PRECISION ECO Compostable / Plant Based Procedural Mask with Earloops
K221209 · Padm Medical, Inc. · FXX · Mar 8, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221209
Device Name
PRECISION ECO Compostable / Plant Based Procedural Mask with Earloops
Applicant
Padm Medical, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Mar 8, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended to be worn to protect both the patient and healthcare professional from transfer of microorganisms, body fluids, and particulate material. The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, non-sterile, disposable device.
Device Story
Single-use, non-sterile, ASTM Level 3 procedural mask; four-layer flat pleated design with earloops and malleable aluminum nose piece. Constructed from polylactide resin nonwoven fabrics (spunbond and meltblown). Used in clinical settings by healthcare professionals and patients to provide a physical barrier against microorganisms, body fluids, and particulates. Components bonded via ultrasonic welding. No drugs, biologics, or nanoparticles included. Device provides protection via filtration and fluid resistance; helps reduce infection transmission risk.
Clinical Evidence
Bench testing only. Performance evaluated against ASTM F2101-19 (BFE ≥ 98%), ASTM F2299 (PFE ≥ 98% at 0.1 µm), ASTM F1862 (160 mm Hg blood penetration), EN 14683 (differential pressure < 6.0 mm H2O/cm²), and 16 CFR Part 1610 (Class 1 flammability). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). Accelerated aging conducted per ASTM 1980-21.
Technological Characteristics
Four-layer construction using polylactide resin nonwoven fabrics; aluminum nose piece; polyester/spandex earloops. Ultrasonic welding assembly. ASTM Level 3 performance. Non-sterile, disposable. No electronic components or software.
Indications for Use
Indicated for use by patients and healthcare professionals to reduce the transfer of microorganisms, body fluids, and particulate material, and to reduce potential exposure to blood and body fluids in infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Altor Safety 4-Ply Surgical Mask (Model: 62232) (K211762)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 11, 2023
PADM Medical Inc. Pablo Batista Vice President of Engineering Operations 1595 Buffalo Place - Unit A Winnipeg, Manitoba R3T 1L9 Canada
Re: K221209
Trade/Device Name: PRECISION ECO Compostable / Plant Based Procedural Mask with Earloops Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX
Dear Pablo Batista:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 8, 2023. Specifically, FDA is updating this SE Letter for a typo in the trade name in the Indications for Use Form as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Brent Showalter, Ph.D., OHT6: Office of Orthopedic Devices, (240) 402-1840, brent.showalter@fda.hhs.gov.
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
March 8, 2023
PADM Medical Inc. Pablo Batista Vice President of Engineering and Operations 1595 Buffalo Place - Unit A Winnipeg, Manitoba R3T 1L9 Canada
Re: K221209
Trade/Device Name: PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 12, 2023 Received: January 17, 2023
Dear Pablo Batista:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K221209
K221209 Page 1 of 1
Device Name
PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops
Indications for Use (Describe)
The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended to be worn to protect both the patient and healthcare professional from transfer of microorganisms, body fluids, and particulate material. The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, non-sterile, disposable device.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Designation Under Part 61 CFR 901 Subpart F | <input type="checkbox"/> |
|----------------------------------------------|-------------------------------------|
| On-The-Ground Use Under 49 CFR 901 Subpart G | <input checked="" type="checkbox"/> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date of Summary Prepared: | March 3rd, 2023 |
|------------------------------|---------------------------------------------------------------|
| 510(K) Number: | K221209 |
| Applicant: | PADM Medical Inc. |
| | Unit A - 1595 Buffalo Place |
| | Winnipeg, MB R3T 1L9 |
| | Canada |
| Primary Contact: | Pablo Batista |
| | Phone: +1 (204) 289 4491 x233 |
| | Email: pbatista@precisionadm.com |
| Device Name | PRECISION ECO™ Compostable / Plant Based Procedural Mask with |
| | Earloops |
| Model Number | Black: 100195 |
| | White: 100196 |
| Device Classification Name | Mask, Surgical |
| Trade Name | PRECISION ECO™ Compostable Procedural Mask with Earloops |
| | PRECISION ECO™ Plant Based Procedural Mask with Earloops |
| Classification | Class II 21 CFR 878.4040 |
| Regulatory Medical Specialty | General and Plastic Surgery |
| Product Code | FXX |
| Primary Predicate Device | K211762 |
| | Company Name: Altor Safety |
| | Device Name: Altor Safety 4-Ply Surgical Mask (Model: 62232) |
## Indications for Use
The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended to be worn to protect both the patient and healthcare professional from transfer of microorganisms, body fluids, and particulate material. The PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, non-sterile, disposable device.
## Device Description
The PRECISION ECO™ Compostable / Plant Based Procedural Masks with Earloops are single use, ASTM Level 3 disposable devices, provided non-sterile. The mask has a four-layer, flat, pleated body with two earloops and a malleable nose piece. All components of the mask design are bonded together via ultrasonic welding, and it does not include any drugs, biologics, or nanoparticles. The mask comes in one configuration with two (2) different color schemes: black (100195), and white (100196).
TABLE / lists all materials used in the mask.
{5}------------------------------------------------
#### TABLE /: MATERIALS
| Raw Material | Product Details |
|------------------------------------------------------------------------|----------------------------------------------------|
| Inner Layer<br>Applicable to both models: 100195 and 100196 | Spunbond nonwoven polylactide resin fabric |
| Middle Layer<br>Applicable to both models: 100195 and 100196 | Meltblown nonwoven polylactide resin fabric |
| Outer Layers (Qty 2)<br>(White)<br>Applicable to white product: 100196 | Spunbond nonwoven natural polylactide resin fabric |
| Outer Layers (Qty 2)<br>(Black)<br>Applicable to black product: 100195 | Spunbond nonwoven black polylactide resin fabric |
| Nose Piece<br>Applicable to both models: 100195 and 100196 | Aluminum |
| Ear Loops<br>(White)<br>Applicable to white product: 100196 | Round woven polyester/spandex |
| Ear Loops<br>(Black)<br>Applicable to black product: 100195 | Round woven polyester/spandex |
The following TABLE // outlines a comparison of the materials used for each model number (color) of mask. As evident from the materials comparison, the only difference between each device is the colorant used in the outer layers and earloops for each model.
#### TABLE //: COMPARISON BLACK (100195) AND WHITE (100196) MODEL(S)
| Raw Material | Black (100195) | White (100196) |
|------------------------------|-------------------------------------------------------------|---------------------------------------------|
| Inner Layer | Spunbond nonwoven polylactide resin<br>fabric | Spunbond nonwoven polylactide resin fabric |
| Middle Layer | Meltblown nonwoven polylactide resin<br>fabric | Meltblown nonwoven polylactide resin fabric |
| Outer Layers<br>(Quantity 2) | Spunbond nonwoven polylactide resin<br>fabric with colorant | Spunbond nonwoven polylactide resin fabric |
| Nose Piece | Aluminum | Aluminum |
| Ear Loops | Round woven polyester/spandex with<br>colorant | Round woven polyester/spandex<br> |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows a diagram of a face mask with its different layers. The inner layer is white, the middle layer is meltblown, and the outer layers are black. There are two outer layers labeled as 'Outer Layer 1 Black' and 'Outer Layer 2 Black'.
FIGURE 1: DEVICE REPRESENTATIVE ILLUSTRATION (Black product 100195) Instructions for use for the mask are located on the box artwork and read as follows. Hold mask by ear loops with the nose clip at the top. Ensure the logo side or colored side is facing outward. Place one ear loop around each ear. Bend the nose clip to match the shape of the nose and ensure there are no gaps. While holding the nose clip at the bridge of your nose, pull the mask down under your chin. Properly dispose of the mask by touching only the ear loops
## Technical Comparison to Primary Predicate Device
TABLE /// provides a classification comparison between the PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops and the Primary Predicate Device manufactured by Altor Safety.
| Characteristic | Subject Device | Primary Predicate Device | Result |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Manufacturer | PADM Medical Inc. | Altor Safety | N/A |
| Device Name | Surgical Mask | Surgical Mask | N/A |
| Model | Black: 100195<br>White: 100196 | 62232 | N/A |
| 510K | Pending. | K211762 | N/A |
| Device Class,<br>product code,<br>regulation # | Class II, FXX (21 CFR878.4040) | Class II Device, FXX (21<br>CFR878.4040) | Same |
| Intended Use | The PRECISION ECOTM<br>Compostable / Plant Based<br>Procedural Mask with Earloops is<br>intended to be worn to protect both<br>the patient and healthcare<br>professional from transfer of<br>microorganisms, body fluids, and<br>particulate material. The<br>PRECISION ECOTM Compostable /<br>Plant Based Procedural Mask with<br>Earloops is intended for use in<br>infection control practices to reduce | The Altor Safety 4-Ply Surgical<br>Mask is intended to be worn to<br>protect both the patient and<br>healthcare professional from<br>transfer of microorganisms,<br>body fluids, and particulate<br>material. The Altor Safety 4-Ply<br>Surgical Mask is intended for<br>use in infection control practices<br>to reduce the potential exposure<br>to blood and body fluids. This is | Same |
|--|
{7}------------------------------------------------
| the potential exposure to blood and<br>body fluids. This is a single use,<br>non-sterile, disposable device. | a single use, non- sterile,<br>disposable device. |
|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
TABLE /V provides a comparison of materials used in the PRECISION ECO™ Compostable / Plant Based Procedural Mask with Earloops and the Primary Predicate Device manufactured by Altor Safety.
| Characteristic | Subject Device | Primary Predicate Device | Result |
|---------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------|-----------|
| Color | Black/White | Blue | Different |
| Disposable | Yes | Yes | Same |
| Mask Style | Flat pleated | Flat pleated | Same |
| Outer Layer 1<br>Material | Polylactide resin spunbond<br>nonwoven fabric | Spunbond polypropylene<br>(SBPP) | Different |
| Outer Layer 2<br>Material | Polylactide resin spunbond<br>nonwoven fabric | Spunbond polypropylene<br>(SBPP) | Different |
| Filter Layer Material | Polylactide resin meltblown<br>nonwoven | Meltblown Polypropylene<br>(MBPP) | Different |
| Inner Layer Material | Polylactide resin spunbond<br>nonwoven fabric | Spunbond Polypropylene<br>(SBPP) | Different |
| Nose Piece Material | Aluminum | Virgin polyethylene plastic 24<br>gauge soft annealed carbon<br>steel, and kraft paper | Different |
| Earloop Material | Polyester/ spandex | Spandex/nylon | Similar |
| Earloop Style | Round knitted earloop | Round knitted earloop | Same |
| Mask Body L X W | (170mm-180mm) x<br>(90mm-100mm) | (171.45mm-177.8mm) x<br>(92mm) | Similar |
| Earloop | Yes | Yes | Similar |
## TABLE /V: MATERIALS COMPARISON TO PRIMARY PREDICATE DEVICE(S)
The difference in materials, nose piece material, and mask dimensions does not impact the safety and effectiveness of the subject device as the subject device's performance is equivalent to the Primary Predicate Device. As evident in TABLE IV, the base materials and configuration for each product color (black/white) are identical. The colorant is not influencing the performance of the device, due to the extremely low concentration present in the device.
The performance requirements and test criteria are listed in TABLE V. The performance testing was conducted on production lots and on artificially aged samples. Accelerated aging was conducted per ASTM 1980-21: Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices. The device met the acceptance criteria for both production lots and artificially aged samples.
{8}------------------------------------------------
| Characteristic | Test Method | Acceptance Criteria |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Bacterial filtration<br>efficiency (BFE) | ASTM F2101-19: Standard Test Method for Evaluating the<br>Bacterial Filtration Efficiency (BFE) of Medical Face Mask<br>Materials, Using a Biological Aerosol of Staphylococcus<br>aureus | ≥ 98% |
| Sub-micron<br>particulate<br>efficiency at 0.1 µm<br>(PFE) | ASTM F2299: Standard Test Method for Determining the<br>Initial Efficiency of Materials Used in Medical Face Masks to<br>Penetration by Particulates Using Latex Spheres | ≥ 98% |
| Blood Penetration | ASTM F1862: Standard Test Method for Resistance of<br>Medical Face Masks to penetration by Synthetic Blood<br>(Horizontal Projection of Fixed Volume at a Known Velocity) | @160 mm Hg |
| Differential<br>Pressure | EN 14683:2019 Annex C: Medical face masks -<br>Requirements and test methods | < 6.0 mm H2O/cm² |
| Flammability | 16 CFR Part 1610: Standard for the Flammability of<br>Clothing Textiles | Class 1 (≥3.5 s flame<br>spread) |
| Cytotoxicity | ISO10993-5: Biological evaluation of medical devices -Part<br>5: Tests for invitro cytotoxicity. | ≤ Grade 2 |
| Irritation | ISO10993-10: Biological evaluation of medical devices -<br>Part 10: Tests for irritation and delayed-type<br>hypersensitivity. | Non-irritant |
| Sensitization | ISO10993-10: Biological evaluation of medical devices -<br>Part 10: Tests for irritation and delayed-type<br>hypersensitivity. | Non-sensitizer |
## TABLE V: STANDARDS AND ACCEPTANCE CRITERIA
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and is substantially equivalent to the legally marketed Primary Predicate Device [K211762].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.