K221224 · Blue Belt Technologies, Inc. · OLO · Aug 19, 2022 · Neurology
Device Facts
Record ID
K221224
Device Name
REAL INTELLIGENCE CORI
Applicant
Blue Belt Technologies, Inc.
Product Code
OLO · Neurology
Decision Date
Aug 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Real Intelligence CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Device Story
Robotic-assisted orthopedic surgical navigation and burring system; uses passive infrared tracking camera to map bone landmarks and implant geometry. Surgeon defines bone surface model and surgical plan; system controls robotic bur cutting engagement via exposure control (retracting bur into guard) and speed control (regulating tool control unit) based on proximity to planned surface. Used in hospital/surgical suite by trained medical professionals. Includes CORI Knee TENSIONER, an electromechanical accessory for quantitative assessment of soft tissue laxity during gap planning. Output provides spatial boundaries and navigation data to assist surgeon in bone preparation and implant placement. Benefits include improved accuracy in bone resection and ligament balancing.
Clinical Evidence
Bench testing only. Included biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, IEC 60601-1-2), software verification (IEC 62304), and usability engineering validation (IEC 62366-1). Testing confirmed design inputs met and performance is equivalent to predicates.
Technological Characteristics
Robotic-assisted navigation system using passive infrared tracking. Includes electromechanical TENSIONER for force feedback. Software version 1.7. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, and IEC 62366-1. Materials evaluated per ISO 10993-1.
Indications for Use
Indicated for surgical procedures where stereotactic surgery is appropriate and reference to rigid anatomical bony structures can be determined, including unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip arthroplasty (THA).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
August 19, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Blue Belt Technologies, Inc. Corrine Herlinger Principal Regulatory Affairs Specialist 2905 Northwest Blvd., Ste. 40 Plymouth, Minnesota 55441
Re: K221224
Trade/Device Name: Real Intelligence Cori (Cori) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 21, 2022 Received: July 21, 2022
Dear Corrine Herlinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K221224
Device Name REAL INTELLIGENCE CORI (CORI)
Indications for Use (Describe)
CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined.
These procedures include unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip atthroplasty (THA).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
# 510(k) Summary
| 510(k) Owner | Blue Belt Technologies, Inc.<br>2905 Northwest Blvd Ste. 40<br>Plymouth, MN 55441 USA<br>Tel: (763) 452-4950<br>Fax: (763) 452-4675 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Corrine Herlinger<br>Principal Regulatory Affairs Specialist<br>Tel: (412) 683-3844 x 4128<br>Email: corrine.herlinger@smith-nephew.com |
| Date of Submission | April 26, 2022 |
| Classification Reference | 21 CFR 882.4560 |
| Product Code | OLO |
| Supported Codes | HSX, JWH, MBH |
| Common/Usual Name | Orthopedic Stereotaxic Instrument |
| Trade/Proprietary Name | REAL INTELLIGENCE® CORI® (CORI) |
| Primary Predicate Device | REAL INTELLIGENCE® CORI® (K220255) |
| Secondary Predicate Device | OMNIBotics Knee System (K200888) |
| Reason for Submission | This Traditional 510(k) submission supports an<br>update to CORI to allow the system to be used with<br>the CORI Knee TENSIONER (TENSIONER) and the<br>associated software upgrade, CORI 1.7. |
K221224 - Page 1 of 8
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.
### Intended Use
Real Intelligence CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
## Indications for Use
CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip arthroplasty (THA).
## Device Description
The subject of this Traditional 510(k) is REAL INTELLIGENCE CORI (CORI), a robotic-assisted orthopedic surgical navigation and burring system. CORI uses established technologies of navigation via a passive infrared tracking camera. Based on intraoperatively-defined bone landmarks and known geometry of the surgical implant, the system aids the surgeon in establishing a bone surface model for the target surgery and planning the surgical implant location. For knee applications, CORI then aids the surgeon in executing the surgical plan by controlling the cutting engagement of the surgical bur.
CORI knee application software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved with two modes:
- . Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the robotic system retracts the bur inside the guard, disabling cutting.
- Speed control regulates the signal going to the tool control unit itself and limits the speed of the drill if the target surface is approached.
Alternatively, the surgeon can disable both controls and operate the robotic drill as a standard navigated surgical drill.
K221224 - Page 2 of 8
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
## Currently Supported Knee Implants
The following Smith+Nephew knee implants are supported on CORI:
### Table 1: Currently Supported Smith+ Nephew Knee Implants
| Implant Model Name | 510(k) Number | Classification<br>Product Code |
|--------------------------------------------|---------------------------|--------------------------------|
| STRIDE Unicondylar Knee | K123380 | HSX |
| ZUK Select Knee System | K160738 | HSX |
| JOURNEY II Unicompartmental Knee System | K191211 | HSX |
| JOURNEY UNI | K102069 | HSX |
| JOURNEY II CR | K121443 | JWH |
| JOURNEY II BCS | K111711 | JWH |
| JOURNEY II XR | K141471, K152726 | JWH |
| LEGION CR/PS | K951987, K962557, K093746 | JWH |
| LEGION Porous CR Femoral Components | K073325, K091543 | MBH |
| LEGION Porous CR Narrow Femoral Components | K210566 | MBH |
| LEGION Porous Tibia | K100897 | MBH |
| Porous Tibia Baseplate | K211221 | MBH |
| GENESIS II CR/PS | K951987, K962557 | JWH |
| ANTHEM | K142807 | JWH |
## Discussion of Similarities and Differences
This Traditional 510(k) submission supports an update to the CORI System to allow the system to be used with the CORI Knee TENSIONER (TENSIONER) and the associated software upgrade, CORI 1.7. The TENSIONER is an electromechanical accessory designed to provide feedback to CORI with respect to the force that the user applies to the joint during gap assessment. It is used by the surgeon in the gap planning stages of a total knee replacement procedure to provide a quantitative assessment of soft tissue laxity.
The modifications made to CORI do not impact the system's intended use, indications for use, or fundamental scientific technology.
Blue Belt Technologies believes that CORI is subject to premarket notification requirements under Section 510(k) of the Federal Food, Drug and Cosmetic Act ("FFDCA" or "the Act") and is substantially equivalent to the previously cleared REAL INTELLIGENCE CORI system, K220255, and the OMNIBotics Knee System, K200888.
K221224 - Page 3 of 8
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. The logo is clean and modern, with a focus on the company name.
### Table 2: Predicate Device
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|------------------------------|------------------------|----------------------|----------------|
| Blue Belt Technologies, Inc. | REAL INTELLIGENCE CORI | K220255 | 03/29/2022 |
| Corin Ltd. | OMNIBotics Knee System | K200888 | 06/27/2020 |
### Table 3: Summary of Technological Similarities with Predicate
| Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use and<br>Indications for Use<br>[Same] | Same as primary<br>predicate. | REAL INTELLIGENCE CORI (CORI)<br>is intended to assist the surgeon<br>in providing software-defined<br>spatial boundaries for<br>orientation and reference<br>information to anatomical<br>structures during orthopedic<br>procedures. | The OMNIBotics Knee<br>System is indicated for<br>stereotaxic surgery to aid the<br>surgeon in locating<br>anatomical structures and<br>aligning endoprostheses with<br>these anatomical structures<br>during Total Knee<br>Arthroplasty. |
| | | CORI is indicated for use in<br>surgical knee procedures in<br>which the use of stereotactic<br>surgery may be appropriate, and<br>where reference to rigid<br>anatomical bony structures can<br>be determined. These<br>procedures include unicondylar<br>knee replacement, total knee<br>arthroplasty, and total hip<br>arthroplasty. | The BalanceBot™ is indicated<br>as a tool for adjustment of<br>soft tissue and the femoral<br>implant to reduce instability<br>from flexion gap asymmetry.<br>The OMNIBotics Knee<br>System supports OMNI Apex<br>Knee™ implants and CORIN<br>Unity Knee™ implants. |
| Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 |
| Technological<br>Characteristics –<br>Operating Principle<br>[Same] | Same as primary<br>predicate. | For knee applications, CORI uses<br>established technologies to<br>prepare bone for attachment of<br>UKR and TKA implant<br>components. In the case of a<br>total knee arthroplasty, the<br>bone surface may also be<br>prepared to receive the femoral<br>and tibial cutting guides with<br>final bone surface for receiving<br>the implant prepared using a<br>standard surgical saw.<br><br>CORI uses intraoperative data<br>collection (image-free or non-CT<br>data generation) to create a<br>model of the patient's femur<br>and/or tibia, dependent on the<br>procedure being performed, and<br>allows the surgeon to prepare a<br>surgical plan.<br><br>The system uses predefined<br>boundaries generated during<br>the planning process to control<br>the motion of the surgical bur<br>and limit the amount of bone<br>removed in order to shape the<br>condyles or tibial plateau in<br>preparation for placement of<br>the surgical implant.<br><br>Bur cutting is controlled either<br>by retracting the bur in a guard,<br>or by controlling the speed of<br>the bur as the target surface is<br>approached.<br><br>To support the hip application,<br>CORI uses a Virtual Machine<br>with hypervisor to enable the<br>Windows-based HIP7 software<br>to run on CORI. | N/A |
| Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 |
| Technological<br>Characteristics -<br>Tensioning<br>Instrument<br>[Substantially<br>Equivalent] | The TENSIONER is an<br>electromechanical<br>tensioning<br>instrument with<br>reverse-action pliers<br>to distract the joint. | Consists of a Gap Planning stage<br>to assist users with (manual)<br>ligament balancing. | The BalanceBot is an electro-<br>mechanical tensioning<br>instrument with integrated<br>force sensors and position<br>actuators. |
| Technological<br>Characteristics -<br>Gap Balancing<br>[Substantially<br>Equivalent] | The TENSIONER is<br>used during the Gap<br>Planning stage of the<br>CORI Total Knee<br>Arthroplasty<br>application and<br>provides feedback to<br>CORI with respect to<br>the force that the<br>user applies to the<br>joint to assist with<br>ligament balancing. | Consists of a Gap Planning stage<br>to assist users with (manual)<br>ligament balancing. | BalanceBot assists the<br>surgeon with ligament<br>balancing during total knee<br>arthroplasty. |
| Knee Implant<br>Product Codes<br>Supported<br>[Same] | Same as primary<br>predicate. | HSX, JWH, MBH | N/A |
| Environment of Use<br>[Same] | Same as primary<br>predicate. | CORI is intended to be used by<br>trained medical professionals in<br>a hospital or clinical setting<br>equivalent to an orthopedic<br>surgery suite. | N/A |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo features a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
K221224 - Page 5 of 8
ReshapingMobility™
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a stylized blue swirl on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
K221224 - Page 6 of 8
ReshapingMobility™
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
## Non-Clinical Testing (Bench)
Design verification and validation testing demonstrated that CORI meets all design requirements and is as safe and effective as its primary predicate device (K220255) and secondary predicate device OMNIBotics Knee System (K200888). Comprehensive performance testing demonstrated that the system meets required design inputs. Performance data consisted of physical performance testing for all system components. Additionally, the following evidence was provided:
- . Biocompatibility evaluation demonstrating that the system satisfies the requirements of ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process.
- . Safety and Electromagnetic Compatibility (EMC) testing demonstrating that the device complies with IEC 60601-1 Medical Electric Equipment – Part 1: General Requirements for Basic Safety and Essential Performance and IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests.
- . Software verification testing, including software integration and workflow testing, was completed. Software was developed in accordance with IEC 62304 Medical device software -Software life cycle processes, and this submission contains documentation per the requirements of FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
- . Usability Engineering Validation Testing demonstrating that representative users were able to safely and effectively use CORI and TENSIONER in a simulated use environment. Human factors and usability engineering processes were followed per IEC 62366-1:2015+A1:2020 Application of Usability Engineering to Medical Devices.
All verification and validation testing concluded with acceptable results. Based on these test results, Blue Belt Technologies has concluded that all design inputs have been met and that the design verification and validation testing performed did not raise any new questions of safety or effectiveness.
## Conclusions
The subject device, CORI, described in this submission has the same intended use and the same fundamental scientific technology as the primary predicate device, K220255. The primary difference between the subject device and CORI (K220255) is the update to allow the system to be used with TENSIONER and the associated software upgrade, CORI 1.7. The main functionality of the Total Knee Arthroplasty (TKA) software is the same as presented in K201022, with modifications to allow for communication with TENSIONER. The addition of TENSIONER does not change the system's method for gap calculation, which remains the same as the previously cleared devices.
While the technological characteristics differ between manual and a tool-assisted gap-balancing measurement, the questions regarding accuracy and implementation of intraoperative gap-balancing
K221224 - Page 7 of 8
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
apply to both the subject and primary predicate device. Additionally, usability testing demonstrated that users are able to successfully perform gap balancing using TENSIONER and the CORI system; therefore, the difference of the technological characteristics does not introduce new questions of safety or effectiveness.
Blue Belt Technologies believes that FDA can find CORI to be substantially equivalent to the predicate devices, REAL INTELLIGENCE CORI system, K220255, and the OMNIBotics Knee System, K200888. The information presented in this 510(k) premarket notification demonstrates that CORI is as safe and effective as the primary predicate CORI system (K220255).
K221224 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.