# Transducer Probe Cover (K221278)

_Hony Medical Co., Ltd. · ITX · Jan 11, 2023 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221278

## Device Facts

- **Applicant:** Hony Medical Co., Ltd.
- **Product Code:** [ITX](/productcode/ITX.md)
- **Decision Date:** Jan 11, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1570
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle guided procedures for external intact skin diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable.

## Device Story

Device is a transparent, thin-walled polyurethane sleeve with rubber ring/tape strip; applied over diagnostic ultrasound transducer heads. Used in clinical settings by healthcare providers during ultrasound scanning and needle-guided procedures on external intact skin. Function: acts as a physical microbial barrier to prevent transfer of body fluids, microorganisms, and particulates between patient and probe; maintains sterile field. Application: user applies sterile coupling gel to transducer face or inside cover, inserts probe, and unrolls cover over length of transducer; secured with bands. Removal: reverse of application. Benefits: facilitates easier probe cleaning/disinfection; protects patient and staff; does not impair ultrasound imaging. Device is sterile, single-use, and disposable.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ISO 10993 (biocompatibility: cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis), ISO 11607-1 (sterilization/shelf life), ASTM F1671-13 (viral penetration), ISO 527-3 (tensile strength/elongation), and ISO 9073-4 (tear resistance). Acoustic properties (impedance, velocity, attenuation) confirmed to have negligible effect on ultrasound imaging.

## Technological Characteristics

Material: 0.05mm thick, 49 GSM polyurethane film. Form factor: tubular sleeve with heat-sealed distal end, various dimensions. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectivity: None (mechanical barrier). Standards: ISO 10993 (biocompatibility), ISO 11607-1 (packaging), ASTM F1671-13 (viral barrier), ISO 527-3/ISO 9073-4 (mechanical properties).

## Regulatory Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

## Predicate Devices

- GENERAL PURPOSE TRANDUCER COVER ([K970513](/device/K970513.md))

## Reference Devices

- CIV-Clear cover ([K211270](/device/K211270.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 11, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Hony Medical Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd. Pudong, Shanghai 200120 CHINA

Re: K221278

Trade/Device Name: Transducer Probe Cover Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: ITX Dated: December 7, 2022 Received: December 15, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221278

Device Name Transducer Probe Cover

#### Indications for Use (Describe)

Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle guided procedures for external intact skin diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable.

| Type of Use (Select one or both, as applicable)                                        |                                                                                       |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <div> <span> </span>         Prescription Use (Part 21 CFR 801 Subpart D)       </div> | <div> <span> </span>         Over-The-Counter Use (21 CFR 801 Subpart C)       </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K221278

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

## 1.0 Submitter's Information

| Name:    | Hony Medical Co., Ltd.                                                |
|----------|-----------------------------------------------------------------------|
| Address: | No.12,9Road,Changlong Ind.Zone,Taishan City,Guangdong<br>China 529222 |
| Tel:     | +86 15916327827                                                       |
| Contact: | Zhu Huina                                                             |

#### Designated Submission Correspondent

| Contact: | Mr. Boyle Wang                                                      |
|----------|---------------------------------------------------------------------|
| Name:    | Shanghai Truthful Information Technology Co., Ltd.                  |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China |
| Tel:     | +86-21-50313932                                                     |
| Email:   | Info@truthful.com.cn                                                |

Date of Preparation: Jan.11,2023

## 2.0 Device Information

| Trade name:          | Transducer Probe Cover                       |
|----------------------|----------------------------------------------|
| Common name:         | Ultrasonic Diagnostic Transducer Probe Cover |
| Classification name: | Transducer, Ultrasonic, Diagnostic.          |
| Model(s):            | Various Dimension                            |
| Production code:     | ITX                                          |
| Regulation number:   | 21 CFR 892.1570                              |
| Classification:      | Class II                                     |
| Panel:               | Radiology                                    |

#### 3.0 Predicate and Reference Device Information

### Predicate#

| Manufacturer: | CIVCO MEDICAL INSTRUMENTS CO., INC. |
|---------------|-------------------------------------|
| Trade Device: | GENERAL PURPOSE TRANDUCER COVER     |

{4}------------------------------------------------

510(k) number: K970513

### Reference#

| Manufacturer:  | CIVCO MEDICAL INSTRUMENTS CO., INC. |
|----------------|-------------------------------------|
| Trade Device:  | CIV-Clear cover                     |
| 510(k) number: | K211270                             |

## 4.0 Device Description

The subject device is composed of transparent thin-walled polyurethane sleeve, rubber ring and (or) tape strip. A conventional 0.05mm thin, 49 GSM (Grams per Square Meter), transparent high strength polyurethane film tube shape, in various dimensions with heat sealed distal end to be applied over a transducer probe to provide a Transducer Cover that can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe.

Ultrasound imaging is not impaired by use of the cover as it is intended. Adequate coupling between the cover and the transducer is required. The Transducer Probe Cover is utilized by applying sterile transmission, coupling, or lubricating gel onto the transducer face or into closed end of cover, inserting ultrasound transducer into closed end of cover and unrolling cover over length of the transducer as desired, and securing open end of cover with bands as necessary the removal process is accomplished by pulling the cover off the transducer in a reverse method from the application.

The subject device is furnished in sterile condition, for single use patient/procedure use, disposable.

As the device is single use device, which is individually packaged sterile devices. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified bacterial film to maintain the sterilization condition of the device.

#### 5.0 Indication for Use Statement

Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle guided procedures for external intact skin diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable.

{5}------------------------------------------------

# 6.0 Summary of Non-Clinical Testing

Summary of non-clinical and performance testing-bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 10993-1, ISO 10993-7, ISO 11607-1. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.

6.1 Sterilization and shelf life - Transducer Probe Cover is delivered sterile and have successfully been tested according to ISO 11607- 1. The label shelf life is 3 years. 6.2 Biocompatibility testing - Transducer Probe Cover has successfully been tested for cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity, material medicated pyrogenicity and Hemolysis Testing. The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. Transducer Probe Cover is non-toxic and biocompatible.

6.3 Performance testing – Bench The performance of Transducer Probe Cover has been verified. Tests as described in table 1 have been completed.

| Test Item           | Acceptance Criteria                                                                                                                                                                                                                                                                                                                | Result Summary                                                                                                                                                                             |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Viral Penetration   | To validate the ability to withstand viral penetration per ASTM F1671-13.                                                                                                                                                                                                                                                          | Meets requirements of ASTM F1671 for prevention of blood-borne pathogens, and the material has been tested and proven to be a viral barrier for microbes 25 nanometers and larger in size. |
| Tensile             | The maximum tensile force for longitudinal breaking of the membrane of the Transducer Probe Cover shall be no less than 20N, the maximum tensile force for transverse breaking shall not be less than 10N, and the elongation at break of the membrane of the disposable ultrasonic inspection sheath shall not be less than 100%. | Meets requirements of ISO 527-3:1995.<br>Tensile strength: 37.41N/25mm, longitudinal tensile strength: 39.25N/25mm, elongation at break: 517.51%                                           |
| Tear resistance     | Under the specified conditions, test the tearing strength of the disposable ultrasonic inspection sheath is not less than 5N                                                                                                                                                                                                       | Meets requirements of ISO 9073-4-2021<br>Longitudinal crack resistance: 16.64N<br>Lateral crack resistance: 17.64N                                                                         |
| Water resistance    | To determine the resistance of the materials to the penetration of water by impact.<br>The time that can withstand 500mm hydrostatic pressure should not be less than 300s.                                                                                                                                                        | Meet the requirements                                                                                                                                                                      |
| Acoustic properties | Sound Attenuation: Measured at 35°C, Sound                                                                                                                                                                                                                                                                                         | Meet the requirements:                                                                                                                                                                     |

Table 1: Performance testing summary - Bench

{6}------------------------------------------------

|                                                     | Attenuation shall be ≤0.1dB/(cm·MHz);<br>Acoustic Impedance: Measured at 35°C ,Acoustic<br>Impedance shall be $1.5x10^6~1.7x10^6$ Pa·s/m<br>Sound Velocity: Measured at 35°C, the Sound<br>Velocity (Acoustic Velocity) shall be 1520-1620m/s. | Acoustic Velocity: 1594.4 m/s;<br>Acoustic Attenuation: 0.01dB/(cm·MHz)<br>Acoustic Impedance:1.60 Pa·s/m<br>The Transducer Probe Cover does not affect<br>the acoustic properties of the ultrasound<br>device. |                                                                                                                                                                                                                                                             |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Force at Break and<br>Tear Resistance<br>Properties | To demonstrate material breaking force property<br>and tear resistance property per ISO 527-3:2018                                                                                                                                             | Meets requirements of ISO 527-3:2018:<br>Force at break:28.85 N/25mm;<br>Tear resistance:515%.                                                                                                                  |                                                                                                                                                                                                                                                             |
|                                                     | Water Leakage                                                                                                                                                                                                                                  | Watertightness Test for Detection of Holes.<br>The subject probe cover shall be free of pinhole                                                                                                                 | No Water Leakage                                                                                                                                                                                                                                            |
|                                                     |                                                                                                                                                                                                                                                | Airburst pressure<br>and Volume                                                                                                                                                                                 | To demonstrate the mechanical strength and<br>durability. The Airburst Pressure shall not be less<br>than 1.0 kpa and the volume shall be not less than<br>28 dm³ for cover with a mid-body width greater<br>than or equal to 65,0 mm and less than 75,0 mm |
| Ultrasound imaging<br>effect                        | Ultrasound imaging is not impaired by use of the<br>cover as it is intended                                                                                                                                                                    |                                                                                                                                                                                                                 | Meets requirements<br>The image is clear.                                                                                                                                                                                                                   |

# 7.0 _Summary of Clinical Testing

No clinical study is included in this submission.

{7}------------------------------------------------

## 8.0 Technological Characteristic Comparison Table

| Item                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Conclusion                                                                                                                                |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No.                            | K221278                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | K970513                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K211270                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | --                                                                                                                                        |
| Product Code                          | ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                      |
| Regulation No.                        | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                      |
| Class                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                      |
| Intended<br>Use/Indication for<br>Use | Transducer Probe Cover placed over<br>diagnostic ultrasound transducer/<br>probe scan head instruments. The<br>cover allows use of the transducer in<br>scanning and needle guided<br>procedures for external intact skin<br>diagnostic ultrasound, while helping to<br>prevent transfer of microorganisms,<br>body fluids, and particulate material to<br>the patient and healthcare worker<br>during reuse of the transducer. The<br>cover also provides a means for<br>maintenance of a sterile field.<br>Transducer Probe Cover are furnished<br>sterile; single use patient/procedure,<br>disposable. | Protective cover or sheath placed<br>over diagnostic ultrasound<br>transducer / probe scan head<br>instruments. The cover allows use of<br>the transducer in scanning and<br>needle guided procedures for body<br>surface, endocavity, and<br>intra-operative diagnostic ultrasound,<br>while helping to prevent transfer of<br>microorganisms, body fluids , and<br>particulate material to the patient and<br>healthcare worker during reuse of<br>the transducer (both sterile and<br>non-sterile covers) . The cover also<br>provides a means for maintenance of<br>a sterile field (sterile covers only) .<br>CIVCO Poly Ultrasound Transducer | The cover is intended as a microbial<br>barrier between the patient and<br>medical imaging equipment.<br>The transducer covers are used for<br>adult of all body sizes in sterile and<br>non-sterile fields and for the following<br>applications.<br>• Abdominal - Diagnostic imaging<br>and minimally invasive puncture<br>procedures.<br>• Small Parts - Diagnostic imaging<br>and minimally invasive puncture<br>procedures.<br>• Regional Anesthesia - Minimally<br>invasive puncture procedures.<br>• Vascular Access - Vessel<br>identification and catheter | Similar with the<br>predicate device                                                                                                      |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                           |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Covers are furnished sterile &<br>non-sterile single use patient /<br>procedure, disposable.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | placement.<br>• Surgical - Diagnostic imaging and<br>puncture procedures.<br>• Transesophageal - Diagnostic<br>imaging and monitoring of heart<br>chamber, valves and vessels.<br>• Transrectal - Diagnostic imaging<br>and minimally invasive puncture<br>procedures.<br>• Transvaginal - Diagnostic imaging<br>and minimally invasive puncture<br>procedures                                                                                                                                                                                                      |                                                                                                                                           |
| Materials &<br>Construction           | Polyurethane, tubular, sealed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Polyurethane and polyethylene<br>extruded thermoplastic film, in<br>one-piece, open on one end, closed<br>on other end.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ethyl Methyl Acrylate<br>(EMA) and Polyethylene blend                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | All the device have<br>passed the<br>biocompatibility<br>evaluation                                                                       |
| Model                                 | Various Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Various Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Various Size                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                      |
| Microbial Barrier                     | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same with the<br>Reference device.                                                                                                        |
| Acoustic<br>Performance               | Acoustic Impedance: $1.60 x 10^6$ Pa<br>s/m                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | $1.54 x 10^6$ Pa s/m                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The subject transducer<br>probe cover has<br>negligible effect on the<br>transmission of<br>ultrasound signal and<br>the transducer probe |
|                                       | Acoustic Velocity: 1594 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1630 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                           |
|                                       | Acoustic Attenuation:<br>0.01dB/(cm-MHz)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                           |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | cover does not degrade<br>signal or image.                                                                                                |
| Sterile                               | EO sterilization, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Both in EO sterilization and<br>non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Both in EO sterilization and<br>non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                      |
| Disposable,<br>Single Use Only        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                      |
| Shelf Life                            | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not Publicly Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same with the Reference<br>Device                                                                                                         |
| Biocompatibility                      | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Conform with ISO 10993 standards                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                      |

## Table 2- Comparison of Technology Characteristics

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:

- Same intended use
- · Similar material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods
- · Same fundamental technology/principal of operation/user interface

The subject device is just provided in sterile condition while the predicate device is provided both in sterile and non-sterile condition, there is no significant risk raised by the difference.

#### 9.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K221278](https://fda-staging.innolitics.com/device/K221278)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).