← Product Code [LYZ](/productcode/LYZ) · K220992

# Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple (K220992)

_Ever Global (Vietnam) Enterprise Corporation · LYZ · Oct 19, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220992

## Device Facts

- **Applicant:** Ever Global (Vietnam) Enterprise Corporation
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Oct 19, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable, non-sterile, powder-free vinyl examination glove; available in black, blue, or purple; sizes S, M, L, XL. Provides single-use barrier protection for examiner's hands; prevents contamination between patient and examiner. Used in clinical settings; worn by healthcare professionals. Device performance verified against ASTM D5250-19 standards for tensile strength, elongation, and barrier integrity. Biocompatibility confirmed via ISO 10993 testing (irritation, sensitization, cytotoxicity).

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D5151-19 (freedom from holes), ASTM D412-16/D573-04 (tensile strength/elongation), ASTM D6124-06 (residual powder), and ISO 10993 series (biocompatibility: irritation, sensitization, and cytotoxicity).

## Technological Characteristics

Material: Polyvinyl chloride (PVC). Non-sterile. Powder-free. Complies with ASTM D5250-19. Dimensions: 230 mm length min; palm/finger thickness 0.08 mm min. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-23.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Vinyl Examination Glove, Clear/Yellow ([K170575](/device/K170575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 19, 2022

Ever Global (Vietnam) Enterprise Corporation % Elizabeth Deng US Agent Elizabeth Deng 5748 Eaglewood Place Ranch Cucamonga, California 91739

Re: K220992

Trade/Device Name: Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: September 15, 2022 Received: September 19, 2022

Dear Elizabeth Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K220992

#### Device Name

Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple

Indications for Use (Describe)

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'Eg' in yellow, with 'E' being larger than 'g'. Behind the letters is a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom, both in white text on a blue background.

LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG FAX: (0251)351 4023 024 BOOK.COM/PMGLOVESSUPERIEURVIETNAM

# 510(k) SUMMARY

The assigned 510(k) Number: K220992

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

#### 0.0 Summary Preparation Date: -Oct 18th, 2022

#### 1.0 Submitter:

| Submitter's name :      | Ever Global (Vietnam) Enterprise Corporation                             |
|-------------------------|--------------------------------------------------------------------------|
| Submitter's address:    | Long Thanh Industrial Zone<br>Taman Village Dong Nai Province, VN 810000 |
| Phone number:           | 84-61-3514022                                                            |
| Fax number:             | 84-61-3514023                                                            |
| Name of contact person: | Jerry Lin                                                                |

Elizabeth Deng

#### 2.0 US Agent:

US Representative Name: Company Address:

Telephone Number: Contact Email Address:

5748 Eaglewood Place Rancho Cucamonga, California Rancho Cucamonga, CA 91739 909 4659188 baxianunited48(@yahoo.com

#### 3.0 Name of the Device

Proprietary/Trade name:

Common Name: Classification Name: Device Classification: Regulation Number: Product Code:

### 4.0 Predicate device

Device Name:

Company name: 510(K) Number: Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple Vinyl Examination Gloves Non-powdered Patient Examination Glove Class I 21 CFR 880.6250 LYZ

Disposable Powder Free Vinyl Examination Glove, Clear/Yellow Ever Global (Vietnam) Enterprise Corporation K170575

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E', 'G', and 'C' in yellow, superimposed on a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom. The logo is framed by a blue border.

LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG NAI 810000 VIETNA AX: (0251)351 4023 024 ACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

# 510(k) SUMMARY

The assigned 510(k) Number: K220992

### 5.0 Device Description:

A Disposable Powder Free Vinyl Exam Glove (with color Black, Blue, or Purple) is a patient examination glove made from poly (vinyl chloride), non-sterile (as per 21 CFR 880.6250, Class I). This device is available in size S, M, L, and XL. The operation principle of this medical device is to provide single use barrier protection for the wearer and the device meets all the requirement specifications for barrier protection, tensile properties as defined in ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

#### 6.0 Device Indications for use:

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Device<br>Characteristic | Predicate Device                                                                                                                                                                                     | Subject Device                                                                                                                                                                                                       | Comparison           |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Product Name             | Disposable Powder Free Vinyl<br>Examination Glove,<br>Clear/Yellow                                                                                                                                   | Disposable Powder Free Vinyl<br>Exam Glove, Black/Blue/Purple                                                                                                                                                        | Same except<br>color |
| 510(K) No.               | K170525                                                                                                                                                                                              | K220992                                                                                                                                                                                                              | n/a                  |
| Product Owner            | Ever<br>Global<br>(Vietnam)<br>Enterprise Corporation                                                                                                                                                | Ever<br>Global<br>(Vietnam)<br>Enterprise Corporation                                                                                                                                                                | same                 |
| Product Code             | LYZ                                                                                                                                                                                                  | LYZ                                                                                                                                                                                                                  | same                 |
| Regulation               | 21 CFR 880.6250                                                                                                                                                                                      | 21 CFR 880.6250                                                                                                                                                                                                      | same                 |
| Class                    | I                                                                                                                                                                                                    | I                                                                                                                                                                                                                    | same                 |
| Indications for<br>Use   | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or finger<br>to prevent contamination<br>between patient and examiner. | A non-powdered patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or finger<br>to prevent contamination<br>between patient and examiner. | same                 |
| Power free               | Yes                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                  | same                 |
| Size                     | Small/ Medium/Large/X Large                                                                                                                                                                          | Small/Medium/Large/X Large                                                                                                                                                                                           | same                 |
| Over-The-<br>Counter Use | Yes (21 CFR 801 Subpart C)                                                                                                                                                                           | Yes (21 CFR 801 Subpart C)                                                                                                                                                                                           | same                 |
| Single Use               | YES                                                                                                                                                                                                  | YES                                                                                                                                                                                                                  | same                 |
| Non-Sterile              | YES                                                                                                                                                                                                  | YES                                                                                                                                                                                                                  | same                 |
| Dimensions-<br>Length    | Complies with ASTM D5250-06<br>230 mm min.                                                                                                                                                           | Complies with ASTM D5250-19<br>230 mm min.                                                                                                                                                                           | same                 |

#### 7.0 Comparison of device technological characteristics:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'F' and 'g' in a stylized font, with the letters overlapping each other. The letters are yellow, and they are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and the words 'ENTERPRISE CORP.' are at the bottom.

Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

# 510(k) SUMMARY

The assigned 510(k) Number: K220992

| Device<br>Characteristic   | Predicate Device                                                                                                                                | Subject Device                                                                                                                                                                                                       | Comparison                                                 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Dimensions -<br>Palm Width | Complies with ASTM D5250-06<br>Small    85 ± 5<br>Medium  95 ± 5<br>Large   105 ± 5<br>X large   115 ± 5                                        | Complies with ASTM D5250-19<br>Small    85 ± 5<br>Medium  95 ± 5<br>Large   105 ± 5<br>X large   115 ± 5                                                                                                             | same                                                       |
| Dimensions -<br>Thickness  | Complies with ASTM D5250-06<br>Palm - 0.08mm min.<br>Finger - 0.1 mm min.                                                                       | Complies with ASTM D5250-19<br>Palm - 0.08 mm min.<br>Finger - 0.08 mm min.                                                                                                                                          | same, both<br>are thicker<br>than<br>standard<br>request   |
| Physical<br>Properties     | Tensile Strength<br>Before Aging: 11 MPa, min.<br>After Aging: 11 MPa, min.<br>Elongation:<br>Before Aging: 300% min.<br>After Aging: 300% min. | Tensile Strength<br>Before Aging: 11 MPa, min.<br>After Aging: 11 MPa, min.<br>Elongation:<br>Before Aging: 300% min.<br>After Aging: 300% min.                                                                      | same                                                       |
| Residual<br>powder         | Complies with D5250-06<br>< 2mg per glove                                                                                                       | Complies with ASTM D5250-19<br>< 2mg per glove                                                                                                                                                                       | same                                                       |
| Freedom from<br>Holes      | In accordance with ASTM<br>D5250-06 (G-1 with AQL 2.5)                                                                                          | In accordance with ASTM<br>D5250-19 (G-1 with AQL 2.5)                                                                                                                                                               | same                                                       |
| Biocompatibili<br>ty       | ISO 10993-10:<br>-Skin Sensitization: Pass<br>Irritation test: Pass                                                                             | ISO 10993-10 (Blue & Purple<br>Gloves) & ISO 10993-23 (Black<br>Glove):<br>-Skin Irritation test: Pass<br><br>ISO 10993-10<br>-Skin Sensitization Test: Pass<br><br>ISO 10993-5<br>-In vitro cytotoxicity test: Pass | similar.<br>Subject<br>device pass<br>cytotoxicity<br>test |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E' and 'G' in a stylized, interconnected design, with the letters 'e' and 'c' below. The letters are in yellow and are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom.

long thanh industrial zone, taman village, dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

# 510(k) SUMMARY

The assigned 510(k) Number: K220992

### 8.0 Assessment of Non-Clinical Performance Data:

| Test                                              | Test Method                                                                                 | Purpose                                                                                                                       | Acceptance Criteria                                                                                                                                                              | Results |
|---------------------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimension                                         | ASTM D3767                                                                                  | Determine the<br>geometrical dimension<br>of gloves                                                                           | Length: 230 mm min.<br>Thickness:<br>Palm - 0.08 mm min.<br>Finger - 0.08 mm min.<br>Palm Width:<br>Small 85 ± 5 mm<br>Medium 95± 5 mm<br>Large 105 ± 5 mm<br>X Large 115 ± 5 mm | Pass    |
| Freedom from holes<br>(Water leak)                | 21 CFR 800.20. &<br>ASTM D5151-19                                                           | Detect the holes on the<br>gloves.                                                                                            | G-I/AQL 2.5                                                                                                                                                                      | Pass    |
| Tensile strength<br>(Before aging/After<br>aging) | ASTM D412-16<br>& ASTM D573-<br>04                                                          | Evaluate the tensile<br>(tenson) properties of<br>the gloves. In addition,<br>it also determines the<br>influence of elevated | Before Aging: 11 MPa,<br>min.<br>After Aging: 11 MPa,<br>min                                                                                                                     | Pass    |
| Elongation (Before<br>aging/After aging)          | ASTM D412-16<br>& ASTM D573-<br>04                                                          | temperature on the<br>physical properties of<br>gloves.                                                                       | Before Aging: 300%<br>min.<br>After Aging: 300% min.                                                                                                                             | Pass    |
| Powder Residual                                   | ASTM D6124-06                                                                               | Determine the average<br>powder mass found on<br>the gloves                                                                   | < 2mg per glove                                                                                                                                                                  | Pass    |
| Biocompatibility-<br>Skin Irritation              | ISO 10993-<br>10:2010 (Blue &<br>Purple Gloves) &<br>ISO 10993-23:<br>2021 (Black<br>Glove) | determine the potential<br>of glove to promote<br>skin sensitization &<br>irritation reactions after<br>repeated applications | Negative Response                                                                                                                                                                | Pass    |
| Biocompatibility-<br>Skin Sensitization           | ISO 10993-<br>10:2010                                                                       |                                                                                                                               | No contact sensitization                                                                                                                                                         | Pass    |
| Biocompatibility-<br>cytotoxicity                 | ISO 10993-<br>5:2009                                                                        | determine the<br>cytotoxicity potential<br>of glove                                                                           | No in vitro cytotoxicity                                                                                                                                                         | Pass    |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'F' and 'g' in a stylized font, with the letter 'F' stacked on top of the letter 'g'. The letters are yellow and are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are written above the letters, and the words 'ENTERPRISE CORP.' are written below the letters.

Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

## 510(k) SUMMARY

The assigned 510(k) Number: K220992

### 9.0 Assessment of Clinical Performance Data:

Clinical data is not needed for this type of device.

#### 10.0 Conclusion:

The conclusion drawn from the nonclinical tests demonstrate that the subject device Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple is as safe, as effective, and performs as well as or better than the legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K220992](https://fda-staging.innolitics.com/device/K220992)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).