The Qraycam PRO is intended to be used as an aid in the diagnosis of dental caries.
Device Story
Qraycam PRO is a fluorescence-based caries detection aid for dentists and dental specialists. It uses Quantitative light-induced fluorescence (QLF-D) technology to assess green auto-fluorescence of tooth structures and red fluorescence of bacterial metabolites. The device features an integrated LED light source (405nm blue light for fluorescence, white LED for illumination) and a variable focus liquid lens. It captures images of the oral cavity, which are displayed on an attached LCD screen or a connected PC/laptop. Dentists use these images to identify and quantify dental caries stages. The device is used in clinical dental settings to assist in oral health examinations and caries diagnosis, with images stored in electronic charts for clinical decision-making and insurance claims. It benefits patients by enabling early or more accurate detection of caries.
Clinical Evidence
Bench testing only. Comparative performance evaluation was conducted between the subject extra-oral camera and the predicate intra-oral camera using the same teeth under identical conditions. The study measured the ΔF average value to confirm clinical performance equivalence. No clinical trials or biocompatibility testing were performed as the device does not contact the patient.
Technological Characteristics
Fluorescence caries detection device using QLF-D technology. Light source: 405nm UV/blue LED and white LED. Features variable focus liquid lens (50mm-200mm). Connectivity: PC/laptop via USB. Standards: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), IEC 62471 (Photobiological Safety), IEC 62304 (Software). Reusable device.
Indications for Use
Indicated for use as an aid in the diagnosis of dental caries in patients undergoing dental examinations. Contraindicated for use on teeth where visual identification is difficult due to malocclusion.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.
September 4, 2024
Aiobio Co., Ltd. % Jeongkeun Kim Regulatory Affairs Consultant KMC,Inc. Room no. 1709, 123, Digital-ro 26-gil, Guro-gu Seoul, 08390 Korea. South
Re: K221275
Trade/Device Name: Qraycam PRO Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: Class II Product Code: NBL Dated: August 6, 2024 Received: August 6, 2024
Dear Jeongkeun Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Bobak Shirmohammadi -S
For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K221275
Device Name
Qraycam PRO
Indications for Use (Describe)
The Qraycam PRO is intended to be used as an aid in the diagnosis of dental caries.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AIOBIO. The letters "AIO" are in gray, and the letters "BIO" are in a blue square. The logo is simple and modern.
#### K221275 510(k) Summary
This summary of 510(K) – safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Auq 06, 2024
## 1. Applicant / Submission Sponsor
AIOBIO Co.,Ltd Address: # 306 Sunil Technopia, 555 Dunchon-Daero, Jungwon-Gu Seongnam-Si Gyeonggi, KR 13215 Tel: +82-2-5615101
#### 2. Submission Correspondent
JeongKeun Kim (Consultant, KMC, Inc.) Address: Room no. 1709, 123, Digital-ro 26-gil, Guro-gu, Seoul, 08390, REPUBLIC OF KOREA Tel: +82-70-8965-5554 Email: jkkim@kmcerti.com
#### 3. Device Identification
Trade/Proprietary Name: Qraycam PRO Common Name: Laser, Fluorescence Caries Detection Classification Regulation: 21CFR 872.17445 Product Code: NBL Device Class: 2
#### 4. Predicate Devices
| | Predicate Device #1 |
|---------------|--------------------------|
| Manufacturer | INSPEKTOR DENTAL CARE BV |
| Device Name | INSPEKTOR™ PRO |
| 510(k) number | K040063 |
## 5. Description
Qraycam PRO is a fluorescence caries detection aid, designed to assist dentists and dental specialists in identifying caries areas during oral health examinations.
The device produces images that can be displayed on an LCD screen attached to the main body or on a computer monitor.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for AIOBIO. The letters 'AIO' are in gray, and the letters 'BIO' are in a blue square. The logo is simple and modern.
- イ Integral LED that allows the LED to be changed without detaching the body tube
- > Ease of use with automatic focusing
- イ Images can be captured and saved in electronic charts and insurance claims programs
#### 6. Indications for use
The Qraycam PRO is intended to be used as an aid in the diagnosis of dental caries.
## 7. Substantial Equivalence
Qraycam PRO is substantially equivalent to the predicate devices (InspektorTM Pro, K040063). The following table is presented to demonstrate substantial equivalence.
| | Subject Device | Predicate Device | |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Manufacturer | AIOBIO CO.,LTD | INSPEKTOR DENTAL CARE BV | |
| Device Name | Qraycam PRO | INSPEKTORTM PRO | |
| 510(k) number | | K040063 | |
| Classification<br>Product<br>Code /<br>Regulatory Number | NBL<br>872.1745 | NBL<br>872.1745 | |
| Regulatory Class | Class II | Class II | |
| Indications for Use | The Qraycam Pro is intended to be<br>used as an aid in the diagnosis of<br>dental caries. | The InspektorTM Pro is intended to<br>be used as an aid in the diagnosis<br>of dental caries. | |
| Prescription or OTC | Prescription use | Prescription use | |
| Operation | The lens and LED lights mounted<br>on the body tube of the device<br>can be used with a special filter<br>to illuminate the oral cavity and<br>check the state of the patient's<br>oral health. This caries detection<br>device contains its own light<br>source and is used through<br>connection to a desktop PC or<br>laptop in dentistry. | The Inspektor Pro intra-oral<br>camera consists of a control box<br>that is connected via a liquid light<br>guide and electrical cable to the<br>camera handpiece. The control box<br>contains a 35W arc lamp with a<br>violet-blue band-pass filter in front<br>of it. Light passing through the filter<br>is coupled into a liquid light guide<br>for delivery to the tooth surfaces,<br>exposing them to blue light. The<br>thus obtained fluorescence images<br>can be visualised on a monitor<br>real-time, using the miniature CCD<br>camera inside the handpiece.<br>Using the Inspektor Pro software<br>images can easily be captured,<br>stored and analysed | |
| Technology | QLF-D | QLF-D | |
| Image sensor | | - 2.1Megapixel image sensor<br>- CMOS Video image sensor | - 1/4-inch colour CCD sensor |
| Resolution | | - LCD: 3.5inch hdmi monitor<br>Signal Input Port: hdmi<br>1920x1080 60p<br>Physical resolution: 320*480<br>Adjustable resolution: 320x480 -<br>1920x1080<br>Play video: 60Fps high speed<br>support | Play video: 480p ~ 1080p, 50Fps<br>speed support |
| LENS | Type | Variable focus Liquid LENS | Single LENS Panasonic video camera |
| | Focus | Trigger AF (50mm ~ 200mm) | N/A |
| | Focal length | 7.5 mm | None |
| | F number | 2.9 | None |
| | Back Focal Length | 4.07 | None |
| Light<br>Source | White | 4EA White chip LED | None |
| | Blue | 12EA UV chip LED | Short-arc micro-discharge XENON<br>with blue band pass filter |
| Light wavelength | | 405 nm<br>(min: 365nm, max: 410nm) | 404 nm |
| Light<br>Power | White | 3.3W (3.3V, 1000mA) | - |
| | Blue | 1.47W (3.2V-4.2V, 350mA) | 35 W |
| Interface | | PC USB 2.0 | PC USB 2.0 |
| Image analysis | | Qraycam PRO software | INSPEKTOR PRO software |
| Single-use | | Reusable | Reusable |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for AIOBIO. The letters "AIO" are in gray, and the letters "BIO" are in a blue square. The logo is simple and modern.
# Technical Characteristics
1) Same Points between subject device and predicate device
| Item | Description |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | The proposed product code of the subject device is NBL for fluorescence caries<br>detection aid device function. It is the same product code as the predicate device<br>in K040063. |
| Regulatory<br>number and<br>classification | The proposed regulatory number of the subject device is 872.1745 and the<br>classification is Class II. It is same regulatory number and classification as the<br>predicate device in K040063. |
| Indications for<br>Use | The subject device is indicated for use in dental diagnosis procedures for<br>detecting the dental caries using a fluorescence technology (QLF-D<br>technologies). |
| Technology | The same of technology, QLF-D technology to detect caries using a light energy<br>delivered to the teeth. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AIOBIO. The letters "AIO" are in gray, and the letters "BIO" are in white inside of a blue box. The logo is simple and modern.
| Item | Description |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Quantitative light-induced fluorescence (QLF-D) technology is with QLF-D<br>the green auto-fluorescence of tooth structures and the red fluorescence of<br>bacterial metabolites can be assessed. |
| Light source | The similar method of lighting energy delivered to teeth as UV LED(blue light). |
| Blue light<br>wavelength | Images are obtained by illuminating the oral cavity with the blue light<br>(Wavelength 405nm) and filter. And also, images can be analyzed to quantify the<br>dental caries stages by dedicated software. |
| Single Use /<br>Reusable | The subject device is reusable. It is same with the predicate device. |
2) Different Points between subject device and predicate device
| Item | Description |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Light source | The subject device has difference Light source about white LED.<br>White LED is used for the purpose only of illumination and better focusing in the<br>oral cavity.<br>And also, we performed photobiological safety of lamps (LED) test complied with<br>IEC 62471. |
| Oral camera<br>type | The subject device is an Outral Oral Camera, whereas the predicate device is an<br>Intral Oral Camera, presenting some differences. However, both devices share<br>the same intended use and similar technology, allowing for the identification of<br>dental caries at an equivalent level within a limited intraoral scope. Additionally,<br>the manufacturer has specified in the user manual:<br>"Please do not use the device on teeth where visual identification is difficult due<br>to malocclusion."<br>Therefore, it has been confirmed that there is no significant difference in clinical<br>safety regarding the use and intended purpose of the products. |
## 8. Performance data
The following performance data were provided to support the substantial equivalence determination.
## Non-clinical performance data
Non-clinical tests relied on this premarket notification submission for a determination of substantial equivalence include testing showing compliance with the following standards:
- Electromagnetic compatibility (EMC) Testing
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for AIOBIO. The letters "AIO" are in gray, and the letters "BIO" are in white on a blue background. The logo is simple and modern.
- Electrical Safety Testing
IEC 60601-1:2005+A1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance IEC 80601-2-60:2019, Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- Software Verification and Validation IEC 62304:2015, Medical Device Software - Software Life Cycle Processes
- Photobiological Safety Testing IEC 62471:2006, Photobiological safety of lamps and lamp systems
- Non-clinical Bench test
The manufacturer conducted a comparative performance evaluation through bench tests to confirm that there are no significant differences in clinical and intended use outcomes between the Outral Oral Camera and the predicate Intral Oral Camera under the same conditions and using the same teeth.
As a result, the ΔF average value, the performance criterion, was found to meet the standard, confirming the clinical performance equivalence of the product.
- Biocompatibility Testing
No biocompatibility testing was conducted.
- No part of this device comes into contact with the patient's body directly or indirectly.
## 9. Conclusion
In comparing between the subject device and the predicate device, there are Product code, Regulatory number, Classification, Indications for Use, Technology, Light source, Blue light wavelength, IEC 60601-1, IEC 60601-1-2, IEC 62471 and ISO 14971.
Although the Light source and Oral camera type are difference, these differences do not raise different questions of safety and effectiveness.
Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.