K220934 · Terumo Medical Products (Hangzhou) Co., Ltd. · MOF · Jun 29, 2022 · Cardiovascular
Device Facts
Record ID
K220934
Device Name
RADIFOCUS Torque Device
Applicant
Terumo Medical Products (Hangzhou) Co., Ltd.
Product Code
MOF · Cardiovascular
Decision Date
Jun 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
RADIFOCUS Torque Device is intended to facilitate guidewire manipulation during interventional procedures.
Device Story
RADIFOCUS Torque Device is a manual, single-use accessory used during interventional procedures to facilitate guidewire manipulation. The device consists of a handle (polyoxymethylene) and a screw-on cap (polypropylene) that applies a clamping force to the guidewire. It is operated by a clinician during procedures to provide a grip on the guidewire, allowing for easier rotation and advancement. The device is compatible with guidewires ranging from 0.010" to 0.038" in diameter. By providing a secure handle, it assists the physician in navigating the guidewire through the vasculature, potentially improving procedural efficiency and control.
Clinical Evidence
No clinical data included; substantial equivalence is supported by non-clinical bench testing. Performance tests included appearance, fixed strength of guidewire, fixed strength of cap, fixable dimension, guidewire damage, collet release, and torque slip force. All tests met predetermined acceptance criteria.
Technological Characteristics
Manual torque device; materials: polyoxymethylene (handle), polypropylene (cap). Compatible with 0.010"-0.038" guidewires. Sterilized via ethylene oxide (EtO) per ISO 11135:2014 to SAL 10^-6. Single-use, disposable.
Indications for Use
Indicated for facilitating guidewire manipulation during interventional procedures for patients requiring such procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
June 29, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below that.
Terumo Medical Products (Hangzhou) Co., Ltd. % Qing Liu Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue. Suite 320 Somerset, New Jersey 08873
Re: K220934
Trade/Device Name: RADIFOCUS Torque Device Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, PTL Dated: March 31, 2022 Received: March 31, 2022
Dear Qing Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220934
Device Name RADIFOCUS Torque Device
Indications for Use (Describe)
RADIFOCUS Torque Device is intended to facilitate guidewire manipulation during interventional procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Terumo logo. The logo features a red swoosh above the word "TERUMO" in green. The font is sans-serif and the letters are bold.
# 510(K) SUMMARY
#### A. SUBMITTER INFORMATION (807.92(a)(1))
| Prepared by: | Qing Liu |
|--------------|-------------------------------|
| | Regulatory Affairs Specialist |
| | Terumo Medical Corporation |
| | Tel. (908) 842-9016 |
| | Fax (410) 398-6079 |
#### Prepared for: Owner/Operator
Terumo Medical Products (Hangzhou) Co., Ltd. M4-9-5 Hangzhou Economic and Technological Development Zone, Hangzhou, 310018, People's Republic of China Registration Number: 3004102031
# Manufacturer and Sterilization Facility (Applicant)
Terumo Medical Products (Hangzhou) Co., Ltd. M4-9-5 Hangzhou Economic and Technological Development Zone, Hangzhou, 310018, People's Republic of China Registration Number: 3004102031
#### Contact Person: Qing Liu
Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Ave, Suite 320 Somerset, NJ 08873, USA Tel. (908)842-9016 Fax (410) 398-6079 E-mail: kyo.ryu@terumomedical.com
Date prepared: 6/29/2022
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red swoosh above the company name, which is written in green, block letters. The swoosh is curved and tapers to a point on the left side of the logo.
#### B. DEVICE NAME (807.92(a)(2))
| Proprietary Name: | RADIFOCUS Torque Device |
|-----------------------|------------------------------|
| Common Name: | Torque Device |
| Classification Name: | Catheter Guide Wire |
| Classification Panel: | Neurology, Cardiovascular |
| Regulation: | 21 CFR 870.1330 |
| Product Code: | Primary: MOF, Secondary: PTL |
| Classification: | Class II |
#### C. PREDICATE DEVICE (807.92(a)(3))
The legally marketed device to which substantial equivalence is claimed is: K910969 - TORQUE DEVICE FOR A GUIDE WIRE, manufactured by Terumo Corporation, Japan.
### D. DEVICE DESCRIPTION (807.92(a)(4))
RADIFOCUS Torque Device is a supportive device intended to facilitate guide wire manipulation during interventional procedures.
This product consists of a cap that screws onto the main body of the device (handle) to apply a clamping force on the guide wire. The cap is made of polypropylene and the handle is made of polyoxymethylene. The torque device can be used for the guide wires with diameters of 0.010" to 0.038". RADIFOCUS Torque Device is sterilized with ethylene oxide.
### E. INDICATIONS FOR USE (807.92(a)(5))
RADIFOCUS Torque Device is intended to facilitate guidewire manipulation during interventional procedures.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green. The red curved line is positioned above and to the left of the word "TERUMO", creating a dynamic and modern look. The font used for "TERUMO" is bold and sans-serif, contributing to the logo's clean and professional appearance.
# F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS (807.92(a)(6)) RADIFOCUS Torque Device, the subject of this 510(k), is substantially equivalent in its intended use, technology, principle of operation, materials, and performance to the TORQUE DEVICE FOR A GUIDE WIRE, manufactured by Terumo Corporation, Japan (K910969).
The comparison of the technological characteristics is summarized in the table below.
| Device<br>Characteristic | Subject Device:<br>RADIFOCUS Torque<br>Device (K220934) | Predicate Device:<br>TORQUE DEVICE FOR A<br>GUIDE WIRE (K910969) |
|---------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Manufacturer | Terumo Medical Products<br>(Hangzhou) Co., Ltd. | Terumo Corporation |
| Operation<br>Principle | Manual | Same |
| Design /<br>Construction | This product consists of a cap and a<br>handle with a clamp. | Same |
| Materials | · Cap - Polypropylene<br>• Handle –Polyoxymethylene | Same |
| Package | • Unit pouch<br>• Shelf box<br>• Large shelf box | Same |
| Specifications | Compatible with guide wires that<br>range from 0.010" (0.26 mm) to<br>0.038" (0.97 mm) in diameter. | Compatible with guide wires that<br>range from 0.014" (0.36 mm) to<br>0.038" (0.97 mm) diameter. |
| Sterilization | Ethylene Oxide (validated in<br>accordance with ANSI / AAMI /<br>ISO 11135-1 to achieve SAL 10-6) | Same |
| Disposable,<br>Single Use | Yes | Same |
## G. NON-CLINICAL TESTS (807.92(b)(1))
### Performance
Performance testing was conducted to ensure the RADIFOCUS Torque Device met the predetermined specifications throughout the shelf life, to verify conformity to the acceptance criteria, and demonstrate substantial equivalence to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Terumo logo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is positioned above and to the left of the word.
No new issues of safety and effectiveness were raised with the testing performed. The following performance tests were performed on the RADIFOCUS Torque Device:
| Performance Test | Results |
|------------------------------|---------------------------|
| Appearance | Meets acceptance criteria |
| Fixed strength of guide wire | Meets acceptance criteria |
| Fixed strength of cap | Meets acceptance criteria |
| Fixable dimension | Meets acceptance criteria |
| Guide wire damage | Meets acceptance criteria |
| Collet release | Meets acceptance criteria |
| Torque slip force | Meets acceptance criteria |
The RADIFOCUS Torque Device tested met the predetermined acceptance criteria. Based on the results of the performance testing, the subject RADIFOCUS Torque Device is substantially equivalent to the predicate.
### Biocompatibility
The RADIFOCUS Torque Device is a non-tissue contacting medical device, therefore, this 510(k) submission does not include biocompatibility data.
### Sterilization
The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices.
The sterilization process was validated utilizing the overkill half cycle approach to provide a Sterility Assurance Level (SAL) of 10-6.
Post sterilization, RADIFOCUS Torque Device product meets the maximum residue limits in accordance with ISO 10993-7: 2008.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red arc above the company name, which is written in green, block letters. The arc is positioned above and to the left of the word "TERUMO."
The maximum allowable residue levels are:
| - Ethylene oxide (EtO): | 4 mg/device |
|--------------------------------|-------------|
| - Ethylene chlorohydrin (ECH): | 9 mg/device |
#### Risk Analysis
A Product Risk Analysis was conducted in accordance with ISO 14971, and it was determined that there were no new issues of safety or effectiveness.
#### H. CLINICAL TESTS (807.92(b)(2))
This 510(k) submission does not include data from clinical tests, which were not deemed necessary.
### I. CONCLUSION (807.92(b)(3))
In summary, the RADIFOCUS Torque Device, subject of this 510(k), is substantially equivalent in its intended use, technology, principle of operation, materials, and performance to the primary predicate device K910969 – TORQUE DEVICE FOR A GUIDE WIRE manufactured by Terumo Corporation, Japan.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.