IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Over-the-counter, home-use IPL device for hair removal; utilizes Xenon arc flashlamp to emit light pulses; powered by external adapter; features finger-switch activation and integrated skin sensor to ensure full contact before pulse emission; includes cooling function for user comfort; intended for single-person use on body and face; device output controlled by microprocessor; user operates device by placing treatment window against skin; light pulses target hair follicles to reduce regrowth; benefits include non-invasive hair reduction with minimal pain.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, 10993-10), electrical safety/EMC (IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, 60601-2-83), photobiological eye safety (IEC 62471), and usability engineering (IEC 60601-1-6).
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults with Fitzpatrick skin types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
iPulse SmoothSkin Gold Hair Removal System (K160968)
Reference Devices
IPL Hair Removal Device Joy Version, CB-027 (K173813)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 1, 2022
Shenzhen Ulike Smart Electronics Co.,Ltd % Tracy Che Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90. Oianhai Road Shenzhen, Guangdong 518052 China
Re: K221002
Trade/Device Name: IPL Hair Removal Device, Model(s): UI04A, UI04B, UI04C Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: March 22, 2022 Received: April 4, 2022
Dear Tracy Che:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221002
Device Name IPL Hair Removal Device Model(s): UI04A, UI04B, UI04C
#### Indications for Use (Describe)
IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K221002
# "510(k) Summary" as required by 21 CFR Part 807.92.
#### I. Submitter
Shenzhen Ulike Smart Electronics Co.,Ltd
No.8, Keyuan Road, Yuehai Subb-district, Nanshan District, Shenzhen, Guangdong, China Post code: 518000 Tel.: +86 15915373017
Xiaoming Lin Regulation Manager Tel: +86 15915373017 Email: rd5(@ulikebeauty.com
### II. Device
Name of Device: IPL Hair Removal Device Model(s): UI04A, UI04B, UI04C Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810
# III. Predicate Device
Primary predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
|----------------|-----------------------------------------------|---------------|---------------|
| CyDen Limited. | iPulse SmoothSkin Gold<br>Hair Removal System | K160968 | Apr.14, 2016 |
Reference devices:
| Manufacturer | Reference Device | 510(k) Number | Approval Date |
|------------------------------------------------------------|------------------------------------------------|---------------|----------------|
| Shen Zhen CosBeauty Co., Ltd | IPL Hair Removal Device<br>Joy Version, CB-027 | K173813 | Sept. 07, 2018 |
| Dongguan Define Beauty<br>Electronic Technology Co.<br>Ltd | IPL HAIR REMOVAL SG-<br>8025 | K212318 | Jan.14, 2022 |
# IV. Device Description
IPL Hair Removal Device (Model: UI04A, UI04B, UI04C), is an over-the-counter, home-use and single-person-use device for hair reduction by using Intense Pulsed Light (IPL). It works below
{4}------------------------------------------------
the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable treatment window and is suitable for multiple hair removal areas (such as: lips, underarms, bikini lines, arms, legs, etc.). It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.
IPL Hair Removal Device, model: UI04A, UI04B, UI04C have the same indication for use, performance, structure design and operation, the only deference is their enclosure color (UI04A is white, UI04B is mint green and UI04C is deep green).
# V. Indications for Use
IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
# VI. Materials
| Component<br>name | Material of Component | Body Contact Category | Contact Duration |
|----------------------------------------------------------------------|---------------------------------|-------------------------------------------|--------------------|
| IPL Hair<br>Removal Device<br>(Enclosure and<br>treatment<br>window) | ABS, PC,<br>Crystal application | Surface-contacting<br>device: Intact skin | Less than 24 hours |
We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to Section 16 "Biocompatibility Discussion".
#### VII.Comparison of Technological Characteristics With the Predicate Device
The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device and reference devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference devices.
IPL Hair Removal Device is compared with the following Predicate Device and Reference Devices in terms of intended use, design, material, specifications, and performance:
{5}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Remark | |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-------------------|
| 510(k)<br>Number | Pending | K160968 | K173813 | K212318 | / | |
| Trade name | IPL Hair Removal<br>Device (UI04A,<br>UI04B, UI04C) | iPulse SmoothSkin<br>Gold Hair Removal<br>System | IPL Hair Removal<br>Device Joy Version, CB-<br>027 | IPL HAIR<br>REMOVAL SG-/<br>8025 | / | |
| Manufacturer | Shenzhen Ulike<br>Smart Electronics<br>Co.,Ltd | Cyden Limited. | Shen Zhen CosBeauty<br>Co., Ltd | Dongguan Define<br>Beauty Electronic<br>Technology Co.<br>Ltd | / | |
| Regulation<br>number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | |
| Product code | OHT | OHT, GEX | OHT | OHT | Same | |
| Device<br>classification | Class II | Class II | Class II | Class II | Same | |
| Indication for<br>use/ Intended<br>use | IPL Hair<br>Removal<br>Device is indicated<br>for the removal of<br>unwanted hair. The<br>device is also<br>indicated for the<br>permanent reduction<br>in hair regrowth,<br>defined as the long-<br>term, stable reduction<br>in the number of hairs<br>regrowing when<br>measured at 6, 9 and<br>12 months after the<br>completion of a<br>treatment regime. | The<br>iPulse<br>SmoothSkin Gold<br>Hair Removal System<br>is indicated for the<br>removal of unwanted<br>hair. The iPulse<br>Smoothskin Gold is<br>also indicated for the<br>permanent reduction<br>in hair regrowth,<br>defined as the long-<br>term, stable reduction<br>in the number of hairs<br>regrowing when<br>measured at 6, 9 and<br>12 months after the<br>completion of a<br>treatment regime. | The IPL Hair Removal<br>Device Joy Version is<br>indicated for the<br>removal of unwanted<br>hair. The device is also<br>indicated for the<br>permanent reduction in<br>hair regrowth, defined<br>as the long-term, stable<br>reduction in the number<br>of hairs regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a<br>treatment regime. The<br>device is used for adults<br>with Fitzpatrick skin<br>types I - IV. | The IPL Hair<br>Removal (Model:<br>SG-8025) is an<br>over the Counter<br>device intended for<br>the removal of<br>unwanted body<br>and/or facial hair<br>in adults. It is also<br>intended for<br>permanent<br>reduction in<br>unwanted hair.<br>Permanent hair<br>reduction is<br>defined as the<br>long-term stable<br>reduction in the<br>number of hairs re-<br>growing when<br>measured at 6. 9.<br>and 12 months<br>after the<br>completion of a<br>treatment regimen. | Same, only<br>wording<br>difference | |
| Prescription<br>or OTC | OTC | OTC | OTC | OTC | Same | |
| Applicable | Fitzpatrick<br>Skin | Unknown | Fitzpatrick skin types I - | Unknown | Different | |
| Comparison<br>Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Remark | |
| skin | Types I-V | | IV | | Note 1 | |
| Treatment<br>area | large areas (e.g. arms,<br>legs, chest) and small<br>areas (e.g. fingers,<br>lip) | Unknown | Large areas (legs, arms,<br>back and abdomen), face<br>(upper lip, chin and<br>sideburns) | Body and face | Similar | |
| Device design | | | | | | |
| Source<br>energy | Supplied by external<br>adapter | External<br>Power<br>supply | Supplied by external<br>adapter | Supplied by<br>external adapter | Same | |
| Power supply | 100-240V~, 50/60Hz | 110V or 230V,<br>50/60Hz | 100-240 VAC, 50/60Hz | Unknown | Same | |
| Dimension | 60.5(W)x38(H)x169.<br>7(L)mm | Unknown | 126*78*200mm | 205*76*56mm<br>(H*W*D) | Different<br>Note 2 | |
| Sterilization | Not required | Not required | Not required | Not required | Same | |
| Output specification | | | | | | |
| Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same | |
| Energy<br>medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon lamp | Same | |
| Wavelength<br>range | 550-1200mm | 510-1100nm | 510-1200nm | 530nm | Similar<br>Note 3 | |
| Energy<br>density | $3.03-5.3J/cm^2$ | $3-6J/cm^2$ | $1.8~5.1J/cm^2$ | $2.5J/cm^2$ | Similar<br>Note 4 | |
| Output<br>energy | $10.0~17.5J$ | $9~18J$ | Body<br>lamp<br>cartridge | 11.77~22.21J<br>510~1200nm | Level 1: 7.5J<br>Level 2: 8.5J<br>Level 3: 9.5J<br>Level 4: 11J<br>Level 5: 12J | Similar<br>Note 4 |
| | | | Facial<br>lamp<br>cartridge | 3.65~7.04J<br>512~1197nm | | |
| | | | Bikini<br>lamp<br>cartridge | 3.84~7.22J<br>511~1200nm | | |
| Spot size | 3.3cm² | 3cm² (3cm by 1cm) | Body: 4.2cm²<br>Bikini and face: 2.0cm² | 3.0cm² | Similar<br>Note 5 | |
| Pulse<br>duration | 7-10ms | 2ms to 10ms | 9.20~11.20ms | 1ms | Similar<br>Note 6 | |
| Pulsing<br>control | Finger switch | Finger switch | Finger switch | Finger switch | Same | |
| Delivery<br>device | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Direct illumination<br>to tissue | Same | |
| Output<br>intensity level | 5 Levels | Unknown | 5 levels | 5 levels | Same | |
| Comparison<br>Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device<br>2 | Remark | |
| Software/<br>Firmware/<br>Microprocess<br>or Control? | Yes | Yes | Yes | Yes | Same | |
| Additional features | | | | | | |
| Electrical<br>safety | IEC 60601-1-2<br>IEC 60601-1-11<br>ANSI<br>AAMI<br>ES60601-1<br>IEC 60601-2-57<br>IEC 60601-2-83 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 | Similar | |
| Eye safety | IEC 62471 | IEC 62471 | Unknown | IEC 62471 | Same | |
| Biocompatibi<br>lity | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.