Email: 3152329212@qq.com
Date Prepared: 06/21/2022
#### II. SUBMISSION CORRESPONDENT
| Primary Contact Person: | Asher |
|---------------------------|----------------------------|
| | Regulatory Affairs Manager |
| | AskWay Innovative Ltd. |
| | Phone: 086-(0) 18925456615 |
| | Email: Asher@cx.cx |
| Secondary Contact Person: | Albert Ou |
General Manager AskWay Innovative Ltd. Phone: 086-(0) 13751890680 Email: winner_link@126.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for PROF BEAUTY. To the left of the text is a circular graphic of a woman's face. The text to the right of the graphic says PROF BEAUTY in large, bold letters. Below that text is the phrase "Innovations creating the beauty" in a smaller font.
### III. PROPOSED DEVICE
| Trade Name: | Diode Laser Hair Removal Device |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Model(s): | EVOLUTION MEDICAL, M-I-X MEDICAL |
| Classification Name: | Laser Surgical Instrument For Use In General And Plastic<br>Surgery And In Dermatology (21 CFR 878.4810) |
| Regulation Class: | II |
| Product Code: | GEX |
| Review Panel: | General & Plastic Surgery |
IV. PREDICATE DEVICE
| 510(k) Number: | K210663 |
|----------------|---------------------------------------------|
| Device Name: | Dermatological Diode Laser Systems |
| Manufacturer: | Beijing HuaCheng Taike Technology Co., Ltd. |
The predicate has not been subject to a design-related recall.
#### DEVICE DESCRIPTION V.
The Diode Laser Hair Removal Device is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a diode laser as a laser source (808 nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a foot switch and a laser handpiece.
There are two models included, EVOLUTION MEDICAL, M-I-X MEDICAL, the two models have same mechanism of action, principle and specification, with only one difference: the adjustable range of Energy density is different, EVOLUTION MEDICAL (1-77 J/cm²), M-I-X MEDICAL (1-70 J/cm²).
### vi. INDICATIONS FOR USE
The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the
{5}------------------------------------------------
number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
## vii. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Diode Laser Hair Removal Device (EVOLUTION MEDICAL, M-I-X MEDICAL) is the same or similar to the cleared predicate device.
The Diode Laser Hair Removal Device (EVOLUTION MEDICAL, M-I-X MEDICAL) has same indications for use with the predicate device, and similar technological characteristics such as fluence, laser wavelength, spot size, and pulse duration, etc. with the predicate device. Please refer to the following table for details:
| Item | Proposed Device<br>K221312 | Predicate Device<br>K210663 | Remark |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device name | Diode Laser Hair Removal Device | Dermatological Diode Laser Systems | / |
| Product model | EVOLUTION MEDICAL/<br>M-I-X MEDICAL | CM01D/CM02D | / |
| K number | K221312 | K210663 | / |
| Product code | GEX | GEX | Same |
| Classification<br>regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Class | 2 | 2 | Same |
| Indications for Use | The Diode Laser Hair removal device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | The Dermatological Diode Laser Systems (Model: CM01D/CM02D) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | Same |
| Prescription use or<br>not | Prescription use | Prescription use | Same |
| Configuration | Main Unit<br>Handpiece | Main Unit<br>Handpiece | Same |
{6}------------------------------------------------
| Item | Proposed Device<br>K221312 | Predicate Device<br>K210663 | Remark |
|---------------------------|------------------------------------------------------------|---------------------------------------------------------------|---------|
| Foot Control | Foot Control | Foot Control | Same |
| Principle of<br>Operation | Diode laser | Diode laser | Same |
| Laser Type | Diode laser | Diode laser | Same |
| Laser Classification | Class IV | Class IV | Same |
| Laser Wavelength | 808 nm | 808 nm | Same |
| Spot size | 12.6mm*20.6mm | CM01D: 10 x 30 mm<br>CM02D: 9 x 12 mm | Similar |
| Fluence | EVOLUTION MEDICAL: 1-77 J/cm²<br>M-I-X MEDICAL: 1-70 J/cm² | CM01D: 5-100 J/cm²<br>CM02D: 3-30 J/cm² | Similar |
| Frequency | 1-10 Hz | CM01D: 1-10 Hz<br>CM02D: 1-3 Hz | Same |
| Pulse Duration | 3-320 ms | CM01D: 15-400 ms<br>CM02D: 35-400 ms | Similar |
| Power Supply | AC 120V/60 Hz | AC 110V/60Hz | Similar |
| Dimension | 393mm x 430mm x 1130mm | CM01D: 650mm x 650mm x 1230mm<br>CM02D: 252mm x 210mm x 193mm | / |
| Weight | 48kg | CM01D: 75kg<br>CM02D: 3kg | / |
### Analysis:
The proposed device is slightly different in spot size, fluence range, pulse duration, power supply, dimension, and weight from the predicate device. Spot size, fluence range, pulse duration can all be covered by predicate device, the differences in power supply, dimension and weight can be accepted by complying with the non-clinical test conducted. Therefore, these differences will not affect the safety and effectiveness, and the proposed device is determined to be the same or similar with predicate device.
| Item | Proposed Device<br>K221312 | Predicate Device<br>K210663 | Remark |
|----------------------------------|----------------------------------------------------|--------------------------------------------|--------|
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with ANSI/AAMI ES60601-1,<br>IEC 60601-2-22 | Comply with IEC 60601-1,<br>IEC 60601-2-22 | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for PROF BEAUTY. To the left of the text is a circular graphic with a line drawing of a face. Below the company name is the tagline "innovations creating the beauty."
| Item | Proposed Device<br>K221312 | Predicate Device<br>K210663 | Remark |
|----------------------------------------------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------|--------|
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC<br>60825 | Same |
| Patient Direct/Indirect Contact Materials and Biocompatibility | | | |
| Patient<br>Direct/Indirect<br>Contact<br>Materials | Tip of Handle (6063 Aluminum<br>& S1 Quartz) | Tip of Handle (6061 Aluminum<br>&Quartz)<br>Device Housing (Acrylonitrile<br>Butadiene Styrene) | Same |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No evidence of sensitization | No evidence of sensitization | Same |
| Irritation | No evidence of irritation | No evidence of irritation | Same |
| Biocompatibility<br>testing<br>standards | Comply with ISO 10993-5,<br>ISO 10993-10 | Comply with ISO 10993-5,<br>ISO 10993-10 | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Prof Beauty. The logo consists of a circular graphic to the left of the text "PROF BEAUTY". The graphic is a line drawing of a face inside of a circle. Below the text "PROF BEAUTY" is the text "innovations creating the beauty".
#### VIII. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was the same or similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ANSI/AAMI A2:2010/(R)2012 (Consolidated Text), Medical electrical equipment — Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
IEC 60601-2-22 Edition 3.1 2012-10, Medical Electrical Equipment - Part 2-22: Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment.
IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)
Performance Testing for Energy Output Accuracy.
Software Verification and Validation Testing was conducted per "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", and the level of concern was determined to be Moderate for the proposed device.
### IX. CLINICAL TEST CONCLUSION
No clinical study is included in this submission.
#### CONCLUSIONS X.
Based on the comparison and analysis above, the proposed device is determined to be as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.