K221042 · Gi Supply, Inc. · BTA · Apr 19, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221042
Device Name
Renova RP Centesis Pump
Applicant
Gi Supply, Inc.
Product Code
BTA · General, Plastic Surgery
Decision Date
Apr 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RenovaRP® Centesis Pump is intended to be used in conjunction with the RenovaRP Tube Set and the RenovaRP Fluid Drainage Bags to remove ascitic fluid from the abdominal cavity for paracentesis, and remove pleural effusion from the thoracic cavity for thoracentesis. The Renova RP Centesis Pump is intended to be used by trained healthcare professionals knowledgeable about paracentesis and thoracentesis.
Device Story
RenovaRP Centesis Pump is a powered peristaltic suction pump; used in hospital/clinic settings by trained healthcare professionals. Device connects to RenovaRP Tube Set and Fluid Drainage Bags to remove ascitic fluid or pleural effusion. Pump operates on AC power (100-230V); provides 700 mL/min flow rate. Healthcare providers use the system to perform paracentesis and thoracentesis procedures. Output is the physical removal of fluid from body cavities into drainage bags. Benefits include efficient, controlled fluid aspiration for patient symptom relief.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via simulated use, functional testing, distribution testing, and standards compliance (IEC 60601-1, IEC 60601-1-2, EN IEC 62366-1). Biocompatibility testing (ISO 10993-1) and cleaning validation (ISO 17664) performed on components.
Technological Characteristics
Powered peristaltic suction pump; AC-powered (100-230V, 50/60Hz). Flow rate 700 mL/min. Dimensions 33x23x33 cm; weight 8.5 lbs. Class I, Type B electrical safety. Reusable pump; single-patient-use sterile disposable tube sets and drainage bags. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for removal of ascitic fluid from the abdominal cavity (paracentesis) and pleural effusion from the thoracic cavity (thoracentesis) in patients requiring fluid drainage. Intended for use by trained healthcare professionals.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
RenovaRP® Tube Set and Fluid Drainage Bags (K970187)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 19, 2023
GI Supply, Inc. Thomas Saladin Global Regulatory Lead 5069 Ritter Road Suite 104 Mechanicsburg, Pennsylvania 17055
Re: K221042
Trade/Device Name: Renova RP Centesis Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: March 20, 2023 Received: March 20, 2023
## Dear Thomas Saladin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
{1}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S -Trumbore -S Date: 2023.04.19 Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221042
Device Name RenovaRP® Centesis Pump System
Indications for Use (Describe)
The RenovaRP® Centesis Pump is intended to be used in conjunction with the RenovaRP Tube Set and the RenovaRP Fluid Drainage Bags to remove ascitic fluid from the abdominal cavity for paracentesis, and remove pleural effusion from the thoracic cavity for thoracentesis.
The Renova RP Centesis Pump is intended to be used by trained healthcare professionals knowledgeable about paracentesis and thoracentesis.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2812 2.71875H2.71875C1.52344 2.71875 0.5 3.74219 0.5 4.9375V13.2812C0.5 14.4766 1.52344 15.5 2.71875 15.5H13.2812C14.4766 15.5 15.5 14.4766 15.5 13.2812V4.9375C15.5 3.74219 14.4766 2.71875 13.2812 2.71875ZM13.2812 13.2812H2.71875V4.9375H13.2812V13.2812ZM6.1875 11.0938L2.71875 7.625L4.28125 6.0625L6.1875 7.96875L11.7188 2.4375L13.2812 4L6.1875 11.0938Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="2.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I.SUBMITTER/510(k) SPONSOR
Sponsor/Manufacturer GI Supply, Inc. 5069 Ritter Road Suite 104 Mechanicsburg, PA 17055 USA Phone: (800)-451-5797
#### Contact Person
Thomas Saladin Sr. Regulatory Affairs Specialist Phone: (952)-471-6665 Email: tsaladin@laborie.com
## Alternate Contact Meghan Moore Manager Regulatory Affairs Email: mmoore@laborie.com
## Preparation Date
April 14, 2023
## II. DEVICE NAME/CLASSIFICATION
| Device Trade/Proprietary Name | RenovaRP® Centesis Pump System |
|----------------------------------|--------------------------------|
| Device Common or Usual Name | Suction Pump, Peristaltic Pump |
| Device Regulatory Classification | Class II |
| Device Classification Regulation | 21 CFR 878.4780 |
| Product Code | BTA |
| Submission Type | Traditional 510(k) |
| Classification Panel | General & Plastic Surgery |
#### III. PREDICATE DEVICE
K970186, RenovaRP® Paracentesis Pump System
#### IV. DEVICE DESCRIPTION
The RenovaRP® Centesis Pump is intended to be used in conjunction with the RenovaRP® Tube Set and RenovaRP® Fluid Drainage Bags (provided separately from the pump) to remove fluid from the abdominal cavity for paracentesis, and pleural effusion from the thoracic cavity for thoracentesis.
## V. INTENDED USE / INDICATIONS FOR USE
#### Intended Use / Indications for Use
The RenovaRP® Centesis Pump is intended to be used in conjunction with the RenovaRP Tube Set and the RenovaRP Fluid Drainage Bags to remove ascitic fluid from the abdominal cavity for paracentesis, and remove pleural effusion from the thoracic cavity for thoracentesis.
The Renova RP Centesis Pump is intended to be used by trained healthcare professionals knowledgeable about paracentesis and thoracentesis.
{4}------------------------------------------------
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The detailed substantial equivalence comparison of the similarities and differences between the RenovaRP® Centesis Pump System [subject device] and the RenovaRP® Paracentesis Pump (K970186) [predicate device] is provided in the table below. Please note that the devices have the same manufacturer, and only differ with regard to indications for use and proprietary/trade name. The fundamental design and operational and technological characteristics of the subject device remain the same.
No performance data was needed to evaluate the expansion of the indications for use to include thoracentesis. The change of the subject device's indications does not increase risks or impact safety or effectiveness when compared to the predicate device.
| Regulatory | | | Similarities / |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Information | [Subject Device] | [Predicate Device] | Differences |
| Manufacturer | GI Supply, Inc. | GI Supply, Inc. | Same |
| Device Trade or<br>Proprietary Name | RenovaRP® Centesis Pump<br>System | RenovaRP® Paracentesis Pump | Different |
| 510(k) Number | K221042 | K970186 | --- |
| Device Class | Class II | Class II | Same |
| Device<br>Classification<br>Name | General & Plastic Surgery | General & Plastic Surgery | Same |
| Device Common<br>Name | Pump, Portable, Aspiration<br>(Manual or Powered) | Pump, Portable, Aspiration<br>(Manual or Powered) | Same |
| Product Code | BTA | BTA | Same |
| Regulation<br>Number | 21 CFR 878.4780 | 21 CFR 878.4780 | Same |
| | Design Features and Capabilities of the Device | | |
| Indications for Use | The RenovaRP® Centesis Pump is<br>intended to be used in<br>conjunction with the RenovaRP<br>Tube Set and the RenovaRP Fluid<br>Drainage Bags to remove ascitic<br>fluid from the abdominal cavity<br>for paracentesis, and remove<br>pleural effusion from the thoracic<br>cavity for thoracentesis.<br>The Renova RP Centesis Pump is<br>intended to be used by medically<br>trained healthcare professionals<br>knowledgeable about<br>paracentesis and thoracentesis. | The RenovaRP® Paracentesis<br>Pump is intended as a peristaltic<br>pump to remove ascitic fluid<br>from the abdominal cavity in<br>conjunction with the RenovaRP®<br>Paracentesis Kit. The RenovaRP®<br>Paracentesis Pump is intended to<br>be used by medically trained<br>healthcare professionals<br>knowledgeable about<br>paracentesis. | Different |
| Intended Use | Suction pump for removal of<br>fluids | Suction pump for removal of<br>fluids | Same |
| Prescription or<br>Over-the-Counter<br>(OTC) Use | Prescription Use | Prescription Use | Same |
| Use Environment | Hospital / Clinic | Hospital / Clinic | Same |
| Regulatory | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| Sterile | Pump is non-sterile.<br>Disposable components are<br>sterile. | Pump is non-sterile.<br>Disposable components are<br>sterile. | Same |
| Disposable versus<br>Non-Disposable | Pump is reusable.<br>Disposable components are<br>single patient use. | Pump is reusable.<br>Disposable components are<br>single patient use. | Same |
| Pump Physical<br>Specifications | 13x9x13 in (33x23x33 cm)<br>8.5 lbs (3.9kg) | 13x9x13 in (33x23x33 cm)<br>8.5 lbs (3.9kg) | Same |
| Pump Power | AC only model | AC only model | Same |
| Pump Input<br>Voltage | 100-230V | 100-230V | Same |
| Pump Input<br>Current | 2.5A | 2.5A | Same |
| Pump Input<br>Frequency | 50/60Hz | 50/60Hz | Same |
| Flow Rate | 700 mL/min | 700 mL/min | Same |
| Pump Electrical<br>Requirements | Class I, Type B | Class I, Type B | Same |
| Packaging | Pump is placed in a polybag in<br>between a top foam and bottom<br>foam insert prior to placement in<br>a 200# shipper box. Power cord<br>and IFU are placed in a polybag<br>on top of the pump under a foam<br>cap prior to box closure.<br><br>Disposable components are<br>provided separately from the<br>pump. Tube sets are individually<br>sterile packaged, with each 200#<br>shipper box containing five (5)<br>tube sets and one (1) IFU. Bags<br>are individually sterile packaged,<br>with each 200# shipper box<br>containing 20 bags. No IFU for<br>the collection bags is required. | Pump is placed in a polybag in<br>between a top foam and bottom<br>foam insert prior to placement in<br>a 200# shipper box. The power<br> | Same |
| Shelf Life | The pump has an expected<br>service life of 7-10 years.<br><br>Tube sets have a shelf life of 24<br>months, while the fluid drainage<br>bags have a shelf life of 25<br>months. | The pump has an expected<br>service life of 7-10 years.<br><br>Tube sets have a shelf life of 24<br>months, while the fluid drainage<br>bags have a shelf life of 25<br>months. | Same |
| Regulatory<br>Information<br>Manufacturer | GI Supply, Inc.<br>Design Features and Capabilities of the Device | GI Supply, Inc.<br>Design Features and Capabilities of the Device | Same |
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| RenovaRP® Tube<br>Set and Fluid<br>Drainage Bags | Part of the proposed device;<br>packaged and shipped separately<br>as they are consumables. | N/A | The RenovaRP®<br>Tube Set and<br>Fluid Drainage<br>Bags were cleared<br>via K970187. They<br>are the same as<br>provided under<br>K970187 except<br>the revised<br>labeling for the<br>expanded<br>indication to<br>include<br>thoracentesis.<br>Including the tube<br>set and fluid<br>drainage bags as<br>part of the<br>RenovaRP®<br>Centesis Pump<br>System does not<br>introduce new<br>questions of<br>safety and<br>efficacy. |
{5}------------------------------------------------
{6}------------------------------------------------
## VII. SUMMARY OF PERFORMANCE DATA AND PERFORMANCE TESTING CONCLUSIONS
The RenovaRP® Centesis Pump system was evaluated using GI Supply's risk management process in accordance with ISO 14971:2019 – Medical devices – Application of risk management to medical devices. All identified risks were adequately mitigated and were previously verified by means of nonclinical performance testing.
No testing was required in support of substantial equivalence to the predicate device as the subject and predicate devices are identical in terms of device design, fundamental technology, physical characteristics, performance, and intended use. The only proposed change to the device is its indications for use, which are being expanded to include thoracentesis.
The expansion of indications did not require any additional nonclinical testing to demonstrate safety and effectiveness nor to support the substantial equivalence to the subject device; however, nonclinical bench testing is being provided as part of this submission to verify device performance and safety.
## Summary of Nonclinical Performance Testing
A series of functional nonclinical performance testing was previously conducted on the RenovaRP® Centesis Pump including simulated use testing, functional testing, and distribution testing. The device was successfully evaluated for standards compliance to IEC 60601-1, Medical electrical equipment – Part 1: General requirements for basic safety and performance and 60601-1-2 – Medical electrical equipment
{7}------------------------------------------------
– Parts 1-2, General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests. Additionally, usability of the device was assessed by applying the evaluation process specified in EN IEC 62366-1:2015, Medical devices – Part 1: Application of usability engineering to medical devices (Annex C, Evaluation of a user interface of unknown provenance - UOUP).
A series of functional nonclinical performance testing was previously conducted on the RenovaRP® Tube Set including simulated use testing, post-accelerated aging testing, real-time aging, biocompatibility testing (cytotoxicity, irritation, and sensitization per ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing), and distribution testing.
A series of functional nonclinical performance testing was previously conducted on the RenovaRP® Fluid Drainage Bag including graduation mark testing, post-accelerated aging testing, real-time aging, and distribution testing.
## Cleaning Validation
A cleaning validation was conducted per ISO 17664, Processing of healthcare products on the RenovaRP® Centesis Pump.
#### VIII. CONCLUSIONS
The RenovaRP® Centesis Pump System [subject device] is substantially equivalent to the predicate device, RenovaRP® Paracentesis Pump (K970186) and its disposable components (K970187) with respect to device design, fundamental technology, physical characteristics, performance, and intended use for the removal of fluids. Device materials and packaging are identical. The disposable components are provided sterile, while both the subject and predicate pump component of the system are non-sterile, reusable, and have the same expected lifetime.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.