K221086 · Tmrw Life Sciences, Inc. · QUJ · Nov 4, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K221086
Device Name
CryoRobot Select System
Applicant
Tmrw Life Sciences, Inc.
Product Code
QUJ · Obstetrics/Gynecology
Decision Date
Nov 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6120
Device Class
Class 2
Indications for Use
The CryoRobot Select System is intended to provide an automated liquid nitrogen storage system for oocytes, embryos, and sperm to facilitate the identification, storage, and retrieval of specimens.
Device Story
The CryoRobot Select System is an automated cryogenic storage platform for oocytes, embryos, and sperm. It comprises the CryoRobot Select (robotic storage tank), CryoBeacons (RFID-enabled storage units), and CryoTransporter (insulated container). Users load up to 24 CryoBeacons into the CryoTransporter, which is then placed into the robot's drawer. The system uses RFID to identify the presence of CryoBeacons and automates the transport of these units between the drawer and the 250L cryogenic storage tank. The system maintains cryogenic temperatures throughout the loading, storage, and retrieval processes. A touchscreen interface allows users to manage specimens and monitor system status. By automating the identification and handling of cryopreserved samples, the device aims to reduce manual handling errors and ensure consistent temperature maintenance, thereby protecting specimen viability during long-term storage.
Clinical Evidence
Bench testing only. Performance testing included auto-positioning accuracy, manual drawer operation, thermal performance during loading/storage/retrieval, RFID traceability, system safety (alarms, emergency LN2 activation, UPS function), and freezer hold-time assessments. Electrical safety and EMC testing were conducted per IEC 60601-1-2 and IEC 61010-1. Wireless and cybersecurity features were evaluated per FDA guidance.
Technological Characteristics
Automated liquid nitrogen storage system; 250L cryogenic tank; robotic arm for specimen transport; RFID-enabled storage units (CryoBeacons); insulated transport container (CryoTransporter). Software-controlled with touchscreen interface. Operates via automated pick-and-place mechanism. Complies with IEC 60601-1-2 (EMC) and IEC 61010-1 (Electrical Safety).
Indications for Use
Indicated for the automated liquid nitrogen storage, identification, and retrieval of oocytes, embryos, and sperm for patients undergoing assisted reproduction.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 4, 2022
TMRW Life Sciences, Inc. Christina Miracle VP, QARA 250 Hudson Street, 6th Floor New York, NY 10013
Re: K221086
> Trade/Device Name: CryoRobot Select System Regulation Number: 21 CFR§ 884.6120 Regulation Name: Assisted Reproduction Accessories Regulatory Class: II Product Code: QUJ Dated: October 3, 2022 Received: October 3, 2022
Dear Christina Miracle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221086
Device Name CryoRobot Select System
## Indications for Use (Describe)
The CryoRobot Select System consists of the CryoRobot Select, CryoBeacons, and CryoTransporter. The indications for use for the CryoRobot Select System is shown below:
The CryoRobot Select System is intended to provide an automated liquid nitrogen storage system for oocytes, embryos, and sperm to facilitate the identification, storage, and retrieval of specimens.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span></span> <label> <input checked="true" type="checkbox"/> </label> </div> |
|----------------------------------------------|-------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span></span> <label> <input type="checkbox"/> </label> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 00<br>00 1 | 400000<br>of of the<br>4494 | 1-222 | Course<br>67<br>00 | @ |
|------------|-----------------------------|-------|--------------------|---|
|------------|-----------------------------|-------|--------------------|---|
# 510(k) Summary – K221086
### I. SUBMITTER:
| Manufacturer: | TMRW Life Sciences, Inc.<br>250 Hudson Street, 6th Floor<br>New York, NY 10013 | |
|---------------|--------------------------------------------------------------------------------|-------------------------------------------------|
| | Phone: | 1-866-733-TMRW |
| | Email: | ra@tmrw.org |
| | Contact Person: | Christina Miracle |
| | Date Prepared: | November 1, 2022 |
| II. | DEVICE | |
| | Trade Name of Device: | CryoRobot Select System |
| | Common Name: | Automated Cryopreservation Storage System |
| | Regulation Name: | Assisted Reproduction Accessories |
| | Regulation Number: | 21 CFR 884.6120 |
| | Regulatory Class: | Class II |
| | Product Code: | QUJ (Automated Cryopreservation Storage System) |
| III. | PREDICATE DEVICE | |
| | Primary Predicate Device: | Kryo ART Controlled Rate Freezer (K032086) |
| | Manufacturer: | PLANER, plc. |
| | Product Code: | MQG (Assisted Reproduction Accessory) |
The predicate device has not been subject to a design-related recall.
The Agency cleared the predicate device with the product code MQG. On July 11, 2017, Federal Register Notice 82 FR 31976 exempted controlled-rate cryopreservation freezers. A 510(k) is not needed to market a controlled-rate cryopreservation freezer if it does not exceed the limitation of
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "TMRUW" spelled out using blue circles. The letters are formed by arranging the circles in the shape of each letter. The letters are evenly spaced and the circles are all the same size.
exemption found in 21 CFR 884.9. Therefore, the predicate device now falls under product code PUB (Accessory, Assisted Reproduction, Exempt).
A 510(k) submission was required for the subject device, as it exceeded the limitations of exemption per 21 CFR 884.9.
## DEVICE DESCRIPTION IV.
The CryoRobot Select System includes the CryoRobot Select (the robot), CryoBeacons, and the CryoTransporter.
The CryoRobot Select System provides an automated liquid nitrogen storage system for oocytes, embryos, and sperm to facilitate the identification, storage, and retrieval of specimens. The system consists of a cryogenic tank, robotic hardware for automation, dedicated robot control software, RFID-enabled storage units ("CryoBeacons") and an insulated container for carrying multiple CryoBeacons ("CryoTransporters").
The robot has an automated container picking system, designed to pick and transport individual CryoBeacons to and from the cryogenic tank. The robot allows users to load 24 CryoBeacons at a time, with a freezer storage capacity of 1,386 through a user-accessible drawer. The system maintains cryogenic temperatures during the specimen storage and retrieval processes.
The CryoBeacon is a radio-frequency identification (RFID) enabled storage unit that holds up to eight dimensionally compatible cryodevices. TMRW does not manufacture cryodevices for use with CryoBeacons. Cryodevices compatible with the CryoBeacon include devices up to 135 mm long with the gametes or embryos located at or below 50 mm when measured from the distal end of the closed device.
The CryoTransporter is an insulated cryogenic container for carrying 24 CryoBeacons to and from the CryoRobot Select system. The CryoTransporter is filled with liquid nitrogen to maintain a cryogenic environment during CryoBeacon transport. For safe carrying, the CryoTransporter has a handle and a vented, transparent lid for viewing its contents. The CryoTransporter is placed in the drawer of the CryoRobot Select to transfer or retrieve CryoBeacons from the cryogenic tank.
### V. INDICATIONS FOR USE
The CryoRobot Select System consists of the CryoRobot Select, CryoBeacons, and CryoTransporter. The indications for use for the CryoRobot Select System is shown below:
The CryoRobot Select System is intended to provide an automated liquid nitrogen storage system for oocytes, embryos, and sperm to facilitate the identification, storage, and retrieval of specimens.
## COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL VI. CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICE
{5}------------------------------------------------
Image /page/5/Figure/0 description: The image shows the word "TMRUW" spelled out using blue circles. The letters are formed by arranging the circles in a grid-like pattern to create the shapes of the letters. The letters are spaced apart from each other. There is a registered trademark symbol in the upper right corner of the letter W.
| Parameter | Subject Device<br>(K221086) | Predicate Device<br>(K032086) | Comparison |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The CryoRobot Select<br>System consists of the<br>CryoRobot Select,<br>CryoBeacons, and<br>CryoTransporter. The<br>indications for use for<br>the CryoRobot Select<br>System is shown<br>below:<br>The CryoRobot Select<br>System is intended to<br>provide an automated<br>liquid nitrogen storage<br>system for oocytes,<br>embryos, and sperm to<br>facilitate the<br>identification, storage,<br>and retrieval of<br>specimens. | The Kryo ART brand<br>of controlled rate<br>freezers are intended to<br>be used to freeze<br>gametes or embryos at<br>a user determined rate. | The indications for use<br>of the subject device is<br>not identical to that of<br>the predicate device.<br>However, the two<br>devices have the same<br>intended use, which is<br>to contain, freeze, and<br>maintain gametes<br>and/or embryos at an<br>appropriate freezing<br>temperature. |
| Software | Software-controlled | Software-controlled | Same |
| Design | The CryoRobot Select<br>System includes an<br>insulated<br>CryoTransporter to<br>transport CryoBeacons<br>holding cryodevices<br>containing gametes and<br>embryos to the<br>loading/retrieval drawer<br>of the CryoRobot<br>Select. The CryoRobot<br>Select identifies the<br>presence or absence of<br>CryoBeacons in the<br>drawer and uses this<br>information to<br>determine whether a<br>storage or retrieval<br>procedure will occur.<br>The CryoRobot Select<br>will then transport the<br>CryoBeacons to or<br>from the storage tank | Controlled rate liquid<br>nitrogen freezers with<br>different chamber sizes<br>to cryopreserve<br>samples. | Different |
{6}------------------------------------------------
| 0.00000<br>*****;<br>1888 | 200 | Course<br>00 | 1 | (<br>X<br>COOL |
|---------------------------|-----|--------------|---|----------------|
| | | 00 | | 00 |
| | CryoRobot Select<br>maintains the samples<br>at cryogenic<br>temperatures by<br>maintaining exposure<br>to liquid nitrogen<br>throughout sample<br>transport. | | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|-----------|
| Gamete/Embryo-<br>Contacting | No. Samples in<br>cryodevices are loaded<br>in device. | No. Samples in<br>cryodevices are loaded<br>in device. | Same |
| | 250 L storage tank | Dewar sizes:<br>1.7 L<br>3.3 L<br>16 L | Different |
| Tank Size | | | |
| | | | |
| Number of Samples | 1,386 CryoBeacons | Not known | Different |
| | Touchscreen and<br>specimen management<br>software | MRV controller system | Different |
| User Interface | | | |
| | CryoRobot Select<br>CryoBeacons<br>CryoTransporter | Not known | Different |
| Components | | | |
| RFID | Yes | No | Different |
As shown in the table, the subject and predicate device do not have identical indications for use statements; however, they have the same intended use (i.e., to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature).
There are similarities in the technologies between the subject and proposed predicate device, including the following:
- Both are software-controlled devices.
- . Both monitor temperature and use liquid nitrogen to freeze and/or maintain samples at cryogenic temperatures for purposes of long-term storage.
- Both include features to provide liquid nitrogen from storage tanks to cool and maintain ● the cryogenic state of samples.
- Both devices are able to provide information on their status/performance to device users. ●
Differences in technology include the following:
- . The subject device includes RFID features to identify the presence of CryoBeacons in the drawer of the CryoRobot Select that inform the system to prepare for an automated loading or retrieval procedure initiated by the user. This represents automation of the manual process done for samples after freezing using the predicate device. Although the predicate does not include these features, they do not raise different questions of S&E as compared to the proposed predicate (i.e., both devices raise similar S&E questions: will samples remain frozen during device use?, will a failure lead to sample damage or loss?, will a sample be mis-identified?, etc.).
{7}------------------------------------------------
Image /page/7/Figure/1 description: The image shows the word "TMRW" spelled out using blue circles. The letters are formed by arranging the circles in the shape of each letter. The letters are evenly spaced and the word is centered in the image. There is a registered trademark symbol to the right of the letter W.
- The subject device CryoTank design is for long-term storage as opposed to the design of the proposed predicate that is for holding samples for a short duration after freezing. This difference does not raise different questions of safety and effectiveness, as both devices are intended to maintain samples in a cryogenic state while within the device (similar questions to that stated above).
- The subject device transfers a maximum of 24 CryoBeacons to or from the storage tank . during a loading or retrieval operation and can hold a maximum of 1,386 CryoBeacons in the 250 L storage tank. The maximum number of samples the predicate device equipped with dewar sizes of 1.7-16 L can freeze and maintain in a single run is not known. Differences in the number of samples that can be processed in a single use of the device or the total number of samples that can be maintained in the device do not raise different questions of S&E between the two devices (i.e., both devices raise similar S&E questions: can the devices maintain the cryogenic state of the maximum number of samples loaded in device?).
- . The subject and predicate devices have different user interface systems that are used to operate and receive information on device function status. Differences in user interface systems do not raise different questions of safety and effectiveness.
### VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
# Reprocessing
Reprocessing instructions were based on validated cleaning and disinfection testing conducted in accordance with the 2015 FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
# Electrical Safety and Electromagnetic Compatibility
Testing was conducted in accordance with:
- . IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- IEC 61010-1 Edition 3.1 2017-01 Consolidated version, Standard for Safety for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements
# Wireless Technology
The wireless technology features of the device were assessed in accordance with the 2007 FDA guidance "Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff."
# Software
Software was evaluated in accordance with the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
# Cybersecurity
Cybersecurity was evaluated in accordance with the 2014 FDA guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "TMRW" spelled out with blue circles. The letters are formed by arranging the circles in a grid-like pattern. The letters are spaced closely together, and the overall effect is a playful and modern design. The "R" has a registered trademark symbol next to it.
## Performance Testing
Performance testing was conducted to demonstrate the system performs as expected in the following areas:
- . Auto position testing to demonstrate capability of the robot to pick and place samples to and from the CryoTransporter and CryoRobot Select storage tank.
- . Manual drawer opening/closing to confirm that the drawer can be opened and closed manually.
- . Thermal performance tests to monitor the temperature of cryodevices during loading, storage, and retrieval to ensure the sample is not compromised and to ensure delivery of liquid nitrogen when necessary.
- RFID traceability verification tests to ensure the reliability of the RFID features. ●
- System safety verification testing to ensure that all safety systems operate as intended (e.g., alarms, emergency liquid nitrogen system activation, uninterruptable power supply function, drawer position, drawer finger trap testing, etc.).
- . Freezer hold time assessment to verify the hold time window in which specimens remain below the acceptable temperature without power and liquid nitrogen supply.
- . CryoTank verification tests to ensure the tank maintains samples during storage.
### CONCLUSIONS VIII.
The results of the testing described above demonstrate that the subject device is as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.