← Product Code [PSQ](/productcode/PSQ) · K221029

# PERI-GUARD Repair Patch, SUPPLE PERI-GUARD Repair Patch (K221029)

_Synovis Life Technologies, Inc. · PSQ · Oct 19, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221029

## Device Facts

- **Applicant:** Synovis Life Technologies, Inc.
- **Product Code:** [PSQ](/productcode/PSQ.md)
- **Decision Date:** Oct 19, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3470
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

PERI-GUARD and SUPPLE PERI-GUARD are intended to be used as an intracardiac patch.

## Device Story

PERI-GUARD and SUPPLE PERI-GUARD are bovine pericardium-derived patches used in cardiac surgery. Tissue is crosslinked with glutaraldehyde and treated with sodium hydroxide. Patches are terminally sterilized via gamma irradiation and supplied in a sterile double-pouch system. Used by surgeons in clinical settings for structural repair, defect closure, and suture-line reinforcement. Device provides a mechanical barrier for tissue reconstruction; benefits include structural integrity and biocompatibility for long-term implantation. MR Safe.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including visual inspection, suture retention, dimensional analysis, burst strength, collagenase digestion, denaturation temperature, water permeability, pliability, chemical/heavy metal residuals, bioburden, pyrogenicity/endotoxins, temperature excursion, and biocompatibility (ISO 10993-1:2018). Sterilization validated per ISO 11137; packaging integrity validated per ISO 11607-1, ASTM F88-15, and ASTM F2096-11.

## Technological Characteristics

Bovine pericardium crosslinked with glutaraldehyde and treated with 1M sodium hydroxide. Gamma-sterilized. MR Safe. Packaging validated per ISO 11607-1; sterile barrier integrity tested per ASTM F88-15 and ASTM F2096-11. Biocompatibility per ISO 10993-1:2018 for long-term implant.

## Regulatory Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

## Predicate Devices

- PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Pericardium Patch ([K142447](/device/K142447.md))
- PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch ([K983602](/device/K983602.md))

## Reference Devices

- Synovis PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology ([K192615](/device/K192615.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Synovis Life Technologies, Inc. A subsidiary of Baxter International, Inc. Megan Sajjad Sr. Manager, Regulatory Affairs 2575 University Avenue West St. Paul, Minnesota 55114

Re: K221029

Trade/Device Name: PERI-GUARD Repair Patch, SUPPLE PERI-GUARD Repair Patch Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene Regulatory Class: Class II Product Code: PSQ Dated: September 20, 2022 Received: September 20, 2022

Dear Megan Sajjad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

Page 2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Rachel Neubrander, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221029

Device Name

PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch

Indications for Use (Describe)

PERI-GUARD is intended for repair of pericardial structures. PERI-GUARD is also intended for use as a patch for intracardiac defects, great vessel, septal defects and annulus repair, and suture-line buttressing.

SUPPLE PERI-GUARD is intended for use as a prosthesis for pericardial closure.

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510k SUMMARY: PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch K221029

#### I. SUBMITTER

Synovis Life Technologies, Inc. (Synovis) (A Subsidiary of Baxter International Inc.) 2575 University Avenue West St. Paul, MN 55114-1024 Phone: 651-796-7410 Fax: 224-270-4119

Contact Person: Megan Sajjad, Sr. Manager, Regulatory Affairs

Date prepared: October 18, 2022

# II. DEVICE

Device Trade Names: PERI-GUARD and SUPPLE PERI-GUARD

Common Name: Repair Patch

Classification Name: 21 CFR 870.3470 - Intracardiac Patch Or Pledget, Biologically Derived

Product Code: PSQ

#### III. PREDICATE DEVICES

Primary: PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Pericardium Patch, K142447

Manufacturer: Synovis Life Technologies, Inc. (A Subsidiary of Baxter International Inc.)

Secondary: PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch, K983602

Manufacturer: Synovis Life Technologies, Inc. (A Subsidiary of Baxter International Inc.)

Additionally, Synovis PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology (K192615) serves as a Reference device for this 510k.

{4}------------------------------------------------

# IV. DEVICE DESCRIPTION

PERI-GUARD Repair Patch (PERI-GUARD) and SUPPLE PERI-GUARD Repair Patch (SUPPLE PERI-GUARD) are derived from bovine pericardium procured from cattle originating in the United States. The pericardium is crosslinked with glutaraldehyde and treated with 1 molar sodium hydroxide for a minimum of 60 minutes at 20-25°C (68-77°F).

PERI-GUARD and SUPPLE PERI-GUARD are terminally sterilized using gamma irradiation and packaged within a sterile double-pouch system. The contents of the unopened, undamaged outer pouch are sterile.

PERI-GUARD and SUPPLE PERI-GUARD are MR Safe.

PERI-GUARD and SUPPLE PERI-GUARD utilize animal tissue; patient must be informed prior to any procedure.

See Table 1 for PERI-GUARD and SUPPLE PERI-GUARD product models and sizes.

| PERI-GUARD Model Number        | Size (cm) |
|--------------------------------|-----------|
| PG0404                         | 4 x 4     |
| PG0608                         | 6 x 8     |
| PG0814                         | 8 x 14    |
| PG1016                         | 10 x 16   |
| PG1225                         | 12 x 25   |
| SUPPLE PERI-GUARD Model Number | Size (cm) |
| SPG0404                        | 4 x 4     |
| SPG0406                        | 4 x 6     |
| SPG0608                        | 6 x 8     |
| SPG0814                        | 8 x 14    |

#### Table 1 - PERI-GUARD and SUPPLE PERI-GUARD Product Models and Sizes

# V. INTENDED USE/INDICATIONS FOR USE

# Statement of Intended Use:

PERI-GUARD and SUPPLE PERI-GUARD are intended to be used as an intracardiac patch.

{5}------------------------------------------------

# Indications for Use

PERI-GUARD is intended for repair of pericardial structures. PERI-GUARD is also intended for use as a patch for intracardiac defects, great vessel, septal defects and annulus repair, and suture-line buttressing.

SUPPLE PERI-GUARD is intended for use as a prosthesis for pericardial closure.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

PERI-GUARD and SUPPLE PERI-GUARD are substantially equivalent to the predicate PERI-GUARD and SUPPLE PERI-GUARD devices based on having the same fundamental technology and intended use. The changes between the subject and predicate devices include packaging and sterilization changes. The safety and performance of the subject PERI-GUARD and SUPPLE PERI-GUARD devices have been evaluated through non-clinical testing.

The subject and predicate devices are identical in the following respects:

- Intended Use (both predicates) ●
- Indication for use (both predicates) ●
- Viral inactivation processing steps (both predicates) ●
- Sterilization method (primary predicate) ●
- . Strength specifications (both predicates)
- . Chemical and physical specifications (secondary predicate)

The following technological differences exist between the subject and predicate devices:

- Modified packaging design .
- Same sterilization method but modified sterilization parameters and . packaging for sterilization compared to the primary predicate

#### PERFORMANCE DATA VII.

The following performance data were provided in support of the substantial equivalence determination:

- . Visual
- Suture retention ●
- Dimensional ●
- Burst strength ●
- . Collagenase digestion
- Denaturation Temperature ●
- Water Permeability ●
- Pliability ●
- Chemical and heavy metal residuals ●
- Bioburden

{6}------------------------------------------------

- Pyrogenicity/Endotoxins
- Temperature Excursion Testing ●
- Biocompatibility Assessment ●

The results of performance testing demonstrate the modified PERI-GUARD and SUPPLE PERI-GUARD devices are substantially equivalent to the predicate devices.

# Biocompatibility

The results of new biocompatibility testing, along with applicable historical testing of the primary predicate devices, demonstrate the biocompatibility of the product in accordance with ISO 10993-1:2018 requirements for a long-term implant with tissue/blood contact and equivalence to the predicate devices.

# Shelf Life

Synovis has performed aging testing to support a 1 year shelf life claim.

#### Sterilization and Packaging

Sterilization validation was conducted according to ISO 11137 parts 1 and 2.

The the modified packaging was designed and evaluated in accordance with ISO 11607-1. The integrity of the sterile barrier is supported by testing conducted in accordance with ASTM F88-15 and ASTM F2096-11.

#### Validation Studies

Human factors testing was conducted to confirm that the re-designed packaging allows for aseptic transfer of the product to the sterile field without compromising sterility, and that the modified packaging design does not impact the surface of the tissue patch.

# VII. CONCLUSION

The subject PERI-GUARD and SUPPLE PERI-GUARD devices share the same intended use and technological characteristics as the predicate devices. The physical, functional and performance specifications for the devices are substantially equivalent. Testing supports that the subject devices are as safe and effective as the predicate devices when used according to their labeling.

---

**Source:** [https://fda-staging.innolitics.com/device/K221029](https://fda-staging.innolitics.com/device/K221029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
