ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System
K221190 · Siemens Medical Solutions USA, Inc. · IYN · Aug 16, 2022 · Radiology
Device Facts
Record ID
K221190
Device Name
ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Aug 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ACUSON Juniper ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. The ACUSON Juniper Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
Device Story
Mobile, software-controlled diagnostic ultrasound systems; acquire harmonic ultrasound echo data; display in B-Mode, M-Mode, PW/CW Doppler, Color Doppler, Tissue Imaging, 3D/4D volume imaging on flat panel display. Operated by trained healthcare professionals in clinical settings. Includes analysis packages (e.g., Arterial Health Package for Intima Media Thickness) to assist physician clinical diagnosis. Provides visual tools for patient communication regarding cardiovascular health. System includes thermal/mechanical indices for bio-effect monitoring. Benefits include non-invasive anatomical visualization and quantitative analysis to support clinical decision-making.
Clinical Evidence
No clinical studies were required or submitted to support substantial equivalence. Bench testing only.
Technological Characteristics
Mobile ultrasound system; flat panel display; supports multiple transducers. Complies with IEC 62359 (acoustic output), AAMI ES60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasonic diagnostic equipment), and ISO 10993-1 (biocompatibility). Connectivity includes DICOM 3.0. Software-controlled with various imaging modes (2D, Doppler, M-mode, 3D/4D).
Indications for Use
Indicated for use by trained healthcare professionals for diagnostic ultrasound imaging and measurement in abdominal, obstetric, gynecologic, small parts, pediatric, neonatal, vascular, urologic, echocardiographic, musculoskeletal, and intraoperative applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
August 16, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
Siemens Medical Solutions USA, Inc. % Sulgue Choi Regulatory Affairs 22010 South East 51st Street ISSAQUAH WA 98029
## Re: K221190
Trade/Device Name: ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OIJ Dated: July 21, 2022 Received: July 22, 2022
# Dear Sulgue Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Michael D. O'Hara, Ph.D. Deputy Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221190
#### Device Name
ACUSON Juniper Diagnostic Ultrasound System, ACUSON Juniper Select Diagnostic Ultrasound System
### Indications for Use (Describe)
For ACUSON Juniper
The ACUSON Juniper ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
#### For ACUSON Juniper Select
The ACUSON Juniper Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography and Musculoskeletal applications using different ultrasound transducers for different applications.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
Operating Modes 2D-Mode · 2D-Mode with Harmonic Imaging Color flow Doppler · Color (Velocity) · Power (Energy) Pulsed Wave Doppler · Pulsed Wave Doppler Tissue Imaging · High Pulsed Repetition Frequency Pulsed Wave Doppler Continuous Wave Doppler · Steerable Continuous Wave Doppler for imaging transducers · Auxiliary Continuous Wave Doppler for pencil transducers M Mode · M-Mode with Harmonic Imaging · Anatomical M-mode 3D/4D Volume Imaging
{3}------------------------------------------------
#### Combined Modes
· 2D-Mode with Color
- 2D-Mode with Doppler
- · 2D-Mode with Color and Doppler
- · 2D-Mode with M-Mode
- 2D-Mode with M-Mode and Color
- · 2D-Mode with Elastography*
- 2D-Mode with Contrast Agent Imaging
- · 2D/Anatomical M-mode
- *Available only for the ACUSON Juniper system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date: | Jul 12, 2022 | | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------|--------|
| 1. Sponsor: | Siemens Medical Solutions USA, Inc.,<br>Ultrasound Division<br>22010 South East 51st Street<br>Issaquah, WA 98029 | | |
| Contact<br>Person: | Sulgue Choi<br>Tel: (425) 281-9898 | | |
| Secondary<br>Contact<br>Person: | Bongsoo Cho<br>Tel: +82 10 4697 2268<br>E-mail: bongsoo.cho@siemens-healthineers.com | | |
| 2. Device<br>Name: | ACUSON Juniper Diagnostic Ultrasound System<br>ACUSON Juniper Select Diagnostic Ultrasound System | | |
| Common<br>Name: | Diagnostic Ultrasound System with Accessories | | |
| Classification: | Regulatory Class: II<br>Review Category: Tier II<br>Classification Panel: Radiology | | |
| | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| | Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYC |
| | Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
| | Biopsy needle guide Kit | 892.1560 | 90-OIJ |
| Manufacturing<br>Site: | Siemens Healthineers Ltd.<br>2nd -3rd floor, 143, Sunhwan-ro,<br>Jungwon-gu, Seongnam-si, Gyeonggi-do,<br>Republic of Korea | | |
#### 510(k) Summary K221190
#### 3. Legally Marketed Predicate Devices
The ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System are the multi-purpose diagnostic ultrasound systems with accessories and proprietary software and is substantially equivalent to the company's own products.
- -Primary Predicate Device: ACUSON Juniper(K201130)
- Reference Device: ACUSON Redwood(K210743) -
{5}------------------------------------------------
## 4. Device Description
The ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System are the multi-purpose mobile, software controlled, diagnostic ultrasound systems with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Their function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Doppler Tissue Image, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging and 3D Imaging, or Harmonic Imaging and 4D imaging on a Flat Panel Display.
### 5. Intended Use/Indications for Use
### For ACUSON Juniper
The ACUSON Juniper ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric. Neonatal. Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
### For ACUSON Juniper Select
The ACUSON Juniper Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Uroloqy, Echocardiography and Musculoskeletal applications using different ultrasound transducers for different applications.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
### Operating Modes
2D-Mode 2D-Mode with Harmonic Imaging . Color flow Doppler
- Color (Velocity) .
- Power (Energy) .
- Pulsed Wave Doppler
- Pulsed Wave Doppler Tissue Imaging .
{6}------------------------------------------------
High Pulsed Repetition Frequency Pulsed Wave Doppler . Continuous Wave Doppler
- Steerable Continuous Wave Doppler for imaging transducers .
- Auxiliary Continuous Wave Doppler for pencil transducers .
M Mode
- . M-Mode with Harmonic Imaging
- . Anatomical M-mode
3D/4D Volume Imaging
Combined Modes
- 2D-Mode with Color
- 2D-Mode with Doppler ●
- 2D-Mode with Color and Doppler .
- 2D-Mode with M-Mode ●
- 2D-Mode with M-Mode and Color ●
- 2D-Mode with Elastography* .
- 2D-Mode with Contrast Agent Imaging ●
- . 2D/Anatomical M-mode
*Available only for the ACUSON Juniper system.
## 6. Summary of Technological Characteristics and Determination of Substantial Equivalence
The ACUSON Juniper Diagnostic Ultrasound System and ACUSON Juniper Select Diagnostic Ultrasound System are substantially equivalent to the company's own products that are already cleared for US distribution. The predicate devices of ACUSON Juniper & ACUSON Juniper Select are ACUSON Juniper (K201130) and ACUSON Redwood (K210743) with reqards to intended use, indications for use, technological characteristics (Transducers, accessories and software features) and safety and effectiveness.
The submission device differs from the predicated devices as following:
1) ACUSON Juniper:
(1) Transducer & Accessories
- The addition of the 6C1, which is substantially equivalent to the 5C1 transducer ● previously cleared on the submission device, ACUSON Juniper (K201130).
- . The addition of the 14L4, which is substantially equivalent to the 13L4 transducer previously cleared on the ACUSON Juniper (K201130).
- The addition of the CW8, which is substantially equivalent to the CW5 ● transducer previously cleared on the ACUSON Juniper (K201130).
- The addition of needle guide kit, 14L4, which is substantially equivalent to the ● needle guide Kit, 13L4 previously cleared on the ACUSON Juniper (K201130).
- Addition of a battery pack and BW printer ●
(2) Software Feature:
- The addition of 'Contrast Agent Imaging' clinical application, which was already cleared on the ACUSON Redwood(K210743)).
- 2) ACUSON Juniper Select: The proposed ACUSON Juniper Select systems is substantially equivalent to ACUSON Juniper (K201130) with regards to intended use, Indications for use, technological characteristics (Transducers, accessories and
{7}------------------------------------------------
software features) and safety and effectiveness. All of technological characteristics are migrated (identical SW & HW platform) from the predicate device, ACUSON Juniper (K201130), and there is no new feature or transducer compared to the predicate.
The ACUSON Juniper and ACUSON Juniper Select are substantially equivalent to the predicate devices with regard to the intended use, indications for use and technical characteristics.
| | ACUSON Juniper<br>This submission | ACUSON Juniper<br>Select<br>This submission | ACUSON Juniper<br>(K201130)<br>Predicate device | ACUSON Redwood<br>(K210743)<br>Reference device |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|---------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| Indications for Use: | | | | |
| ■ Abdominal<br>(Abdominal, Abdominal<br>Difficult, Renal, Bowel,<br>Focused Assessment<br>with Sonography for<br>Trauma, Lung) | √ | √ | √ | |
| ■ Obstetrics<br>(Obstetrics, Early<br>Obstetrics, Fetal<br>Echocardiography,<br>Advanced Obstetrics) | √ | √ | √ | |
| ■ Gynecology<br>(Gynecology, Pelvic<br>Floor) | √ | √ | √ | |
| ■ Small Parts<br>(Breast, Testicles,<br>Penile, Thyroid) | √ | √ | √ | |
| ■ Pediatric<br>(Pediatric Hip Joint,<br>Pediatric Abdomen) | √ | √ | √ | |
| ■ Neonatal<br>(Neonatal Head) | √ | | √ | |
| ■ Vascular<br>(Carotid, Peripheral<br>Venous, Peripheral<br>Arterials, Transcranial<br>Doppler) | √ | √ | √ | |
| ■ Urology<br>(Pelvis, Prostate) | √ | √ | √ | |
| ■ Echocardiography<br>(Adult<br>Echocardiography,<br>Pediatric<br>Echocardiography,<br>Neonatal<br>Echocardiography,<br>Trans-esophageal<br>Echocardiography) | √ | √ | √ | |
| ■ Musculoskeletal<br>(Spine,<br>Musculoskeletal,<br>Digital, Nerve) | √ | √ | √ | |
| | ACUSON Juniper<br>This submission | ACUSON Juniper<br>Select<br>This submission | ACUSON Juniper<br>(K201130)<br>Predicate device | ACUSON Redwood<br>(K210743)<br>Reference device |
| ■<br>Intraoperative<br>(Intraoperative<br>Abdomen,<br>Intraoperative Vascular) | √ | | √ | |
| Modes: | | | | |
| ■<br>2D (Brightness mode) | √ | √ | √ | |
| ■<br>C (Color Flow Doppler) | √ | √ | √ | |
| ■<br>D (Doppler) | √ | √ | √ | |
| ■<br>M (Motion Mode) | √ | √ | √ | |
| ■<br>CW (Continuous Waver<br>Doppler) | √ | √ | √ | |
| Features: | | | | |
| ■<br>DICOM (3.0<br>Connectivity, Worklist,<br>MPPS) | √ | √ | √ | |
| ■<br>DICOM SR OB/GYN | √ | √ | √ | |
| ■<br>DICOM SR Cardiac | √ | √ | √ | |
| ■<br>DICOM SR Vascular | √ | √ | √ | |
| ■<br>syngo Arterial Health<br>Package (AHP) | √ | √ | √ | |
| ■<br>syngo Auto Follicle | √ | √ | √ | |
| ■<br>syngo Auto OB | √ | √ | √ | |
| ■<br>eSie Left Heart (eSie<br>LH) | √ | | √ | |
| ■<br>Stress Echo | √ | | √ | |
| ■<br>3-Scape Imaging | √ | √ | √ | |
| ■<br>fourSight 4D | √ | √ | √ | |
| ■<br>Advanced fourSight 4D<br>Imaging | √ | √ | √ | |
| ■<br>eSie Touch Elasticity<br>Imaging | √ | | √ | |
| ■<br>eSie Measure Workflow<br>Acceleration Package | √ | √ | √ | |
| ■<br>eSieScan Workflow<br>Protocol | √ | √ | √ | |
| ■<br>Virtual Touch<br>Quantification (VTQ) | √ | | √ | |
| ■<br>SieScape Panoramic<br>Imaging | √ | √ | √ | |
| ■<br>Dynamic Tissue<br>Contrast Enhancement | √ | √ | √ | |
| | ACUSON Juniper<br>This submission | ACUSON Juniper<br>Select<br>This submission | ACUSON Juniper<br>(K201130)<br>Predicate device | ACUSON Redwood<br>(K210743)<br>Reference device |
| (DTCE) | | | | |
| ■ Advanced SieClear<br>(ASC) | √ | √ | √ | |
| ■ Clarify Vascular<br>Enhancement (Clarify<br>VE) | √ | √ | √ | |
| ■ eSielmage (TEQ/TGO) | √ | √ | √ | |
| ■ Anatomical M-mode<br>(AMM) | √ | √ | √ | |
| ■ HD Zoom (Res) | √ | √ | √ | |
| ■ Needle Visualization | √ | | √ | |
| ■ Tissue Harmonic<br>Imaging (THI) | √ | √ | √ | |
| ■ Doppler Tissue Imaging<br>(DTI) | √ | √ | √ | |
| ■ Custom Tissue Imaging<br>(CTI) | √ | √ | √ | |
| ■ eSieCalcs | √ | √ | √ | |
| ■ Wireless | √ | √ | √ | |
| ■ Veterinary(VET)<br>Imaging | √ | √ | √ | |
| ■ US Security (Virus<br>Protection) | √ | √ | √ | |
| ■ VVI(Velocity Vector<br>Image) | √ | | √ | |
| ■ LVO(Left Ventricular<br>Opacification) | √ | √ | √ | |
| ■ Contrast Agent Imaging | √ | | - | √ |
{8}------------------------------------------------
### ACUSON Juniper & Juniper Select Diagnostic Ultrasound System 510(k) Submission
{9}------------------------------------------------
### ACUSON Juniper & Juniper Select Diagnostic Ultrasound System 510(k) Submission
# 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- IEC 62359:2010 /A1(2017), Ultrasonic - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic field / This document and its separate amendments continue to be valid together with the consolidation version.
- Safety and EMC Requirements for Medical Equipment
{10}------------------------------------------------
- AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and O A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005)
- IEC 60601-1:2005/A1(2012), Medical electric equipment Part 1: General O requirements for basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment Part 1-2: o General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-18 Edition 3.0 2009-08. Medical electrical equipment Part 2-18: O Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment Part 2-37: O Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- . ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
## 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
The subject of this premarket submission, ACUSON Juniper and ACUSON Juniper Select, did not require clinical studies to support substantial equivalence.
## 9. Summary
Subject devices and predicate devices have the same intended use and substantially equivalent key technological features, and the non-clinical data support the safety of the device and demonstrate that the ACUSON Juniper and ACUSON Juniper Select should perform as intended in the specified use conditions. Therefore, it is the opinion of Siemens Medical Solutions USA. Inc. that the ACUSON Juniber and ACUSON Juniper Select to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.