K221066 · Sichuan Prius Biotechnology Co., Ltd. · FMI · Oct 4, 2022 · General Hospital
Device Facts
Record ID
K221066
Device Name
Sterile Hypodermic Needles for Single Use
Applicant
Sichuan Prius Biotechnology Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Oct 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Needles for Single Use are intended to be used with a luer slip or luer lock syringe for general purpose fluid injection/ aspiration.
Device Story
Sterile hypodermic needles for single use; consist of stainless steel needle, polypropylene hub, and protective cap. Used manually by clinicians to aspirate or inject fluids. Compatible with luer slip or luer lock syringes. Available in various gauges (16G-30G) and lengths (13mm-38mm). Lubricated with polydimethylsiloxane. Device is sterile, single-use, and intended for limited contact (<24 hours) with the blood path. Healthcare providers use the device to deliver medications or withdraw fluids, directly impacting patient treatment and diagnostic procedures.
Clinical Evidence
No clinical data. Bench testing only. Conformance demonstrated with ISO 7864, ISO 9626, ISO 80369-7, ISO 80369-20, and USP <788>. Biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, hemolysis, material-mediated toxicity) and sterilization validation (ISO 11135) were performed.
Technological Characteristics
Materials: Stainless Steel SUS304 (needle), Polypropylene (hub/cap), Polydimethylsiloxane (lubricant), Epoxy adhesive. Dimensions: 16G-30G gauges; 13mm-38mm lengths. Wall types: Thin Wall (TW) and Regular Wall (RW). Connectivity: None (manual). Sterilization: Ethylene Oxide (EO) to SAL 10^6. Standards: ISO 7864, ISO 9626, ISO 80369-7/20, ASTM F1886, ASTM F88, ASTM F1929, ASTM D4169, ASTM F1980.
Indications for Use
Indicated for general purpose fluid injection or aspiration in patients requiring such procedures, using a luer slip or luer lock syringe.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Sterile Hypodermis Needles for Single Use (K180417)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 4, 2022
Sichuan Prius Biotechnology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai. 200120 China
Re: K221066
Trade/Device Name: Sterile Hypodermic Needles for Single Use Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: August 26, 2022 Received: September 8, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221066
Device Name
Sterile Hypodermic Needles for Single Use
Indications for Use (Describe)
The Sterile Hypodermic Needles for Single Use are intended to be used with a luer slip or luer lock syringe for general purpose fluid injection/ aspiration.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221066 -510 K Summary
- 1. Date of Preparation: 10/08/2022
- 2. Sponsor Identification
## Sichuan Prius Biotechnology Co., Ltd.
No.2 Prius Road, Luo Long Industrial Park Nanxi District, 644104 Yibin City, Sichuan Province, PEOPLE'S REPUBLIC OF CHINA
Establishment Registration Number: Not registered yet
Contact Person: Yan Liu Position: Management Representative Tel: +86-831-3839889 Fax: +86-831-3839887 Email: 48363603@qq.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### Identification of Proposed Device 4.
Trade Name: Sterile Hypodermic Needles for Single Use Common Name: Syringe Needle
Regulatory Information Classification Name: Needle, Hypodermic, Single Lumen Classification: II; Product Code: FMI; Regulation Number: 21CFR 880.5570; Review Panel: General Hospital;
- న. Predicate
K180417, Sterile Hypodermis Needles for Single Use
#### 6. Device Description
The Sterile Hypodermic Needles for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of needle and a needle hub. The proposed device is available in variety of needle length and needle gauge. The proposed device is compatible for use with a luer slip or luer lock syringe.
| Needle Sizes | | |
|---------------|--------------|------|
| Needle Length | Needle gauge | Wall |
| 32mm | 16G | TW |
| 38mm | 16G | TW |
| 32mm | 18G | TW |
| 38mm | 18G | TW |
| 32mm | 19G | TW |
| 38mm | 19G | TW |
| 32mm | 20G | TW |
| 38mm | 20G | TW |
| 32mm | 21G | TW |
| 38mm | 21G | TW |
| 32mm | 22G | TW |
| 38mm | 22G | TW |
| 19mm | 23G | TW |
| 25mm | 23G | TW |
| 32mm | 23G | TW |
Needle specification:
{5}------------------------------------------------
| 19mm | 24G | RW |
|------|-----|----|
| 25mm | 24G | RW |
| 32mm | 24G | RW |
| 16mm | 25G | RW |
| 19mm | 25G | RW |
| 25mm | 25G | RW |
| 13mm | 26G | RW |
| 16mm | 26G | RW |
| 19mm | 26G | RW |
| 13mm | 27G | RW |
| 16mm | 27G | RW |
| 19mm | 27G | RW |
| 13mm | 28G | RW |
| 16mm | 28G | RW |
| 13mm | 29G | RW |
| 16mm | 29G | RW |
| 13mm | 30G | RW |
| 16mm | 30G | RW |
#### 7. Indication for Use
| Characterization | Proposed device | Predicate Device | | |
|-----------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | Sterile Hypodermic Needles for<br>Single Use<br>K221066 | Sterile Hypodermic Needles for<br>Single Use<br>K180417 | | |
| | Indication for Use | The Sterile Hypodermic Needles<br>for Single Use are intended to be<br>used with a luer slip or luer lock<br>syringe for general purpose fluid<br>injection/ aspiration. | The Sterile Hypodermic Needles for<br>Single Use are intended to be used<br>with a luer slip or luer lock syringe<br>for general purpose fluid injection/<br>aspiration. | |
| | | Prescription or OTC (over<br>the counter) | Prescription use | Prescription use |
| | | Technological<br>Characteristic | Proposed Device | Predicate Device |
| | | Sterile Hypodermic Needles for<br>Single Use | Sterile Hypodermic Needles for<br>Single use | |
| | K221066 | K180417 | | |
| Configuration | Needle hub<br>Protective cap<br>Needle | Needle hub<br>Protective cap<br>Needle | Same | |
| Operation Mode | For manual use only | For manual use only | Same | |
| Single Use | Single Use | Single Use | Same | |
| Needle<br>Performance | Complied with<br>ISO 7864,<br>ISO 9626 | Complied with<br>ISO 7864,<br>ISO 9626 | Same | |
| Needle Gauge | 16G, 18G, 19G, 20G, 21G,<br>22G, 23G, 24G, 25G, 26G,<br>27G, 28G, 29G, 30G | 14G, 15G, 16G, 17G, 18G,<br>19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G,<br>29G, 30G | Different<br>See comment #<br>1 | |
| Needle Length | 13mm, 16mm, 19mm, 25mm,<br>32mm, 38mm | 6mm, 9mm, 12mm, 13mm,<br>16mm, 20mm, 25mm, 38mm,<br>50mm, 60mm | Different<br>See comment #<br>2 | |
| Wall type | TW: 16G, 18G, 19G, 20G, 21G,<br>22G, 23G<br>RW: 24G, 25G, 26G, 27G, 28G,<br>29G, 30G | TW: 14G, 15G, 16G, 17G, 18G,<br>19G, 20G, 21G, 22G, 23G,<br>29G, 30G<br>RW: 14G, 15G, 16G, 18G, 19G,<br>20G, 21G, 22G, 24G, 25G,<br>26G, 27G, 29G, 30G | Different<br>See comment #<br>3 | |
| Needle | Stainless Steel SUS304 | Stainless Steel SUS304 | Same | |
| Needle hub | Polypropylene (PP) | Polypropylene (PP) | Same | |
| Needle cap | Polypropylene (PP) | Polypropylene (PP) | Same | |
| Lubricants | Polydimethylsiloxane | Polydimethylsiloxane | Same | |
| Adhesive | Epoxy adhesive | HTU-3312 Glue | Different<br>See comment #<br>4 | |
#### 8. Technological Characteristic
The table below includes a comparison of the technological characteristics between the new device and those of the predicate.
# Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
Discussions of differences in technological characteristics
Comment # 1 – Needle Gauge
The proposed device has the less gauges compared to the predicate device and the proposed gauge can be covered by the range of needle gauge available in the predicate device. In addition, the needle performance has been tested and results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626.
{7}------------------------------------------------
## Comment # 2 – Needle Length
The proposed device has less needle lengths compared to the proposed needle length specifications can be covered by the range of available needle lengths for the predicate device. In addition, the needle performance has been tested and the results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626.
# Comment # 3 – Wall Type
The proposed device is available in thin wall and normal wall. The predicate device also has thin wall and normal wall. The performance data results demonstrate that the needs the requirements of ISO 7864 and ISO 9626.
### Comment # 4 – Material of Adhesive
The proposed device has different adhesive compared to the predicate device. This difference does not impact intended use and does not raise new questions of safety and effectiveness. However, biocompatibility test has been conducted on the proposed device and the test result does not show anv adverse effect. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.
#### 9. Non- Clinical Testing
The device described in this summary the Sterile Hypodermic Needle for Single Use. were tested and demonstrated to be in conformance with the following FDA recognized standards:
- > ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- > ISO 80369-20:2015 Small-bore connectors for liguids and gases in healthcare applications-Part 20: Common test methods
- > ISO 7864: 2016, Sterile Hypodermic Needles for Single Use.
- > ISO 9626:2016 Stainless Steel Needle Tubing for the Manufacture of Medical Devices
### Biocompatibility
In accordance with ISO10993-1 the needle are classified as: Externally Communicating Device, Blood Path direct, Limited Contact (< 24hrs). The following testing was conducted:
- > Cytotoxicity
- A Irritation
- A Skin Sensitization
- > Acute Systemic Toxicity
- > Pyrogenicity
- Hemolvsis A
- A Material Mediated Toxicity
{8}------------------------------------------------
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
Sterility, Shipping, and Shelf -life
The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 106. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 20 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 5 year shelf-life.
- . Sterile Barrier Packaging performed on the proposed device: ○ Visual Inspection ASTM F1886 / F1886M-16 0 Seal Strength ASTM F88/F88-15 o Dye penetration ASTM F1929-15
- . Simulated transportation testing in accordance to ASTM D4169-16 on final, packaged, and sterile device.
Shelf-life of 5-years is validated using FDA recognized standard ASTM F1980-16 Standard . Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
### 10. Clinical Test Conclusion
No clinical study is included in this submission.
### 11. Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Sterile Hypodermic Needles for Single Use is as safe, as effective, and performs as well as the legally marketed predicative device cleared under K180417.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.