K221043 · Mectronic Medicale S.R.L. · PBX · Feb 17, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221043
Device Name
Doctor Tecar Plus, Doctor Tecar Smart
Applicant
Mectronic Medicale S.R.L.
Product Code
PBX · General, Plastic Surgery
Decision Date
Feb 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Doctor Tecar Plus and Doctor Tecar Smart devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.
Device Story
Doctor Tecar Plus and Doctor Tecar Smart are diathermy devices providing topical heating via radiofrequency (RF) energy. The device uses capacitive and resistive delivery modes to elevate tissue temperature up to 45°C. It consists of a touchscreen, multi-function knob (Plus model), plate, and handpiece. Operated by clinicians in a professional setting, the device allows manual or automatic energy delivery. The clinician monitors the treatment via the interface to manage intensity and duration. By increasing local tissue temperature, the device aims to alleviate pain, reduce muscle spasms, and improve local circulation. Safety features include a patient-operated switch to deactivate energy emission.
Clinical Evidence
Bench testing only. Testing evaluated the device's ability to maintain skin surface temperature (40-45°C) for 12-20 minutes. Study included 3 subjects with varying skin tones and somatotypes across multiple body sites. Testing utilized both automatic and manual modes with nine different applicators. Results confirmed consistent temperature maintenance with no discrepancies found.
Technological Characteristics
Radiofrequency diathermy device; capacitive and resistive energy delivery. Frequencies: 300 kHz–1.5 MHz (Plus), 448 kHz–750 kHz (Smart). Power: 200 VA (Plus), 70 VA (Smart). Standards: CEI EN 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6, IEC 62304, IEC 62366, ISO 14971. Biocompatibility per ISO 10993-1, -5, -10. User interface: Touchscreen and knob. Patient safety switch included.
Indications for Use
Indicated for patients requiring topical heating for pain relief, muscle spasms, or increased local circulation. No specific age or gender restrictions; intended for use on intact skin.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
February 17, 2023
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Mectronic Medicale S.r.l. Gloria Aloisini Quality Manager Via Orio al Serio 15 Grassobbio, BG 24050 Italy
Re: K221043
Trade/Device Name: Doctor Tecar Plus, Doctor Tecar Smart Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX Dated: January 16, 2023 Received: January 18, 2023
Dear Gloria Aloisini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Colin K. Chen -S
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K221043
Device Name Doctor Tecar Plus Doctor Tecar Smart
#### Indications for Use (Describe)
The Doctor Tecar Plus and Doctor Tecar Smart devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="1" width="15" x="0" y="0"></rect> <line stroke="black" stroke-width="2" x1="2" x2="13" y1="2" y2="13"></line> <line stroke="black" stroke-width="2" x1="13" x2="2" y1="2" y2="13"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="1" width="15" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a logo for a company called "Electronic Medicale". The logo features a blue star on the left side, followed by the word "Electronic" in a stylized font. Below "Electronic", the word "Medicale" is written in a smaller, sans-serif font, also in blue. The overall design has a slightly three-dimensional effect, with subtle shading and depth.
# 510(K) SUMMARY
510(k) Number K221043 510(k) Application Traditional
### CONTACT DETAILS
Mectronic Medicale S.r.l. Via Orio al Serio, 15 24050 Grassobbio (BG) ITALY Phone: +39 035 656080, FAX: +39 035 65736 Official Correspondent Gloria Aloisini Mectronic Medicale S.r.l. gloriaaloisini@mectronicmedicale.it Phone: +39 035 656080, FAX: +39 035 65736
### TRADE NAME
Doctor Tecar Plus Doctor Tecar Smart Common Name Device Common name: Radiofrequency Device Classification name Classification name: Electrosurgical, cutting and coagulation accessories Classification: Class II Regulation Nos .: 21 CFR 878.4400 Product Codes: 79 PBX - Massager, Vacuum, Radio Frequency Induced Heat
### PREDICATE DEVICE
The device under submission is substantially equivalent to the predicate devices:
- NuEra Tight RF Family, K200359, Bios S.r.l.
### DEVICE DESCRIPTION
Doctor Tecar Plus and Doctor Tecar Smart are devices for diathermy. This therapy develops an increase in temperature in the area of the body which is treated through the passage of a particular electrical current. Through twin delivering modes, capacitive and resistive, the operator can set a specific treatment. The purpose of the treatment is to raise the temperature inside the tissue at a value maximum to 45 °C. The device is composed by a touch screen and a Multi-
{4}------------------------------------------------
K221043 510(k) Summary
Image /page/4/Picture/2 description: The image shows a logo for a company called "Electronic Medicale". The logo features a stylized star shape in dark blue, positioned to the left of the company name. The text "ELECTRONIC" is in a stylized font, with the letters connected and a horizontal line above them. Below "ELECTRONIC" is the word "MEDICALE" in a smaller font.
Function knob (only Plus model) for managing all the device's functions, a plate, an handpiece for the delivery of radiofrequency energy, software and an on/off button to activate and deactivate the energy emission.
## DEVICE INTENDED USE
The Doctor Tecar Plus and Doctor Tecar Smart devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.
# PREDICATE DEVICE COMPARISON
The product specification, functionality, indications for use, and treatment parameters of the Doctor Tecar Smart and Doctor Tecar Plus are the same or very similar to the legally marketed predicate and reference devices. The technological similarities and differences between the doctor Tecar Smart/Plus and the predicate devices are described in the following table.
| Feature | Predicate Device | Doctor Tecar Device | Comments |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------|
| Regulation and product<br>Classification Code | 21 CFR 878.4400<br>PBX | 21 CFR 878.4400<br>PBX | - |
| Intended Use / Indication<br>For Use | The device is is intended to<br>provide topical heating for the<br>purpose of elevating tissue<br>temperature for the<br>treatment of selected medical<br>conditions such as relief of<br>pain, muscle spasms, and<br>increase in local circulation. | Same | - |
| Type of use | Prescription Use (21 CFR 801,<br>Subpart D) | Same | - |
| Target Population | Whole population | Same | - |
| Anatomical Site | The devices are developed as<br>portable devices indicated in<br>therapy for muscular &<br>skeletal pathologies. | Same | - |
| Clinical use | Prescription type | Same | - |
| Energy Used And/Or<br>Delivered | 250 W | PLUS : 200 VA<br>SMART: 70 VA | No significant<br>difference |
| Frequencies | 470 kHz; 1 MHz; 2 MHz; 4<br>MHz ; 6 MHz | PLUS : 300 kHz - 448 kHz - 500<br>kHz - 750 kHz - 1 MHz - 1.2 MHz<br>- 1.5 MHz | No significant<br>difference |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for a company called "Electronic Medicale". The logo features a blue star on the left side, followed by the company name in a stylized font. The word "Electronic" is larger and more prominent, while the word "Medicale" is smaller and positioned below it.
#### K221043 510(k) Summary
| Feature | Predicate Device | Doctor Tecar Device | Comments |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-------------------------------|
| | | SMART : 448 kHz - 750 kHz | No significant<br>difference |
| Type of energy | Radiofrequency Waves | Radiofrequency Waves | - |
| Active Electrode | Round shape | Same | - |
| Neutral Electrode | Rectangular shape | Same | - |
| Design | Radiofrequency Energy with<br>intensity adjustment from 0%<br>to 100% | Same | - |
| Patient Safety Switch | Available | Same | - |
| Human Factors | Graphical user interface<br>allows a better user<br>experience and proper<br>therapy selection | Same<br>IEC 60601-1-6 | - |
| Standards Met | IEC 60601-1;<br>IEC 60601-1- 2<br>IEC 60601-2-2<br>IEC 60601-1- 6<br>IEC 62304<br>ISO 14971 | CEI EN 60601-1<br>CEI EN 60601-1-2<br>IEC 60601-2-2<br>IEC 60601-1-6<br>IEC 62304<br>IEC 62366<br>UNI EN ISO 14971 | No significant<br>differences |
| Biocompatibility | IEC 60601 - 1<br>ISO 10993 | IEC 60601 - 1<br>ISO 10993 | - |
| Compatibility With The<br>Environment And Other<br>Devices | IEC 60601-1-2 | IEC 60601-1-2 | - |
| Sterility | N.A. | N.A. | - |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | - |
| Mechanical Safety | IEC 60601-1 | IEC 60601-1 | - |
| Chemical Safety | IEC 60601-1 | IEC 60601-1 | - |
| Thermal Safety | IEC 60601-1 | IEC 60601-1 | - |
#### SAFETY TEST
The device has been tested and found in compliance with the following standards
- CEI EN 60601-1 Electro-medical equipment – General requirements for safety.
- . CEI EN 60601-1-2 Electro-medical equipment - General requirements for safety – Secondary standard: Electromagnetic compatibility - Prescriptions and tests.
- . IEC 60601-2-2 Medical electrical equipment. Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
{6}------------------------------------------------
K221043 510(k) Summary
Image /page/6/Picture/2 description: The image shows a logo for a company called "Electronic Medicale". The logo features a blue star on the left side, followed by the word "Electronic" in blue, stylized font. Below "Electronic" is the word "Medicale" in a smaller, blue, stylized font. The logo has a slight shadow effect, giving it a three-dimensional appearance.
- IEC 60601-1-6 Medical electrical equipment- Part 1: General requirements for basic safety and essential Performance. Collateral standard: Usability.
- IEC 62304 Medical device software - Software life-cycle processes.
- IEC 62366 Medical devices -Application of usability engineering to medical devices .
- . UNI EN ISO 14971 Medical devices - Application of risk management to medical devices.
#### BENCH TEST
A test was performed to in order to to demonstrate the capability of Doctor Tecar Smart/Plus to maintain the temperature on the surface of the human skin. Based on the treatment recommendations, the test report demonstrates that Doctor Tecar Plus is capable of maintaining a skin surface temperature of 40℃ - 45℃ for 10-20 minutes after the temperature of the skin has reached 40℃. The test recorded actual skin temperature data to demonstrate safety when using the device as intended. The temperature test was done evaluating the device on 3 different people and for each people it was done the test on three different place on the body that are consistent with our instructions and indication for use. In this test we have included people with light and dark skin and people with different somatotype. For each person and for each place the test was performed in automatic mode, using electrodes and in manual mode with nine different applicators: small flat applicator, small rounded applicator, medium flat applicator, medium rounded applicator, big flat applicator, big rounded applicator, medium capacitive applicator, small bipolar applicator and big bipolar applicator. The temperature of the skin was measured every minutes for a treatment period of 12 minutes after the temperature of the skin has reached the 40°C. No discrepancies were found.
### BIOCOMPATIBILITY
The Doctor Tecar Smart and Doctor Tecar Plus patient-contacting components have surfaces that, in accordance with the instructions for use, are categorized as 'surface devices', contacting 'intact skin', for a 'limited' time period (< 24 h), resulting in biocompatibility evaluation as follows, in accordance with Table A.1 of ISO 10993-1 and FDA guidance:
- Cytotoxicity: ISO 10993-5
- . Sensitization: ISO 10993-10
- . Irritation: ISO 10993-10
These patient-contacting components have all previously been cleared by FDA by means of successful 510(k) submissions. No further biocompatibility data is therefore included within this submission.
# CONCLUSIONS
Doctor Tecar Plus and Doctor Tecar Smart generate radio frequency for treatment of selected tissues. They share the same basic characteristics and the same intended use as the predicate device. Moreover the safety and bench tests demonstrate that the the device is safe and performs as intended, and do not raise any concern about safety and effectiveness. There are no substantial differences between the product defined in this 510(k) submission and the predicate device that is,
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo for a company called "Electronic Medical". The logo features a large, dark blue star on the left side, with the company name written in a stylized font to the right of the star. The text is also in dark blue, and the word "MEDICAL" is stacked below "ELECTRONIC".
#### K221043 510(k) Summary
they are essentially equivalents for the requested intended use. Therefore, the proposed device is substantially equivalent to the NuEra Tight RF Radiofrequency Device, cleared under 510(k) registration K200359.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.