1) When using Electrical Muscle Stimulation, Transcutaneous Electrical Nerve Stimulator is intended: for users with conditions or disease that are associated with impaired (poor) blood flow in the legs/ ankles/ feet, the device through the foot-pads is intended for use as an adjunctive treatment (as an addition to your existing treatment) to temporarily reduce lower extremity pain, swelling and cramping · to temporarily increase local blood circulation in healthy leg muscles · to stimulate healthy muscles in order to improve and facilitate muscle performance · to temporarily relieve pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties · to relax muscle spasm to increase blood flow circulation · for prevention of retardation of disuse atrophy for muscle re-education · for maintaining or increasing range of motion · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis · Provide quadricep strengthening · Improve knee stability secondary to quadricep strengthening 2) When using Transcutaneous Electrical Nerve Stimulator to deliver Transcutaneous Electrical Nerve Stimulation (TENS), it is intended to provide: · symptomatic relief and management of chronic, intractable pain · relief of pain associated with arthritis · temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
Device Story
Device (KTR-405) is a transcutaneous electrical nerve stimulator (TENS) and electrical muscle stimulator (EMS) for OTC use. Input: user-selected modes and intensity levels via wireless remote or device controls. Operation: device generates biphasic, pulsed symmetric, rectangular electric pulses delivered to skin via footpads or electrode patches. Used by patients in home/clinic settings while seated or reclining. Output: electrical stimulation to muscles/nerves. Clinical effect: temporary pain relief, increased local blood circulation, muscle conditioning, and prevention of disuse atrophy. Healthcare provider/patient uses output to manage chronic pain or post-surgical recovery. Benefits: non-invasive pain management, improved muscle performance, and reduced swelling.
Clinical Evidence
No clinical data. Bench testing only. Verification included software validation, waveform testing per FDA guidance, and biocompatibility testing (ISO 10993-5, ISO 10993-10) for footpads. Electrical safety and EMC testing performed per IEC 60601 series.
Indicated for adults with conditions/diseases associated with impaired blood flow in legs/ankles/feet; adjunctive treatment to reduce lower extremity pain, swelling, cramping; increase local blood circulation; stimulate healthy muscles; improve muscle performance; relieve pain from sore/aching muscles (shoulder, waist, back, upper/lower extremities) due to strain; relax muscle spasms; prevent disuse atrophy; muscle re-education; maintain/increase range of motion; post-surgical calf stimulation to prevent venous thrombosis; quadricep strengthening; improve knee stability; symptomatic relief of chronic intractable pain; relief of arthritis pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 24, 2022
Shenzhen Kentro Medical Electronics Co., Ltd. % Yvonne Liu Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90. Oianhai Road Shenzhen, Guangdong 518052 China
Re: K220998
Trade/Device Name: Transcutaneous Electrical Nerve Stimulator, Model: KTR-405 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX, IPF, NYN, GZJ Dated: July 12, 2022 Received: July 18, 2022
Dear Yvonne Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K220998
#### Device Name
Transcutaneous Electrical Nerve Stimulator, Model: KTR-405
Indications for Use (Describe)
- 1) When using Electrical Muscle Stimulation, Transcutaneous Electrical Nerve Stimulator is intended:
- for users with conditions or disease that are associated with impaired (poor) blood flow in the legs/ ankles/ feet, the device through the foot-pads is intended for use as an adjunctive treatment (as an addition to your existing treatment) to temporarily reduce lower extremity pain, swelling and cramping
- · to temporarily increase local blood circulation in healthy leg muscles
- · to stimulate healthy muscles in order to improve and facilitate muscle performance
- · to temporarily relieve pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
- (arms) and lower extremities(legs) due to strain from exercise or normal household duties
- · to relax muscle spasm
- to increase blood flow circulation
- · for prevention of retardation of disuse atrophy
- for muscle re-education
- · for maintaining or increasing range of motion
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- · Provide quadricep strengthening
- · Improve knee stability secondary to quadricep strengthening
2) When using Transcutaneous Electrical Nerve Stimulator to deliver Transcutaneous Electrical Nerve Stimulation (TENS), it is intended to provide:
- symptomatic relief and management of chronic, intractable pain
- · relief of pain associated with arthritis
· temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K220998
This "510(k) Summary" of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.
# (1) Applicant information:
| 510(k) owner's name: | SHENZHEN KENTRO MEDICAL ELECTRONICS CO., LTD |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 2nd Floor No 11, Shanzhuang Road, Xikeng Village, Yuanshan<br>Street, Longgang District, Shenzhen City, Guangdong<br>Province,China |
| Contact person: | Zewu Zhang |
| Phone number: | +86 755 3382 5998 |
| Fax number: | +86 755 3382 5996 |
| Email: | 522378976@qq.com |
| Date of summary prepared: | August 24, 2022 |
## (2) Reason for submission
New device, there were no prior submissions for the device.
## (3) Proprietary name of the device
| Trade name/model: | Transcutaneous Electrical Nerve Stimulator/ KTR-405 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name: | Stimulator, Nerve, Transcutaneous, Over-The-Counter;<br>Stimulator, Muscle, Powered, For Muscle Conditioning;<br>Stimulator, Muscle, Powered;<br>Stimulator, Electrical, Transcutaneous, For Arthritis;<br>Stimulator, Nerve, Transcutaneous, For Pain Relief |
| Regulation number: | 21 CFR 882.5890, 890.5850 |
| Product code: | NUH, NGX, IPF, NYN, GZJ |
| Review panel: | Neurology, Physical Medicine |
| Regulation class: | Class II |
## (4) Predicate device
| | Predicate device |
|-----------------------|---------------------------------------|
| Sponsor | ACTEGY LTD |
| Device Name and Model | Revitive Medic Coach<br>Model: 5575AQ |
{4}------------------------------------------------
| 510(k) Number | K210825 |
|-------------------|---------------------------|
| Product Code | NGX, NUH, IPF, NYN, GZJ |
| Regulation Number | 21 CFR 890.5850, 882.5890 |
| Regulation Class | II |
#### (5) Description/ Design of device:
Transcutaneous Electrical Nerve Stimulator is a product that adopts modern electronic science and technology to deliver electric pulses generated to the user's skin through the electrodes.
The proposed model KTR-405 provides a combination of transcutaneous electrical nerve stimulation (TENS) and electrical muscle stimulation (EMS). It has the following basic characteristics: 1) 10 modes to satisfy different demands. Modes 1-8 are for foot and can be switched to another mode for a minute automatically, while mode 9 is for knee and mode 10 for body. Modes for foot, knee and body can be switched manually during a session; 2) wonderful electric pulse combination, 0~150 levels can be adjusted and chosen according to personal preference; 3) rechargeable battery which can work up to 36 hours 4) wireless and rechargeable remote control allows easy control of the device 5) use while seated or reclining with Relax and Recline mode; 6)foot cover helps keep feet warm and cosy; 7) LCD display and power display make the operation simple and easy; 8) carry handle for easy transportation; 9) magnetic power cable for simple and safe connection.
Transcutaneous Electrical Nerve Stimulator is mainly composed of the host and electrode patches. And it can be run either from the mains power supply or via its internal rechargeable battery. To start therapy, you need to connect the device to the mains power supply. This will charge the battery, allowing you to use the device wirelessly in subsequent sessions. Then while seated, place both your bare feet onto the footpads of the device or paste the electrode pads onto painful areas and press on/off button to power on. The modes and intensity can be selected according to needs. And the current status is displayed on LCD.
#### (6) Intended use / indications:
1) When using Electrical Muscle Stimulation, Transcutaneous Electrical Nerve Stimulator is intended:
· for users with conditions or disease that are associated with impaired (poor) blood flow in the legs/ ankles/ feet, the device through the foot-pads is intended for use as an adjunctive treatment (as an addition to your existing treatment) to temporarily reduce lower extremity pain, swelling and cramping
- · to temporarily increase local blood circulation in healthy leg muscles
- · to stimulate healthy muscles in order to improve and facilitate muscle performance
{5}------------------------------------------------
· to temporarily relieve pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
- to relax muscle spasm
- to increase blood flow circulation
- for prevention of retardation of disuse atrophy
- for muscle re-education
- · for maintaining or increasing range of motion
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- · Provide quadricep strengthening
- · Improve knee stability secondary to quadricep strengthening
2) When using Transcutaneous Electrical Nerve Stimulator to deliver Transcutaneous Electrical Nerve Stimulation (TENS), it is intended to provide:
- · symptomatic relief and management of chronic, intractable pain
- relief of pain associated with arthritis
· temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
#### (7) Materials
| Component name | Material of<br>Component | Body Contact<br>Category | Contact Duration |
|----------------|--------------------------------------|--------------------------|--------------------|
| Electrode pads | EVA foam, carbon film, hydrogel, PET | Surface skin contact | Less than 24 hours |
| Foot pads | Silica gel | Surface skin contact | Less than 24 hours |
We have directly purchased the electrode pads from qualified supplier which has obtained FDA clearance with a 510(k) number of K171381 and been legally marketed to US market. Also we've conducted biocompatibility tests for foot pads and all pass. For details, please refer to "Biocompatibility Discussion".
## (8) Technological characteristics and substantial equivalence
{6}------------------------------------------------
| Item | Subject device | Predicate device | Remark |
|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Trade name | Transcutaneous Electrical<br>Nerve Stimulator<br>Model: KTR-405 | Revitive Medic Coach<br>Model: 5575AQ | / |
| 510 (k) number | K220998 | K210825 | / |
| Regulation number | 21 CFR 882.5890, 890.5850 | 21 CFR 882.5890, 890.5850 | Same |
| Regulation description | Transcutaneous electrical nerve<br>stimulator for pain relief;<br>powered muscle stimulator | Transcutaneous electrical<br>nerve stimulator for pain<br>relief; powered muscle<br>stimulator | Same |
| Product code | NUH, NGX, IPF, NYN, GZJ | NGX, NUH, IPF, NYN, GZJ | Same |
| Class | II | II | Same |
| Indications for use/<br>Intended use | 1) When using Electrical<br>Muscle Stimulation,<br>Transcutaneous Electrical<br>Nerve Stimulator is intended:<br>• for users with conditions or<br>disease that are associated<br>with impaired (poor) blood<br>flow in the legs/ ankles/ feet,<br>the device through the<br>foot-pads is intended for use<br>as an adjunctive treatment (as<br>an addition to your existing<br>treatment) to temporarily<br>reduce lower extremity pain,<br>swelling and cramping<br>• to temporarily increase<br>local blood circulation in<br>healthy leg muscles<br>• to stimulate healthy<br>muscles in order to improve<br>and facilitate muscle<br>performance<br>• to temporarily relieve pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back, upper<br>extremities (arms) and lower<br>extremities (legs) due to strain<br>from exercise or normal<br>household duties<br>• to relax muscle spasm | When using Electrical Muscle<br>Stimulation (NMES),<br>Revitive is intended:<br>• for users with conditions or<br>disease that are associated<br>with impaired (poor) blood<br>flow in the legs/ ankles/ feet,<br>NMES though the foot-pads<br>is intended for use as an<br>adjunctive treatment (as an<br>addition to your existing<br>treatment) to temporarily<br>reduce lower extremity pain,<br>swelling and cramping<br>•to temporarily increase local<br>blood circulation in healthy<br>leg muscles<br>•to stimulate healthy muscles<br>in order to improve and<br>facilitate muscle performance<br>•to temporarily relieves pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back, upper<br>extremities (arms) and lower<br>extremities(legs)<br>due to strain from exercise or<br>normal household duties<br>•for relaxation of muscle<br>spasm,<br>•for increase of blood flow | Same |
| | to increase blood flow circulation | circulation, for prevention of retardation | |
| | for prevention of | of disuse atrophy, | |
| | retardation of disuse atrophy | for muscle re-education, | |
| | for muscle re-education | for maintaining or | |
| | for maintaining or | increasing range of | |
| | increasing range of motion | motion,and | |
| | Immediate post-surgical | Immediate post-surgical | |
| | stimulation of calf muscles to | stimulation of calf muscles to | |
| | prevent venous thrombosis | prevent venous thrombosis | |
| | Provide quadricep | Provide quadricep | |
| | strengthening | strengthening | |
| | Improve knee stability | Improve knee stability | |
| | secondary to quadricep | secondary to quadricep | |
| | strengthening | strengthening | |
| | 2) When using | (2) When using Revitive to | |
| | Transcutaneous Electrical | deliver Transcutaneous | |
| | Nerve Stimulator to deliver | Electrical Nerve Stimulation | |
| | Transcutaneous | (TENS), it is intended to | |
| | Nerve Stimulation (TENS), it | provide: | |
| | is intended to provide: | symptomatic relief and | |
| | symptomatic relief and | management of chronic, | |
| | management of chronic, | intractable pain | |
| | intractable pain | relief of pain associated with | |
| | relief of pain associated | arthritis | |
| | with arthritis | temporarily relieves pain…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.