P20 Elite Series Digital Color Doppler Ultrasound System
Applicant
Sonoscape Medical Corp.
Product Code
IYN · Radiology
Decision Date
Sep 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic. The system is applicable for people who need clinical ultrasound examination. Age, weight, health condition and race are unlimited. The system is intended for use in the following clinical applications: Fetal, Abdominal, Pediatric, Small Organ(breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superfcial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/Gyn and Urology. Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Directional Power Doppler, Tissue Harmonic Imaging, Tissue Doppler Imaging, 3D/4D Imaging mode, Elastography, Contrast and Combined modes: B/M. B/PWD, B/THI, M/Color M. B/Color Doppler, B/Color Doppler/PWD, B/Power Doppler/PWD.
Device Story
Mobile, software-controlled digital color Doppler ultrasound system; transmits ultrasonic energy into patient body; receives and processes echoes to generate real-time anatomical and fluid flow images. Used in hospitals/clinics by trained physicians. Input: ultrasonic signals via linear, convex, or phased array probes. Output: B-mode, M-mode, Doppler (PW/CW/Color/Power/Directional), Tissue Harmonic, Tissue Doppler, Elastography, Contrast, 3D/4D imaging. Includes automated measurement/analysis tools (e.g., Auto OB, Auto NT, Auto EF, Auto bladder, Auto IMT) and assistant functions (S-MSK, Sono-Help, Sono-Assistant, Sono-Synch). Physician reviews images/measurements on integrated LCD monitor; user retains control to edit, accept, or reject automated measurements. Facilitates clinical diagnosis and patient management.
Clinical Evidence
No clinical testing was required. Bench testing only, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), acoustic output (IEC 60601-2-37, AIUM/NEMA UD 2), and software verification/validation (IEC 62304). Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Mobile console with touch screen, keyboard, and color LCD monitor. Employs linear, convex, and phased array probes. Energy source: electrical. Connectivity: supports various imaging modes and automated measurement functions. Software-controlled. Biocompatibility per ISO 10993. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37.
Indications for Use
Indicated for ultrasound imaging, measurement, display, and analysis of human body and fluid in hospital or clinic settings. Applicable to all ages, weights, health conditions, and races. Clinical applications include Fetal, Abdominal, Pediatric, Small Organ, Cephalic, Trans-rectal, Trans-vaginal, Peripheral/Cerebral Vascular, Musculo-skeletal, Cardiac, Laparoscopic, OB/Gyn, and Urology.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
P60 Series Digital Color Doppler Ultrasound System (K171000)
Reference Devices
S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System (K201059)
{0}------------------------------------------------
September 13, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sonoscape Medical Corp. % Toki Wu Regulatory Affairs Manager Room 201&202, 12th Building, Shenzhen Software Park Phase II 1 Keji Middle 2nd Road, Yuehai Subdistrict, Nanshan Shenzhen, Guangdong 518057 CHINA
Re: K221140
Trade/Device Name: P20 Elite Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 5, 2022 Received: August 15, 2022
Dear Toki Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221140
#### Device Name
P20 Elite Series Digital Color Doppler Ultrasound System
#### Indications for Use (Describe)
The P20 Elite Series Digital Color Doppler Ultrasound System (P25 Exp, P20 Exp, P20 Elite, P22 Elite, P22 Exp, P15 Exp and P15 Elite) is a general-purpose ultrasonic imaging instrument intended for use by a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital clinic. The system is applicable for people who need clinical ultrasound examination. Age, weight, health condition and race are unlimited.
The system is intended for use in the following clinical applications: Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/ Gyn and Urology.
Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Directional Power Doppler, Tissue Harmonic Imaging, Tissue Doppler Imaging mode, Elastography, Contrast and Combined modes: B/M, B/PWD, B/THI, M/Color Doppler, B/Color Doppler/PWD, B/Power Doppler/PWD.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K221140
## 1. Submitter [21 CFR807.92 (a) (1)]
| Submitter: | SONOSCAPE MEDICAL CORP. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Room 201 & 202, 12th Building, Shenzhen Software Park Phase II, 1<br>Keji Middle 2nd Road, Yuehai Subdistrict, Nanshan District,<br>Shenzhen, 518057, Guangdong, China |
| Contact Person: | Toki Wu |
| Tel: | +86 755 26722890 |
| Fax: | +86 755 26722850 |
| Email: | ra@sonoscape.net |
| Date Prepared | September 9, 2022 |
## 2. Device [21 CFR807.92 (a) (2)]
| Trade Name: | P20 Elite Series Digital Color Doppler Ultrasound System | | | |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Models: | P25 Exp, P25 Elite, P20 Exp, P20 Elite, P22 Elite, P22 Exp, P15 | | | |
| | Exp, P15 Elite | | | |
| | A Same Career Station Comparis Comparis Comments of Canada Comments of Canada Comments of Canada Comments of Canada Comments of Canada Comments of Canada Comments of Canada C | | | |
Common Name: Diagnostic Ultrasound System and Transducers
Classification Regulatory:
| | CFR Number | Product Code |
|-------------------------------------------------------|------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System<br>(Primary) | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
| Classification Panel: Radiology | | |
Device Class: ll
## 3. Predicate Device(s) [21 CFR 807.92(a) (3)]
The identified predicate device within this submission is as follows:
| Type | Manufacturer | Device | 510 (K) Number |
|-------------------|-------------------------------|-------------------------------------------------------|----------------|
| Primary Predicate | SONOSCAPE<br>MEDICAL<br>CORP. | P60 Series Digital Color<br>Doppler Ultrasound System | K171000 |
{4}------------------------------------------------
| Reference | SONOSCAPE | S60 Elite Series/S70 Series K201055 | | |
|-----------|-------------------|-------------------------------------|-------|---------|
| Devices | MEDICAL | Digital | Color | Doppler |
| CORP. | Ultrasound System | | | |
#### 4. Device Description [21 CFR 807.92(a) (4)]
This SonoScape P20 Elite Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.
This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.
This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.
This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler, Power Doppler and Directional Power Doppler Imaging, or the combination of these modes, Contrast Imaging, Elastography, 3D/4D.
#### 5. Intended Use [21 CFR 807.92(a) (5)]
The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic. The system is applicable for people who need clinical ultrasound examination. Age, weight, health condition and race are unlimited.
The system is intended for use in the following clinical applications: Fetal, Abdominal, Pediatric, Small Organ(breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superfcial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/Gyn and Urology.
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Directional Power Doppler, Tissue Harmonic Imaging, Tissue Doppler Imaging, 3D/4D Imaging mode, Elastography, Contrast and Combined modes:
{5}------------------------------------------------
B/M. B/PWD, B/THI, M/Color M. B/Color Doppler, B/Color Doppler/PWD, B/Power Doppler/PWD.
#### 6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]
P20 Elite Series Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate device:
| Type | Manufacturer | Device | 510 (K) Number |
|--------------------------|-------------------------|---------------------------------------------------------------------|----------------|
| Primary Predicate Device | SONOSCAPE MEDICAL CORP. | P60 Series Digital Color Doppler Ultrasound System | K171000 |
| Reference Devices | SONOSCAPE MEDICAL CORP. | S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System | K201059 |
P20 Elite Series Digital Color Doppler Ultrasound System has almost the same intended uses, complies with the same regulation and safety standards, has the consistent acoustic output levels, and has similar probes and technical characteristics with the predicate device legally marketed P60 Series (K171000).
#### Intended Use Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series has almost the same intended use. There are some additional information for subject device, including patient population and use environment, but all of them are also the same between the subject device and the primary predicate device.
#### Regulation and Safety Standards Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series comply with the same regulation and safety standards.
#### Acoustic Output Levels Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have the consistent acoustic output levels.
#### Probes Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have the similar probes. There are six new probes, including C1-5, 2P1, 3P-A, 6V1A, L741 and L752; except the probe C1-5, all the others have been cleared with other Digital Color Doppler Ultrasound Systems manufactured by SONOSCAPE MEDICAL CORP., such as P10 Series (K173058), S50 Series (K152854), P20 Series
{6}------------------------------------------------
(K172993) and etc.
The new probe C1-5 has the same technical characteristics as the probe 3C-A, including probe type, central frequency, indications for use, operation mode, functions, material and manufacturing process. There is only small difference in size, which doesn't affect the safety, effectiveness and clinical use. There is no new risk raised in safety and effectiveness of the new probes. Therefore they can be considered Substantially Equivalent in safety and effectiveness, so the SE is not affected.
#### Biopsy bracket Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have five new biopsy brackets NGBC1-5, NGB2P1, NGB3P-A, NGBL741-2 and NGBL752. Except the biopsy brackets NGBC1-5 and NGBL741-2, all the others have been cleared with P10 Series and P20 Series Digital Color Doppler Ultrasound System (K173058, K172993), manufactured by SONOSCAPE MEDICAL CORP.
Compared with NGB3C-A, the new biopsy brackets NGBC1-5 and NGBL741-2 have the same material, manufacturing process and reprocessing procedures, and have similar design and appearance; there is only small difference in the size between them, which doesn't affect reprocessing procedures. There is no new risk raised in safety and effectiveness of the new biopsy brackets. Therefore they can be considered Substantially Equivalent in safety and effectiveness, so the SE is not affected.
#### Technical Characteristics Comparison:
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have almost the same main technical features, including Design, Operation Controls, Operation Mode, Display Modes and Power Supply.
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have some newly added features in Function, including contrast imaging, Color 3D, S-Live/S-Live Silhouette/S-Live Contour, S-Depth, STIC, 3D/4D HyCoSy, SPI, Micro F, SR Flow, Auto Face, VCI and etc., which are the same as or similar with the imaging functions of the reference device S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System (K201059); at the same time, the Subject Device P20 Elite Series have four accessibility functions, including S-MSK, Sono-Help, Sono-Assistant and Sono-Synch, which are an assistant function or used for teaching purpose.
Compared with the predicate device P60 Series (K171000), the Subject Device P20 Elite Series have some new features (Measurement Items), including AVC Follicle, Auto
{7}------------------------------------------------
OB. Auto NT, Auto EF, Auto bladder, Auto IMT and etc., which are the same as or equivalent with the measurement functions of the reference device S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System (K201059). The end user can edit, accept, or reject the measurements and can modify the measurement results at any point. The final output and its accuracy is controlled by the end user.
The differences in function and measurement items will not raise new risk and different questions of safety and effectiveness. Therefore they can be considered Substantially Equivalent in safety and effectiveness, the SE is not affected.
#### Summary of the comparison
Compared with the predicate device legally marketed SonoScape P60 Series (K171000) and S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System (K201059), the subject device (P20 Elite Series Digital Color Doppler Ultrasound System) are all analyzed with the predicate device. The comparison showed that they can be considered Substantially Equivalent in safety and effectiveness. There is no new risk raised, and the SE is not affected.
#### 7. Non-Clinical Tests [21 CFR 807.92(b) (1)]
Non-clinical testing to assure compliance with electrical, mechanical, thermal and electromagnetic compatibility safety, acoustic output and biocompatibility were performed and have been found to conform to applicable standards. The P20 Elite Series has been designed and manufactured to meet the following standards:
IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance [08/20/2012];
IEC 60601-1-2:2014. Medical electrical equipment - Part 1-2 General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests [2014-02];
IEC 60601-2-37:2015 Medical Electrical Equipment-Part 2-37: Particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [Edition 2.1, 2015];
ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity [06/01/2009]:
ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for irritation and skin sensitization [08/01/2010]:
AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for
{8}------------------------------------------------
diagnostic ultrasound equipment [08/21/2009].
The Digital Color Doppler Ultrasound System is verified through the relevant summarized information as followed:
| Performance test | Testing protocols and<br>fail/acceptance criteria | Testing results |
|-----------------------------------------|---------------------------------------------------|-----------------|
| Electrical safety testing | IEC 60601-1:2005+A1:2012 | Passed |
| EMC testing | IEC 60601-1-2:2014 | Passed |
| Acoustic testing | IEC 60601-2-37:2007+A1:2015 | Passed |
| | AIUM/NEMA UD 2:2004 (R2009) | |
| Software Verification and<br>Validation | IEC 62304:2006 +A1:2015 | Passed |
Laboratory tests (including Phantom tests) were conducted to verify that the P20 Elite Series met all design specifications and the P20 Elite Series conformed to applicable medical device standards.
#### 8. Clinical Test [21 CFR 807.92(b) (2)]
No clinical testing was required.
#### 9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]
In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SONOSCAPE MEDICAL CORP. concludes that P20 Elite Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.