Scopis ENT Software, Scopis ENT Software with TGS, Scopis Planning Software, Electromagnetic Navigation Unit
Applicant
Stryker Corporation
Product Code
PGW · Neurology
Decision Date
Jul 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Scopis ENT Software is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computerassisted surgery may be appropriate The Scopis ENT Software with TGS is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computer-assisted planning and surgery may be appropriate. The Scopis Planning Software is intended as an aid for precisely locating anatomical structures in preparation for either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computer-assisted pre-operative planning may be appropriate. The Electromagnetic Navigation Unit is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
Device Story
Computer-assisted image-guided surgery system; uses electromagnetic tracking to display instrument position on preoperative CT/MRI patient models. System components: Electromagnetic Navigation Unit (computer platform), software, patient/instrument trackers, pointers, suction tubes, calibration body. Field generator creates electromagnetic field for localization. Registration via anatomical landmarks or surface matching. Surgeon uses system in OR to navigate instruments relative to patient anatomy; supports Augmented Reality overlay of endoscopic video. Benefits: precise localization of anatomical structures during ENT procedures. Software variants: Scopis ENT Software (navigation), Scopis ENT Software with TGS (navigation + target path planning), Scopis Planning Software (pre-operative planning only).
Clinical Evidence
No clinical data. Bench testing only: accuracy per ASTM F2554, software verification/validation per IEC 62304, electrical safety/EMC per ANSI/AAMI ES 60601-1 and IEC 60601-1-2, and cybersecurity verification per NIST 800-53.
Technological Characteristics
Electromagnetic tracking system; Windows 10 IoT Enterprise LTSC 2019 OS. Accuracy testing per ASTM F2554. Compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 62304, ISO 14971, and NIST 800-53. Networked via hospital infrastructure for data transfer; standalone navigation unit.
Indications for Use
Indicated for any medical condition where stereotactic surgery is appropriate and rigid anatomical structures (e.g., paranasal sinuses, mastoid anatomy) can be identified relative to CT/MR-based models. Includes transsphenoidal access, intranasal, sinus (maxillary antrostomies, ethmoidectomies, sphenoidotomies/explorations, turbinate resections, frontal sinusotomies), and anterior skull-based procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
TGS Guidewire and updated Scopis ENT Software (K193118)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
July 12, 2022
Stryker Corporation Maya Schiel Senior Staff Regulatory Affairs Specialist Boetzinger Strasse 41 Freiburg, Baden-Wuerttemberg D-79111 Germany
Re: K221098
Trade/Device Name: Scopis ENT Software, Scopis ENT Software with TGS, Scopis Planning Software, Electromagnetic Navigation Unit Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: April 12, 2022 Received: April 14, 2022
Dear Maya Schiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K221098
#### Device Name
Stryker ENT Navigation System
with Scopis® ENT Software, Scopis® ENT Software with TGS®. Scopis® Planning Software and Electromagnetic Navigation Unit
#### Indications for Use (Describe)
The Stryker ENT Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- Transphenoidal access procedures;
- Intranasal procedures;
- Sinus procedures, such as maxillary antrostomies, sphenoidotomies/sphenoid explorations, turbinate resections, and frontal sinusotomies;
- ENT related anterior skull based procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# st
### 510(k) Summary
## Traditional 510(k): Scopis ENT Software 510(k) Number: K221098
#### Submitter Information
| 510(k) Submitter | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>79111 Freiburg, Germany |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------|
| FDA Establishment Number | 8010177 |
| Contact Person | Maya Schiel<br>Senior Staff Regulatory Affairs Specialist<br>Phone: +49 30 311 911 89<br>Mail: maya.schiel@stryker.com |
| Date Submitted | April 12, 2022 |
| Subject Device | |
| Trade / Proprietary Name | Scopis ® ENT Software<br>Scopis ® ENT Software with TGS ®<br>Scopis ® Planning Software<br>Electromagnetic Navigation Unit |
| Common Name | Ear, Nose, And Throat Stereotaxic Instrument |
| Classification | Class II, 21 CFR 882.4560 |
| Classification Product Code | PGW |
| Classification Name | Stereotaxic Instrument |
#### Legally Marketed Predicate Device
| Predicate Device Name<br>and 510(k) Number | TGS Guidewire and updated Scopis ENT<br>Software (K193118) |
|--------------------------------------------|------------------------------------------------------------|
| | Scopis ENT Software<br>Scopis ENT Software with TGS |
| | Scopis Hybrid Navigation System EM (K161491) |
- -Electromagnetic Navigation Unit
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are all capitalized and black. The word is centered and takes up most of the frame. The background is white.
The purpose of this Traditional 510(k) Submission is to gain clearance for the new version of the Scopis software variants and the updated Electromagnetic Navigation Unit as parts of the Stryker ENT Navigation System.
The initial clearance of this navigation system was achieved via K161491 in 2017 by Scopis GmbH. Scopis GmbH was acquired by Stryker in November 2017. In July 2018, the legal manufacturer for the devices was changed to Stryker Leibinger GmbH & Co. KG. Since October 2019, the devices have also been physically manufactured in Freiburg under the Stryker Leibinger Quality Management System. The 510(k) ownership was also transferred.
As part of the post-acquisition integration the Scopis name of the system was changed from Scopis Hybrid Navigation System to Stryker ENT Navigation System.
The Scopis software applications were most recently cleared under K193118 in 2020. The computer platform called Electromagnetic Navigation Unit was cleared within the first submission for the system with K161491.
#### Device Description
The Stryker ENT Navigation System is a computer-assisted image-guided surgery system. The system displays the position of navigated instruments on a model of the patient's anatomy based on preoperative images (CT or MRI) using electromagnetic tracking technology. The position of the instruments and the patient are localized within an electromagnetic field produced by a field generator. The navigation of instruments relative to the patient's anatomy is established via registration of the patient's anatomy to the preoperative images via anatomical landmarks, or surface matching. The position of navigated instruments is then displayed with respect to preoperative images.
The system comprises a computer platform, the software, patient trackers, pointers and suction tubes, instrument trackers, and a dedicated calibration body.
This submission is about the new version 3.6 of the Scopis software application, and the updated computer platform called Electromagnetic Navigation Unit with the new operating system version 3.6.A.
#### Electromagnetic Navigation Unit
The Electromagnetic Navigation Unit serves as platform of the Stryker ENT Navigation System. This is the computer platform where the navigation software is installed on, and where the main components required for navigation are connected to, like Field Generator or electromagnetic instruments.
Optionally, the Electromagnetic Navigation Unit allows to display the output of external endoscopic imaging devices from Stryker or other manufacturers. Instrument trackers allow for navigation of corresponding endoscope optics. A dedicated calibration body is utilized to calibrate endoscopes and enable Augmented Reality overlay of endoscopic video images.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and connected, giving the word a modern and sleek appearance. The color of the text is black, contrasting with the white background.
#### Scopis Software Application
The software application provides the functionality necessary to import the patient image data sets, build the image-based model of the anatomy, plan a surgical procedure, and display location of navigated instruments relative to the model of the anatomy.
There are two navigation software variants: Scopis ENT Software and Scopis ENT Software with TGS. In comparison to Scopis ENT Software, Scopis ENT Software with TGS offers the ability to create more advanced surgical plans such as target paths to anatomy and display these during navigation. Both Scopis navigations software variants are installed on the Electromagnetic Navigation Unit.
In addition, there will be one designated planning software called Scopis Planning Software. This variant offers identical functionality for creation of advanced surgical plans like Scopis ENT Software with TGS but does not include any navigation functionality. This software is intended to be used on a planning system, like a personal computer of the surgeon, to prepare the application of one of the navigation software variants. Accordingly, after planning the data are to be transferred to the Electromagnetic Navigation Unit for surgical navigation with one of the navigation software variants.
#### Intended Use
The Scopis ENT Software is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computerassisted surgery may be appropriate
The Scopis ENT Software with TGS is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computer-assisted planning and surgery may be appropriate.
The Scopis Planning Software is intended as an aid for precisely locating anatomical structures in preparation for either open or percutaneous procedures together with the Stryker ENT Navigation System. The software application is indicated for any medical condition in which the use of computer-assisted pre-operative planning may be appropriate.
The Electromagnetic Navigation Unit is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
#### Indications for Use
The Stryker ENT Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
{6}------------------------------------------------
Traditional 510(k) K221098 Scopis Software Application
Image /page/6/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and the letters are closely spaced together. The color of the text is black, and the background is white.
- -Transsphenoidal access procedures;
- -Intranasal procedures;
- -Sinus procedures, such as maxillary antrostomies, ethmoidectomies, sphenoidotomies/sphenoid explorations, turbinate resections, and frontal sinusotomies;
- -ENT related anterior skull-based procedures
#### Comparison of Technological Characteristics
This submission is about the new version of the Scopis software application and the updated operating system running on the corresponding computer platform.
The general way the devices are used, and function has not been changed. So, the subject and predicate devices are identical in most characteristics, such as:
- FDA product codes and regulation
- Intended users and target patient population ●
- Indications for Use ●
- Application Area ●
- Type of Use and type of Patient Contact
- Intended Environment
- Tracking Technology ●
- Programming Language
- Operating Principles
- Modes of Operation ●
- . Image Import
- Reference Images ●
- Patient Registration ●
- Planning Objects .
- Control Mechanism ●
The new software version has been developed to add some new features to the software. These new features include, but are not limited to:
- Support tracking multiple instruments of the same kind ●
- Improved guidance during patient registration ●
- . Field generator alignment support
- . Introducing a new software variant, the Scopis Planning Software
- Cybersecurity improvements .
Further, the operating system running on the computer platform of the navigation system called Electromagnetic Navigation Unit has been updated to Windows 10 IoT Enterprise LTSC 2019.
{7}------------------------------------------------
#### Performance Data
#### Non-Clinical Testing
Non-clinical design verification and validation activities were conducted to provide evidence that the subject devices successfully meet their requirements and perform according to their specifications and intended uses. Table 2 provides a summary of the performance data
Table 2: Summary of Performance Data
| Test Item | Test summary |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Accuracy | To ensure accuracy, subject devices are tested in accordance<br>with ASTM F2554. The devices meet state-of-the-art<br>acceptance criteria for point accuracy, precision, and distance<br>accuracy. |
| Software | Scopis software applications were subject to verification and<br>validation testing in compliance with IEC 62304 and FDA<br>guidances. All testing criteria were met. |
| Electrical Safety and<br>Electromagnetic Compatibility | Applicable devices in scope were checked and certified for<br>compliance with ANSI/AAMI ES 60601-1 and IEC 60601-1-2. |
| Safety | Verification of risk controls specified in the corresponding risk<br>analysis is in accordance with ISO 14971. |
| Security | Verification of the security of the operating system with respect<br>to technical requirements (NIST 800-53 and FDA Guidances)<br>and operating system updates. All testing criteria were met. |
#### Clinical Testing
No clinical testing was deemed necessary for this Traditional 510(k) submission.
#### Conclusion
The subject devices have the same main functionalities and follow the same operating principles as their predicates. The fundamental technological characteristics do not change. The only differences are caused by the introduction of additional features of the new software version and some improvements, e.g., for cybersecurity. The performance data demonstrate that the subject devices meet their specifications and the requirements on safety and effectiveness.
In conclusion, the information provided demonstrates that the Electromagnetic Navigation Unit and the three software variants are at least as safe and effective as their predicates (K161491, K193118) and do not raise different issues of safety or effectiveness, and therefore supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.