K221154 · International Biomedical · MPO · Oct 7, 2022 · Physical Medicine
Device Facts
Record ID
K221154
Device Name
Infant Heel WarmerTM; Heel Snuggler®
Applicant
International Biomedical
Product Code
MPO · Physical Medicine
Decision Date
Oct 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5710
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
The Infant Heel Warmer™ Heel Snuggler® are instant warm packs intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, nontoxic, non-sterile, disposable device. The Infant Heel Warmer™, Heel Snuggle® have two configurations, the Heel Snuggler®, which is designed to conform to the shape of the infant's heel, and the standard Infant Heel Warmer™. The devices are primarily used in hospitals, Doctor's offices, and other healthcare settings.
Device Story
Single-use, disposable, non-sterile instant warm pack; contains supersaturated sodium acetate solution and catalyst disk. User flexes internal disk to initiate exothermic crystallization; generates heat (104-107°F). Used in hospitals/clinics by healthcare providers to warm infant heels prior to blood draws. Device secured to infant foot/ankle via adhesive straps; straps perforated for easy removal. Heat increases local blood circulation; aids blood collection. Benefits include improved blood flow for easier sampling.
Clinical Evidence
Bench testing only. Evaluated temperature profiles (activation, surface, duration, skin temperature), shelf-life stability (24 months), and burst strength (200 lb compression). Biocompatibility evaluation leveraged from K936084 per ISO-10993 standards.
Technological Characteristics
Single-use, non-sterile, disposable pouch. Materials: Poly/nylon pouch, food-grade sodium acetate and water solution. Activation: Mechanical (flexing metal catalyst disk). Exothermic crystallization principle. Dimensions: 5.5" x 3.5". Adhesive strap attachment. No software or electronic components.
Indications for Use
Indicated for use on infants to increase blood circulation to the heel to facilitate blood collection for analysis.
Regulatory Classification
Identification
A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2022
International Biomedical Amy Pieper Director of Regulatory Affairs 8206 Cross Park Drive Austin, Texas 78754
Re: K221154
Trade/Device Name: Infant Heel Warmer"]11; Heel Snuggler® Regulation Number: 21 CFR 890.5710 Regulation Name: Hot Or Cold Disposable Pack Regulatory Class: Class I, reserved Product Code: MPO Dated: September 6, 2022 Received: September 6, 2022
Dear Amy Pieper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part Part 809): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221154
Device Name Infant Heel Warmer™; Heel Snuggler®
#### Indications for Use (Describe)
The Infant Heel Warmer™ Heel Snuggler® are instant warm packs intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, nontoxic, non-sterile, disposable device. The Infant Heel Warmer™, Heel Snuggle® have two configurations, the Heel Snuggler®, which is designed to conform to the shape of the infant's heel, and the standard Infant Heel Warmer™. The devices are primarily used in hospitals, Doctor's offices, and other healthcare settings.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
#### Submitter Information:
#### Regulatory Affairs Contact:
International Biomedical 8206 Cross Park Drive Austin, TX 78754 U.S.A.
Amy Pieper Director of Regulatory Affairs (512) 873-0033 - phone (512) 873-9090 - fax
Date Summary Prepared: October 7, 2022
### Device Identification:
Trade Names: Infant Heel Warmer™; Heel Snuggler® Common Name: Infant Heel Warmer Regulatory Class: I Regulatory Name: Hot or cold disposable pack Regulatory Number: (21 CFR 890.5710) Product Code: MPO 510(k) number: K221154
### Predicate Device:
CooperSurgical Inc. – Infant Heel Warmer (Model 24401), WarmGel® Infant Heel Warmer
(Model 20418)— K151845
### Reference Device:
Omni Therm Inc.- Omni Warm Gel Packs - K936084
### Device Description:
The Infant Heel Warmer™/Heel Snuggler® is a single use, disposable pouch that is filled with a super saturated solution of food grade sodium acetate and a catalyst disk. When the catalyst disk inside the packet is flexed by the user, the catalyst disk reacts with the non-toxic solution and initiates a chemical exothermic crystallization of the sodium acetate, generating heat. When activated at 75°F, both models will reach between 104-107°F and then gradually decrease in temperature over time.
The Heel Snuggler® is a butterfly shape with two straps and the Infant Heel Warmer™ is rectangular with one strap. Straps with adhesive ends are attached to the packet for securing the heel warmer to the infant's foot, if desired. Immediately following activation of the user
510(K) Summary
{4}------------------------------------------------
rests the bottom of the infant's foot on one portion of the packet and holds the other portion of the packet to the ankle. The straps with adhesive tape on the end are then fitted across the infant's foot and ankle and secured to the opposite side of the adhesive portion of the strap is affixed to the packet only, not the infant's skin. The straps on the heel warmer are perforated, which enables the user to easily tear the straps and remove the heel warmer from the infant's foot after use.
## Indications for Use:
The Infant Heel Warmer™; Heel Snuggler are instant warm packs intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, nontoxic, non-sterile, disposable device. The Infant Heel Warmer™, Heel Snuggler have two configurations, the Heel Snuggler , which is designed to conform to the shape of the infant's heel, and the standard Infant Heel Warmer". The devices are primarily used in hospitals, Doctor's offices, and other healthcare settings.
## Substantial Equivalence:
The substantial equivalence of the International Biomedical Infant Heel Warmer™; Heel Snuggler to the predicate is shown by similarity in intended use, indications for use, materials and performance. The table below provides a comparison of the technological characteristics of the subject device to the predicate.
{5}------------------------------------------------
| System<br>Specification | International Biomedical Infant<br>Heel WarmerTM; Heel Snuggler® | CooperSurgical Inc.<br>Infant Heel Warmer<br>(Model 24401)<br>WarmGel® Infant<br>Heel Warmer<br>(Model<br>20418)K151845 | Comparison |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Infant Heel WarmerTM; Heel<br>Snuggler® are instant warm packs<br>intended to be used on an infant's<br>heel to increase blood circulation<br>to the area to aid in the drawing of<br>blood for analysis. It is a single<br>use, nontoxic, non-sterile,<br>disposable device. The Infant Heel<br>WarmerTM; Heel Snuggler® have<br>two configurations, the Heel<br>Snuggler®, which is designed to<br>conform to the shape of the<br>infant's heel, and the standard<br>Infant Heel WarmerTM. The<br>devices are primarily used in<br>hospitals, Doctor's offices, and<br>other healthcare settings.<br>K221154<br>Subject Device | An instant warm pack<br>intended to be used on an<br>infant's heel to increase<br>blood circulation to the area<br>to aid in the drawing of<br>blood for analysis. It is a<br>single use, non-toxic,<br>disposable warmer.<br>The device is primarily used<br>in hospitals, doctor's offices,<br>and other healthcare<br>facilities.<br>Predicate Device | Similar – The subject device has<br>two different shapes. The Slight<br>Hour-glass shape of the subject<br>device reduces the overall volume<br>of the device, but the temperature<br>characteristics are unaffected.<br>The shape change allows the<br>device to contour to the small<br>patient foot. Therefore, the slight<br>shape difference does not<br>introduce any new issues of safety<br>and effectiveness. |
| Environment of<br>Use | Hospitals, doctor's offices and<br>other healthcare settings. | Hospitals, doctor's offices,<br>and other healthcare<br>facilities. | Same |
| Prescriptive | Yes | Yes | Same |
| Pouch Material | Poly/nylon | Poly/nylon | Same |
| Solution<br>Material | Sodium acetate (food grade) and<br>water | Liquid version – Sodium<br>acetate (food grade) and<br>water<br>Gel Version - Sodium<br>Acetate (food grade), water,<br>and hydroxyethylcellulose | Same (for the liquid version). The<br>subject device does not offer a<br>"Gel Version" like the predicate.<br>Only offering the liquid version<br>does not introduce any new issues<br>of safety and effectiveness as it is<br>merely reducing the number of<br>products offered. |
| Activation<br>Method | Flexing Disc | Flexing metal Disc | Same |
| Average<br>Maximum Device<br>Surface<br>Temperature and<br>Maximum Device | 104-107°F | 104°F, maximum peak 107°F | Same |
| surface Peak<br>Temperature | | | |
| Average Skin<br>Surface<br>Temperature | 102°F | Not publicly Available | Similar – The predicate device<br>does not publish the skin surface<br>temperature, but the subject<br>device temperature is in line with<br>or slightly lower than the<br>predicate for device surface<br>temperature and there is a strong<br>correlation in this type of product<br>between the device surface<br>temperature and the skin surface<br>temperature. Therefore the<br>stated skin surface temperature of<br>the subject device doesn't<br>introduce any new issues of safety<br>and effectiveness. |
| Size of Pouch | 5.5″ × 3.5″ | 5″ × 3.5″ | Similar – subject device is 0.5"<br>larger. Slight change in size does<br>not introduce any new issues of<br>safety and effectiveness. |
| Shape of Pouch<br>(Heel Warmer) | Rectangular (Heel Warmer)<br>Rectangular with Hour-glass cut<br>out in center (Heel Snuggler) | Not Publicly Available | Similar - The Slight Hour-glass<br>shape of the subject device<br>reduces the overall volume of the<br>device, but the temperature<br>characteristics are unaffected.<br>The shape change allows the<br>device to contour to the small<br>patient foot. Therefore, the slight<br>shape difference does not<br>introduce any new issues of safety<br>and effectiveness. |
| Method of<br>Attachment to<br>Heel | Adhesive Strap | Adhesive Strap | Same |
| Sterility, Number<br>of Uses | Non-sterile, Single Use,<br>Disposable | Non-sterile, Single Use,<br>Disposable | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for International Biomedical. The logo consists of a stylized letter "B" with a swirl at the top left, followed by the words "international" in a smaller font size. Below "international" is the word "BIOMEDICAL" in a larger, bolder font, with the registered trademark symbol to the right.
{7}------------------------------------------------
## Bench Testing:
The following tests/evaluations were performed
- Temperature Profile Testing .
- Temperature profile testing was performed on the International Biomedical O Infant Heel Warmer™; Heel Snuggler and compared to the data from the predicate device. Testing included activation temperature testing, surface temperature, temperature duration testing and skin temperature. All tests were a pass.
- Stability Testing ●
- o Stability Testing was performed on the International Biomedical Infant Heel Warmer™; Heel Snuggler in conjunction with the temperature profile testing to confirm its shelf life of 24 months.
- Burst Testing ●
- o The Subject device has been evaluated for burst strength testing by using a 200 lb compression for 10 seconds. We consider the burst testing to be safe and effective.
- Biocompatibility Testing ●
- The biocompatibility evaluation for the International Biomedical Infant Heel o Warmer™; Heel Snuggler was leveraged from K936084.. Therefore, the evaluation is considered to be in accordance with the FDA guidance on Biocompatibility on the International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing".
### Conclusion:
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate, the International Biomedical Infant Heel Warmer™ ; Heel Snuggler® has been shown to be substantially equivalent to the predicate device identified, and does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.