K221075 · Sichuan Prius Biotechnology Co., Ltd. · FPA · Nov 21, 2022 · General Hospital
Device Facts
Record ID
K221075
Device Name
Infusion Sets for Single Use
Applicant
Sichuan Prius Biotechnology Co., Ltd.
Product Code
FPA · General Hospital
Decision Date
Nov 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Infusion Sets for Single Use is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.
Device Story
Infusion Sets for Single Use are gravity-fed devices designed to deliver fluids from a container to a patient's vascular system. The device consists of a spike (with protective cap), drip chamber, drip tube, flow regulator, flexible tubing, and a conical fitting (luer lock or luer slip). Optional components include an air vent, fluid filter, clamp, and injection site. The device is operated manually by healthcare professionals in clinical settings. By regulating flow through the tubing, the device enables controlled fluid administration, supporting patient hydration or medication delivery. The device is provided sterile for single use.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with ISO 8536-4 (infusion performance), ISO 10993 series (biocompatibility including cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, and pyrogenicity), and ASTM standards for packaging integrity and seal strength.
Technological Characteristics
Gravity-fed infusion set. Materials: ABS, PVC, HDPE. Components: spike, drip chamber, flow regulator, flexible tubing, conical fitting (luer lock/slip). Optional: air vent, 15μm fluid filter, clamp, injection site. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 8536-4, ISO 80369-7/20, ISO 10993, ASTM F756, ASTM F88, ASTM F1929, ASTM F1886, USP <151>, USP <85>, USP <788>.
Indications for Use
Indicated for patients requiring fluid administration from a container to the vascular system via an intravenous needle or catheter.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2022
Sichuan Prius Biotechnology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K221075
Trade/Device Name: Infusion Sets for Single Use Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: October 21, 2022 Received: October 21, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221075
Device Name Infusion Sets for Single Use
Indications for Use (Describe)
The Infusion Sets for Single Use is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221075 - 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K221075
- Date of Preparation: 11/21/2022 1.
- 2. Sponsor Identification
#### Sichuan Prius Biotechnology Co., Ltd.
No.2 Prius Road, Luo Long Industrial Park Nanxi District, 644104 Yibin City, Sichuan Province, PEOPLE'S REPUBLIC OF CHINA
Establishment Registration Number: Not registered yet
Contact Person: Yan Liu Position: Management Representative Tel: +86-831-3839889 Fax: +86-831-3839887 Email: 48363603@qq.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
#### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### Identification of Proposed Device 4.
Trade Name: Infusion Sets for Single Use Common Name: Intravascular Administration Set
Regulatory Information Classification Name: Intravascular Administration Set Classification: II Product Code: FPA Regulation Number: 21 CFR 880.5440 Review Panel: General Hospital
Indication for Use Statement:
The Infusion Sets for Single Use is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.
#### Device Description
The proposed device consists of basic configuration and optional components. The basic configuration consists of protective cap of spike, drip chamber, drip tube, flow regulator, flexible tube, and conical fitting. Air vent, fluid filter, clamp, and injection site are optional components. The device is provided sterile, single use.
- న. Identification of Predicate Device
510(k) Number: K163160 Product Name: Sterile Single-use Infusion Set
- 6. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device<br>K163160 | Remark |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | FPA | FPA | Same |
| Regulation Number | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| Class | Class II | Class II | Same |
| Indication for Use | The Infusion Sets for Single Use is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into | The device is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein. | Same |
#### Table 1 General Comparison
{5}------------------------------------------------
| | the vein. | | | | |
|-------------------------------|---------------------------|---------------------------------------------|------------------------------------------|---------------------------------------|-----------|
| Configuration | Basic<br>Configuration | Protective cap of spike | Protector Cap of Spike | Same | |
| | | Spike | Spike | Same | |
| | | Drip chamber | Drip chamber | Same | |
| | Flexible Tube | Flexible Tube | Same | | |
| Configuration | | Conical fitting (luer lock<br>or luer slip) | Conical fitting (luer lock or luer slip) | Same | |
| | Optional<br>configuration | Injection site | Injection Site | Same | |
| | | Fluid filter | Fluid filter | Same | |
| | | Air vent | Air vent | Same | |
| | | Clamp | Clamp | Same | |
| Operation Mode | | For manual use only | For manual use only | Same | |
| Label/Labeling | | Conform with 21 CFR Part 801 | Conform with 21 CFR Part 801 | Same | |
| Filter Characteristics | | 15μm | 15μm | Same | |
| Tubing Diameter | | 3.9mm | 3.9mm | Same | |
| Length (mm) | | 1635mm, 1665mm | 1610, 1900mm | Different | |
| Capacity (ml) | | 18ml | 19, 21ml | Different | |
| Infusion<br>Performance | Set | Comply with ISO 8536-4 | Comply with ISO 8536-4 | Same | |
| | | | | Same | |
| Patient contact Material | | | | | |
| Spike | | Acrylonitrile Butadiene Styrene (ABS) | High-density Polyethylene (HDPE) | Different | |
| Drip chamber | | Poly Vinyl Chloride (PVC) | Acrylonitrile Butadiene Styrene (ABS) | Different | |
| Flexible Tube | | Poly Vinyl Chloride (PVC) | Poly Vinyl Chloride (PVC) | Same | |
| conical fitting | | | Poly Vinyl Chloride (PVC) | Poly Vinyl Chloride (PVC) | Same |
| Injection site | | | Poly Vinyl Chloride (PVC) | Acrylonitrile Butadiene Styrene (ABS) | Different |
| Fluid filter | | | Acrylonitrile Butadiene Styrene (ABS) | Acrylonitrile Butadiene Styrene (ABS) | Same |
| Biocompatibility | | | | | |
| Cytotoxicity | | | No Cytotoxicity | No Cytotoxicity | |
| Intracutaneous<br>Reactivity | | | No Intracutaneous Reactivity | No Intracutaneous Reactivity | |
| Skin Sensitization | | | No Skin Sensitization | No Skin Sensitization | |
| Acute Systemic Toxicity | | No Acute Systemic Toxicity | No Acute Systemic Toxicity | Same | |
| Hemolysis | | No Hemolysis | No Hemolysis | | |
| Pyrogen | | No Pyrogen | No Pyrogen | | |
| Subacute Systemic<br>Toxicity | | No subacute systemic toxicity | / | | |
{6}------------------------------------------------
| Method | EO sterilized | EO sterilized | Same |
|-----------------|------------------|------------------|------|
| SAL | $10^{-6}$ | $10^{-6}$ | Same |
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same |
#### Different-Length
The total length for the proposed device is different from predicate device. However, the performance test has been conducted on the proposed device and the test result can meet the requirements of ISO 8536-4. Therefore, this difference does not raise new questions of safety and effectiveness for the proposed device.
#### Different-Capacity
The total capacity for the proposed device is different from predicate device. However, the performance test has been conducted on the proposed device and the test result can meet the requirements of ISO 8536-4. Therefore, this difference does not raise new questions of safety and effectiveness for the proposed device.
#### Different -Patient Contact Material
The patient contact materials for proposed device are different from predicate device. However, the biocompatibility testing per ISO 10993-1 was performed and the results show there are no adverse effect. Therefore, the differences do not raise new questions of safety and effectiveness for the proposed device.
## Similar- Biocompatibility
The biocompatibility test was performed on the proposed device and one additional endpoint was evaluated compared to the predicate device, which is subacute toxicity. The test results for the endpoints show there is no adverse effect on the material. Therefore, the provided biocompatibility testing supports substantial equivalence to the predicate device.
- Non-Clinical Test Conclusion 7.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-5:2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization
- A ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity
- ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials >
- ISO 10993-7:2008 Biological Evaluation of Medical Devices-Part 7: Ethylene Oxide >
{7}------------------------------------------------
Sterilization Residuals
- > ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- > ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for intravascular or hypodermic applications
- > ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-Part 20: Common test methods.
- > ISO 8536-4:2019 Infusion equipment for medical use- Part 4: Infusion Sets for Single Use, gravity feed.
- > ISO 8536-14:2016 Clamps and flow regulators for transfusion and infusion equipment without fluid contact
- > USP <151>Pyrogen Test
- > USP <85> Bacterial Endotoxins Test
- > USP <788> Particular Matter in Injections.
- 8. Clinical Test Conclusion
No clinical study is included in this submission.
- Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.