← Product Code [IZL](/productcode/IZL) · K221081

# iQFlex M Mobile X-ray System, iQFlex MD Mobile X-ray System (K221081)

_First Source, Inc. · IZL · Jun 13, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221081

## Device Facts

- **Applicant:** First Source, Inc.
- **Product Code:** [IZL](/productcode/IZL.md)
- **Decision Date:** Jun 13, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1720
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The iQFlex M/iQFlex MD Mobile X-ray System is a medical device intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). This device is not intended for mammography.

## Device Story

Mobile X-ray system; mains-powered; consists of LED display, X-ray generator, X-ray tube, collimator, and control board. Used in clinical facilities; operated by physicians or technicians. Input: user-selected kV/mAs parameters via control panel. Transformation: system control software manages high-voltage tank to excite X-ray tube for emission; monitors device status. Output: X-ray radiation for imaging; requires external film-cassette or flat-panel detector (not included). Benefits: compact, lightweight, portable design for diagnostic imaging at point-of-care. Healthcare providers use images for clinical diagnosis; device supports handheld or stand-mounted configurations.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety, EMC, and software verification/validation. Testing confirmed device performance meets specifications and conforms to IEC 60601-1, 60601-1-2, 60601-1-3, and 60601-2-54 standards.

## Technological Characteristics

Mains-powered mobile X-ray system. Components: high-voltage tank, inverter, collimator, system control board. Generator frequency: 50 kHz. Output: 40-100 kV. Dimensions: X-ray generator 365x240x230mm (M) or 185mm (MD). Connectivity: supports external flat-panel detectors (e.g., PIXXGEN, DRTECH, VIEWORKS). Software: embedded control software for real-time circuit interaction. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 62304, IEC 62366-1.

## Regulatory Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- MinXray HF120/60HPowerPlus ([K040046](/device/K040046.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 13, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

First Source Inc. % Mr. Woo Sung Park Consultant MEDMONTS Co., Ltd. Life officetel 320, 40, 63-ro, Youngdeungpo-gu Seoul, 07345 REPUBLIC OF KOREA

Re: K221081

Trade/Device Name: iQFlex M, iQFlex MD Mobile X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: February 28, 2022 Received: April 27, 2022

Dear Mr. Park:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221081

Device Name iQFlex M/iQFlex MD Mobile X-Ray System

Indications for Use (Describe)

The iQFlex M/iQFlex MD Mobile X-ray System is a medical device intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities).

This device is not intended for mammography.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text 'K221081' in a simple, sans-serif font. The text is presented in a horizontal arrangement, with each character clearly legible. The overall impression is clean and straightforward, with the focus solely on the alphanumeric sequence.

| Image: Logo | 005_510(k) Summary | Page: | 005-1 |
|-------------|--------------------|-------|-------|
|-------------|--------------------|-------|-------|

005 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number:

#### SUBMITTER I.

First Source Inc.

Address: 3495 Winton Place, Building E, Suite #1, Rochester, NY 14623 Phone: 585.272.1690, Toll Free: 800.349.5980, Fax: 585.272.7678 https://1stsourceimaging.com

Email: sales@fsimed.com

Person: Ronald Viola

Position: President

Date Prepared: Feb.28, 2022

### II. DEVICE

Name of Device: i9Flex M/i9Flex MD Mobile X-ray System

Requlation Name: Mobile X-ray system

Regulation Number: 892.1720

Regulatory Class: II

Product Code: IZL

### III. PREDICATE DEVICE

MinXray HF120/60HPowerPlus™, K040046

Name of Device: Min Xray, MODEL HF120/60H PowerPlus™

Regulation Name: Mobile X-ray system

{4}------------------------------------------------

#### K221081

| 17<br>SOURCE |
|--------------|
|--------------|

Page:

Regulation Number: 892.1720

Regulatory Class: II

Primary Product Code: IZL

This predicate has not been subject to a design-related recall.

No reference devices were used in this submission.

# IV. DEVICE DESCRIPTION

This Mobile X-ray System (Model: i9Flex M/iºFlex MD) consists of a LED display with up and down soft-keys for controlling kV, an X-ray generator (linepowered transformer), an X-ray tube assembly, and a collimator. A cart or a stand can be used with the iºFlex MD. In addition, this unit has preset memory keys to store and select kV/mAs, The iQFlex M/iQFlex MD is used with a film-cassette or flat-panel detector. The image detectors (a necessary component of a fully-functional diagnostic system) are not part of this submission.

This device is a mains-powered mobile X-ray system, designed and manufactured by First Source Inc.

Compared with traditional X-ray products, this device has exquisite structure, compact design, light weight and easy operation.

The major components of the X-ray main unit include: handle, enclosure, control panel, system control board, high-voltage tank, inverter, collimator (beam limiter), and system control software running on the system control board.

The system control software is for real-time interaction and control with various circuit modules inside the X-ray generator. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters.

The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board, and control the highvoltage tank to generate high-voltage to excite the X-ray tube inside to emit Xrays, control the switch of the collimator indicator, and monitor the working status of the device, and control the display of the status indicators.

{5}------------------------------------------------

| 1<br>20<br>ל<br>/<br>1 |
|------------------------|
|------------------------|

| 17<br>SOURCE |
|--------------|
|--------------|

Page:

The system is for X-ray imaging and diagnosis in facilities with mobile or fixing sites.

The iQFlex M/iºFlex MD is not intended for mammography.

The device can be used with an X-ray flat panel detector, a computer for receiving and detecting signal results and an image processing software. This mobile X-ray system is designed for handheld or stand-mounted imaging. Model i®Flex MD can be configured to an optional portable stand that complies with IEC 60601- 1 safety standard. The recommended maximum load that the stand can safely carry is 30kgs to ensure the mechanical stability and effectiveness of the device.

The cybersecurity risks of the i9Flex M/iºFlex MD have been addressed to assure that no new or increased cybersecurity risks were introduced as a part of device risk analysis. These risks are defined as sequence of events leading to a hazardous situation, and the controls for these risks were treated and implemented as proposed in the risk analysis (e.g., requirements, verification).

# V. INDICATIONS FOR USE

The iPFlex M/iPFlex MD Mobile X-ray System is a medical device intended for use by a qualified/trained physician or technician forthe purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). This device is not intended for mammography.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The comparison between the overall specifications of predicate device (MinXray HF120/60HPowerPlus™) and the subject device (i9Flex M/i9Flex MD mobile X-ray system) is shown in Table 1.

{6}------------------------------------------------

| Image: logo | 005_510(k) Summary | Page: | 005-4 |
|-------------|--------------------|-------|-------|
|-------------|--------------------|-------|-------|

| Item                                          | Predicate Device                                                                                                                                                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                               | MinXray HF120/60HPowerPlus             TM (K040046)                                                                                                                                               | ioFlex M/ioFlex MD (without digital imaging)                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for use                           | This radiographic system is intended for use by a<br>qualified/trained physician or technician on both<br>adult and pediatric subjects for taking diagnostic<br>x-rays. Not for mammographic use. | The ioFlex M/ioFlex MD Mobile X-ray System is a<br>medical device intended for use by a<br>qualified/trained physician or technician for the<br>purpose of acquiring X-ray images of the desired<br>parts of patient's anatomy (including head, cervical<br>spine, chest, abdomen, lumbar spine, pelvis and<br>extremities).<br>The device may be used for handheld diagnostic<br>imaging of body extremities. This device is not<br>intended for mammography. |
| Weight                                        | 17.94 kg                                                                                                                                                                                          | X-ray Generator: 11.0 kg(24.25lbs)<br>Mobile Cart: 38.1kg (84lbs)                                                                                                                                                                                                                                                                                                                                                                                              |
| Size                                          | 453*292*224 mm                                                                                                                                                                                    | X-ray Generator:<br>- ioFlex M : 365(L) x 240(W) x 230(H) (mm)<br>- ioFlex MD : 365(L) x 240(W) x 185(H) (mm)<br>Mobile Cart with X-ray Generator<br>0.74m x 1.36m(2.41ft. x 4.47ft.);<br>with maximum height positioning: 1m x 2m<br>(3.27ft. x 6.56ft.)                                                                                                                                                                                                      |
| Output                                        | 60 mA(0.01-0.1sec), 42 mA (0.11 – 5.0sec) @<br>40 - 50 kVDC<br>50 mA(0.01-0.1sec), 35 mA (0.11 – 5.0sec) @<br>52 - 60 kVDC                                                                        | 40kV ~ 60kV, 30mA, 0.4mAs ~ 120mAs<br>61kV ~ 70kV, 35mA, 0.4mAs ~ 63mAs<br>61kV ~ 70kV, 30mA, 80mAs ~ 100mAs<br>71kV ~ 80kV, 30mA, 0.4mAs ~ 50mAs                                                                                                                                                                                                                                                                                                              |
| Predicate Device                              | Subject Device                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| MinXray HF120/60HPowerPlus™ (K040046)         | iQFlex M/iQFlex MD (without digital imaging)                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 45 mA(0.01-0.1sec), 31.5 mA (0.11 – 5.0sec) @ | 71kV ~ 80kV, 25mA, 63mAs ~ 80mAs                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 62 - 70 kVDC                                  | 81kV ~ 90kV, 25mA, 0.4mAs ~ 40mAs                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 38 mA(0.01-0.1sec), 26.6 mA (0.11 – 5.0sec) @ | 81kV ~ 90kV, 20mA, 50mAs ~ 100mAs                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 72 - 80 kVDC                                  | 91kV ~ 100kV, 25mA, 0.4mAs ~ 32mAs                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 33 mA(0.01-0.1sec), 23.1 mA (0.11 – 5.0sec) @ | 91kV ~ 100kV, 20mA, 40mAs ~ 80mAs                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 82 - 90 kVDC                                  |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 30 mA(0.01-0.1sec), 21 mA (0.11 – 5.0sec) @   |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 92 - 100 kVDC                                 |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 20 mA(0.01-0.1sec), 14 mA (0.11 – 5.0sec) @   |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 102 - 120 kVDC                                |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Up-Down pushbuttons for kVp selections and    | Up/Down pushbuttons for kV/exposure time(mAs)                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| exposure time selections with LED indicators  | selections and LED indicators for selected                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| mAs indicator                                 | kV/exposure time(mAs)                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| (0.01-0.2sec) in 0.01sec. Step                | 0.010~6.000sec.                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| (0.2-0.4sec) in 0.02sec. Step                 | 0.010, 0.013, 0.014, 0.016, 0.017, 0.020, 0.024.                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Item                                          | Predicate Device                                                                                                                                                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                               | MinXray HF120/60HPowerPlus™ (K040046)                                                                                                                                                             | i9Flex M/i9Flex MD (without digital imaging)                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                               |                                                                                                                                                                                                   | 1.142, 1.250, 1.280, 1.333, 1.428, 1.600, 1.666,<br>1.800, 2.000, 2.100, 2.285, 2.500, 2.520, 2.666,<br>2.857, 3.150, 3.200, 3.333, 3.428, 4.000, 4.800,<br>5.000, 6.000 (86 Steps)<br>(0.4~120mAs)                                                                                                                                                                                                                                                            |
| Memory<br>Settings<br>(technique)             | 5 memories                                                                                                                                                                                        | 5 memories                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| HF Generator                                  | 85 kHz                                                                                                                                                                                            | 50 kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| kW                                            | 2.4 kW peak -> 3.15                                                                                                                                                                               | 2.5 kW                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| kVp                                           | 40-120 kVp                                                                                                                                                                                        | 40-100 kV                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| X-ray Tube                                    | Superior X-ray Tube Company SXR-130 1.2 mm, 65 kHU                                                                                                                                                | OX/125-1.2(CEI) 1.2mm, 35kHU                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Collimator                                    | Advantech -> Collimare                                                                                                                                                                            | Manual Type, Double Slit                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Flat-panel<br>detector                        | None                                                                                                                                                                                              | PIXXGEN PRUDENT 1212/1417/1717 declared in<br>K211108<br>DRTECH EVS 4343W/WP, 3643W/WP declared in<br>K193017<br>VIEWORKS VIVIX-S 1717V declared in K181003<br>or alternatives U.S. 510(k) cleared                                                                                                                                                                                                                                                             |
| Flat-panel<br>detector                        | None                                                                                                                                                                                              | PRUDENT 1212 : 140µm, 2,048 x 2,048 pixels, a-<br>Si/IGZO                                                                                                                                                                                                                                                                                                                                                                                                      |
| Item                                          | Predicate Device<br>MinXray HF120/60HPowerPlus™ (K040046)                                                                                                                                         | Subject Device<br>ioFlex M/ioFlex MD (without digital imaging)                                                                                                                                                                                                                                                                                                                                                                                                 |
| specifications                                |                                                                                                                                                                                                   | PRUDENT 1417 : 140µm, 2,560 x 3,072 pixels, a-Si/IGZO<br>PRUDENT 1717 : 140µm, 3,072 x 3,072 pixels, a-Si/IGZO<br>EVS 4343W/WP: 140µm, 3,072 x 3,072 pixels, IGZO<br>EVS 3643W/WP: 140µm, 2,560 x 3,072 pixels, IGZO<br>VIVIX-S 1717V: 140µm, 3,072 x 3,072 pixels, a-Si                                                                                                                                                                                       |
| Photo                                         | Image: MinXray HF120/60HPowerPlus                                                                                                                                                                 | Image: iFlex M<br>Image: iFlex MD                                                                                                                                                                                                                                                                                                                                                                                                                              |

# Table 1 Comparison of Technology Characteristics

{7}------------------------------------------------

Image /page/7/Picture/6 description: The image shows the logo for 1st Source. The logo features the number "1st" in black, stacked on top of the word "SOURCE" in a larger, also black font. Behind the text is a teal crescent shape that curves around the right side of the text, creating a semi-circular frame. The logo is simple and modern, using a clean font and a limited color palette.

Use Interface

Exposure time

(0.4-1.0sec) in 0.05sec Step

(1.0-5.0sec) in 0.1sec Step

ltem

0.180, 0.200, 0.210, 0.228, 0.250, 0.252, 0.266, 0.285, 0.315, 0.320, 0.333, 0.371, 0.400, 0.433,

0.025, 0.028, 0.030, 0.033, 0.037, 0.040, 0.043,

0.045, 0.050, 0.052, 0.053, 0.057, 0.064, 0.065, 0.066, 0.071, 0.080, 0.083, 0.091, 0.100, 0.106, 0.114, 0.125, 0.128, 0.133, 0.142, 0.160, 0.166,

0.457, 0.500, 0.520, 0.533, 0.571, 0.640, 0.650, 0.666, 0.714, 0.800, 0.833, 0.914, 1.000, 1.066,

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for 1st Source. The logo features the text "1st SOURCE" in a stylized font. The "1st" is smaller and positioned above the word "SOURCE". A teal crescent shape partially surrounds the text, adding a visual element to the design. The logo is simple and professional, likely representing a financial institution or related service.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for 1st Source. The logo features the number "1st" stacked on top of the word "SOURCE". The text is positioned inside of a teal crescent shape. The background of the image is white.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for 1st Source. The logo features the text "1st SOURCE" in bold, black letters. Above the text is a teal, crescent-shaped graphic that partially encircles the "1st" part of the text. The overall design is simple and professional.

#### PERFORMANCE DATA VII.

#### Performance Testing:

The iQFlex M/i9Flex MD conforms to the following standards:

| ISO 14971:2012                                                | Medical device - application of risk management to<br>medical device                                                                                                                                                |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-<br>1:2005/AMD1:2012<br>EN 60601-<br>1:2006/A1:2013 | Medical electrical equipment -- Part 1: General<br>requirements for basic safety and essential<br>performance                                                                                                       |
| IEC 60601-1-2:2014 /<br>EN60601-1-2:2015                      | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>disturbances - Requirements and tests                      |
| EN 60601-1-3:2008/<br>A11:2016                                | Medical electrical equipment - Part 1-3: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Radiation<br>protection in diagnostic X-ray equipment (IEC<br>60601-1-3:2008) |
| IEC 60601-2-54:2015/<br>EN 60601-2-54:2015                    | Medical electrical equipment -- Part 2-54: Particular<br>requirements for the basic safety and essential<br>performance of X ray equipment for radiography and<br>radioscopy                                        |
| IEC<br>62304:2006/AMD1:2015                                   | Medical device software - Software life-cycle<br>processes                                                                                                                                                          |
| IEC 62366-1:2015/<br>EN 62366:2008 AMD1:<br>2015              | Medical devices - Application of usability<br>engineering to medical devices                                                                                                                                        |
| IEC 60601-1-<br>6:2010+A1:2013                                | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability                                                                     |

### Discussion of Testing

The performance characteristics and operation / usability of the i9Flex MD were evaluated in non-clinical (bench) testing. These studies have demonstrated the intended workflow, related performance, overall function, shipping performance, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and test results have

{11}------------------------------------------------

demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.

The following performance data were provided in support of the substantial equivalence

determination:

# · Electrical safety and electromaqnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the i9Flex M/19Flex MD Mobile X-ray System in compliance to IEC 60601-1-2:2014 - Medical electrical equipment Part 1-2, Electromagnetic Compatibility.

· Software Verification and Validation Testing Software verification and validation testing was conducted and documentation has been provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered to present a "moderate" level of concern. This is based on the fact that a malfunction of or latent design flaw in the software component could lead to an erroneous diagnosis, or to a delay in delivery of appropriate medical care that may lead to a minor injury.

### · Management of Cybersecurity

Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (issued on October 2, 2014) was also followed to consider issues related to cybersecurity in the design and development process of this device.

# Clinical Studies

Not applicable. Clinical testing was not deemed to be required to show substantial equivalence. We relied on non-clinical testing and compliance with standards.

#### VIII. CONCLUSIONS

After analyzing bench tests, it is the conclusion of First Source Inc. that the i®Flex M/i9Flex MD Mobile X-ray System is as safe and effective as the predicate device, has the same indications for use, has few technological differences, which are addressed through performance testing and compliance with the standards listed above, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K221081](https://fda-staging.innolitics.com/device/K221081)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
